K133580 · Siemens Medical Solutions USA, Inc. · OWB · Apr 28, 2014 · Radiology
Device Facts
Record ID
K133580
Device Name
ARTIS ONE
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Apr 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis one include cardiac angiography, neuro-angiography, general angiography, rotational multipurpose body angiography, radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by x-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.
Device Story
Artis one is an angiographic X-ray system for diagnostic and interventional procedures. It utilizes a C-arm stand, X-ray tube (MEGALIX Cat Plus), flat panel detector (Trixell pixium 2630S), and digital imaging system. The device acquires X-ray images, which are processed by the AXIS Imaging System (SW VA10) to provide visualization for clinicians. It supports 3D imaging, position-triggered imaging for spatial data synthesis, and image fusion/navigation. Operated by healthcare professionals in clinics or ORs, the system assists in real-time guidance during interventions. Output is displayed for clinical decision-making, aiding in diagnostic accuracy and procedural guidance. Benefits include improved visualization for complex interventions and support for diverse patient populations, including pediatric and obese patients.
Clinical Evidence
Clinical evidence included a 'Concurrent Study' (bench test) and a Clinical Use Test performed on normal and obese adult patients. No adverse effects or complications were observed. The study confirmed diagnostic image quality adequate for interventional guidance. No new clinical applications were introduced; therefore, no formal clinical trial was required.
Technological Characteristics
System components: C-arm, MEGALIX Cat Plus 125/40/90-125GW X-ray tube, Trixell pixium 2630S flat panel detector, MFD collimator, and digital imaging system. Software: Version VA10. Conformance to IEC 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, 60601-2-54, NEMA PS 3.1-3.20, ISO 14971, and AAMI ANSI ISO 10993-1. Connectivity: Integrated 3D imaging and spatial data synthesis.
Indications for Use
Indicated for diagnostic imaging and interventional procedures, including cardiac, neuro, general, and rotational body angiography, and radiographic/fluoroscopic procedures. Suitable for pediatric and obese patients. Used for image-guided surgery, fusion, and navigation in clinical or OR settings.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Artis zee/ zeego with CSX-10 Detector SW VC21 (K122644)
Artis Q and Q.zen - Modular Angiographic System (K123529)
{0}------------------------------------------------
K133580 APR 2 8 2014
## 510(k) Summary: Artis one
Siemens Medical Solutions USA, Inc. Company: 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: March 25, 2014
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Solution USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Establishment Registration Number: 2240869
## Manufacturing Site:
Siemens Shenzhen Magnetic Resonance Ltd. Siemens MRI Center, Gaoxin C. Ave., 2nd, Hi-Tech Industrial Park 518057 Shenzhen, China Establishment Registration Number: 3004754211
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc., 51 Valley Stream Parkway Malvern, PA 19355 Fax: (610) 448-1787 Phone: (610) 448 -3536 Email: patricia.d.jones@siemens.com
#### Device Name and Classification: 3.
| Trade Name: | Artis one Angiographic System |
|----------------------------|------------------------------------------------|
| Classification Name: | Angiographic X-Ray System |
| | Image intensified fluoroscopic x-ray<br>system |
| Common Name: | Angiographic X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1600 |
| | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Code | OWB. JAA. IZI |
{1}------------------------------------------------
- Legally Marketed Predicate Devices 4. Trade Name:
510(k) Clearance: Clearance Date: Classification Name:
Classification Panel: CFR Section: Device Class: Product Code:
Trade Name:
510(k) Clearance: Clearance Date: Classification Name:
Classification Panel: Classification Regulation:
Device Class: Product Code:
Trade Name: 510(k) Clearance: Clearance Date: Classification Name:
Classification Panel: Classification Requlation: Device Class: Product Code:
Artis zee/ zeego with CSX-10 Detector SW VC21 K122644 May 20, 2013 Image intensified fluoroscopic x-ray system Radiology 21 CFR §892.1600 Class II OWB
Artis Q and Q.zen - Modular Angiographic System K123529 February 26, 2013 Angiographic X-Ray System Image intensified fluoroscopic x-ray svstem Radiology 21 CFR §892.1600 21 CFR §892.1650 Class II OWB, JAA, IZI
X-LEONARDO Workstation K042995 November 24, 2004 Picture archiving and communications svstem Radiology 21 CFR 8892.2050 Class II LLZ
#### 5. Device Description:
Siemens Medical Solutions USA, Inc. intends to market an angiography x-ray system the Artis one Angiographic System, SW Version VA10. This 510(k) submission describes several modifications to the previously cleared predicate devices the Artis zee/ zeego K122644; Artis Q and Q.Zen- Modular Angiographic System K123529 and the X-LEONARDO Workstation K042995. The following modifications are made to the cleared predicate devices and developed into the Subject Device the Artis one:
1). A modified Flat Panel detector Trixell pixium 2630S; 2). A modified collimator MFD; 3). A modified x-ray tubes assembly MEGALIX Cat Plus 125/40/90-125GW; 4). A modified Software Version VA10; 5). A modified C-Arm stand; 6). A new patient table; 7). A modified imaging processing system;
{2}------------------------------------------------
8). A modified display; 9). Modified user Control Modules; 10). Integrated 3D imaging; 11). Proposed product claims associated with the above device modifications (see Section 13 Labeling).
#### 6. Indication for Use:
Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis one include cardiac angiography, neuro-angiography, general angiography, rotational multipurpose bodv angiography, radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by x-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions.
The Artis one can also support the acquisition of position triggered imaging for spatial data synthesis.
