K190550 · Broda Seating (Broda Lp) · IOR · Jul 1, 2019 · Physical Medicine
Device Facts
Record ID
K190550
Device Name
IKON 40 Manual Wheelchair
Applicant
Broda Seating (Broda Lp)
Product Code
IOR · Physical Medicine
Decision Date
Jul 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The standard IKON 40 is a manual wheelchair device that is intended to be used to provide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is to be used as a means of mobility for persons limited to a sitting position.
Device Story
IKON 40 is a foldable, ultra-lightweight manual wheelchair; provides mobility for individuals limited to a sitting position. Device features aluminum folding frame, adjustable seat dimensions, swing-away/elevating legrests, and various wheel options. Operated by user or caregiver in clinical or home settings. Device provides physical mobility support; does not involve electronic processing, AI, or software. Substantial equivalence established via bench testing against ANSI/RESNA standards for strength, stability, and durability.
Clinical Evidence
No clinical data. Bench testing only. Device performance validated against ANSI/RESNA and ISO 7176 standards for static stability, braking effectiveness, dimensions, mass, maneuvering space, and structural strength (static, impact, fatigue). Flammability testing conducted per CAL 117 and ISO 8191.
Technological Characteristics
Aluminum folding frame; manual operation. Adjustable seat width (16-22"), depth (16-22"), and height (14-19"). Features swing-away/elevating legrests, flip-up footrests, and adjustable armrests. Compatible with 22"/24" spoke or composite mag wheels. Complies with ANSI/RESNA Wheelchair Standards (Vol 1, Sections 1, 3, 5, 7, 8, 15, 16) and flammability standards (CAL 117, ISO 8191). No electronic components, software, or energy source.
Indications for Use
Indicated for adolescents and adults (ages 12+) with mobility limitations requiring a sitting position, with a maximum weight capacity of 300 lbs.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs in the United States.
July 1, 2019
Broda Seating (Broda LP) Jeremy Ballard Product Manager 560 Bingemans Centre Drive Kitchener, ON, N2B3X9 CA
Re: K190550
Trade/Device Name: IKON 40 Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 28, 2019 Received: March 28, 2019
Dear Jeremy Ballard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek Pinto, Ph.D. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190550
Device Name IKON 40 Manual Wheelchair
Indications for Use (Describe)
The standard IKON 40 is a manual wheelchair device that is intended to be used to provide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is to be used as a means of mobility for persons limited to a sitting position.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
Date Prepared (revised): June 28,2019
K190550
Submitter:
Broda Seating (Broda LP)
560 Bingemans Centre Drive
Kitchener, ON, N2B3X9
Registration Number 8043392
Phone: 844-552-7632
Fax: 844-442-7632
#### Device:
Device Trade Name: IKON 40 Manual Wheelchair Common or Usual Name: Folding Ultra-Lightweight, Manual Wheelchair Classification Name: Mechanical Wheelchair 21 CFR §890.3850 Regulatory Class: I Product Code IOR: Wheelchair, mechanical
Predicate Device:
MyOn HC™ Manual Wheelchair (K152536); manufactured by Invacare®
(1525712)
{4}------------------------------------------------
### Device Description:
### Technical Specifications (* Standard Configuration):
The IKON 40 is a foldable, ultra-lightweight and fully adjustable manual wheelchair.
| Specification | IKON 40 | |
|-----------------------------------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Cross bar | *1 x 1 | 2 x 1 |
| Seat widths | 16", *18" | 20", 22" |
| Seat depths | 16"- 22" Adjustable | 16"- 22" Adjustable |
| Seat height | 14", 15", 16", 17", 18", 19"<br>Adjustable | 14", 15", 16", 17", 18", 19"<br>Adjustable |
| Max user weight | 300 lbs. | 300 lbs. |
| Chair length | 40" | 40" |
| Chair width | 26" | 26" |
| Length (no legrests) | 30" | 30" |
| Folded height | 34.5" | 34.5" |
| Folded width | 13.5" | 13.5" |
| Total weight (Including legrests, armrests and<br>anti-tippers) | 34 lbs. | 34 lbs. |
| Front seat height | 16" | 16" |
| Backrest height | 16.5" | 16.5" |
| Legrests | *Swing Away<br>Elevating | Swing Away<br>Elevating |
| Legrest range | 13.5"- 19" Adjustable | 13.5"- 19" Adjustable |
| Legrest angle (°) | 70° | 70° |
| Footrests | Flip Up and Angle Adjustable | Flip Up and Angle Adjustable |
| Armrests | *Flip Back Adjustable Height<br>Fixed Non-Adjustable | *Flip Back Adjustable Height<br>Fixed Non-Adjustable |
| Armrest height range | 9"- 11" Adjustable | 9"- 11" Adjustable |
| Rear wheel | *22", 24" Spoke<br>22", 24" Composite Mag | *22", 24" Spoke<br>22", 24" Composite Mag |
| Front wheel | 5", 6", *7", 8" Castors | 5", 6", *7", 8" Castors |
# Indications for Use:
The standard IKON 40 is a manual wheelchair device that is intended to brovide mobility to persons ages 12 and over (adolescents and adults) with a weight capacity of 300 lbs. The device is a means of mobility for persons limited to a sitting position.
