ARTIS icono

K190768 · Siemens Medical Solutions USA, Inc. · OWB · Sep 12, 2019 · Radiology

Device Facts

Record IDK190768
Device NameARTIS icono
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeOWB · Radiology
Decision DateSep 12, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Angiographic image reconstruction———————
Vessel roadmap visualization———————
Automatic exposure control———————
Stent visualization———————

Indications for Use

ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis. The ARTIS family include also the software option DynaCT with following indications for use: DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up.

Device Story

Modular angiography system (floor or biplane) for visualization of vessels and internal structures; utilizes multi-axis motorized C-arm, X-ray tube, flat panel detector, and high-voltage generator. Operated by physicians in interventional suites or ORs via table-side control modules and footswitches. Inputs include X-ray projections; system performs 2D/3D image reconstruction (DynaCT), roadmap generation, and automated exposure control (AEC). Outputs displayed on ceiling-mounted monitors. Enhances clinical decision-making by providing real-time navigation, 3D soft-tissue visualization, and standardized procedure workflows (Case Flow). Benefits include reduced radiation dose, faster door-to-groin times in stroke settings, and improved ergonomic workflow for interventionalists.

Clinical Evidence

Bench testing only. Performance assessed via verification and validation testing, including electrical safety (AAMI ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and software validation (IEC 62304). Human factors/usability validation conducted via clinical use tests and customer feedback.

Technological Characteristics

Modular angiography system; multi-axis C-arm; flat panel detectors (as21HDR, as40HDR); high-voltage generator; ceiling-mounted displays. Connectivity via remote service connection. Software-based image processing (DynaCT, Roadmap, AEC). Antimicrobial coating. Sterilization not applicable (non-invasive).

