Zirkonzahn COLOUR LIQUID is used for coloring pre-sintered zirconia structures.
Device Story
Zirkonzahn COLOUR LIQUID consists of aqueous solutions of transition and lanthanoid metal salts used to color milled, pre-sintered zirconia dental structures. Used by specialized dental technicians in a laboratory setting; application via dipping or metal-free brush. After coloring, structures are dried and sintered at temperatures above 1400 °C. The device provides various VITA and intensive shades to achieve desired aesthetic outcomes for dental prosthetics. Benefits include improved aesthetic matching of zirconia restorations to natural dentition.
Clinical Evidence
Bench testing only. Testing evaluated physical properties of colored zirconia versus untreated zirconia to ensure no negative impact on functionality. Biocompatibility established per ISO 10993-1:2010 via testing of representative and worst-case mixtures. Shelf life and storage conditions validated through real-time aging protocols evaluating color, workability, and appearance.
Technological Characteristics
Aqueous solutions of transition and lanthanoid metal salts. Available in 20 ml, 50 ml, and 100 ml bottles. Application via manual dipping or brush. Sintering required at >1400 °C. Biocompatibility evaluated per ISO 10993-1:2010.
Indications for Use
Indicated for coloring pre-sintered zirconia structures in dental prosthetics by specialized dental technicians.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
ArgenZ Liquid Shade, Argen Z Incisal Effect, ArgenZ Color Modifier, ArgenZ Pontic Reducer (K182833)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and consists of the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
November 22, 2019
Zirkonzahn srl Sandra Leitner Regulatory Affairs Via An der Ahr 7 Gais, 39030 ITALY
Re: K190518
Trade/Device Name: Color Liquid, Vita Liquid Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: October 21, 2019 Received: October 24, 2019
Dear Sandra Leitner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190518
Device Name Zirkonzahn COLOUR LIQUID
| Indications for Use (Describe) | Zirkonzahn COLOUR LIQUID is used for coloring pre-sintered zirconia structures. |
|--------------------------------|---------------------------------------------------------------------------------|
|--------------------------------|---------------------------------------------------------------------------------|
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div style="display:inline-block;"> <input checked="true" type="checkbox"/> </div> Registration Use (Part 21 CFR 201.2 Subpart D) </td><td><div style="display:inline-block;"> <input type="checkbox"/> </div> Over-The-Counter Use (21 CFR 201.2 Subpart C) </td></tr></table> | <div style="display:inline-block;"> <input checked="true" type="checkbox"/> </div> Registration Use (Part 21 CFR 201.2 Subpart D) | <div style="display:inline-block;"> <input type="checkbox"/> </div> Over-The-Counter Use (21 CFR 201.2 Subpart C) |
| <div style="display:inline-block;"> <input checked="true" type="checkbox"/> </div> Registration Use (Part 21 CFR 201.2 Subpart D) | <div style="display:inline-block;"> <input type="checkbox"/> </div> Over-The-Counter Use (21 CFR 201.2 Subpart C) | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 05 510(k) Summary
#### K190518
#### 510 (k) SUMMARY
### APPLICANT
ZIRKONZAHN SRL Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 660 Fax: +39 0474 066 661 E-mail: info@zirkonzahn.com
#### CONTACT PERSON
Sandra Leitner Regulatory Affairs ZIRKONZAHN SRL Via An der Ahr 7 Gais, ITALY 39030 Phone: +39 0474 066 784 Fax: +39 0474 066 661 E-mail: sandra.leitner@zirkonzahn.com
### Date Summary Prepared: November 22, 2019
### DEVICE IDENTIFICATION
| Trade/Proprietary Name: | Zirkonzahn COLOUR LIQUID |
|-------------------------|---------------------------------------|
| Generic / Common Name: | Liquid for Dental Zirconia Prosthesis |
| Regulation Number: | 872.6660 |
| Classification Name: | Porcelain powder for clinical use |
| Class: | II |
| Product Code: | EIH |
| Panel: | Dental |
{4}------------------------------------------------
### LEGALLY MARKETED PREDICATE DEVICE
| Company: | The Argen Corporation |
|----------------|----------------------------------------------------------------------------------------------|
| Device Name: | ArgenZ Liquid Shade, Argen Z Incisal Effect, ArgenZ Color<br>Modifier, ArgenZ Pontic Reducer |
| Product Code: | EIH |
| 510(k) Number: | K182833 |
## INDICATIONS FOR USE
Zirkonzahn COLOUR LIQUID is used for coloring pre-sintered zirconia structures.
## DEVICE DESCRIPTION
Zirkonzahn COLOUR LIQUID devices are used for coloring milled and finished presintered zirconia structures. Coloring is performed by dipping the framework into the liquid or by using a metal-free brush. Afterwards, the structure is dried and subsequently sintered at temperatures above 1400 ℃.
