K190904 · Conformis, Inc. · LPH · Aug 28, 2019 · Orthopedic
Device Facts
Record ID
K190904
Device Name
Conformis BeneFIT Hip System
Applicant
Conformis, Inc.
Product Code
LPH · Orthopedic
Decision Date
Aug 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ConforMIS BeneFIT Hip System is indicated for use in skeletally mature individuals undergoing total hip replacement due to: - A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, avascular necrosis, or congenital hip dysplasia. - Treatment of non-displaced non-unions of the hip, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement, manageable by other techniques. - Revision procedures for failed previous hip surgery (excluding situations where hardware is present). The ConforMS BeneFIT Hip System implants are intended for cementless fixation using an anterior or posterior surgical approach.
Device Story
BeneFIT Hip System is an uncemented, primary total hip replacement prosthesis. Components include a monoblock femoral stem with a 12/14 taper neck, femoral heads (cobalt chromium or Biolox Delta ceramic), and an acetabular shell with a polyethylene liner and cancellous screws. The system is used by orthopedic surgeons in clinical settings for total hip arthroplasty. The device is implanted via anterior or posterior surgical approaches; fixation is achieved through press-fit design and hydroxyapatite coating on the femoral stem. The acetabular liner features circumferential scallops for rotational security and orientation adjustment. By replacing damaged joint surfaces, the system aims to restore hip function, reduce pain, and improve mobility for patients with degenerative or traumatic hip conditions.
Clinical Evidence
Bench testing only. Performance evaluated via femoral stem/neck fatigue testing, taper junction testing, acetabular liner-cup disassembly and impingement testing, wear testing, bone screw testing, HA coating characterization (microstructure, static/shear fatigue), and range of motion testing. Cadaveric evaluation also performed.
Technological Characteristics
Uncemented, semi-constrained hip prosthesis. Materials: Cobalt chromium alloy, Biolox Delta ceramic, polyethylene, hydroxyapatite coating. Design: Monoblock femoral stem, 12/14 taper, press-fit fixation. Acetabular component includes shell with screw holes and rotational locking liner. Sterilized.
Indications for Use
Indicated for skeletally mature individuals requiring total hip replacement due to osteoarthritis, traumatic/rheumatoid arthritis, avascular necrosis, congenital hip dysplasia, non-displaced non-unions, femoral neck/trochanteric fractures, or revision of failed prior hip surgery (excluding hardware presence).
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Exactech Novation Crown Cup with InteGrip Acetabular Shell (K102795)
Submission Summary (Full Text)
{0}------------------------------------------------
August 28, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ConforMIS, Inc. Emmanuel Nyakako Sr. Vice President, Regulatory and Quality Affairs 600 Technology Park Billerica, Massachusetts 01821
Re: K190904
Trade/Device Name: Conformis BeneFIT Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH, OQG Dated: July 29, 2019 Received: July 30, 2019
Dear Emmanuel Nyakako:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190904
### Device Name ConforMIS, Inc. BeneFIT Hip System
#### Indications for Use (Describe)
The ConforMIS BeneFIT Hip System is indicated for use in skeletally mature individuals undergoing total hip replacement due to:
- · A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, theumatoid arthritis, avascular necrosis, or congenital hip dysplasia.
- · Treatment of non-displaced non-unions of the hip, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement, manageable by other techniques.
- Revision procedures for failed previous hip surgery (excluding situations where hardware is present). . The ConforMS BeneFIT Hip System implants are intended for cementless fixation using an anterior or posterior surgical approach.
Type of Use (Select one or both, as applicable):
| <div> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 6.0 510(K) SUMMARY (PAGE 1 OF 5)
| 510(K) SUMMARY (PAGE 1 OF 5) | | |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Submitter's Name and Address: | ConforMIS, Inc.<br>600 Technology Park<br>Billerica, MA 01821 | |
| Establishment Registration Number: | 3009844603 and 3004153240 | |
| Date of Summary: | July 29, 2019 | |
| Contact Person: | Emmanuel Nyakako, Sr. Vice President, Regulatory and Quality Affairs | |
| Telephone Number: | (781) 345-9164 | |
| Fax Number: | (781) 583-5006 | |
| Name of the Device: | ConforMIS BeneFIT Hip System | |
| Common Name: | Total Hip Replacement System | |
| Regulatory Status and Regulation Number: | Class II<br>21 CFR 888.3358<br>21 CFR 888.3353 | |
| Classification Name: | Uncemented, primary total hip replacement composed of femoral and acetabular components. | |
| Device Classification: | Product Code:<br>LPH: Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented<br><br>LZO: Hip joint metal/ceramic/polymer semi constrained cemented or nonporous uncemented<br><br>MEH: Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calcium phosphate | |
