MariGen Wound Extra

K190528 · Kerecis Limited · KGN · Jul 10, 2019 · SU

Device Facts

Record IDK190528
Device NameMariGen Wound Extra
ApplicantKerecis Limited
Product CodeKGN · SU
Decision DateJul 10, 2019
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The subject device is indicated for the management of wounds including: - Partial and full-thickness wounds - Pressure ulcers - Venous ulcers - Chronic vascular ulcers - Diabetic ulcers - Trauma wounds (abrasions, lacerations, second-degree burns, skin tears) - Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence) - Draining wounds

Device Story

MariGen Wound Extra is a sterile, pliable, non-crosslinked wound dressing derived from processed cod fish dermal matrix (collagen). Supplied as solid, fenestrated, or meshed lyophilized sheets in sizes up to 20 x 30 cm. Used by clinicians to manage various wound types; applied directly to the wound bed. Acts as a scaffold to support healing; fully resorbable, requiring no removal. Provides a physical barrier and wound environment management. Benefits include support for tissue regeneration in chronic and acute wounds.

Clinical Evidence

No clinical data provided. Evidence consists of biocompatibility testing, elemental impurities/chemical residual analysis, and bench testing. Safety and effectiveness supported by data from the identical predicate device and animal studies.

Technological Characteristics

Material: Cod fish dermal matrix (collagen). Form: Lyophilized sheet (solid, fenestrated, or meshed). Dimensions: Up to 20 x 30 cm. Sterilization: Ethylene Oxide (SAL 10^-6). Resorbable. Single-use. No electronic components or software.