The Zirkonzahn COLOUR LIQUIDS are available in different colors and in different bottle sizes (20 ml, 50 ml, 100 ml).
| VITA tooth shades | A1, A2, A3, A3,5, A4<br>B1, B2, B3, B4<br>C1, C2, C3, C4<br>D2, D3, D4<br>1M1, 1M2<br>2L1.5, 2M2<br>3M2, 3M3<br>4M2 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| VITA Bleach shades | 0M1, 0M2, 0M3 |
| ZZ Intensive shades | Tissue A, Tissue B, Tissue C<br>Orange 1, Brown 2<br>Incisal grey, Incisal violet, Incisal blue<br>LuNe<br>A-Cervical, B-Cervical, C-Cervical, D-Cervical |
| ZZ Special liquids | Translucent<br>Opaque<br>Fluorescence |
The following colors are included in this submission:
The devices are used in the dental field by specialized technicians only.
{5}------------------------------------------------
## DISCUSSION OF NON CLINICAL TESTS
Non-clinical testing was performed to evaluate the physical properties of colored zirconia compared to untreated zirconia, to demonstrate that the subject device does not negatively impact the functionality of the zirconia. The results show that the devices are well suited for their intended use.
Furthermore, to establish the shelf life and storage conditions of the subject devices, real time aging was performed (including evaluation of color, workability and appearance) according to Zirkonzahn internal test protocols.
Biocompatibility was established in consideration of the International Standard 10993-1:2010 'Biological Evaluation of Medical Devices' Part 1: Evaluation and Testing. Zirkonzahn performed biocompatibility testing for representative mixtures and Worst Case mixtures of the subject devices or established biocompatibility through evaluation of the respective substances. Therefore Zirkonzahn divided the Colour Liquid devices into 6 different groups:
| Colour Liquid groups | Biocompatibility |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Main group | Tested on representative mixtures:<br>- "lamelle COLOUR LIQUID"<br>- "Campioni realizzati in polvere di zirconia<br>(Prettau Zirkon) infiltrate con un liquido”<br>Tested on Worst Cases:<br>- Worst Case 1<br>- Worst Case 2<br>- Worst Case 3 |
| Colour Liquid Aquarell Bio-<br>Pigment | Evaluated |
| Colour Liquid Prettau<br>Aquarell - Translucent | Evaluated |
| Colour Liquid Prettau<br>Aquarell - Opaque | Evaluated |
| Thinner | Evaluated |
| Stabilizer | Evaluated |
{6}------------------------------------------------
# SUBSTANTIAL EQUIVALENCE
| | Zirkonzahn<br>COLOUR LIQUID<br>(new devices) | ArgenZ Liquids<br>(K182833) | Comparison |
|------------------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Company | Zirkonzahn srl | The Argen Corporation | |
| Product Code | EIH | EIH | Same |
| Regulation<br>Number | 872.6660 | 872.6660 | Same |
| Regulation<br>Name | Porcelain powder for<br>clinical use | Porcelain powder for<br>clinical use | Same |
| Indications for<br>use | Zirkonzahn<br>COLOUR<br>LIQUID is used for<br>coloring pre-sintered<br>zirconia structures. | ArgenZ Liquids are<br>intended to be used<br>by trained dental<br>technicians as an<br>accessory for<br>shading ArgenZ<br>frameworks and<br>ArgenZ allzirconia,<br>monolithic<br>restorations for<br>anterior and<br>posterior dental<br>prosthetics. | Similar |
| Chemical<br>Composition | Aqueous solutions of<br>transition and<br>lanthanoid metal salts | Aqueous solutions of<br>transition and<br>lanthanoid metal<br>salts | Similar |
| Biocompatibility | Established | Established | Same |
| Principles of<br>Operation | Brushing or<br>Dipping<br>Technique | Brushing or Dipping<br>Technique | Same |
| Type of<br>Packaging and<br>Volume | Bottle; 20 ml,<br>50 ml,<br>100 ml | Bottle; 30 ml, 100 ml | Similar |
| Shade | Various, including A1-<br>D4 | Various, including<br>A1-D4 | Similar |
| Storage<br>Conditions | Max. 25 °C<br>5 years | 25 - 75 °C<br>2 years | Minor<br>difference |
The table above provides a detailed comparison of the submitted devices and the predicate devices. Both devices are liquids used for coloring pre-sintered zirconia structures. They only slightly differ in their storage conditions. However, Zirkonzahn evaluated the storage conditions for the Zirkonzahn COLOUR LIQUID devices with real life functional testing and concluded that these indications are appropriate. Therefore, Zirkonzahn concludes that this difference can be rated as minor.
{7}------------------------------------------------
# CONCLUSION
Based on the available information, the new devices and the predicates are similar in function, production technology and intended use. Therefore, we conclude that the proposed devices are substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.