K152536 · Invacare Corporation · IOR · Jan 5, 2016 · Physical Medicine
Device Facts
Record ID
K152536
Device Name
MyOn HC Manual Wheelchair
Applicant
Invacare Corporation
Product Code
IOR · Physical Medicine
Decision Date
Jan 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.
Device Story
MyOn HC™ is a foldable, manually operated, adjustable lightweight wheelchair. It provides mobility and support to users limited to a sitting position, ranging from limited to full-day usage. The device features a horizontal folding mechanism, adjustable center of gravity (5"), and stepless backrest angle adjustments (-15° to +15°) operable while the user is seated. It accommodates various riggings (swing-in/out, detachable, fixed 80°). Accessories include posture belts, tipper aids, air pumps, passive illumination, cane holders, and luggage carriers. The device is intended for use by adolescents and adults (ages 12+) with weight capacities of 220 or 290 lbs depending on configuration. It is operated by the user or a caregiver to facilitate mobility in various environments.
Clinical Evidence
No clinical data. Bench testing only. Testing included ANSI/RESNA WC/Volume 1 & 2 (2009) for static stability, dimensions, maneuvering space, seating dimensions, static/impact/fatigue strength, information disclosure, and ignition resistance. Flammability testing performed per CAL117:2013 and ISO 8191-1:1987/8191-2:1988.
Technological Characteristics
Manual, foldable, lightweight wheelchair. Features horizontal folding frame, adjustable center of gravity, and stepless backrest adjustment. Materials include standard wheelchair components. Testing conducted per ANSI/RESNA WC/Volume 1 & 2 (2009) and ISO 8191 standards. No electronic or software components.
Indications for Use
Indicated for adolescents and adults (ages 12+) with mobility limitations who are restricted to a sitting position and require a manual wheelchair for support and mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5, 2016
Invacare Corporation Elijah Wreh Regulatory Affairs Manager (Pre-Market) One Invacare Way Elyria, Ohio 44035
Re: K152536
Trade/Device Name: MyOn HCTM Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: December 4, 2015 Received: December 7, 2015
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152536
Device Name MyOn HCTM Manual Wheelchair
Indications for Use (Describe)
The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
| SUBMITTER: | Invacare Corporation<br>One Invacare Way<br>Elyria, OH 44035 |
|-----------------------|--------------------------------------------------------------------------------------------------|
| | Phone: (440) 329-6840<br>Email: <a href="mailto:ewreh@invacare.com">ewreh@invacare.com</a> |
| CONTACT PERSON: | Elijah N. Wreh<br>Regulatory Affairs Manager (Pre-Market) |
| Date Prepared: | January 5, 2016 |
| DEVICE | |
| Name of Device: | MyOn HCTM Manual Wheelchair |
| Common or Usual Name: | Wheelchair, mechanical |
| Classification Name: | Mechanical wheelchair 21 CFR §890.3850 |
| Regulatory Class: | I |
| Product Code: | IOR: Wheelchair, mechanical |
| PREDICATE DEVICE: | Action Patriot Manual Wheelchair (K930803)<br>No reference devices were used in this submission. |
### DEVICE DESCRIPTION
This Traditional [510(k)] submission is being supplied to the U.S. FDA to obtain authorization to market the MyOn HC™ Manual Wheelchair. The MyOn HC™ Manual Wheelchair is a foldable, manually operated fully adjustable lightweight wheelchair. It is indicated to provide mobility to persons which have limitations in mobility. It provides support and mobility to users that are seated in the wheelchair, for limited time up to permanent, full day usage. The design incorporates a horizontal folding mechanism connected to a left & right side frame on which many adjustments can be made to meet the individual user needs. The frame can be either equipped with swing in or swing out, detachable, riggings with an angle of 70°, 80° or the frame which has a fixed 80° front. It allows 5" of center of gravity (CG) adjustments and stepless backrest angle adjustments from -15° to + 15° which can be achieved whist the user is seated in the wheelchair. The subject device intended use is to provide mobility to persons ages 12 and up
{4}------------------------------------------------
(adolescents and adults) with a weight capacity of 220 & 2901bs depending on the seat width. There is no prior submission for the subject device.
