K162719 · Conformis, Inc. · LPH · Jun 14, 2017 · Orthopedic
Device Facts
Record ID
K162719
Device Name
iTotal® Hip Replacement System
Applicant
Conformis, Inc.
Product Code
LPH · Orthopedic
Decision Date
Jun 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The iTotal® Hip Replacement System is designed from a patient's pre-operative CT scan which must include certain necessary anatomic landmarks that are clearly identifiable. Total hip replacement using the iTotal® Hip Replacement System is indicated for use in skeletally mature individuals undergoing total hip replacement due to: · A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis. or congenital hip dysplasia. · Treatment of non-displaced non-unions of the hip, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. · Revision procedures for failed previous hip surgery (excluding situations where hardware is present). The iTotal® Hip Replacement System includes standard hip replacement components as well as the following patient specific components: femoral neck, acetabular cup, single use instrumentation. The iTotal® Hip Replacement implants are intended for cementless fixation using an anterior or posterior surgical approach.
Device Story
Patient-specific hip replacement system; utilizes pre-operative CT scans to design custom femoral neck and acetabular components; includes standard femoral stems, femoral heads, and polyethylene liners. System includes patient-specific instrumentation (iJigs) for intra-operative positioning and bone cutting/reaming. Used by orthopedic surgeons in clinical settings for total hip arthroplasty. Design process involves proprietary software to translate patient anatomy into custom implant geometry. Benefits include patient-specific fit for cementless fixation. Output is a set of implants and guides tailored to patient anatomy, facilitating surgical execution and potentially improving joint function.
Clinical Evidence
Bench testing only. Testing included femoral stem/neck fatigue, taper junction testing, acetabular liner-cup disassembly, impingement, wear testing (adhesive/abrasive), HA coating characterization (microstructure, bonding, shear fatigue), range of motion, iJig drop/resection simulation, iPoly XE characterization, and cadaveric evaluation. Software verification and validation were also performed.
Technological Characteristics
Materials: Titanium alloy, Cobalt Chromium alloy, Vitamin E infused highly cross-linked UHMWPE (iPoly® XE), hydroxyapatite (HA) coating. Design: Semi-constrained, cementless, patient-specific femoral neck and acetabular cup. Sterilization: Sterile (SAL 10^-6), single-use. Connectivity: Proprietary software for design based on CT imaging.
Indications for Use
Indicated for skeletally mature individuals requiring total hip replacement due to osteoarthritis, traumatic/rheumatoid arthritis, avascular necrosis, congenital hip dysplasia, non-displaced non-unions, femoral neck fractures, trochanteric fractures with head involvement, or revision of failed previous hip surgery (excluding hardware presence).
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
iTotal® CR & PS Knee Replacement Systems (K160025 & K161668)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2017
ConforMIS, Inc. Emmanuel Nyakako Sr. Vice President, Regulatory And Quality Affairs 600 Technology Park Drive Billerica, MA 01821
Re: K162719 Trade/Device Name: iTotal® Hip Replacement System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH, OQG Dated: May 15, 2017 Received: May 16, 2017
Dear Mr. Nyakako:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K162719
Device Name iTotal® Hip Replacement System
#### Indications for Use (Describe)
The iTotal® Hip Replacement System is designed from a patient's pre-operative CT scan which must include certain necessary anatomic landmarks that are clearly identifiable. Total hip replacement using the iTotal® Hip Replacement System is indicated for use in skeletally mature individuals undergoing total hip replacement due to:
· A severely painful and/or disabled joint from osteoarthritis, theumatoid arthritis, avascular necrosis. or congenital hip dysplasia.
· Treatment of non-displaced non-unions of the hip, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
· Revision procedures for failed previous hip surgery (excluding situations where hardware is present).
The iTotal® Hip Replacement System includes standard hip replacement components as well as the following patient specific components: femoral neck, acetabular cup, single use instrumentation.
