The MasteRad MX30 is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest. abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
Device Story
Stationary digital diagnostic X-ray system; utilizes X-ray tube, high voltage generator, and solid-state digital acquisition panels to capture radiographic images. Operated by qualified doctors or technicians in clinical settings. System components include floor-mounted tube stand, collimator, and digital panels (Varex or Toshiba/Canon models). Input consists of X-ray radiation passing through patient anatomy; output is digital radiographic image processed via Voyance software. Healthcare providers use images for diagnostic assessment of skeletal and soft tissue structures. System supports patient positioning (sitting, standing, prone, supine). Benefits include high-quality diagnostic imaging for various body parts. Device integrates with existing clinical workflows via DICOM standards.
Clinical Evidence
No clinical trials performed. Clinical evidence limited to evaluation of images acquired from the new Toshiba FDXA4343R digital panel. Images reviewed by a board-certified radiologist and determined to be of excellent diagnostic quality, satisfying requirements for new solid-state X-ray imaging devices.
Technological Characteristics
Stationary digital X-ray system. Components: Toshiba X-ray tube, CPI CMP 200DR generator, Collimare collimator, and various Varex/Toshiba digital panels (CsI with a-Si photodiode). Connectivity: DICOM 3.0 compliant, wired/wireless (IEEE 802.11n). Standards: IEC 60601-1, IEC 60601-1-2, UL 60601-1. Software: Voyance (updated version).
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic radiographic imaging of the skull, spinal column, chest, abdomen, and extremities. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 25, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an eagle-like figure, representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
MedicaTech USA % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
## Re: K190601
Trade/Device Name: MasteRad MX30 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MOB Dated: March 29, 2019 Received: April 2, 2019
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190601
Device Name MasteRad MX30
Indications for Use (Describe)
The MasteRad MX30 is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
Type of Use (Select one or both, as applicable)
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary, 510(k) K190601
Submitter: MedicaTech USA 50 Maxwell Irvine , CA , 92618 Toll Free : +1 800 817 5030 Phone : +1 949 679 2881 FAX : +1 949 679 2882 Registration Number 3004989804 Contact: George Makar, Predsident Date Prepared: January 23, 2019
#### 1. Identification of the Device:
Trade/Device Names: MasteRad MX30 Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: KPR and MQB Common/Usual Name: Stationary Digital Diagnostic X-Ray System
2. Equivalent legally marketed device: K143257 Trade/Device Name: KrystalRad 1100 and KrystalRad 3000 Manufacturer: MedicaTech USA Regulation Number: 21CFR892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: II Product Code: KPR and MQB Common/Usual Name: Stationary Digital Diagnostic X-Ray System
- 3. Indications for Use The MasteRad MX30 is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest. abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.
- 4. Description of the Device: This device represents a new combination of already cleared solid state digital x-ray acquisition panels (plus one NEW panel) and software with the diagnostic x-ray compnents required to make a complete system. The purchaser may select their digital panel from this list:.
| Manufacturer Model Number | MedicaTech Model Number | 510(k) Number |
|---------------------------|-------------------------|---------------|
| Varex PaxScan 2530Wv4 | Master X CW25 | K161459 |
| Varex PaxScan 4336Wv4 | Master X CW36 | K161459 |
| Varex PaxScan 4343R | Master X FC43 | K172951 |
| Varex PaxScan 4343RC | Master X PT43 | K172951 |
| Varex XRpad 4343F | Master X HDF43 | K142698 |
| Varex XRpad 3025 | Master X WPH25 | K161942 |
| Varex XRpad 4336 | Master X WHD36 | K161966 |
| Toshiba* FDXA4343R | Master X FA43 | New |
{4}------------------------------------------------
As compared to our predicate system, the tube stand is floor mounted instead of ceiling mounted. The collimator is different. Instead of the Ralco collimator, a Collimare model is supplied. The purchaser selects one of the following FDA certified models: CML-150-0001-C; CTL-150-0001-C; CML-125-0001-C. The x-ray tube and the high voltage generator remains the same as our predicate. The x-ray tube is a Toshiba model and the generator is the CPI CMP 200DR. The image acquisition software is a newer version of our Voyance software originally cleared in K130377
The system complies with the CDRH Radiological Health performance standard in the Code of Federal Regulations, as well as the voluntary IEC standards IEC 60601-1 and IEC 60601-1-2. *Toshiba panel division has recently been purchased by Canon.
- 5. Safety and Effectiveness, comparison to predicate device. This combination device has the same indications for use and very similar technological characteristics as the predicate device, and employs already 510(k) cleared digital panels (except for the Toshiba unit) and software.
