K142698 · Perkinelmer, Inc. · MQB · Oct 23, 2014 · Radiology
Device Facts
Record ID
K142698
Device Name
PerkinElmer, XRpad 4343 F MED Flat Panel Detector
Applicant
Perkinelmer, Inc.
Product Code
MQB · Radiology
Decision Date
Oct 23, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The XRpad 4343 F MED, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
Device Story
XRpad 4343 F MED is a cassette-sized, flat panel X-ray detector for digital radiography. It captures X-ray radiation using an amorphous silicon flat panel with a directly deposited CsI:T1 scintillator. The device integrates into fixed room X-ray systems, receiving X-ray exposure and converting it into digital image data. It is operated by radiology technicians or clinicians within a clinical setting. The detector connects to an X-ray system via cable (GigE interface) and uses dedicated electronics to process and transmit images to the X-ray unit's imaging software. Healthcare providers view these digital images to make diagnostic decisions. The device benefits patients by enabling high-quality digital radiography, replacing traditional film-based systems, and facilitating faster diagnostic workflows.
Clinical Evidence
No new clinical study was required. Substantial equivalence is supported by non-clinical performance testing (DQE, MTF) and the applicability of clinical data previously submitted for the predicate device (K140551) and its predecessor (K122495).
Technological Characteristics
Amorphous silicon flat panel with directly deposited CsI:T1 scintillator. Pixel pitch 100 µm. Active area 430 mm x 430 mm. Enclosure: carbon-fiber and aluminum. Connectivity: Gb Ethernet. X-ray synchronization via trigger lines or Automatic Exposure Detection. Complies with UL and IEC standards.
Indications for Use
Indicated for generating radiographic images of human anatomy for diagnostic X-ray procedures in patients where conventional screen-film, digital radiography, or computed radiography systems are used. Not intended for mammographic use.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
PerkinElmer XRpad 4336 MED Flat Panel Detector (K140551)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three figures facing to the right. The figures are connected and appear to be in motion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2014
Perkin-Elmer, Inc. % Ms. Dawn Spooner Associate Director, Regulatory Affairs 940 Winter Street WALTHAM MA 02451
Re: K142698
Trade/Device Name: XRpad 4343F MED Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: September 22, 2014 Received: September 23, 2014
Dear Ms. Spooner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Smh
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K142698
Device Name
XRpad 4343 F MED
Indications for Use (Describe)
The XRoad 4343 F MED, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary XRpad 4343 F MED Flat Panel Detector
This summary of 510(k) safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K142698
October 15, 2014
| Submitted By: | PerkinElmer Medical Imaging<br>2175 Mission College Blvd.<br>Santa Clara, CA 95054<br>U.S.A. |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dawn Spooner<br>Associate Director, Regulatory Affairs<br>Tel. 781-663-6071<br>Fax. 781-663-5969 |
| Device Name: | PerkinElmer XRpad 4343 F MED Flat Panel Detector |
| Classification: | Product Code: MQB<br>Classification Name: Stationary X-ray System<br>Classification Regulation: 21 CFR 892.1680 |
| Predicate Device: | PerkinElmer XRpad 4336 MED Flat Panel Detector |
| | 510(k) Clearance: K140551; August 1, 2014<br>Product Code: MQB<br>Classification Name: Stationary X-ray System<br>Classification Regulation: 21 CFR 892.1680 |
{4}------------------------------------------------
# Device Description:
The XRpad 4343 F MED is a lightweight, cassette-sized, flat panel X-ray detector for digital radiography. The X-ray detector consists of an amorphous silicon flat panel with a directly deposited CsI:T1 scintillator and dedicated read-out, scan, and control electronics, all packaged in a carbon-fiber and aluminum enclosure.
The outside dimensions of the detector are 460 mm × 15 mm, which fits into a standard X-ray cassette Bucky. The active area is 430 mm × 430mm at a pixel pitch of 100um.
The detector can be integrated into a fixed room X-ray system to enable digital radiography. The following accessories are available for the XRpad 4343 F MED:
- XRpad IPU (Interface Power Unit) ●
- XRpad LPT Detector Cable 3 m / 10 ft. ●
- AC Cable IEC 60320 C13 DE ●
- AC Cable IEC 60320 C13 US ●
- Trigger Cable 5 m / 16.5 ft. ●
- Trigger Cable 20 m / 65.5 ft.
- XRD GigE Interface Cable 7.6 m / 25 ft. ●
- XRD GigE Interface Cable 15.25 m / 50 ft. ●
- XRD GigE Interface Cable 30.5 m / 100 ft.
The XRpad 4343 F MED detector is designed to work with any X-ray system (consisting of an X-ray source, generator, collimator, and positioner) intended for use in generating radiographic images of human anatomy for diagnostic X-ray procedures.