Indications for Use

Indicated for management of partial/full-thickness wounds, pressure ulcers, venous/chronic vascular ulcers, diabetic ulcers, trauma wounds (abrasions, lacerations, 2nd-degree burns, skin tears), surgical wounds (donor sites/grafts, post-Mohs/laser surgery, podiatric, dehiscence), and draining wounds.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ July 10. 2019 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Kerecis Limited Gudmundur Sigurjonsson CEO Eyrargata 2 400 Isafjordur, Iceland Re: K190528 Trade/Device Name: MariGen Wound Extra Regulatory Class: Unclassified Product Code: KGN Dated: March 29, 2019 Received: April 5, 2019 Dear Gudmundur Sigurjonsson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) {1}------------------------------------------------ regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Cynthia Chang Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190528 Device Name MariGen Wound Extra Indications for Use (Describe) MariGen Wound Extra is indicated for the management of wounds, including: - Partial and full-thickness wounds - Pressure ulcers - Venous ulcers - Chronic vascular ulcers - Diabetic ulcers - Trauma wounds (second degree burn, abrasions, lacerations, skin tears), - Surgical wounds (donor sites/grafts, post-Mohs surgery, post laser surgery, podiatric, wound dehiscence), - Draining wounds. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510k Summary – K190528 # SUBMITTER/510K HOLDER | Company Address: | Kerecis Limited<br>Eyrargata 2<br>400 Isafjordur<br>Iceland | |------------------|-------------------------------------------------------------| | Contact Person: | Gudmundur Fertram Sigurjonsson<br>Executive Chairman | | Telephone: | 011 354 562 2601 | | Date Prepared: | February 20th, 2019 | # DEVICE NAME | Proprietary Name: | MariGen Wound Extra | |--------------------|---------------------------| | Common/Usual Name: | Collagen, Wound, Dressing | | Class: | Unclassified | | Product code: | KGN | # PREDICATE DEVICE MariGen Wound (K132343), predicate device SecureMesh (K153364), reference device # DEVICE DESCRIPTION The subject device is processed fish dermal matrix composed of fish collagen and is supplied as a sterile intact, or meshed sheet ranging in sizes up to 20 x 30 cm. The subject device is obtained from fish skin via standardized controlled GMP manufacturing process and supplied in terminally sterile packaging. The subject device is biocompatible, pliable, and non-cross linked. The device is intended for single use only. {4}------------------------------------------------ ### INTENDED USE The subject device is indicated for the management of wounds including: - Partial and full-thickness wounds - Pressure ulcers - Venous ulcers - Chronic vascular ulcers - Diabetic ulcers - Trauma wounds (abrasions, lacerations, second-degree burns, skin tears) - Surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence) - Draining wounds # TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE MariGen Wound Extra and the predicate device are made of the same material and share the same functional design, principles of operation, and user safety and efficacy. The subject device is substantially equivalent to the legally marketed predicate device with respect to: Indications for use, material composition, device characteristics, manufacturing processes, packaging material, and shelf life. The subject device performs as well as its predicate device and is offered in a greater variety of sizes, including larger sizes. Both devices can be cut to defect size and shape for better fit. In addition, the subject device shares the same safety as the implantable reference device K153364 SecureMesh, which is made of the same fish skin material (the "MariGen Material"). All devices are fully resorbable and do not require a second intervention for removal. | Device<br>Features | Subject Device:<br>MariGen Wound Extra | Predicate Device:<br>MariGen Wound | SUBSTANTIAL EQUAIVLENCE | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------| | 510(k) Number | K190528 | K132343 | | | Manufacturer | Kerecis Limited | Kerecis Limited | | | Product Codes | KGN | KGN | Equivalent | | Intend. Use | Wound Dressing | Wound Dressing | Equivalent | | Indications | Management of wounds<br>including:<br>- Partial and full-thickness<br>wounds<br>- Pressure ulcers<br>- Venous ulcers<br>- Chronic vascular ulcers | Management of wounds<br>including:<br>- Partial and full-thickness<br>wounds<br>- Pressure ulcers<br>- Venous ulcers<br>- Chronic vascular ulcers | Equivalent | #### Summary Table of Substantial Equivalence {5}------------------------------------------------ | | - Diabetic ulcers | - Diabetic ulcers | | |---------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------------|----------------------------------------------------------------------------| | | - Trauma wounds (abrasions, | - Trauma wounds (abrasions, | | | | lacerations, second-degree | lacerations, second-degree | | | | burns, skin tears) | burns, skin tears) | | | | - Surgical wounds (donor | - Surgical wounds (donor | | | | sites/grafts, post-Mohs | sites/grafts, post-Mohs | | | | surgery, post-laser surgery, | surgery, post-laser surgery, | | | | podiatric, wound dehiscence) | podiatric, wound dehiscence) | | | | - Draining wounds | - Draining wounds | | | Resource<br>origin | Cod Fish skin | Cod Fish skin | Equivalent | | Tissue<br>resource /<br>Scaffold base | Identical fish skin tissue | Identical fish skin tissue | Equivalent | | Nominal sizes | Sized up to<br>20 x 30 cm ( $600 cm²$ ) | 3 x 3.5 cm ( $10.5 cm²$ )<br>3 x 7 cm ( $21 cm²$ )<br>7 x 10 cm ( $70 cm²$ ) | Equivalent,<br>Subject device has more<br>size variety, incl. larger sizes | | Presentation | Solid, fenestrated, or meshed<br>lyophilized skin sheet in a peel<br>pouch | Solid or meshed lyophilized<br>skin sheet in a peel pouch | Equivalent | | Sterilization | Ethylene Oxide SAL 10-6,<br>single use only | Ethylene Oxide SAL 10-6,<br>single use only | Equivalent | | Shelf life | 3 years | 3 years | Equivalent | # PERFORMANCE DATA The subject device is a size modified version of the predicate device. Both are made from the MariGen Material. Biocompatibility tests and product characterization studies performed on the predicate device fully apply to the subject device as those are identical materials. Present performance data includes data from: elemental impurities and chemical residual analysis. The subject device is identical to the predicate device apart from being offered in sizes up to 600cm². The subject device does not raise new safety questions for patients. Kerecis's risk-based approach, executed according to EN ISO 14971: Medical devices - Application of risk management to medical devices, shows that all sizes of the modified device are safe for use. The provided data from biocompatibility testing combined with bench testing and animal studies exhibits product safety and effectiveness. # CONCLUSION The subject device is identical to the predicate device apart from being offered in sizes up to 600cm². {6}------------------------------------------------ Based on the data provided within this submission, the subject device is substantially equivalent to the predicate device with regards to intended use and indications for use, technological characteristics including principles of operation, and performance characteristics and device safety. It is concluded that the subject device is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...