#### 7. Substantial Equivalence:
The Artis one Angiography System is substantially equivalent to the commerically available Siemens' Artis zee/ zeego with CSX-10 Detector SW VC21( K122644) angiographic x-ray system with same indication for use for single plane x-ray systems and similar components. The Artis one contains a similar post processing (AXIS Imaging system SW Version VA10); Clear and Care dose reduction features and similar product labeling claims as cleared in the Artis Q and Zen- Modular Angiographic System 510(k) (K123529). The Artis one also produces 3D imaging similar to the cleared X-LEONARDO Workstation 510(k) (K042995). Provided in Table 1 is a list of all predicate devices for the subject device the Artis one.
| Predicate Device Name<br>and Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
|---------------------------------------------------|------------------|-------------------|------------------------------------------------------------------------------------------------------|
| Artis zee/ zeego with CSX-<br>10 Detector SW VC21 | K122644 | 05/20/2013 | • Indication for use<br>• Hardware components |
| Artis Q and Q.zen- Modular<br>Angiographic System | K123529 | 02/26/2013 | • Post processing<br>(AXIS Imaging System)<br>• Clear and Care features<br>• Product Labeling Claims |
| X-LEONARDO Workstation | K042995 | 11/24/2004 | • 3D imaging |
## Table 1 List of the predicate devices
{3}------------------------------------------------
### Summary of Technological Characteristics of the Subject Device as 8. Compared with the Predicate Device:
Artis one Angiography System is designed as a set of components (C-arm, Xray tube and housing, flat detector, digital imaging system, collimator, generator etc.) that are configured to provide specialized angiography systems. These components are further development of the predicate ones and include minor modifications to existing components.
## Discussion:
The components of the subject device have the same technological characteristics as the ones from the predicate device. The technological characteristics do not differ from the predicate systems.
Many comparable parameters remain the same, some differ slightly as shown in the comparison tables (indicated as "similar").
Testing and validation have been completed. A "Concurrent Study" (bench test) as well a Clinical Use Test have been performed on normal and obese adult patients. The image quality has been considered of diagnostic quality and adequate for guidance during interventional procedures. During these studies no adverse effects or complications have been observed and therefore the device is considered safe and effective.
The tests did show that the subject device Artis one with all its components is comparable to the predicate ones and therefore substantial equivalent to the predicate device.
The subject device Àrtis one does not affect the indication for use nor the intended use of the device nor does it alter its fundamental scientific technology from the 510(k) cleared predicate device Artis zee/ zeego K122644.
#### Non- Clinical Performance Testing 9.
Non-clinical tests were conducted for the Artis one configured with software version VA10 during product development. The modifications described in this Premarket Notification were supported with verification/validation testing and a concurrence study was performed with phantom images.
Siemens claims conformance to a signed statement of conformance to performance standards as follows: IEC 60601-1-2; IEC 60601-1-3; IEC 60601-2-28; IEC 60601-2-43; IEC 60601-2-54; NEMA PS 3.1 - 3.20; ISO 14971; AAMI ANSI ISO 10993-1.
Software documentation for a Major Level of Concern per FDA's Guidance Document "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted on the Artis one software during product development.
{4}------------------------------------------------
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
EMC/electrical safety was evaluated according to the IEC standards. Siemens certify to conform to Voluntary Standards covering Electrical and Mechanical Safety. In conclusion, the identified risk of electrical hazards was mitigated and is substantially equivalent to the predicate devices in terms of safety and effectiveness. All testing and validation have been completed.
Clinical testing was not applicable as the Artis one has no new indication for use nor new clinical applications were introduced to the system.
#### General Safety and Effectiveness Concerns: 10.
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
#### Conclusion as to Substantial Equivalence: 11.
The Artis one Angiography System has the same indication for use as the predicate device Artis zee floor from the Artis zee / zeego family K122644.
The components of the subject device have the same technological characteristics as the ones from the predicate device. The technological characteristics do not differ from the predicate systems.
Testing and validation have been completed. A "Concurrent Study" (bench test) as well a Clinical Use Test did show that the subject device Artis one with all its components is comparable to the predicate ones and therefore substantial equivalent to the predicate device.
The Artis one Angiography System with SW VA10 is substantially equivalent to the commercially available Artis zee/zeego K122644 with all its components as described in the Device Description.
The CARE and CLEAR features are substantially equivalent to the commercially available Artis Q and Artis Q.zen Modular Angiography System
{5}------------------------------------------------
K123529. Postprocessing and 3D Imaging is substantially equivalent to the commercially available X-LEONARDO Workstation K042995.
It is Siemens opinion, that the Artis one Angiography System is substantially equivalent to the listed predicate devices in Table 1.
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center WO66-G609 Silver Spring, MD 20993-0002
April 28, 2014
Siemens Medical Solutions USA, Inc. Patricia Jones 51 Valley Stream Pkwy. Malvern, PA 19355 US
Re: K133580
> Trade/Device Name: Artis one Regulation Number: 21 CFR 892.1650 Regulation Name: Interventional Fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA Dated: March 21, 2014 Received: March 26, 2014
### Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that I DA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sm.h.7)
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number (if known) K133580
Device Name Artis One .
Artis one is an angiography system developed for diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Procedures that can be performed with the Artis one include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whold body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities.
Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination able as an integrated part of the system can be used for Xray imaging, surgery and interventions.
The Artis one can also support the acquisition of position triggered imaging for spatial data systhesis.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.P.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the The burden une for this onlooken of micrination is sources, gather and maintain the data needed and complete unte to review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.