Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis: - Inflammatory degenerative joint disease including rheumatoid arthritis; - Correction of functional deformity; - Revision procedures where other treatments or devices have failed; and - Treatment of fractures that are unmanageable using other techniques. Physica knee system is intended for cemented fixation. The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement. The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to: - non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - inflammatory degenerative joint disease such as rheumatoid arthritis; - treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods: - revision of a failed primary implant; - cuff tear arthropathy (CTA Heads only); - glenoid arthrosis without excessive glenoid bone loss: A1. A2 and B1 according to Walch classification (SMR Hybrid Glenoid only). The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
Device Story
Physica total knee and SMR shoulder systems comprise modular orthopedic implants for joint replacement. This submission introduces an alternative sterilization method (gamma radiation, min 25 kGy) for Physica KR and SMR Reverse Humeral UHMWPE liners. Devices are implanted by orthopedic surgeons in clinical/hospital settings. Implants restore joint function and alleviate pain caused by degenerative disease, fractures, or previous implant failure. Clinical benefit derived from mechanical joint reconstruction. No changes to device design, function, or materials; UHMWPE remains conventional (non-crosslinked).
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including sterilization validation, UHMWPE characterization, and mechanical testing.
Technological Characteristics
Materials: Ti6Al4V (ISO 5832-3, ASTM F1472), CoCrMo (ISO 5832-12, ASTM F1537), UHMWPE (ISO 5834-2, ASTM F648), PoroTi Titanium Coating (ASTM F1580), Tantalum (ISO 13782, ASTM F560). Modular orthopedic prosthesis components. Sterilization: Gamma radiation (min 25 kGy).
Indications for Use
Indicated for skeletally mature patients requiring knee arthroplasty for degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis), functional deformity, revision, or fracture management. Shoulder systems indicated for partial/total/revision replacement in patients with degenerative joint disease, rheumatoid arthritis, acute humeral head fractures, failed primary implants, cuff tear arthropathy, or glenoid arthrosis. Reverse shoulder indicated for rotator cuff deficient joints with severe arthropathy; requires functional deltoid muscle.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 26, 2019
LimaCorporate % David McGurl Director, Regulatory Affairs Musculoskeletal Clinical Regulatory Advisers, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001
Re: K190911
Trade/Device Name: Physica KR Liner and SMR Reverse Humeral Liner Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, HRY, PHX, KWS, MBF, KWT Dated: October 29, 2019 Received: October 29, 2019
Dear David McGurl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Ting Song, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K190911
Device Name
Physica KR Liner and SMR Reverse Humeral Liner
Indications for Use (Describe)
Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or . avascular necrosis:
- Inflammatory degenerative joint disease including rheumatoid arthritis; ●
- Correction of functional deformity; ●
- Revision procedures where other treatments or devices have failed; and
- Treatment of fractures that are unmanageable using other techniques.
Physica knee system is intended for cemented fixation
The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement.
The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:
- non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; ●
- inflammatory degenerative joint disease such as rheumatoid arthritis;
- . treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods:
- revision of a failed primary implant;
- . cuff tear arthropathy (CTA Heads only);
- glenoid arthrosis without excessive glenoid bone loss: A1. A2 and B1 according to Walch . classification (SMR Hybrid Glenoid only).
The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The Hybrid Glenoid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.
The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented use as specified in the following table.
In the Anatomic shoulder the humeral construct consists of the humeral stem, the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral consists of the humeral
{3}------------------------------------------------
stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.
The Anatomic glenoid construct consists of an all polyethylene glenoid, a polyethylene glenoid with metal peg or a metal back assembled with a liner, the Reverse glenoid consists of a metal back/connector/glenosphere construct or of a peg/baseplate/glenosphere construct.
On the glenoid side, the fixation of the all polyethylene glenoid, the polyethylene glenoid with metal peg or the metal back determines if the construct is cemented or uncemented.