# Indications For Use:
The XRpad 4343 F MED, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-rav procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
{5}------------------------------------------------
# Comparison Chart
| Characteristics | Model<br>PerkinElmer<br>XRpad 4336 MED (K140551) | Proposed Model<br>PerkinElmer<br>XRpad<br>4343 F MED |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | The XRpad 4336 MED when used with a<br>radiographic imaging system, is indicated<br>for use in generating radiographic images<br>of human anatomy for diagnostic X-ray<br>procedures, wherever conventional screen-<br>film (SF), digital radiography (DR), or<br>computed radiography (CR) systems may<br>be used. It is not intended for<br>mammographic use. | Same |
| Customer<br>Applications | Single energy, single shot RAD in 100 µm | Same |
| Panel | Single substrate amorphous silicon active<br>TFT/diode array | Same |
| Scintillator | Direct deposition CsI:T1 | Same |
| Pixel matrix | 3556 × 4320 pixels | 4318 × 4320 pixels |
| Pixel pitch | 100 µm | Same |
| Limiting<br>resolution | 5 lp/mm | Same |
| Binning<br>capable | 2 × 2 binning for 200 µm | Same |
| Active area | 355 mm × 430 mm | 430 mm × 430 mm |
| External<br>dimensions<br>(w×l×h) | 384 mm × 460 mm × 15 mm | 460 mm × 460 mm × 15 mm |
| Weight | Approximately 4 kg | Approximately 5 kg |
| Housing<br>material | Aluminum with carbon-fiber | Same |
| Communication<br>interface | Gb Ethernet or 802.11n WiFi | Gb Ethernet<br>No wireless capability |
| WiFi band | 5.1-5.3 GHz | None |
| X-ray<br>synchronization<br>interface | Dedicated trigger in/out signal lines or<br>Automatic Exposure Detection | Same |
| Power | External power supply or battery | External power supply |
| Battery<br>capacity | 53Wh | No battery option |
| Connector | For fixed installation | Same |
| Software | Windows OS | Same |
{6}------------------------------------------------
### Summary of Studies:
The proposed new device and the XRpad predicate device utilize similar technology and materials, are similar in design and construction, and have the same intended use. The construction and physical characteristics of the PerkinElmer XRpad 4343 F MED compared to the predicate PerkinElmer XRpad 4336 MED are similar. The difference is related to the increase in dimensions and does not impact image quality.
The PerkinElmer XRpad 4343 F MED flat panel detector has successfully completed internal non-clinical testing, complies with standards and regulations such as UL and IEC. A new clinical study was not required for the XRpad 4343 MED device. The predicate device, XRpad 4336 MED, was cleared using clinical data derived from testing to support K122495 (XRD 1622 AP3 MED) and applicable to the predicate device. The XRpad 4343 MED device differences do not invalidate the applicability of the clinical study data submitted in K122495. Through non-clinical testing, we have demonstrated the clinical data collected with the (XRD 1622 AP3 MED) in K122495 is also applicable to the XRpad 4343 MED device as it was applicable for the XRpad 4336 MED device.
### Summary of Design Control and Risk Management:
The XRpad 4343 F MED flat panel X-ray detector is a modification of the XRpad 4336 MED (K140551). The 43 cm × 43 cm XRpad 4343 F MED, as compared to the 43 cm × 36 cm XRpad 4336 MED, is larger in one dimension, is heavier, and does not have a battery, or wireless capability.
The risks and hazardous impacts of the device modification were analyzed by FMEA methodology. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented as part of product design. The overall assessment concluded that all identified risks and hazardous conditions were successfully mitigated and accepted.
### Summary of Non-Clinical Data:
To demonstrate the equivalence of the PerkinElmer XRpad 4343 F MED to the predicate device, PerkinElmer has performed internal non-clinical testing and demonstrated compliance with accepted standards and regulations such as UL and IEC following the non-clinical considerations outlined in the Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, August, 1999. The non-clinical testing was conducted and demonstrated the main physical values of the XRpad 4343 F MED such as DQE and MTF are comparable to the predicate device. The non-clinical performance testing conducted and resulting data demonstrate substantial equivalence.
{7}------------------------------------------------
## Substantial Equivalency:
The proposed device and predicate device (XRpad 4336 MED flat panel detector) both utilize similar technology and materials, are similar in design and construction, and have been shown to produce images of equivalent diagnostic quality. Both devices are intended for use in generating radiographic images of human anatomy for diagnostic Xray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. The devices are not intended for mammographic use. Both devices produce digital images which can be transmitted to imaging software of the X-ray unit.
### Conclusion:
Similar to the predicate device, the XRpad 4343 F MED has comparable performance and operational standards. Potential hazards have been studied and controlled by a Risk Management Plan. The non-clinical software verification and validation test results demonstrate that the PerkinElmer XRpad 4343 F MED complies with international and FDA recognized consensus standards and meets the acceptance criteria and is adequate for its intended use.
Based on the information supplied in this 510(k) PerkinElmer concludes, the XRpad 4343 F MED is substantially equivalent to the currently marketed device. XRpad 4336 MED (K140551) in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.