Argyle¿ Polyvinyl Chloride (PVC) and Kangaroo¿ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector
Applicant
Covidien
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jul 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector are intended for enteral feeding to deliver enteral nutrition, liquid or medication to patient from an enteral feeding syringe or feeding set designed with a connector for enteral applications. The device is intended for pediatric patients who require enteral feeding.
Device Story
Sterile, disposable single-lumen feeding tubes; constructed of Polyvinyl Chloride (PVC) or Polyurethane (PU). Features proximal male ENFit small-bore connector compliant with ISO 80369-3. Used in clinical settings for pediatric patients unable to obtain nutrition orally. Device delivers enteral nutrition, liquids, or medications from a compatible syringe or feeding set. Healthcare providers operate the device to provide nutritional support. Benefits include safe, reliable enteral delivery and reduced risk of misconnection via standardized ENFit interface.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, stability, dimensional analysis, and performance testing (fluid leakage, stress cracking, tensile strength) per ISO 80369-3 and ISO 80369-20. Human factors validation and FMEA risk assessment for ENFit misconnection were performed.
Technological Characteristics
Single-lumen PVC or PU tubes; proximal male ENFit small-bore connector. Dimensions and connector design conform to ISO 80369-3. Biocompatible materials. Sterile, disposable. No energy source. No software.
Indications for Use
Indicated for pediatric patients requiring enteral feeding due to illness preventing normal chewing or swallowing; used to deliver nutrition, liquids, or medications.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 13, 2015
Covidien Sona Manickam Senior Regulatory Affairs Specialist 15 Hampshire Street Mansfield, MA 02048
Re: K150084
> Trade/Device Name: Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF Dated: June 12, 2015 Received: June 15, 2015
Dear Sona Manickam,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150084
Device Name
Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector
#### Indications for Use (Describe)
The Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector are intended for enteral feeding to deliver enteral nutrition, liquid or medication to patient from an enteral feeding syringe or feeding set designed with a connector for enteral applications.
The device is intended for pediatric patients who require enteral feeding.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b>☑</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 2.0 510(k) Summary
Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit Connector
In accordance with section 513(i) of the SMDA and as defined in 21CFR Part 807.92 this summary is submitted by:
Covidien 15 Hampshire Street Mansfield, MA 02048 Date Prepared: July 09, 2015
a. Contact Person
Sona Manickam Senior Regulatory Affairs Specialist Covidien Telephone: (508) 261-8147 Fax: (508) 261-8461
- b. Name of Medical Device
Common Name: tube, feeding
U.S. FDA Classification Product Code: PIF
U.S. Regulation Description: Gastrointestinal tube and accessories, 21 CFR 876.5980
Proprietary / Trade Name: Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit Small Bore Connector
- c. Identification of Legally Marketed Device(s)
Argyle Polyvinyl Chloride (PVC) Feeding Tubes, K820441 Argyle Indwell Polyurethane Feeding Tubes, K820442
d. Device Brief Description
The Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit Connector are sterile, disposable medical devices designed to deliver nutrition to patients who cannot obtain nutrition by mouth, are unable to chew or swallow safely, or need nutritional supplementation. The device is designed with a single lumen Polyvinyl Chloride (PVC) or Polyurethane (PU) tubes with ENFit male small bore connector at proximal end.
{4}------------------------------------------------
## e. Device Intended Use
The Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit connector are intended for enteral feeding to deliver enteral nutrition, liquid or medication to patient from an enteral feeding syringe or feeding set designed with a connector for enteral applications.
The device is intended for pediatric patients who require enteral feeding.
- f. Product Comparison Summary
The proposed and predicate enteral feeding devices are all intended for pediatric patients who require enteral feeding due to illness which prevents normal chewing and swallowing. These products have the same intended use, the same function, and the same general technological characteristics. The majority of the parts for both the proposed and predicate feeding tubes are made with the same material. The variation of the materials for certain components has been proven biocompatible and effective. The major technological enhancement is the incorporation of new ENFit connector which is compliant to ISO 80369-3, in the proposed Feeding Tubes. A male ENFit connector is offered for the proposed feeding tubes. Other additions to the proposed products include additional Fr sizes in both PVC and Polyurethane feeding tubes and an added length to the Polyurethane feeding tubes. These variations do not alter the intended use of the device, nor impact the safety and effectiveness of the product.
- g. Nonclinical testing
- Biocompatibility testing has demonstrated the biological safety of parts of the proposed feeding tubes which directly contact the patient.
- . Stability testing evaluated the properties of the proposed Feeding Tubes after accelerated aging in support of the product expiration date.
- Dimensional analysis was conducted for critical dimensions of the proposed Feeding Tubes, in accordance with ISO 80369-3, Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications. The testing demonstrates the proposed devices conform to the criteria in Table B.2 of ISO 80963-3.
- Study performed on the proposed Feeding Tubes with ENFit Connector included the ● tests listed below, in accordance with ISO 80369-3, Small- bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications, using the test methods provided in ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods. The testing demonstrates the proposed devices conform to the requirements of ISO 80963-3.
{5}------------------------------------------------
| Individual Test Defined<br>in ISO 80369-3 | Requirement Defined<br>in ISO80369-3 | Test Method Defined<br>in ISO 80369-20 |
|---------------------------------------------|--------------------------------------|----------------------------------------|
| Fluid Leakage | Clause 6.2 | Annex C |
| Stress Cracking | Clause 6.3 | Annex E |
| Resistance to separation<br>from axial load | Clause 6.4 | Annex F |
| Resistance to separation<br>from unscrewing | Clause 6.5 | Annex G |
| Resistance to overriding | Clause 6.6 | Annex H |
| Disconnection by<br>unscrewing | Clause 6.7 | Annex I |
- Flow testing was conducted and has demonstrated that the proposed feeding tubes are ● substantially equivalent to the performance of of the predicate device.
- Liquid Leakage ●
- Tensile Strength ●
- The risk associated with the misconnection of the ENFit connector has been assessed.
- ENFit Misconnection Data with Failure Modes and Effects 트 Analysis (FMEA)
- l Enteral Connector Misconnection Assessment
- l Enteral Connection Risk Management Report
- I ENFit Misconnection Risk Assessment Report
- Human Factors Validation Study ●
## h. Clinical testing
Clinical evaluations were not relied upon for the determination of substantial equivalence to the predicate device based on the device classification, sufficient safety and functional performance information provided in the submission.
- i. Conclusions
This information provided within this pre-market notification demonstrates that the Argyle™ Polyvinyl Chloride (PVC) and Kangaroo™ Polyurethane (PU) Neonatal and Pediatric Feeding Tubes with ENFit Connector have no difference that would affect the safety or effectiveness of the devices as compared to the predicate devices and demonstrate substantial equivalency.
## End of Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.