Reagents for the measurement of transferrin concentration in human serum or plasma. The obtained values are useful as an aid in the diagnosis of iron deficiency anemia and hemochromatosis. This reagent is for use in the BioSystems BA analyzers.
Device Story
Transferrin assay uses turbidimetric principle to measure transferrin concentration in human serum or plasma samples. Input consists of patient sample mixed with goat anti-human transferrin antibodies; antigen-antibody complexes precipitate, causing light scattering proportional to transferrin concentration. Device is intended for use on BioSystems BA400 analyzers in clinical laboratory settings. Healthcare providers use quantitative output to assist in diagnosing iron deficiency anemia and hemochromatosis. System requires external protein calibrators for curve preparation.
Clinical Evidence
No clinical trials performed. Evidence consists of analytical bench testing: precision (CLSI EP05-A3), linearity (CLSI EP06-A), interference (CLSI EP07-A3), and detection limits (CLSI EP17-A2). Method comparison against predicate using 84 serum and 69 Li-heparin plasma samples showed high correlation (R² 0.993 and 0.997 respectively). Matrix equivalency confirmed with 40 paired samples.
Indicated for the quantitative measurement of transferrin in human serum or plasma to aid in the diagnosis of iron deficiency anemia and hemochromatosis. For prescription use only.
Regulatory Classification
Identification
A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
I Background Information:
A 510(k) Number
K190495
B Applicant
BioSystems S.A.
C Proprietary and Established Names
Transferrin
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DDG | Class II | 21 CFR 866.5880 - Transferrin immunological test system | IM - Immunology |
II Submission/Device Overview:
A Purpose for Submission:
New Device
B Measurand:
Transferrin
C Type of Test:
Quantitative, Turbidimetry
III Intended Use/Indications for Use:
K190495 - Page 1 of 7
{1}
K190495 - Page 2 of 7
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Reagents for the measurement of transferrin concentration in human serum or plasma. The obtained values are useful as an aid in the diagnosis of iron deficiency anemia and hemochromatosis. This reagent is for use in the BioSystems BA analyzers.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
BioSystems BA400 analyzer
IV Device/System Characteristics:
A Device Description:
The Transferrin assay consists of the following reagents:
A Reagent: 60 mL/vial, imidazole buffer 0.05 mol/L, sodium azide 0.95 g/L, pH 7.5.
B Reagent: 15 mL/vial, goat anti-human transferrin antibodies, sodium azide 0.95 g/L
Additional materials required but not provided are protein calibrators (BioSystems cod. 31075). The set contains 5 different levels and it should be used to prepare the calibration curve. The calibrators are supplied ready to use.
These reagents are intended to be used with the BioSystems BA analyzers.
B Principle of Operation:
The Transferrin reagent is a turbidimetric assay in which the transferrin in the sample precipitates in the presence of anti-human transferrin antibodies. The light scattering of the antigen-antibody complexes is proportional to the transferrin concentration and can be measured by turbidimetry.