Indications for Use

Indicated for single and biplane diagnostic imaging and interventional procedures, including pediatric and obese patients. Procedures include cardiac, neuro, general, rotational, and multipurpose angiography, and whole-body radiographic/fluoroscopic procedures. Supports image-guided surgery, image fusion, navigation, and 3D reconstruction (DynaCT) for hard/soft tissue imaging.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below. Siemens Medical Solutions USA, Inc. % Ms. Patricia D. Jones Sr. Regulatory Affairs Specialist 40 Liberty Boulevard, 65-1A MALVERN PA 19355 Re: K190768 Trade/Device Name: ARTIS icono (VE2) Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, IZI, JAA, JAK Dated: August 9, 2019 Received: August 13, 2019 Dear Ms. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see September 12, 2019 {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K190768 Device Name ARTIS icono Indications for Use (Describe) ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/filloroscopic procedures as well as procedures next to the table for i e. patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-1ay, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis. The ARTIS family include also the software option DynaCT with following indications for use: DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 F PSC Publishing Services (301) 443-6740 Page 52 of 304 Siemens Medical Solution, USA, Inc. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots. # 510(k) Summary: ARTIS icono (VE2) Siemens Medical Solutions USA, Inc. Company: 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Date Prepared: March 22, 2019 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. #### 1. General Information: Importer / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Reqistration Number: 2240869 ### Manufacturing Site: Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335 #### Contact Person: 2. Ms. Patricia D. Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Email: patricia.d.jones@siemens-Healthineers.com - 3. Device Name and Classification: - Trade Name: Classification Name: Classification Panel: Regulation Number: Device Class: Product Codes: ARTIS icono (VE2) Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II OWB, IZI, JAA, JAK - 4. Legally Marketed Primary Predicate Device Trade Name: 510(k) Clearance Clearance Date Classification Name: Classification Panel: Regulation Number: Device Class: Artis zee/zeego & Artis Q/Q.zen (Software VD11D) K181407 August 15, 2018 Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logos for Siemens and Healthineers. The Siemens logo is in teal, and the Healthineers logo is in orange. There are also orange dots to the right of the Healthineers logo. Product Code: Subsequent Product Codes: Total Product Life Cycle: OWB IZI, JAA, JAK All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues. # Legally Marketed Secondary Predicate Device Trade Name: 510(k) Clearance Clearance Date Classification Name: Classification Panel: Requlation Number: Device Class: Product Codes: Subsequent Product Codes: Total Product Life Cycle: Artis one K133580 April 28, 2014 Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II OWB, IZI. JAA All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues. # Legally Marketed Secondary Predicate Device | Trade Name: | ARTIS pheno | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Clearance | K163286 | | Clearance Date | March 09, 2017 | | Classification Name: | Image-intensified fluoroscopic x-ray System | | Classification Panel: | Radiology | | Regulation Number: | 21 CFR §892.1650 | | Device Class: | Class II | | Product Codes: | OWB, | | Subsequent Product Codes: | JAA | | Total Product Life Cycle: | All product Recall incidents are considered<br>during the Design Input phase of development<br>to ensure the latest models will not be affected<br>by any of the applicable issues. | #### 5. Device Description: Siemens is introducing a revolutionary new family of angiography sytems, the ARTIS icono (VE2) a new modular angiography system designed to help healthcare professionals in today's stroke centers, deal with a double challenge: to treat more patients, and to treat them faster. This is literally expanding percision medicine to advance therapy outcomes. The new ARTIS icono (VE2) system is a medical device that allows visualization of vessels within the human body. It is of the utmost importance to find the right projections so physician can navigate catheters and other devices safely. The ARTIS icono (VE2) system consist of a patient table and a multi-axis motorized c- {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots in a circular pattern. arm that can be positioned around the patient and angulated in a double-oblique fashion iso-centering the region of interest between the x-ray tube and the flat panel detector. The x-ray generator is