| Indications for<br>Use: | The Conformis BeneFIT Hip System is indicated for use in skeletally<br>mature individuals undergoing total hip replacement due to:<br>• A severely painful and/or disabled joint from osteoarthritis,<br>traumatic arthritis, rheumatoid arthritis, avascular necrosis, or<br>congenital hip dysplasia.<br>• Treatment of non-displaced non-unions of the hip, femoral<br>neck fractures, and trochanteric fractures of the proximal femur with<br>head involvement, unmanageable by other techniques.<br>• Revision procedures for failed previous hip surgery<br>(excluding situations where hardware is present).<br><br>The Conformis BeneFIT Hip System implants are intended for<br>cementless fixation using an anterior or posterior surgical<br>approach. | |
| Identification of<br>the Legally<br>Marketed<br>Devices<br>(Predicate<br>Devices): | Primary Predicate: | |
| | Conformis iTotal Hip System | |
| | Device Class: | II |
| | Product Code: | LPH, JDI, OQG, OQH |
| | Regulation Number: | 21 CFR 888.3358 |
| | 510(k) Number: | K162719 |
| | Secondary Predicates: | |
| | Corail AMT Hip Prosthesis | |
| | Device Class: | II |
| | Product Code: | LZO |
| | Regulation Number: | 21 CFR 888.3353 |
| | 510(k) Number: | K042992 |
| | Exactech Novation Crown Cup with InteGrip Acetabular ShellDevice | |
| | Class: | II |
| | Product Code: | |
| | Regulation Number: | 21 CFR 888.3353 |
| | 510(k) Number: | K102795 |
{4}------------------------------------------------
# 210(K) SUMMARY (PAGE 2 OF হা
{5}------------------------------------------------
### 510(K) SUMMARY (PAGE 3 OF 5)
The Conformis BeneFIT Hip System is an uncemented, primary total Device hip replacement composed of femoral and acetabular components. Description: All components are provided sterile. The femoral component consists of a standard monoblock femoral stem body and neck, which mates with a standard femoral head. The stem and neck are manufactured as one piece and hence are not modular. The proximal neck surface with a 12/14 taper is highly polished and transitions to hydroxyapatite coating in the main stem body and is indicated for uncemented press fit fixation only. The femoral head is designed to connect to the femoral stem neck. All femoral heads are polished and have a 12/14 taper to match the femoral stem. The femoral heads are available in either cobalt chromium alloy (CoCr) or ceramic (Biolox® Delta). The acetabular component consists of a standard size shell in 1mm increments with standard screw hole placement, a mating polyethylene liner, and cancellous screws. The acetabular component is designed for uncemented use; initial implant fixation is achieved through press-fit design. The 6.5mm diameter cancellous screws with low profile head fits through the acetabular shell screw holes and is driven using a 3.5mm hex drive recess. The acetabular component has matching circumferential scallops on the shell and liner that rotationally secure the liner in the shell and allow for dialing the liner in a desired orientation. The liner is provided with an impactor made of a biocompatible nylon material.
{6}------------------------------------------------
## 510(K) SUMMARY (PAGE 4 OF 5)
The Conformis BeneFIT Hip System, subject of this premarket Substantial notification is substantially equivalent to the Conformis iTotal Hip Equivalence: Replacement System (K162719, cleared June 14, 2017), the Corail AMT Hip Prosthesis (K042992, cleared February 11, 2005), and the Exactech Exactech Novation Crown Cup with InteGrip Acetabular ShellDevice (K102795, cleared February 2, 2011). Non-clinical testing was conducted in accordance with applicable FDA guidance documents to confirm that the Conformis BeneFIT Hip System is substantially equivalent to the predicate hip systems.
{7}------------------------------------------------
## 510(K) SUMMARY (PAGE 5 OF 5)
| 510(K) SUMMARY (PAGE 5 OF 5)<br>Description and<br>Assessment of<br>Nonclinical<br>Testing: | Nonclinical Testing: The determination of substantial equivalence for<br>this device was based on a detailed device description and non-clinical<br>laboratory testing. Testing on the Conformis BeneFIT Hip System<br>included functional testing in compliance with the applicable FDA<br>guidance documents.<br><br>Specifically, the following tests were conducted to evaluate the<br>performance of the device:<br>Femoral Stem Fatigue Testing Femoral Neck Fatigue Testing Femoral Taper-CoCr head Junction Testing Femoral Taper-Ceramic Head Junction Testing Acetabular Liner-Cup Disassembly Testing Acetabular Liner Impingement Testing Wear Testing Acetabular Bone Screw Testing HA Coating Microstructure Characterization HA Coating Static Fatigue Testing HA Coating Shear Fatigue Range of Motion Testing Chemical and Mechanical Characterization of iPoly XE Cadaveric Evaluation Test results demonstrated that the device can be considered substantially equivalent to the predicate devices for the specified intended use. |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | Based on the testing conducted, it is concluded that the Conformis<br>BeneFIT Hip System is substantially equivalent to the predicate<br>devices: Conformis iTotal Hip Replacement System (K162719, cleared<br>June 14, 2017), the Corail AMT Hip Prosthesis (K042992, cleared<br>February 11, 2005), and the Exactech Exactech Novation Crown Cup<br>with InteGrip Acetabular ShellDevice (K102795, cleared February 2,<br>2011) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.