The cassette COBAS Integra Tina-quant Transferrin ver.2 contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative immunological determination of human transferrin in serum and plasma. A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum and plasma. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
Device Story
Tina-quant Transferrin ver.2 is an in vitro diagnostic reagent system for use on COBAS Integra clinical chemistry analyzers. It utilizes immunological agglutination; human transferrin in patient serum or plasma samples forms a precipitate with specific antiserum. This precipitate is measured turbidimetrically at 340 nm. The system provides quantitative results to clinicians to assist in diagnosing malnutrition, inflammation, infection, and iron deficiency anemia. It is intended for professional use in clinical laboratory settings. The device includes a 'High Activity' check to identify potential interference from IgM gammopathy (Waldenstroem), requiring post-dilution for accurate results.
Clinical Evidence
Bench testing only. Performance compared to predicate using clinical samples. Intra-assay precision CV 0.77–0.86%; between-day precision CV 1.8–1.9%. Analytical sensitivity (LDL) 0.013 g/L. Method comparison yielded y = 1.06x + 0.03 (r=0.996). Interference testing showed no significant impact from icterus, hemolysis, lipemia (up to 500 mg/dL), or rheumatoid factors.
Technological Characteristics
Immunoturbidimetric assay; utilizes specific antiserum for agglutination; measured at 340 nm. Standardized against CRM 470 (RPPHS). Designed for use on COBAS Integra clinical chemistry analyzers. Measuring range 1.3–520 mg/dL.
Indications for Use
Indicated for quantitative measurement of human transferrin in serum and plasma to aid in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders (e.g., iron deficiency anemia).
Regulatory Classification
Identification
A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
{0}------------------------------------------------
---
11 1
1
| 510(k) Summary | K012393 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. |
| Submitter name, address, contact | Roche Diagnostics Corporation<br>9115 Hague Road<br>Indianapolis, IN 46250<br>(317) 521 - 3831<br><br>Contact Person: Sherri L. Coenen<br><br>Date Prepared: July 25, 2001 |
| Device Name | Proprietary name: Tina-quant Transferrin ver.2<br><br>Common name: Transferrin<br><br>Classification name: Transferrin immunological test system |
| Device Description | The Tina-quant Transferrin ver.2 Assay is based on the principle of immunological agglutination. Human transferrin forms a precipitate with a specific antiserum which is determined turbidimetrically at 340 nm. |
| Intended use | The cassette COBAS Integra Tina-quant Transferrin ver.2 contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative immunological determination of human transferrin in serum. |
| Indications for Use | A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum and plasma. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia. |
.
{1}------------------------------------------------
## 510(k) Summary, Continued
Substantial Equivalence The Tina-quant Transferrin ver.2 is substantially equivalent to other devices legally marketed in the United States. We claim equivalence to the COBAS Integra Tina-quant Transferrin (K951595).
Substantial equivalence similarities
The following table compares the Tina-quant Transferrin ver.2 Assay with the predicate device.
| Feature | Tina-quant Transferrin<br>ver.2 | Transferrin |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The cassette COBAS<br>Integra Tina-quant<br>Trasferrin ver.2 (TRSF2)<br>contains an in vitro<br>diagnostic reagent system<br>intended for use on<br>COBAS Integra systems<br>for the quantitative<br>immunological<br>determination of human<br>transferrin in serum. | The cassette COBAS<br>Integra Transferrin<br>(TRSF) contains an in<br>vitro diagnostic reagent<br>system intended for use<br>on COBAS Integra<br>systems for the<br>quantitative<br>immunological<br>determination of human<br>transferrin in serum. |
| Indication for Use | A transferrin<br>immunological test system<br>is a device that consists of<br>the reagents used to<br>measure by<br>immunological<br>techniques the transferrin<br>(an iron-binding and<br>transporting serum protein)<br>in serum and plasma.<br>Measurement of<br>transferrin levels aids in<br>the diagnosis of<br>malnutrition,acute<br>inflammation, infection,<br>and red blood cell<br>disorders,such as iron<br>deficiency anemia. | A transferrin<br>immunological test<br>system is a device that<br>consists of the reagents<br>used to measure by<br>immunochemical<br>techniques the<br>transferrin (an iron-<br>binding and<br>transporting serum<br>protein) in serum,<br>plasma, and other body<br>fluids. Measurement of<br>transferrin levels aids in<br>the diagnosis of<br>malnutrition, acute<br>inflammation, infection,<br>and red blood cell<br>disorders, such as iron<br>deficiency anemia. |
{2}------------------------------------------------
## 510(k) Summary, Continued
Substantial equivalence similarities
The following table compares the Tina-quant Transferrin ver.2 Assay with the predicate device.