placed separately. The displays for visualizing the x-ray images are mounted at the ceiling with a movable display suspension system. System operation is executed via control modules table side so that the physician can move and position the table and c-arm adequately for best imaging while manipulate the catheters or other devices during x-ray. X-ray release is table side via a footswitch. The ARTIS icono (VE2), modular angiography systems are designed as sets of components that may be combined into two different configurations (Biplane or Floor) to provide specialized angiography systems. In general they are equipped with C-arm, stand, flat panel detector, x-ray tube, collimator, high voltage generator, patient table, and image post processing. The ARTIS icono (VE2) covers the complete range of angiographic applications, cardiac angiography, neuro-angiography, general angiography, surgery and surgical angiography, and radiographic/fluoroscopic procedures. The following components are configured to create a Floor or Biplane configuration: - (1) Floor stand with C-arm, X-ray tube assembly and FD - (2) Patient table - (3) Display ceiling suspension with displays - (4) Footswitch for releasing radiation - (5) Control console for controlling the stand, patient table and imaging system Images and operating elements are displayed on screens. Depending on the ARTIS icono (VE2) system configuration, different display variants are used to visualize image and information content. Displays that visualize single images or large displays that are configurable to visualize multiple images and information content in variaous layouts are used. Post processing can be done in the exam room or in the control room that offers monitors as well; with a footswitch location in the exam room or the control room. The ARTIS icono (VE2) System is capable of 2D and 3D imaging. The c-arms can be mounted on the floor or for biplane systems on the floor and on the ceiling. Other systems and software synqo Application Software, syngo X Workplace, Sensis, and or third-party systems may also be integrated into the ARTIS icono (VE2) screen configuration. Different screen configurations and layouts are possible in the examination room and in the control room. The Subject device ARTIS icono with software version VE2 will support the following categories of modifications. These modifications were made to the Subject Device in comparison to the Predicate Devices: # 1) New/Modified Software & IFU: {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern. - Table 1: Overview of Software Modifications supported by software ● version VE2 | Table 1. New Software Modifications for ARTIS icono (VE2) System | | | |-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--| | Software changes specific to New System Software VE2 (VE20) | | | | Device Software Modification & Applicable Configuration | | | | 1 | Modified Indications For Use Statement (Floor & Biplane) | | | 2 | New System Software VE2 (also known as VE20), software modification/features. | | | 3 | Improved Roadmap (Also known as OPTIQ) (Floor & Biplane) | | | | A. Increased image quality dose ratio and faster workflow via new Architecture and<br>Organ Program parametrization due to improved algorithms | | | | B. Improved Image Quality due to improved algorithim | | | | C. Subtracted fluoro mode: Dose, Time, and Contrast Agent Savings. | | | | D. Automap Integration in DSA Roadmap Workflow | | | 4 | Improved Automatic Exposure Control: (Floor & Biplane) | | | | A. New Automatic Exposure Control (AEC) incl. Structure Scout) | | | | B. CNR Driven Exposure Control (AEC) | | | 5 | 3D Imaging: (Floor & Biplane) | | | | A. syngo Dyna3D HighSpeed | | | | B. syngo Dyna3D Sine Spin | | | | C. TwinSpin (3D with Frontal plane while lateral plane idle (no x-ray) | | | | D. syngo Dyna3D Multiphase | | | 6 | Improved ClearStent Live: (Floor & Biplane) | | | 7 | Updated User Interface: (Floor & Biplane) | | | | A. Case Flow | | | | B. Favorites in control room and examination room | | | 8 | Lateral Plane Switch (Biplane) | | | 9 | Generic Interface for 3rd parties for data transference (Floor & Biplane) | | # 2) New/Modified Hardware: - . Table 2: Overview of Hardware Modifications supported by software Version VE2 | Device Hardware Modifications | | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <b>10</b> | New Multiaxis floor stand (This hardware components enables the following features:<br>High Speed, syngoDyna3D & syngo DynaCT Sine Spin ( <i>Floor &amp; Biplane</i> ) | | <b>11</b> | Agile lateral plane ( <i>Biplane</i> ) Increased angulation (135->208) Improved repositioning precision Laser cross on Lateral Plane Cabling: energy chains in lateral plane | | <b>12</b> | Tables: ( <i>Floor &amp; Biplane</i> ) New Standard Table (Basic Table) New Siemens multi-tilt table (OR Table) (similar to secondary predicate ARTIS pheno) | | <b>13</b> | New