The iTotal® Hip Replacement implants are intended for cementless fixation using an anterior or posterior surgical approach.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 6.0 510(K) SUMMARY (PAGE 1 OF 5)
| Submitter's Name<br>and Address: | ConforMIS, Inc.<br>600 Technology Park Drive<br>Billerica, MA 01821 |
|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration<br>Number: | 3009844603 and 3004153240 |
| Date of Summary: | June 9, 2017 |
| Contact Person: | Emmanuel O. Nyakako, Sr. Vice President, Regulatory and<br>Quality Affairs |
| Telephone Number:<br>Fax Number: | (781) 345-9164<br>(781) 345-0147 |
| Name of the Device: | iTotal® Hip Replacement System |
| Common Name: | Total Hip Replacement System |
| Regulatory Status<br>and Regulation<br>Number:<br>Classification<br>Name: | Class II<br>21 CFR 888.3358<br>21 CFR 888.3353<br>Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis and<br>Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis. |
| Device<br>Classification: | Product Code:<br>LPH: Prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented |
| | LZO: Hip joint metal/ceramic/polymer semi constrained cemented<br>or nonporous uncemented |
| | MEH: Prosthesis, hip, semi-constrained, uncemented,<br>metal/polymer, non-porous, calcium phosphate |
| | OQG: Hip Prosthesis, semi-constrained, cemented,<br>metal/polymer, + additive, porous, uncemented |
{4}------------------------------------------------
#### 510(K) SUMMARY (PAGE 2 OF 5)
| Indications for<br>Use: | The iTotal® Hip Replacement System is designed from a patient's<br>pre-operative CT scan which must include certain necessary<br>anatomic landmarks that are clearly identifiable. Total hip<br>replacement using the iTotal® Hip Replacement System is indicated<br>for use in skeletally mature individuals undergoing total hip<br>replacement due to:<br>• A severely painful and/or disabled joint from osteoarthritis,<br>traumatic arthritis, rheumatoid arthritis, avascular necrosis,<br>or congenital hip dysplasia.<br>• Treatment of non-displaced non-unions of the hip, femoral<br>neck fractures, and trochanteric fractures of the proximal<br>femur with head involvement, unmanageable by other<br>techniques.<br>• Revision procedures for failed previous hip surgery<br>(excluding situations where hardware is present). | |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| | The iTotal® Hip Replacement System includes standard hip<br>replacement components as well as the following patient specific<br>components: femoral neck, acetabular cup, single use<br>instrumentation. | |
| | The iTotal® Hip Replacement implants are intended for<br>cementless fixation using an anterior or posterior surgical<br>approach. | |
| Identification of the<br>Legally Marketed<br>Devices (Predicate<br>Devices): | Pipeline Total Hip System<br>Device Class:<br>Product Code:<br>Regulation Number:<br>510(k) Number: | II<br>LPH, JDI, OQG, OQH<br>21 CFR 888.3358<br>K112802 |
| | Corail AMT Hip Prosthesis<br>Device Class:<br>Product Code:<br>Regulation Number:<br>510(k) Number: | II<br>LZO<br>21 CFR 888.3353<br>K042992 |
| | Signature Planner, Signature Guides<br>Device Class:<br>Product Code:<br>Regulation Number:<br>510(k) Number: | II<br>LPH, LZO, KWZ, JDI, KWY, MAY MEH<br>21 CFR 888.3358<br>K111863 |
| Identification of the<br>Legally Marketed<br>Devices (Reference<br>Devices): | iTotal® CR & PS Knee Replacement Systems<br>Device Class:<br>Product Code:<br>Regulation Number:<br>510(k) Number: | II<br>JWH, OOG, OIY<br>21 CFR 888.3560<br>K160025 & K161668 |
{5}------------------------------------------------
#### 510(K) SUMMARY (PAGE 3 OF 5)
| 510(K) SUMMARY (PAGE 3 OF 5) | |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description: | The iTotal® Hip Replacement System is a patient specific hip<br>replacement system which consists of femoral and acetabular<br>components and patient specific instrumentation (iJigs). The<br>femoral component consists of a standard femoral stem body with<br>an integrated (non-modular) patient specific neck, which connects<br>with a standard femoral head. The acetabular component<br>consists of a metal acetabular cup with two screw holes and<br>polyethylene liners in standard sizes. Standard bone screws and<br>apex hole plug may also be provided with the iTotal® Hip<br>Replacement System. |
| | The iTotal® Hip Replacement System is intended to treat<br>skeletally mature patients who are candidates for total hip<br>replacement surgery. |