| System | | | | Use | |
|--------|----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-----|------------|
| A | R | Components | Material | Cem | Not<br>Cem |
| ● | ● | SMR Stems (Cemented, Cemented Revision) | Ti6Al4V | X | |
| ● | ● | SMR Stems (Cementless Finned, Cementless Revision) | Ti6Al4V | | X |
| ● | | SMR Humeral Bodies (Trauma, Finned) | Ti6Al4V | X | X |
| ● | ● | SMR Reverse Humeral Body | Ti6Al4V | X | X |
| ● | ● | Humeral Extension | Ti6Al4V | X | X |
| ● | | SMR Humeral Heads (Standard*, CTA) | CoCrMo | X | X |
| ● | | SMR Adaptor Tapers (Neutral, Eccentric) | Ti6Al4V | X | X |
| ● | | SMR CTA Head Adaptor for Reverse Humeral Body | Ti6Al4V | X | X |
| | ● | SMR Glenospheres | CoCrMo | | X |
| | ● | SMR Connectors* | Ti6Al4V | | X |
| | ● | Reverse Liners | UHMWPE | X | X |
| ● | | SMR Cemented Glenoids | UHMWPE | X | |
| ● | | SMR 3 Pegs Cemented Glenoids | UHMWPE | X | |
| ● | ●* | SMR Hybrid Glenoid | UHMWPE+Ti6Al4V+Ta | X | X |
| | ● | SMR Hybrid Glenoid Reverse Baseplate + Screw | Ti6Al4V | | X |
| ● | ● | SMR Metal Back Glenoids | Ti6Al4V+PoroTi | X* | X* |
| ● | ● | SMR TT Metal Back Baseplate | Ti6Al4V | X* | X* |
| ● | ● | SMR TT Metal Back Peg | Ti6Al4V | X | X |
| ● | | SMR Metal Back Liner | UHMWPE | X* | X* |
| ●* | ● | SMR Bone screws | Ti6Al4V | | X |
| | | Material Standards | | | |
| | | Ti6Al4V (ISO 5832-3 - ASTM F1472) - CoCrMo (ISO 5832-12 - ASTM F1537) - UHMWPE (ISO 5834-2 - ASTM F648) -<br>PoroTi Titanium Coating (ASTM F1580) - Ta (ISO13782 - ASTM F560) | | | |
## A= Anatomic / R=Reverse
*NOTE:
- In the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR Anatomic ● Shoulder Replacement, is intended for use with bone cement and should be used without bone screws.
- The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR ● Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.
- SMR lateralized connectors are not indicated for use with glenoid bone grafting techniques. ●
- In the US the SMR TT Metal Back Baseplate used as part of the SMR Anatomic Shoulder ● Replacement, is intended for use with bone cement and should be used without bone screws; while when used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.
- If a SMR Hybrid Glenoid is in place and revision to a reverse prosthesis is required, the patient ● can be revised by removing the polyethylene baseplate, leaving the metal peg in place and by connecting it to the SMR Hybrid Glenoid Reverse Baseplate. The SMR Hybrid Glenoid Reverse Baseplate is intended for uncemented use with the addition of screws for fixation.
{4}------------------------------------------------
- The Dia. 50, 52 and 54 mm Humeral Heads with + 3mm increased height cannot be coupled ● to the Long Adaptor Tapers (both concentric and eccentric).
- The Dia. 52 and 54 mm Humeral Heads with + 2mm increased height cannot be coupled to the Long Adaptor Tapers (both concentric and eccentric).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) SUMMARY
| Device Trade Name: | Physica KR Liner and SMR Reverse Humeral Liner |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | LimaCorporate<br>Via Nazionale, 52<br>33038 Villanova di San Daniele del Friuli<br>Udine, Italy |
| Contact: | Mr. Roberto Gabetta<br>Regulatory Manager<br>Phone: +39 338 6439379<br>Email: roberto.gabetta@limacorporate.com |
| Prepared by: | Mr. Dave McGurl<br>Director, Regulatory Affairs<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street NE, Suite 1000<br>Washington, DC 20005<br>Phone: 202.552.5800<br>dmcgurl@mcra.com |
| Date Prepared: | November 26, 2019 |
| Classification: | 21 CFR 888.3560 Knee joint patellofemorotibia polymer/metal/polymer semi-constrained cemented prosthesis.<br>21 CFR 888.3530 Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis.<br>21 CFR 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.<br>21 CFR 888.3670 Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis.<br>21 CFR 888.3650 Shoulder joint metal/polymer non-constrained cemented prosthesis |
| Class: | II |
| Product Code: | JWH, HRY, PHX, KWS, MBF, KWT |
| Predicate Devices: | Physica KR Knee System (K141934)<br>SMR Shoulder System (K110598, K142139, K163397) |
{6}------------------------------------------------
# Indications for Use:
Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:
- Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, . or avascular necrosis:
- Inflammatory degenerative joint disease including rheumatoid arthritis;
- Correction of functional deformity:
- Revision procedures where other treatments or devices have failed; and
- Treatment of fractures that are unmanageable using other techniques. ●
Physica knee system is intended for cemented fixation.
The SMR Shoulder System is intended for partial or total, primary or revision shoulder joint replacement.
The SMR Anatomic Shoulder System is indicated for partial or total, primary or revision shoulder joint replacement in patients suffering from disability due to:
- non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- inflammatory degenerative joint disease such as rheumatoid arthritis;
- treatment of acute fractures of the humeral head that cannot be treated with other fracture ● fixation methods:
- revision of a failed primary implant;
- cuff tear arthropathy (CTA Heads only);
- glenoid arthrosis without excessive glenoid bone loss: A1, A2 and B1 according to . Walch classification (SMR Hybrid Glenoid only).
The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The Hybrid Glenoid Reverse Baseplate must not be used in cases of excessive glenoid bone loss and/or when bone graft is needed.
The Modular SMR Shoulder System allows the assembly of components in various humeral and glenoid constructs. The constructs are intended for cemented and uncemented use as specified in the following table.
In the Anatomic shoulder the humeral construct consists of the humeral body, the adaptor taper and the humeral head. In the Reverse shoulder the humeral construct consists of the humeral stem, the reverse humeral body and the reverse liner. On the humeral side the fixation of the humeral stem determines if the construct is cemented or uncemented.