Flat Detectors ( <i>Floor &amp; Biplane</i> ) Introduction of as21HDR Flat Detector aircooled (Trixell 2121CV) Introduction of as40HDR Flat Detector aircooled (Trixell 3040F) | | <b>14</b> | New Collimator ( <i>Floor &amp; Biplane</i> ) New Collimator (LFD) rotates within sealed housing (same as secondary predicate | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. | | ARTIS pheno) | |----|-----------------------------------------------------------------------------------| | | B. New Collimator (MFD) rotates within sealed housing (same as secondary | | | predicate Artis one) | | 15 | Universal Adapter for Megalix X-ray Tube (Floor & Biplane) | | 16 | Anti-microbial Coating (Floor & Biplane) | | 17 | New Control Modules (Floor & Biplane) | | 18 | New Optional Touch Control Displays (Floor & Biplane) | | 19 | New Variations of Display Ceiling Mounts (Floor & Biplane) | | | A. DCS-2x32" rail mounted (in rails with stand) | | | B. Large Display DCS with additional (Artis) display and ACUSON Freestyle display | | | C. Optional rail (movable) mounted DCS on own rails for two 32" display | | 20 | New 3rd Party Accessory Heatable Mattress (Floor & Biplane) | | 21 | Product Claims for the ARTIS icono (VE2) (Floor Biplane) | # 3) List of Product Claims: - Table 3: Product Claims # Table 3: Product Claims | Claim # | Feature / Component | Labeling Claim | |------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Lateral Plane Switch | Preferred cardiology and neuroradiology lateral plane setup can be reached by an automated drive in less than 90 sec. | | 2 | Lateral Plane Switch | Reduced scattered radiation dose to the user by positioning the tube opposite to the user | | 3 | Lateral Plane Switch | The new biplane is engineered to accomodate two specialties (cardiology and neuroradiolgy) in one room instead of two rooms | | 4 | Multi-axis floor stand | System supports the possibility to visualize anatomy with extreme CAUD angulations (up to 49°) with floor stand | | 5 | Agile lateral plane | System supports the possibility to visualize anatomy with extreme CAUD angulations without mechanical limitations, except collision with the patient. | | 6 | Reliability | Smart components increase uptime due to higher reliability. The Hepton multiaxis floorstand, the agile lateral plane and the table come with new smart Siemens industry motor gears and controllers. These components are significantly more reliable (compared to previous system) and can report their status via a remote service connection. 60% longer time between hardware component exchanges. | | 7 | Multiaxis floor stand | Fully motorized stand positioning, no manual interaction necessary like in previous version (MULTISPACE.F). | | 8 | Multiaxis floor stand | 7% faster movement from head to left side/right side (13 sec from head to left side). Fully motorized stand positioning supports the workflow in sterile environment. | | 9 | Multiaxis floor stand | Full Patient coverage of 2.10 m without repositioning the patient. | | 10 | syngo Dyna3D High Speed | 122% faster c-arm angulation for 3D, 100 degrees per sec instead of 45 degrees.*<br>* Compared to Artis zee and Artis Q | | 11 | Roadmap: increased image quality due to improved algorithms | Roadmap creation time reduced by 50% | | 12 | Roadmap: increased image quality due to improved algorithms | Easy workflow with directly accessible Dynamic Progress via pilot module. | | Claim # | Feature / Component | Labeling Claim | | 13 | Roadmap: increased image quality due to improved algorithms | Fading out the vessel map during fluoro break by the touch of a button leads to visualization of the device. | | 14 | Agile lateral plane | Full body coverage (210 cm versus previously 112 cm with frontal plane) with the lateral plane for 2D imaging. | | 15 | Agile lateral plane | 2D Imaging with the lateral plane during Prostate Artery Embolization and Uterine Fibroid Embolization procedures is possible with the biplane system. | | 16 | syngo DynaCT multiphase | With syngo DynaCT Multiphase it is for the first time possible to assess the collateral status with time resolved DynaCT, depicting 8 different time points within a period of 50 seconds. | | 17 | syngo DynaCT Sine Spin | syngo DynaCT Sine Spin helps interventionalists visualize bleedings in the interventional suite with a soft tissue resolution of 5 HU (@ 10 mm), respectively 10 HU (@ 5 mm), even in the basal part of the brain and close to the skull. | | 18 | syngo DynaCT Sine Spin | syngo DynaCT Sine Spin is reducing cone beam CT artifacts in the basal part of the brain and close to the skull. | | 19 | syngo DynaCT Sine Spin | Raising the bar in consistent 3D image quality for whole brain imaging from cranium to basal. | | 20 | syngo DynaCT Sine Spin | In a stroke emergency setting the door-to-groin time can be shortened by up to 60 minutes by using syngo DynaCT and the angio only approach. | | 