| Characteristi<br>C | K143257 KrystalRad 1100 and<br>KrystalRad 3000 Digital Stationary<br>Radiographic Systems | MastRad MX30 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use: | The KrystalRad Systems are intended for<br>use by a qualified/trained doctor or<br>technician on both adult and pediatric<br>subjects for taking diagnostic<br>radiographic exposures of the skull,<br>spinal column, chest. abdomen,<br>extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the<br>prone or supine position. Not for<br>mammography. | MasteRad MX30 is intended for use by a<br>qualified/trained doctor or technician on<br>both adult and pediatric subjects for taking<br>diagnostic radiographic exposures of the<br>skull, spinal column, chest. abdomen,<br>extremities, and other body parts.<br>Applications can be performed with the<br>patient sitting, standing, or lying in the prone<br>or supine position. Not for mammography.<br>SAME |
| Configuration<br>of Digital<br>Panels | Battery or AC operated wireless IEEE<br>802.11n or Wired Ethernet | Battery or AC operated wireless IEEE 802.11n<br>or Wired Ethernet (depending on the model<br>chosen) SAME |
| Digital Panel<br>Models | Vieworks FXRD-1717SA/SB, or<br>Vieworks FXRD-1417SA/SB or<br>Vieworks FXRD-1417WA/WB or<br>Toshiba: FDX4343R/RPW or<br>Toshiba FDX3543RP/RPW or<br>Varex XRpad 4336 | Varex PaxScan 2530Wv4 or<br>Varex PaxScan 4336Wv4 or<br>Varex PaxScan 4343R or<br>Varex PaxScan 4343RC or<br>Varex XRpad 4343F or<br>Varex XRpad 3025 or<br>Varex XRpad 4336 or<br>Toshiba (Now Canon) FDXA4343R |
| DICOM | DICOM 3 | SAME |
#### Substantial Equivalence Chart: 6.
{5}------------------------------------------------
| Characteristi<br>c | K143257 KrystalRad 1100 and<br>KrystalRad 3000 Digital Stationary<br>Radiographic Systems | MastRad MX30 |
|----------------------------------|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Image<br>acquisition<br>software | Voyance | Voyance. Updated. |
| Power Source | AC Line, various voltages available | SAME |
| Photo | Image: Krystalrad 3000 | Image: MasteRad MX30<br>Difference: Floor Mount instead of Ceiling<br>Mount. |
| Generator | CPI CMP 200DR | SAME |
| X-Ray Tube | Toshiba | SAME |
| Collimator | Ralco R301A | Collimare LLC, one of the following FDA<br>certified models: CML-150-0001-C,<br>CTL-150-0001-C, CML-125-0001-C |
A new digital x-ray receptor panel not previously cleared by FDA is offered. The predecessor to this panel was cleared in K130883. The characteristics are compared in the table below. The images were evaluated by a Board Certified Radiologist, and found to be of diagnostic quality.
| Characteristic | Sedecal Digital Radiogniphic Upgrade<br>Model SDRU-T K130883 (Toshiba Model<br>FDX4343R) | Toshiba Model FDXA4343R (One of the<br>available models for MasteRad.) |
|------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Image Format | | |
| X-ray<br>Conversion<br>Layer | Cesium lodide (Csl) with Amorphous<br>Silicon (a-Si) Photodiode | SAME |
| Active Area | 430(H)×439(V)mm (16.9×17.3 inch) | 426 (H)×425 (V) mm (16.8×16.7 inch)<br>(Not a meaningful difference) |
| Pixel Matrix | 3008(H)×3072(V) | 3040 (H)×3036 (V) (Not a meaningful<br>difference) |
| Pixel Pitch | 143 $ \mu $ m | 140 $ \mu $ m (Not a meaningful difference) |
{6}------------------------------------------------
| Characteristic | Sedecal Digital Radiogniphic Upgrade<br>Model SDRU-T K130883 (Toshiba Model<br>FDX4343R) | Toshiba Model FDXA4343R (One of the<br>available models for MasteRad. |
|------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Cycle Time | Shot to Shot 6sec | Shot to Shot 6sec No difference |
| Performance | | |
| Limiting<br>Resolution | 3.5 Lp/mm typ | 3.7 Lp/mm Max. Slightly better |
| MTF (2.0<br>Lp/mm, 70<br>kVp, 1x1) | 36 % Typ | 50 % typ Slightly better |
| DQE (0),<br>Quantum -<br>Limited | > 70 % | > 56 % (Typ.) (Not a meaningful<br>difference. Images found to be of<br>diagnostic quality, however) |
| A/D<br>Conversion | 14 bit | 16 bit Slightly better |
- 7. Summary of non-clinical testing: We performed software validation and risk management for the updated software version. The following FDA guidances were employed in those activities: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices and Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. The software comlies with NEMA PS 3.1 - 3.18 (2009) Digital Imaging and Communications in Medicine We revised our labeling in light of the FDA guidance: Pediatric Information for X-ray (DICOM) Set. Imaqinq Device Premarket Notifications. We performed integration testing to show that each panel performed correctly in the full system. The device conforms to US Performance Standards and the hardware is UL Listed to US Standards for safety for medical devices (UL 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety standard by Underwriters Laboratories, 04/25/2003.) There is one new digital receptor panel, and we were provided test information per Guidance for the Submission of 510(k)s for Solid State X-ray Imaginq Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016 The software comlies with NEMA PS 3.1 - 3.18 (2009) Digital Imaging and Communications in Medicine (DICOM) Set. The results of a review of bench, safety test, and software validation documentation indicates that the new device is as safe and effective as the predicate device
- 8. Summary of clinical testing: Clinical images are not required for the previously cleared panels, but the new Toshiba panel has not previously received FDA clearance. Clinical images were acquired from the new Toshiba panel and reviewed by a board certified radiologist, as recommended in the FDA document: Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices The images were found to be of excellent diagnostic quality.
- 9. Conclusion: After analyzing software integration validation, safety testing data, and clinical images, it is the conclusion of Medicatech USA that the "MedRad" system is as safe and effective as the predicate device, has insignificant technological differences, and has identical indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.