{7}------------------------------------------------
The Anatomic glenoid construct consists of an all polyethylene glenoid, a polyethylene glenoid with metal peg or a metal back assembled with a liner; the Reverse glenoid consists of a metal back/connector/glenosphere construct or of a peg/baseplate/glenosphere construct.
On the glenoid side, the fixation of the all polyethylene glenoid, the polyethylene glenoid with metal peg or the metal back determines if the construct is cemented or uncemented.
| System | | | | Use | |
|--------|----|----------------------------------------------------|-------------------|-----|------------|
| A | R | Components | Material | Cem | Not<br>Cem |
| • | • | SMR Stems (Cemented, Cemented Revision) | Ti6Al4V | X | |
| • | • | SMR Stems (Cementless Finned, Cementless Revision) | Ti6Al4V | | X |
| • | | SMR Humeral Bodies (Trauma, Finned) | Ti6Al4V | X | X |
| • | • | SMR Reverse Humeral Body | Ti6Al4V | X | X |
| • | • | Humeral Extension | Ti6Al4V | X | X |
| • | | SMR Humeral Heads (Standard*, CTA) | CoCrMo | X | X |
| • | | SMR Adaptor Tapers (Neutral, Eccentric) | Ti6Al4V | X | X |
| • | | SMR CTA Head Adaptor for Reverse Humeral Body | Ti6Al4V | X | X |
| | • | SMR Glenospheres | CoCrMo | | X |
| | • | SMR Connectors* | Ti6Al4V | | X |
| | • | Reverse Liners | UHMWPE | X | X |
| • | | SMR Cemented Glenoids | UHMWPE | X | |
| • | | SMR 3 Pegs Cemented Glenoids | UHMWPE | X | |
| • | •* | SMR Hybrid Glenoid | UHMWPE+Ti6Al4V+Ta | X | X |
| | • | SMR Hybrid Glenoid Reverse Baseplate + Screw | Ti6Al4V | | X |
| • | • | SMR Metal Back Glenoids | Ti6Al4V+PoroTi | X* | X* |
| • | • | SMR TT Metal Back Baseplate | Ti6Al4V | X* | X* |
| • | • | SMR TT Metal Back Peg | Ti6Al4V | X | X |
| • | •* | SMR Metal Back Liner | UHMWPE | X* | X* |
| •* | • | SMR Bone screws | Ti6Al4V | | X |
T6A14V (ISO 5832-3 - ASTM F1472) - CoCrMo (ISO 5832-12 - ASTM F1537) - UHMWPE (ISO 5834-2 - ASTM F648) -PoroTi Titanium Coating (ASTM F1580) - Ta (ISO13782 - ASTM F560)
A= Anatomic / R=Reverse
*NOTE:
- In the US, the SMR Metal Backed Glenoid/Liner construct, used as part of the SMR ● Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws.
- . The SMR Metal Backed Glenoid/Connector/Glenosphere construct, used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.
- SMR lateralized connectors are not indicated for use with glenoid bone grafting techniques.
- In the US the SMR TT Metal Back Baseplate used as part of the SMR Anatomic Shoulder Replacement, is intended for use with bone cement and should be used without bone screws; while when used as part of the SMR Reverse Shoulder replacement, is intended for uncemented use with the addition of screws for fixation.
{8}------------------------------------------------
- If a SMR Hybrid Glenoid is in place and revision to a reverse prosthesis is required, ● the patient can be revised by removing the polyethylene baseplate. leaving the metal peg in place and by connecting it to the SMR Hybrid Glenoid Reverse Baseplate. The SMR Hybrid Glenoid Reverse Baseplate is intended for uncemented use with the addition of screws for fixation.
- . The Dia. 50, 52 and 54 mm Humeral Heads with + 3mm increased height cannot be coupled to the Long Adaptor Tapers (both concentric and eccentric).
- The Dia. 52 and 54 mm Humeral Heads with + 2mm increased height cannot be coupled to the Long Adaptor Tapers (both concentric and eccentric).
## Device Description:
The Physica System is a total knee system and the SMR Reverse is a reverse shoulder system.
The purpose of the Special 510(k) is to add an alternative sterilization method for the Physica KR Liner and SMR Reverse Humeral Liners. Both systems utilize conventional (non-crosslinked) UHMWPE liners (ISO 5834-2 ASTM F648). The liners can be sterilized using a minimum gamma radiation dose of 25 kGy.
#### Substantial Equivalence:
The Physica KR Liner and SMR Reverse Humeral Liner is substantially equivalent to the predicate devices cited on the previous page with respect to indications, design, function, and performance.
#### Preclinical Testing:
The following activities were performed:
- Sterilization validation
- UHMWPE Characterization
- Mechanical Testing ●
## Clinical Testing:
Clinical testing was not necessary to support equivalence.
#### Conclusion:
The Physica KR Liner and SMR Reverse Humeral Liner possesses the same intended use and technological characteristics as the predicate devices. Therefore, the Physica KR Liner and SMR Reverse Humeral Liner is substantially equivalent for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.