21 | Favorites in Toolbars in control room | Image-type specific controls allow for 66% less clicks<br>1 click instead of 3 to get to the functionality you want Optimize/ Organize your workflow | | 22 | Case Flow | Save 1 min, 64% shorter time (33 sec instead of 90 sec) during a Femoral Access Case Flow while keeping the focus on your patient. | | 23 | Case Flow | Save 15 clicks (2 instead of 17) during a Femoral Access Case Flow while keeping the focus on your patient. | | 24 | Case Flow | Case Flows allow reproducible system settings over labs and thereby can lead to standardization over labs. | | 25 | Case Flow | With Case Flows the system can be positioned with zero joystick interaction using the pilot module. | | 26 | Case Flow | Case Flows allow for standardized procedure execution with potential to reduce imaging variations. | | 27 | Case Flow | Case Flows support new team members and rotating staff to get faster up to speed. | | 28 | Shorter Control Module (Exam Room) | Shorter control modules/less buttons (19% / 14 cm shorter), lead to more ergonomic system operation. | | 29 | Shorter Control Module (Exam Room) | The control modules are within easier/shorter reach of the user. | | 30 | Shorter Control Module (Exam Room) | 14 cm shorter control modules better fit on lower radiation protection. | | 31 | Shorter Control Module (Exam Room) | User stands 6 cm closer to the patient due to new table design. More ergonomic comfort leads to better working environment. | | 32 | syngo DynaCT Sine Spin | Engineered for optimized stroke treatment | | 33 | syngo DynaCT Sine Spin | Optimized DynaCT scan trajectory for an artefact reduced grey value resolution in the entire volume | | 34 | Agile lateral plane | Engineered for better utilization between multiple / different modalities (combo use) | | 35 | TwinSpin | Designed to seamlessly integrate 2D and 3D workflows | | 36 | Lateral Plane Switch | Biplane system can switch via motorized movements between settings preferred in cardiology, radiology and neuroradiology. | | Claim<br># | Feature /<br>Component | Labeling Claim | | | | setup. | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. ## 6. Indications for Use: ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image quided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis. The ARTIS family include also the software option DynaCT with following indications for use: DynaCT is an X-ray imaging software option, which allows the reconstruction of two-dimensional images acquired with a standard angiographic C-arm device into a three-dimensional image format. DynaCT is intended for imaging both hard and soft tissues as well as other internal body structures for diagnosis, surgical planning, interventional procedures and treatment follow-up. #### 7. Substantial Equivalence: The ARTIS icono (VE2) System is substantial equivalent to the legally marketed (primary) predicate listed in the table below: | Predicate Device Name<br>and Manufacturer | 510(k) Number | Clearance Date | Comparable Properties | |--------------------------------------------------|---------------|----------------|--------------------------------------------------------------------------------------------------| | <b>Primary Predicate</b><br>Artis Q/Q.zen (VD11) | K181407* | 08/15/2018 | • Indications for use<br>• Detector 3040CV<br>• Software Version VD11D<br>• Cabling Energy Chain | | <b>Secondary Predicates</b><br>Artis one | K133580 | 04/28/2014 | • Floor Stand with Swivel base<br>• CLEARstent Live<br>• Artis basic table | Table 4: Predicate Comparable Properties {10}------------------------------------------------ Image /page/10/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. | Predicate Device Name<br>and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties | |--------------------------------------------------------------------------------------------------------------|------------------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | ARTIS pheno | K163286 | 03/09/2017 | • AEC Dose regulation including structure scout<br>• Gigalix tube<br>• Antimicrobial coating<br>• Collimator rotates in sealed housing<br>• OR Tables<br>• Ergonomic control modules | | Artis zee/zeego (VD11) | K181407* | 08/15/2018 | • Megalix tube<br>• Detector 3040CV<br>• Cabling Energy Chain | | * K181407 was a bundled submission and applicable to both Artis Q/Q.zen (VD11) and Artis<br>zee/zeego (VD11) | | | | #### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device: The ARTIS icono (VE2) System is designed as a set of components (C-arm, X-ray tube and housing, flat panel detector, digital imaging system, collimator, generator etc.) that may be combined into two different configurations (Floor & Biplane to provide specialized angiography systems. Components used with ARTIS icono (VE2) System are either commercially available with current Siemens systems or include modifications