{5}------------------------------------------------
# Technological Characteristics Comparison to the Predicate Device:
The following device characteristics comparison demonstrates that the IKON 40 is substantially equivalent in design, construction, operational uses and functionality for persons limited to a sitting position as the previously cleared MyOn HC™ Manual Wheelchair (K152536).
| Characteristics | IKON 40 | MyOn HC™ (K152536) |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(IFU) | The IKON 40 is a manual wheelchair device<br>that is intended to be used to provide<br>mobility to persons ages 12 and over<br>(adolescents and adults) with a weight<br>capacity of 300 lbs. The device is to be used<br>as a means of mobility for persons limited to<br>a sitting position. | The MyOn HC™ Manual Wheelchair is<br>intended to provide mobility to persons<br>over 12 years of age (adolescents and<br>adults) with a weight capacity of 220 & 290<br>lbs. depending on the seat width. The<br>device is indicated to provide mobility to<br>persons restricted to a sitting position. |
| Cross brace bars | 1 x 1 Aluminum Folding Frame | 1 x 1 Aluminum Folding Frame<br>2 x 1 Aluminum Folding Frame |
| Folding Method | Collapsible Cross-Brace | Collapsible Cross-Brace |
| Warranty | Lifetime on Frame | Lifetime on Frame |
| Seat widths | 16", 18", 20", 22" | 15", 16", 17", 18",19", 20", 22", 24" |
| Seat depths | 16"- 22"<br>Adjustable | 16"- 20"<br>Adjustable |
| Max user weight | 300 lbs. (1 x 1)<br>300 lbs. (2 x 1) | 275 lbs. (1 x 1)<br>350 lbs. (2 x 1) |
| Wheelchair weight<br>(Including legrests,<br>armrests and anti-<br>tippers) | 34 lbs. | 31 lbs. |
| Seat height | 14", 15", 16", 17", 18", 19"<br>Adjustable | 14", 15", 16", 17", 18",19", 20" Adjustable |
| Legrests | Swing Away<br>Elevating | Swing Away<br>Elevating |
| Legrest range | 13.5"- 19" Adjustable | 13.25"- 18.25" Adjustable |
| Legrest angle (°) | 70° | 70°, 80°, 90° |
| Footrests | Flip Up and Angle Adjustable | Flip Up and Angle Adjustable |
| Armrests | Flip Back Adjustable Height<br>Fixed Non-Adjustable | Flip Back Adjustable Height<br>T Arm Adjustable Height |
| Armrest height range | 9"- 11" Adjustable | 9"- 11" Adjustable |
| Rear wheel | 22", 24" Spoke<br>22", 24" Composite Mag | 24", 25", 26" - Spoke<br>20", 22", 24" - Composite Mag<br>20", 24", 25", 26" - Performance |
| Front wheel | 5", 6", 7", 8" Castors | 5", 6", 8" Castors |
| Anti-tippers | Yes, Optional | Yes, Optional |
| Standards - Strength,<br>Stability and Durability | ANSI / RESNA Wheelchair Standard Vol. 1<br>Section 1,3,5,7,8,15 &16 | ANSI / RESNA Wheelchair Standard Vol. 1<br>Section 1,3,5,7,8,15 &16 |
| Standards — FR | CAL 117 / ISO 8191 | CAL 117 / ISO 8191 |
{6}------------------------------------------------
### Performance Testing Data:
### Non-Clinical Testing:
Testing was performed by a non-clinical laboratory on the IKON 40 to determine substantial equivalence:
- ANSI / RESNA - ISO 7176-1:2014 Wheelchairs - Section 1: Determination of static stability
- ANSI / RESNA ISO 7176-3:2012 Wheelchairs Section 3: Determination of effectiveness of brakes ●
- ANSI / RESNA - ISO 7176-5:2008 Wheelchairs - Section 5: Determination of dimensions, mass and maneuvering space
- ANSI / RESNA - ISO 7176-7:1998 Wheelchairs — Section 7: Measurement of seating and wheel dimensions
- ANSI / RESNA - ISO 7176-8:2014 Wheelchairs - Section 8: Requirements and test methods for static, impact and fatigue strengths
- ANSI / RESNA - ISO 7176-11 Wheelchairs- Section 11: Test Dummies
- ANSI / RESNA – ISO 7176-13 Wheelchairs- Section 13: Determination of coefficient of friction
- ANSI / RESNA - ISO 7176-15:1996 Wheelchairs — Section 15: Requirements for information disclosure, documentation and labelling
- ANSI / RESNA WC/Volume 1 2009, Section16: Resistance to Ignition of Upholstered Parts .
- CAL117:2013, Section 1: Flammability Testing
- ISO 8191-1:1987 & 8191-2:1988: Flammability Testing
Animal Study:
Animal testing was not required for this submission
Clinical Testing:
Clinical testing was not required for this submission
#### Conclusions:
The IKON 40, this 510(k) subject device, has the same intended use, safety and similar technological characteristics as the predicate device, the MyOn HC™. Non-clinical laboratory test results support the subject device and demonstrate the IKON 40 should perform as intended in the specified use conditions. Broda has determined that therefore, the IKON 40 Manual Wheelchair is substantially equivalent to the predicate device, MyOn HC™, identified above throughout this document.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.