V Substantial Equivalence Information:
A Predicate Device Name(s):
Tina-quant Transferrin Ver.2
B Predicate 510(k) Number(s):
K012393
C Comparison with Predicate(s):
{2}
| Device & Predicate Device(s): | K190495 Device | K012393 Predicate |
| --- | --- | --- |
| Device Trade Name | Transferrin | TRSF2 Tina-quant Transferrin ver.2 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Reagents for the measurement of transferrin concentration in human serum or plasma. The obtained values are useful as an aid in the diagnosis of iron deficiency anemia and hemochromatosis. This reagent is for use in the BioSystems BA analyzers. | In vitro test for the quantitative determination of transferrin in human serum and plasma on Roche/Hitachi cobas c systems. |
| Measurand | Transferrin | Same |
| Method | Immunological agglutination | Same |
| Assay type | Quantitative | Same |
| Traceability | ERM-DA470k/IFCC | Same |
| Sample Type | Serum and Li-heparin plasma | Same |
| General Device Characteristic Differences | | |
| Analyzer | BioSystems BA400 | Cobas c 311, 501/502 |
| Antibody soruce | Goat anti-human transferrin | Rabbit anti-human transferrin |
| Measuring range | 12 – 520 mg/dL | 10 – 520 mg/dL |
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Third Edition
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition
CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Third Edition
CLSI EP07-A3 Interference Testing in Clinical Chemistry
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
K190495 - Page 3 of 7
{3}
Precision of the Ferritin assay was evaluated according to CLSI EP05-A3. Five different levels of human serum patient samples were used to carry out precision, with concentrations of 20.4 mg/dL, 123 mg/dL, 225 mg/dL, 402 mg/dL and 466 mg/dL. The study was conducted over 20 days, with two runs per day and two replicates per specimen, for a total of 80 measurements. The same five samples were used to assess lot-to-lot and instrument-to-instrument imprecision. For the lot-to-lot imprecision assessment, the study was conducted over five days, with two runs per day and with two replicates per one, using three reagent lots and one analyzer. For the instrument-to-instrument imprecisions assessment, the study was conducted over five days, with two runs per day and with two replicates per run, using three analyzers and one reagent lot. The results are summarized in the table below:
| Sample | Mean (mg/dL) | Within-Run | | Between-Run | | Between-Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| 1 | 20.4 | 0.52 | 2.6 | 1.3 | 6.2 | 1.4 | 6.7 | 0.73 | 3.6 |
| 2 | 123 | 1.1 | 0.9 | 2.3 | 1.9 | 2.6 | 2.1 | 1.3 | 1.1 |
| 3 | 225 | 1.5 | 0.7 | 4.5 | 2.0 | 4.7 | 2.1 | 2.9 | 1.3 |
| 4 | 402 | 7.2 | 1.8 | 8.9 | 2.2 | 11.4 | 2.8 | 7.4 | 1.8 |
| 5 | 466 | 8.1 | 1.7 | 9.3 | 2.0 | 12.3 | 2.7 | 7.0 | 1.5 |
| Sample | Mean (mg/dL) | Between-lot | | Between-Instrument | |
| --- | --- | --- | --- | --- | --- |
| | | SD | CV (%) | SD | CV (%) |
| 1 | 20.4 | 2.2 | 6.6 | 1.2 | 6.3 |
| 2 | 123 | 4.2 | 7.2 | 1.5 | 1.3 |
| 3 | 225 | 6.2 | 4.3 | 7.7 | 3.6 |
| 4 | 402 | 7.1 | 2.3 | 19.2 | 5.2 |
| 5 | 466 | 11.6 | 2.4 | 22.0 | 5.0 |
2. Linearity:
Linearity was evaluated according to CLSI guideline EP6-A. A dilution series with nine levels was prepared by diluting a high sample with a concentration of 641 mg/dL with the low concentration sample of 0 mg/dL to cover the measuring range. The results are summarized as follows:
| Analyte | Test Range | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- | --- |
| Transferrin | 0–641 mg/dL | 1.01 (0.999–1.029) | -0.83 (-5.50–3.84) | 0.999 |
3. Analytical Specificity/Interference:
An interference study was performed in accordance with CLSI EP07-A3 guideline. Two human serum samples with two different analyte concentrations (178 mg/dL and 425 mg/dL) were spiked with increasing concentrations of potential endogenous and exogenous interferents and analyzed in duplicate.
K190495 - Page 4 of 7
{4}
No significant assay interference effects were observed in serum when tested with hemoglobin (500 mg/dL), unconjugated bilirubin (30 mg/dL), conjugated bilirubin (30 mg/dL), triglycerides (1625 mg/dL), rheumatoid factor (300 IU/mL), acetaminophen (20 mg/dL), ascorbic acid (30 mg/dL), N-acetyl-L-cysteine (10 mg/L), heparin (3000 U/L), deferoxamine (250 mg/L), dobesilate (200 mg/L), ampicillin (152 μmol/L), levodopa (20 mg/L), acetylsalicylic acid (3.62 mmol/L), ferrous sulfate (1.0 mmol/L), rifampicin (60 mg/L), EDTA (3.4 μmol/L), valproic acid (3467 μmol/L), phenylbutazone (400 mg/L), theophylline (222 μmol/L), cholesterol (13 mmol/L), metronidazole (701 μmol/L), prednisone (0.84 μmol/L), and methotrexate (2 mmol/L).
