K110598 · Lima Corporate S.P.A. · PHX · Aug 18, 2011 · Orthopedic
Device Facts
Record ID
K110598
Device Name
SMR REVERSE SHOULDER SYSTEM
Applicant
Lima Corporate S.P.A.
Product Code
PHX · Orthopedic
Decision Date
Aug 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The components are intended for use in cemented and uncemented applications, as specified in the following table: [Table omitted for brevity, see source]
Device Story
SMR Reverse Shoulder System is a semi-constrained cemented/uncemented shoulder prosthesis; replaces humeral head and glenoid surface in rotator cuff deficient joints. System comprises modular humeral stems, reverse humeral bodies, UHMWPE liners, metal-backed glenoids, glenospheres, and connectors. Humeral bodies couple to stems via Morse taper and screw; glenospheres attach to metal-backed glenoids via connector and screw. Components manufactured from Ti6Al4V, CoCrMo, and UHMWPE. Used by orthopedic surgeons in clinical settings for total shoulder arthroplasty. Provides joint stability and articulation in absence of functional rotator cuff; relies on deltoid muscle for function. Surgeon selects component sizes and thicknesses to ensure proper joint tension and alignment. Benefits include restoration of shoulder function and pain relief in patients with severe arthropathy.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including fatigue testing of humeral stems and modular connections, and wear testing of reverse liners and glenospheres using worst-case constructs.
Technological Characteristics
Materials: Ti6Al4V (ASTM F1472/ISO 5832-3), CoCrMo (ASTM F1537/ISO 5832-12), UHMWPE (ASTM F648/ISO 5834-2). Modular design with Morse taper connections and screw fixation. Metal-backed glenoids feature Ti plasma spray coating. Fixed 150° cervico-diaphyseal angle. Cemented and uncemented fixation options.
Indications for Use
Indicated for primary, fracture, or revision total shoulder replacement in patients with grossly rotator cuff deficient joints and severe arthropathy. Requires anatomically/structurally suitable joint and functional deltoid muscle.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Limacorporate S.p.A Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590-696
Re: K110598
Trade/Device Name: SMR Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS Dated: August 5, 2011 Received: August 12, 2011
Dear Ms. Hastings:
This letter corrects our substantially equivalent letter of August 18, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
### Page 2 -- Ms. Cheryl Hastings
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
(21 Of RT att 603) produces/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): Unknown K | 0598
Device Name: SMR Reverse Shoulder System
Indications for Use:
### SMR Reverse Shoulder System Indications for Use
The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The components are intended for use in cemented and uncemented applications, as specified in the following table:
| COMPONENT | USE | |
|-------------------------|----------|------------|
| | Cemented | Uncemented |
| Cemented stems | X | |
| Cementless Finned stems | | X |
| Reverse Humeral Bodies | X | X |
| Reverse Liners | X | X |
| Glenospheres | | X |
| Connectors | | X |
| Metal Back Glenoid | | X |
| Bone Screws | | X |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Igor M. Melkerson
(Division/Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110598
Page 1 of 1
Traditional 510(k) - SMR Reverse Shoulder System February 18, 2011
Section 4 - Page 15 of 522
{3}------------------------------------------------
K110598
## AUG 1 8 2011
## Summary of Safety and Effectiveness
Date: June 10, 2011
Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy
U.S. Contact Person: Cheryl Hastings Principal Consultant Phone: 574-527-4220
| Product | Product Code | Regulation and Classification Name |
|--------------------------------|--------------|------------------------------------------------------------------------------------------|
| SMR Reverse Shoulder<br>System | KWS | Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis per 21 CFR 888.3660 |
### Description:
The SMR Reverse Shoulder System consists of a humeral stem, a reverse humeral body, a reverse liner, a metal-back glenoid, a glenosphere and a connector with screw. Bone screws are used for the fixation of the metal-back glenoid to the scapula. Humeral stems are provided for both cemented (cleared via 510(k): K100858) and cementless (cleared via 510(k): K101263) fixation as well as reverse humeral bodies (object of this submission). The SMR Reverse Shoulder System metal back glenoids and glenospheres are intended for uncemented press-fit use only with the addition of screws for fixation.
Two designs of humeral stems are available: the first one (cleared via 510(k): K100858) is intended for cemented use only while the second one (cleared via 510(k): K101263) is intended for uncemented use. The stems are provided with a male Morse taper to allow coupling with the reverse humeral bodies.
All reverse humeral bodies for SMR Reverse Shoulder Systems are made from Ti6Al4V (ASTM F1472 - ISO 5832-3). They are intended to be coupled by means of a Morse taper with the humeral stem; further stabilization is achieved with a screw that joins the humeral body to the stem. SMR reverse humeral bodies have a fixed cervico-diaphyseal angle of 150° to provide the correct varus-valgus alignment of the joint. The humeral body has a female taper for coupling with the reverse liner. Two versions of the SMR reverse liners are available: standard (available in Standard and Short sizes) and finned for trauma applications (fins with holes for tuberosities reconstruction provided).