| Feature | Tina-quant Transferrin<br>ver.2 | Transferrin |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Assay Protocol | Immunoturbidimetric<br>assay | Immunoturbidimetric<br>assay |
| Instrument | COBAS Integra Clinical<br>Chemistry Analyzers | COBAS Integra<br>Clinical Chemistry<br>Analyzers |
| Traceability /<br>Standardization | Standardized against the<br>reference preparation<br>CRM 470, corresponding<br>to RPPHS (Reference<br>Preparation Protein in<br>Human Serum) | Standardized against<br>the reference<br>preparation CRM 470,<br>corresponding to<br>RPPHS (Reference<br>Preparation Protein in<br>Human Serum) |
#### Substantial equivalence differences
The following table compares the Tina-quant Transferrin ver.2 Assay with the predicate device.
| Feature | Tina-quant<br>Transferrin ver.2 | Transferrin |
|-----------------|---------------------------------|-----------------|
| Sample Type | Human serum and<br>plasma | Human serum |
| Measuring Range | 1.3 - 520 mg/dL | 80 – 1280 mg/dL |
{3}------------------------------------------------
Substantial equivalence – performance characteristics, cont.
The performance characteristics of the Tina-quant Transferrin ver.2 Assay and the predicate device are compared in the table below. .
| Feature | Tina-quant Transferrin<br>ver.2 | Transferrin |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Intra-assay<br>precision (% CV) | 0.86% at 1.35 g/L<br>0.77% at 3.36 g/L | 1.5% at 1.10 g/L<br>0.83% at 3.32 g/L |
| Between Day<br>Precision (% CV) | 1.8% at 1.32 g/L<br>1.9% at 3.70 g/L | 1.6% at 1.10 g/L<br>0.97% at 3.32 g/L |
| Limitations | Icterus: No significant interference Hemolysis: No significant interference Lipemia: No significant interference up to an Intralipid level of 500 mg/dL Rheumatoid factors: No significant interference Interference of Gammopathy type IgM (Waldenstroem) sera is recognized by the “High Activity” check. In case samples are flagged “High Act”, correct results can be obtained after post-dilution. | Icterus: No significant interference Hemolysis: No significant Lipemia: No significant interference Rheumatoid factors: No significant interference |
| Analytical<br>sensitivity (LDL) | 0.013 g/L | 0.58 g/L |
| Method<br>comparison | Tina-quant Transferrin ver.2 (Y) / COBAS Integra Transferrin (X)<br>$y = 1.06x + 0.03$ $r = 0.996$ | Transferrin (Y) / nephelometric determination (X)<br>$y = 1.06x + 0.01 g/L$ $r = 0.958$ |
{4}------------------------------------------------
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
and the comments of the country of
:
:
Substantial equivalence performance characteristics, cont.
.
:
:
:
. ·
:
:
.
:
The performance characteristics of the Tina-quant Transferrin The porcennance the predicate device are compared in the table below.
.
11:11:1
.
| Feature | Tina-quant Transferrin<br>ver.2 | Transferrin |
|--------------------------|---------------------------------|---------------------------------|
| Calibration<br>frequency | • after reagent lot change | • after reagent lot change |
| Expected values | 2.0 – 3.6 g/l (200 – 360 mg/dl) | 2.0 – 3.6 g/l (200 – 360 mg/dl) |
:
:
:
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
# SEP 1 9 2001
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Sherri L. Coenen Regulatory Submissions, Centralized Diagnostics Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, Indiana 46250-0457
K012393 Re:
> Trade Name: Roche Diagnostics Tina-quant® Transferrin ver.2 Regulatory Class: 21 CFR § 866.5880 Regulatory Class: II Product Code: DDG Dated: July 25, 2001 Received: July 27, 2001
Dear Ms. Coenen:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific darices), please contact the Office of Compliance at additionally 607.10 for in This angliestions on the promotion and advertising of your device, (201) 59 rttact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mormation on your responsibility on sumer Assistance at its toll-free number (800) 638-2041 or 1101) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use Statement
Roche Diagnostics Corp.
510(k) Number (if known): K012393
Device Name: Tina-quant Transferrin ver.2
Indications For Use:
The cassette COBAS Integra Tina-quant Transferrin ver.2 contains an in vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative immunological determination of human transferrin in serum and plasma. A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum and plasma. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Soumen S. Altare
Ivision Sian-Off Division of Clinical Laboratory Devices
510(k) Number K012393
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.