## Antigen Excess/Prozone Effect:
A study was performed to verify the prozone performance on the BioSystems BA400 analyzer. A total of 11 samples with different analyte concentrations were evaluated. No prozone effect was observed up to 3000 mg/dL.
## 4. Assay Reportable Range:
Transferrin: 12–520 mg/dL
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Traceability:
The calibration of the assay is traceable to ERM-DA470k/IFCC.
### Stability:
Real-time stability testing was conducted using human serum level samples with concentrations of 200 mg/dL and 400 mg/dL. Eight time points were tested, 0, 1, 3, 6, 10, 14, 20, 26 and 41 months. The results from this study support a shelf-life of 26 months when stored at the recommended storage temperature of 2–8°C.
Real-time stability study was carried out with open-vial reagents. Therefore, the results support the open-vial reagent shelf-life of 26 months when stored at the recommended storage temperature of 2–8°C.
On-board stability study supports reagent stability of up to 2 months when kept in the refrigerated compartment of the analyzer.
## 6. Detection Limit:
Detection capabilities studies for each analyte were evaluated based upon CLSI EP17-A2 guideline.
Limit of blank (LoB) study was performed by testing five blank samples comprised of albumin matrix. Samples were tested in replicates of four per run, one run per day for three days using two reagent lots, for a total of 60 observations per reagent lot, and for a total of 120 observations.
K190495 - Page 5 of 7
{5}
Limit of detection (LoD) study was performed by testing five native serum samples with analyte concentrations close to expected detection limit for each analyte. Samples were tested in replicates of four per run, one run per day for three days using two reagent lots, for a total of 60 observations per reagent lot, and for a total of 120 observations.
Limit of Quantitation (LoQ) study was performed using five native serum samples. Samples were tested in replicates of four per run, one run per day for three days using two reagent lots, for a total of 60 observations per reagent lot, and for a total of 120 observations.
The results are as follows:
| Detection Limit Parameter | Transferrin (mg/dL) |
| --- | --- |
| LoB | 0.49 |
| LoD | 1.18 |
| LoQ | 12.0 |
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
A total of 84 human serum samples with transferrin concentrations across the assay measuring range were tested compared to the predicate device. Another study was conducted using a total of 69 human Li-heparin plasma samples with transferrin concentration covering the measuring range compared to the predicate device.
Passing Bablok regression analysis was performed to evaluate the analytical equivalence of the transferrin assay when compared to the predicate. The results are summarized as follows:
| Matrix | N | Range | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- | --- | --- |
| Serum | 84 | 75.4–504 | 0.95 (0.92–0.98) | 7.91 (1.48–14.33) | 0.993 |
| Li-heparin plasma | 69 | 61.1–494.0 | 0.97 (0.95–0.99) | -0.01 (-4.81–4.31) | 0.997 |
2. Matrix Comparison:
Matrix comparison study was carried out using 40 paired specimens of serum and lithium-heparin plasma, across the assay measuring range. The results of the Passing-Bablok regression analysis support the equivalency of the two matrices, and are as follows:
| N | Slope (95% CI) | Intercept (95% CI) | R² |
| --- | --- | --- | --- |
| 40 | 0.94 (0.93–0.95) | 2.18 (0.42–5.63) | 0.999 |
K190495 - Page 6 of 7
{6}
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
Transferrin (serum or plasma): 200–360 mg/dL
The reference range was verified by testing 20 apparently healthy subjects.
This range is given for orientation only, and each lab should establish their own reference range.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K190495 - Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.