| | Using patient imaging (CT scan) and a combination of proprietary<br>and off the shelf software, a patient specific hip replacement<br>system is designed. The iTotal® Hip Replacement System<br>consists of the following components:<br>a) Femoral Component:<br>The femoral stem, with an integrated patient specific<br>neck, is manufactured from titanium alloy and has a<br>plasma sprayed hydroxyapatite (HA) coating. The<br>femoral stem is available in various sizes. Both metal (Cobalt Chromium alloy) and Ceramic femoral<br>heads are available for use with the iTotal® Hip<br>Replacement System. The femoral heads are available in<br>various sizes and offsets. |
| | b) Acetabular Component:<br>The acetabular cup is manufactured from titanium alloy.<br>The acetabular cup features a plasma sprayed outer<br>surface. The acetabular cups are available in various<br>sizes. The acetabular liner is manufactured from highly cross-<br>linked Vitamin E infused ultra-high molecular weight<br>polyethylene (iPoly® XE). The liners are available in a<br>range of sizes with varying internal diameters and offsets. Acetabular screws and the apex hole plug are<br>manufactured from titanium alloy. |
| | c) Ancillary orthopedic manual surgical instruments are provided<br>with the iTotal® Hip Replacement system to assist with<br>implantation. The ancillary instruments are provided sterile<br>and for single-use only. These patient specific instruments<br>are provided to assist in the positioning of total hip<br>replacement components intra-operatively and in guiding the<br>cutting/reaming of bone. |
{6}------------------------------------------------
## 510(K) SUMMARY (PAGE 4 OF 5)
| Summary of<br>Technological<br>Characteristics: | The rationale for substantial equivalence is based on consideration<br>of the following device use and characteristics: | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • | Intended Use: Similar to the predicate devices, the<br>proposed iTotal® Hip Replacement System is intended to<br>be as a total hip prosthesis. |
| | • | Indications for Use: The proposed indications for use for the<br>iTotal® Hip Replacement System are similar to the<br>predicate devices as they have the same intended use. |
| | • | Operating Principle/Fundamental Technology: Similar to the<br>predicate devices the proposed iTotal® Hip Replacement<br>System is a semi-constrained, cementless artificial hip<br>replacement system; it consists of a femoral stem with an<br>integrated neck and a standard femoral head, an<br>acetabular cup with polyethylene liner, with optional<br>acetabular screws/apex holes plug, and ancillary<br>instrumentation. |
| | • | Materials/Coatings: The proposed iTotal® Hip Replacement<br>system uses the same biocompatible materials (i.e. titanium<br>alloy) and coatings (i.e. hydroxyapatite) as the predicate<br>devices |
| | • | Sterilization: Similar to the predicates, the proposed iTotal®<br>Hip is intended to be provided sterile (SAL 1.0x10-6) and for<br>single use. |
{7}------------------------------------------------
### 510(K) SUMMARY (PAGE 5 OF 5)
| Substantial<br>Equivalence: | The iTotal® Hip Replacement System, subject of this premarket<br>notification is substantially equivalent to the Pipeline Total Hip<br>System (K112802, cleared March 09, 2012), the Corail AMT Hip<br>Prosthesis (K042992, cleared February 11, 2005), and the<br>Signature Planner and Signature Guides (K111863, cleared June<br>11, 2012). Non-clinical testing was conducted in accordance with<br>applicable FDA guidance documents to confirm that the iTotal® Hip<br>Replacement System is substantially equivalent to the predicate hip<br>systems.<br><br>Specifically, the following testing was performed to establish<br>substantial equivalence:<br>Femoral Stem Fatigue Testing Femoral Neck Fatigue Testing Femoral Taper-CoCr head Junction Testing Femoral Taper-Ceramic Head Junction Testing Acetabular Liner-Cup Disassembly Testing Acetabular Liner Impingement Testing Wear Testing: Adhesive and Abrasive Wear Acetabular Bone Screw Testing HA Coating Microstructure Characterization HA Coating Bonding Strength (Tensile and Static Fatigue) Testing HA Coating Shear Fatigue Testing Range of Motion Testing iJig Drop Testing iJig Femoral Neck Resection Simulation Characterization of iPoly XE Cadaveric Evaluation Software Verification/Validation Bacterial Endotoxin Testing All testing has demonstrated that the device is substantially<br>equivalent to the predicate devices. |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | Based on the testing conducted, it is concluded that the iTotal® Hip<br>Replacement System is substantially equivalent to the predicate<br>devices: The Pipeline Total Hip System (K112802, cleared March<br>09, 2012), Corail AMT Hip Prosthesis (K042992, cleared February<br>11, 2005), and Signature Planner, Signature Guides (K111863,<br>cleared June 15, 2012). |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.