Reverse liners are made from standard ultra-high molecular weight polycthylene (ASTM F648 - ISO 5834-2); they are coupled via the internal tapcr of the reverse humeral bodies. Two designs of reverse liners are available: standard and retentive (characterized
Traditional 510(k) -SMR Reverse Shoulder System June 10, 2011
Page 1 of 4
{4}------------------------------------------------
## KII 0598
by deeper spherical concavity, higher lateral rim to provide a larger articulating surface and greater congruency between the liner and the glenosphere than the standard liners).
Three thicknesses are available: STD, +3mm and +6mm. As reported for standard liners, the choice of the thickness is made by the surgeon to ensure the appropriate tension and prevent laxity in the joint.
Metal-backed glenoids are made from Ti6Al4V (ASTM F1472 - ISO 5832-3). The surface of the metal-backed glenoid which comes in contact with bone is coated with Ti plasma spray coating. These devices are intended to be press-fitted into a hole drilled into the glenoid cavity during surgery; the peg surface has several wings to aid in fixation. Metal backed glenoids are available in four sizes (small-R, Small, Standard and Large).
The plate has curved surfaces to adapt to the spherical shape of the glenoid cavity. The plate is provided with holes to allow the insertion of screws to fix the device in the glenoid cavity: Small-R, Small and Standard glenoids have two holes while the Large size has four holes.
Glenospheres are made from CoCrMo alloy that conforms to ASTM F1537 - ISO 5832-12. Two different designs are available: a standard glenosphere that can be centered with respect to the glenoid component or eccentrical glenosphere to provide offset. Both designs have a spherical shape of 36mm diameter and articulate with the standard UHMWPE liners of the humeral body.
Glenospheres are intended to be coupled to the metal-back glenoid through the use of a connector manufactured of Ti6Al4V (ASTM F1472 - ISO 5832-3). It incorporates a double male taper: one side is connected to the glenosphere while the other is coupled with the glenoid component. To increase the solidity of the system, a screw is used to link the glenosphere to the glenoid component. The screw is inserted through a hole on the surface of the glenosphere, passing through the internal cavity of the connector and then screwed to the metal-backed liner. Two types of connectors are used depending on the metal backed glenoid component: Small-STD and Small-R.
### Intended Use:
ﻣ ..
The SMR Reverse Shoulder System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy.
The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.
The components are intended for use in cemented and uncemented applications, as specified in the following table:
Traditional 510(k) -SMR Reverse Shoulder System June 10, 2011
Page. 2 of 4
{5}------------------------------------------------
# KII 2598
| COMPONENT | USE | |
|-------------------------|----------|------------|
| | Cemented | Uncemented |
| Cemented stems | X | |
| Cementless Finned stems | | X |
| Reverse Humeral Bodies | X | X |
| Reverse Liners | X | X |
| Glenospheres | | X |
| Connectors | | X |
| Metal Back Glenoid | | X |
| Bone screws | | X |
### Predicate Devices:
Promos Reverse Shoulder System (Plus Orthopedics, K081016); Inverse/reverse Shoulder System (Zimmer, K053274); Delta Xtend Reverse Shoulder (DePuy, K091751).
### Comparable Features to Predicate Device(s):
The SMR Reverse Shoulder System is similar to the predicate devices in terms of intended use, indications, design and materials. The SMR Reverse Shoulder System and the predicates are all intended for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy.
The SMR humeral stems are intended for cemented or uncemented use, depending on the design, as are the predicate humeral stems. Reverse bodies are intended for cemented or uncemented use, they follow the indication of the humeral stems. The SMR and predicate glenoid baseplates are all intended for uncemented use only with the use of bone screws for additional fixation.
Like the Promos Modular Shoulder System, the SMR Shoulder System provides modular humeral stem and humeral body components. The Inverse/reverse and Delta Xtend Shoulder Systems are designed as a one-piece humeral stem / body component and have humeral cups to host reverse liners. All the systems are provided with reverse liners and glenospheres of similar diameters that are intended to articulate together. In all the systems the glenospheres are fixed to the glenoid bone through the use of a glenoid baseplate (also called metal back glenoid).
The components of the SMR Shoulder System are manufactured from the same or similar materials as the predicate devices.
Non-Clinical Testing: The SMR Reverse Shoulder System has undergone fatigue testing to demonstrate both the strength of the humeral stem and the post-fatigue strength of the modular connections; and wear testing of the reverse liner and glenospheres. All mechanical testing was done on worst case components or constructs. Where possible, standard test methods were used to allow comparison to testing of the predicate dcvices. The testing results demonstrated the device's ability to perform under expected clinical conditions.
Traditional 510(k) -SMR Reverse Shoulder System June 10, 2011
Page 3 of 4
{6}------------------------------------------------
## 111 2098
:
:
. .
. .
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Reverse Shoulder System to the predicate device(s). . ... ... ... ...
・
. . . . . . .
. .
Traditional 510(k) -SMR Reverse Shoulder System June 10, 2011
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.