to existing components. Technological differences between the Subject Device and the Predicate Deivce is provided in the Table 5 below for all modifications. | Modification | | Subject Device<br>ARTIS icono (VE2) | Primary Predicate<br>Device<br>Artis Q/Q.zen<br>(VD11D) K181407 | Comparison<br>Results | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | New System<br>Software<br>Changes 1-8 | 1. | New System Software VE2 (also<br>known as VE20), software<br>modification/features | Software (VD11D) | Modified: This feature is<br>modified from the Primary<br>Predicate Device. All<br>software changes are<br>supported with System<br>Validation testing is<br>provided in Appendix K<br>Attachment 1-8 | | | 2. | Improved Roadmap (Also known as<br>OPTIQ) (Floor & Biplane)<br>A. Increased image quality dose<br>ratio and faster workflow via<br>new Architecture and Organ<br>Program parametrization due<br>to improved algorithms<br>B. Improved Image Quality due to<br>improved algorithim<br>C. Subtracted fluoro mode: Dose,<br>Time, and Contrast Agent<br>Savings.<br>D. Auto Map Integration in DSA | Advanced Roadmap | Modified: This feature is<br>modified from the Primary<br>Predicate Device. All<br>software changes are<br>supported with Bench<br>Testing provided in<br>Section 19 and System<br>Validation testing is<br>provided in Appendix K<br>Attachment 1-8 | | Modification | Subject Device<br>ARTIS icono (VE2) | Primary Predicate<br>Device | Comparison<br>Results | | | | 3. Improved Automatic Exposure<br>Control: (Floor & Biplane) | Artis Q/Q.zen<br>(VD11D) K181407<br>Secondary<br>Predicate Device<br>Atris pheno<br>(K163286) | Same: This feature<br>remains unchanged from<br>the Secondary Predicate<br>Device and is only new for | | | | A. New Automatic Exposure<br>Control (AEC) incl. Structure<br>Scout)<br>B. CNR Driven Exposure Control<br>(AEC) | Automatic Exposure<br>Control | the Subject Device. All<br>software changes are<br>supported with Bench<br>Testing. provided in<br>Section 19. System<br>Validation testing is<br>provided in Appendix K<br>Attachment 1-8 | | | | 4. 3D Imaging: (Biplane Floor)<br>A. syngoDyna3D HighSpeed | 3D Imaging | Same: This feature<br>remains unchanged from<br>the Primary Predicate<br>Device.<br>All software changes are<br>supported with Bench<br>Testing provided in<br>Section 19. System<br>Validation testing is<br>provided in Appendix K<br>Attachments, 1-8 | | | | B. syngo DynaCT Sine Spin | syngo DynaCT | Modified: This feature is<br>modified from the<br>previously cleared syngo<br>DynaCT in the Primary<br>Predicate Device. All<br>software changes are<br>supported with Bench<br>Testing provided in<br>Section 19. System<br>Validation testing is<br>provided in Appendix K<br>Attachment 1-8 | | | | C. TwinSpin (3D with Frontal<br>plane while lateral plane idle<br>(no x-ray) | syngo DynaCT | Modified: This feature is<br>modified from the<br>previously cleared syngo<br>DynaCT in the Primary<br>Predicate Device. All<br>software changes are<br>supported with Bench<br>Testing provided in<br>Section 19. System<br>Validation testing is<br>provided in Appendix K<br>Attachment 1-8 | | | | D. syngo DynaCT Multiphase | syngo DynaCT | Modified: This feature is<br>modified from the<br>previously cleared syngo<br>DynaCT in the Primary<br>Predicate Device. All<br>software changes are<br>supported with Bench<br>Testing provided in<br>Section 19. System<br>Validation testing is<br>provided in Appendix K | | | Modification | Subject Device<br>ARTIS icono (VE2) | Primary Predicate<br>Device<br>Artis Q/Q.zen<br>(VD11D) K181407 | Comparison<br>Results | | | | 5. Improved ClearStent Live: (Floor & Biplane) | ClearStent Live | Attachments1-8<br><b>Modified:</b> This feature is modified from the Primary Predicate Device. All software changes are supported with Bench Testing provided in <b>Section 19</b> . System Validation testing is provided in <b>Appendix K</b><br>Attachments1-8 | | | | 6. Updated User Interface: (Floor & Biplane)<br>A. Case Flow<br>B. Favorites in control room and examination room | User Interface | <b>Modified:</b> This feature is modified from the previously cleared Primary Predicate Device. All software changes are supported with Bench Testing provided in <b>Section 19</b> . System Validation testing is provided in <b>Appendix K</b><br>Attachments1-8 | | | | 7. Lateral Plane Switch (Biplane) | Lateral Plane | <b>Modified:</b> This feature is modified from the previously cleared Primary Predicate Device. All software changes are supported with Bench Testing provided in <b>Section 19</b> . System Validation testing is provided in <b>Appendix K</b><br>Attachments1-8 | | | | 8. Generic Interface for 3rd parties for data transference (Floor & Biplane) | User Interface | <b>Modified:</b> This feature is modified from the previously cleared feature in the Primary Predicate Device. All software changes are supported with Bench Testing