The associated accessories include:
- Posture belt .
- Tipper aid
- Air pump .
- Passive illumination ●
- . Cane holder
- . Luggage carrier
### INDICATIONS FOR USE
The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The device comparison showed that the subject device is substantially equivalent in intended use, design and operational principles to the previously cleared Action Patriot Manual Wheelchair (K930803). The subject device is substantially equivalent to the predicate device in regards to intended use, design, materials, and operational principles to provide mobility to persons limited to a sitting position.
{5}------------------------------------------------
### Design Characteristics Comparison
| Design | Action Patriot Manual Wheelchair<br>Predicate Device<br>(K930803) | MyOn HCTM Manual Wheelchair<br>Subject Device<br>Pending Submission |
|---------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Weight Limit | 250Ibs | 220 & 290lbs* |
| Cross-brace Configuration | X-design | Folds lengthwise and backwards* |
| Seat Width | 14"-20" | 12"-21" * |
| Seat Depth | 16"& 18" | 14"-20" *<br>14"-16" adjustable*<br>16"-18" adjustable* |
| Back Style | Fixed or adjustable | Fixed or adjustable |
| Back Angle<br>Adjustment | -5 to +20 degrees in 5 degree increments | -15 to +15 degrees* |
| Back Height | 15-19", 16", 20" | 12"-20" |
| Arm Types | Flip-back arm rests | Flip-back arm rests, adjustable, desk length |
| Anti-Tippers | Optional | Optional |
| Wheel Locks | Push-to-Lock, Pull-to-Lock, Hill Holder | Push-to-Lock, Pull-to-Lock |
| Rear Wheel<br>Sizes | 20", 22", 24" | 22"-25" * |
| Caster Sizes | 6", 8" | 4"-7" * |
| Front<br>riggings | 60° straight<br>70° straight | Pivot inwards and outwards<br>70° straight<br>80° straight*<br>90° straight*<br>Elevate 0-80°* |
### PERFORMANCE DATA
### Non-Clinical Test
Non-clinical laboratory testing was performed on the subject MyOn HCTM Manual Wheelchair to
determine substantial equivalence. The following testing was performed:
- ANSI / RESNA WC/Volume 2 2009, Section 1: Determination of Static Stability .
- ANSI / RESNA WC/Volume 1 2009, Section 5: Determination of Dimensions, Mass ● and Maneuvering Space
- ANSI / RESNA WC/Volume 1 2009, Section 7: Measurement of Seating and Wheel ● Dimensions
- ANSI / RESNA WC/Volume 1 2009. Section 8: Static. Impact And Fatigue Strengths ● Tests
- ANSI / RESNA WC/Volume 1 2009, Section15: Requirements for Information . Disclosure, Documentation and Labeling
- ANSI / RESNA WC/Volume 1 2009, Section16: Resistance to Ignition of Upholstered ● Parts
- CAL117:2013, Section 1: Flammability Testing ●
- ISO 8191-1:1987 & 8191-2:1988: Flammability Testing ●
{6}------------------------------------------------
Testing demonstrated that the subject MyOn HC™M Manual Wheelchair is substantially equivalent to the marketed predicate device.
## Animal Study
Animal testing was not required for this submission.
# Clinical Testing
Clinical testing was not required for this submission.
# CONCLUSIONS
The subject device has the same intended use and similar technological characteristics as the predicate device. The non-clinical laboratory data support the safety of the subject MyOn HC™ Manual Wheelchair and demonstrate that the subject device should perform as intended in the specified use conditions. Therefore, the subject MyOn HCTM Manual Wheelchair is substantially equivalent to the predicate device identified throughout this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.