provided in <b>Section 19</b> . System Validation testing is provided in <b>Appendix K</b><br>Attachments1-8 | | ## Table 5: Comparison of Technological Characteristics {11}------------------------------------------------ Image /page/11/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots. {12}------------------------------------------------ Image /page/12/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots. #### Nonclinical Performance Testing: 9. Non-clinical tests were conducted for the ARTIS icono (VE2) during product development. The ARTIS icono (VE2) was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility: - AAMI ANSI ES60601-1:2005/(R)2012 . - IEC 60601-1-2:2014 . {13}------------------------------------------------ Image /page/13/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots. - IEC 60601-1-3:2013 ● - IEC 60601-1-6:2010/A1:2013 - IEC 60825-1:2007 ● - TR 60878:2015 ● - IEC 62304:2015 . - IEC 80001-1:2010 ● - IEC 60601-2-28:2017 ● - IEC 60601-2-43:2017 ● - IEC 60601-2-54:2009/A1:2015 ● - ISO 10993-1:2009 ● - . ISO 14971:2007 - German national standard DIN 6868-157 . # Table 6: FDA Guidance Documents | | FDA Guidance Document and Effective Date | |-----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1. | Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket<br>Notification Submissions 510(k)<br>Document issued on October 2, 2017 | | 2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policy<br>for 510(k)s<br>Document issued on January 30, 2018 | | 3. | Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s<br>- Guidance for Industry and FDA Staff<br>Document issued on August 12, 2005 | | 4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a change<br>to an existing device.<br>Document issued on October 25, 2017 | | 5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket Notifications [510(k)]<br>Document Issued on July 28, 2014 | | 6. | Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s for<br>Solid State X-ray Imaging Devices<br>Document issued on September 1, 2016 | | 7. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices<br>Document issued on May 11, 2005 | | 8. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in<br>Medical Devices<br>Document issued on September 9, 1999 | | 9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices.<br>Document issued February 3, 2016 | | 10. | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging Device<br>Premarket Notifications.<br>Document issued on November 28, 2017 | | 11. | Guidance for Industry and FDA Staff: Content of Premarket Submissions for<br>Management of Cybersecurity in Medical devices.<br>Document issued on October 2, 2014 | | 12. | Guidance for Industry and FDA Staff: Appropriate Use of Voluntary Consensus<br>Standards in Premarket Submission for Medical Devices<br>Document issued on September 14, 2018 | | 13. | Guidance for Industry and FDA Staff: Medical Device Accessories Describing<br>Accessories and Classification Pathways | {14}------------------------------------------------ Image /page/14/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and below it, "Healthineers" is written in orange. To the right of the text is a graphic of orange dots arranged in a pattern. #### Document issued on December 20, 2017 The modifications described in this Premarket Notification were supported with verification and validation testing. ## Verification and Validation: Software Documentation for a Major Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on ARTIS icono System software (VE2) during product development. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence. ARTIS icono System software (VE2) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according ot the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment. Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 is the hospital. Provided in the Software Section, is the required cybersecurity information. ### Summarv: Performance tests were conducted to test the functionality of ARTIS icono (VE2) System. These tests have been performed to assess the functionality of the subject. device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness. ## 10. General Safety and Effectiveness Concerns: Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all {15}------------------------------------------------ Image /page/15/Picture/1 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is an orange graphic of several circles. equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images. # 11. Conclusion as to Substantial Equivalence: The predicate devices were cleared based on non-clinical supportive information and clinical images and data. Similar non-clinical test results demonstrate that the ARTIS icono (VE2) System acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Devices that is currently marketed for the same intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...