K180847 · Johnson & Johnson Consumer, Inc. · OLP · Jun 19, 2018 · General, Plastic Surgery
Device Facts
Record ID
K180847
Device Name
Neutrogena Light Therapy Acne Mask+
Applicant
Johnson & Johnson Consumer, Inc.
Product Code
OLP · General, Plastic Surgery
Decision Date
Jun 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neutrogena Light Therapy Acne Mask+ is intended to emit energy in the red and blue region of the spectrum, specifically indicated to treat mild to moderate acne on the face.
Device Story
The Neutrogena Light Therapy Acne Mask+ is an OTC wearable device for home use. It utilizes non-coherent light-emitting diode (LED) technology in the blue (425-450nm) and red (620-640nm) spectrums. Blue light targets acne-causing bacteria; red light reduces inflammation. The user wears the mask like glasses and initiates a 10-minute treatment session via an integrated button on a rechargeable module. The device features an automatic shut-off after 10 minutes. The module is recharged via micro-USB. The device provides a non-invasive therapeutic regimen for acne management, allowing patients to self-administer light therapy at home.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluation included electrical safety (IEC 60601-1), home healthcare environment requirements (IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), photobiological safety (IEC 62471), battery safety (IEC 62133-1), and biocompatibility (ISO 10993-1). A Human Factors evaluation was conducted per FDA guidance.
Technological Characteristics
Non-coherent LED light source (Blue 425-450nm, Red 620-640nm). Powered by rechargeable Ni-MH batteries. Form factor is a wearable mask with an integrated control module. Biocompatible medical-grade plastic applied parts. Connectivity via micro-USB for charging. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-57, IEC 62471, IEC 62133-1, ISO 10993-1.
Indications for Use
Indicated for patients with mild to moderate acne on the face.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 19, 2018
Johnson & Johnson Consumer Inc. Timothy Kline Associate Manager, Regulatory Affairs 7050 Camp Hill Rd Fort Washington, Pennsylvania 19034
Re: K180847
Trade/Device Name: Neutrogena Light Therapy Acne Mask+ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: March 30, 2018 Received: April 2, 2018
Dear Timothy Kline:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180847
#### Device Name
Neutrogena Light Therapy Acne Mask+
Indications for Use (Describe)
The Neutrogena Light Therapy Acne Mask+ is intended to emit energy in the red and blue region of the spectrum, specifically indicated to treat mild to moderate acne on the face.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### ട. 510(K) SUMMARY
# 510(k) Summary
### General Information
JOHNSON & JOHNSON CONSUMER INC. Skillman, NJ 08558 Contact Person: Timothy Kline Email: tkline3@its.jnj.com Telephone: 215-273-7241 FAX: 215-273-4123 Summary Preparation Date: March 16, 2018
### Device
Common Name: Acne Light Therapy System Classification Name: Over the Counter powered Light Based Laser for Acne Classification Regulation: 21 CFR 878.4810 Classification Panel: General and Plastic Surgery Product Code: OLP
### Predicate Devices:
Predicate Device(s): Pulsaderm Acne Mask 28 (K170111) Reference Device(s): Illumask Acne Light Therapy (K123999)
### Device Description:
The Light Therapy Acne Mask + uses a known light emitting diode (LED) therapy, noncoherent light, in the red and blue range to treat mild to moderate acne bearing similar methods of operation, treatment regimen, and the ability to be recharged as the predicate K170111. The reference device bears the same technological characteristics including anatomical design and biocompatible materials as the Light Therapy Acne Mask +.
Consistent with the intended use of the predicate and reference device, the Light Therapy Acne Mask + uses blue light to kill acne causing bacteria and the red light to decrease inflammation. The user wears the mask over their face, like a pair of glasses, and presses the button on the integrated module - holding for one second - to initiate the treatment. The device has a continuous working time of 10 minutes and shuts off automatically. Each use counts as a single dose. The module contains rechargeable batteries. The module is then recharged prior to the next treatment via a micro USB cable.
### Indications for Use:
The Light Therapy Acne Mask + is intended to emit energy in the red and blue region of the spectrum, specifically indicated to treat mild to moderate acne on the face.
### Technological Characteristics and substantial equivalence:
The Light Therapy Acne Mask+ share the same methods of operation, a similar treatment regimen, and the ability to be recharged as the predicate (K170111). The reference device
{4}------------------------------------------------
(K123999) is included because the Light Therapy Acne Mask+ shares the same anatomical position to support the scientific methodology of the Light Therapy Acne Mask+.
Table 1 shows that the subject device and the predicate device share similar output and technology characteristics
Table 1. Device Comparison Table
| Feature | Subject Device | Primary Predicate (K170111) |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Classification- Regulation | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Classification- Product Code | Primary: OLP - Laser surgical<br>instrument for use in general<br>and plastic surgery and in<br>dermatology. | Primary: OLP - Laser surgical<br>instrument for use in general<br>and plastic surgery and in<br>dermatology. |
| Indications for Use | Intended to emit energy in the<br>red and blue region of the<br>spectrum, specifically<br>indicated to treat mild to<br>moderate acne on the face | Intend for over the counter use<br>for the treatment of mild to<br>moderate acne |
| Environment of Use | Home | Home |
| Wavelength (nm) | Blue 425-450nm | Blue: 440+/-5nm |
| Wavelength (nm) | Red 620-640nm | Red: 630+/-5nm |
| Total 30-day Energy Dose (J/cm²) | 38.38 J/cm2 | 23.76 J/cm2 |
| Treatment Regimen | 10 minutes/day | 10 minutes/day |
| Power Source(s) | Ni-MH Batteries | Ni-MH / AA Batteries |
### Performance Data:
The bench performance tests demonstrated that the Light Therapy Acne Mask + is as safe, and as effective as the predicate device. Applicable recognized consensus standards included the following:
- . IEC 60601-1: 2005 + A1: 2012 – Medical Electrical Equipment part 1: General requirements for basic safety and essential performance
{5}------------------------------------------------
- . IEC 60601-1-11 Edition 2.0 2015-01 - Medical Electrical Equipment; part 1-11 General Requirements for Basic safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment And Medical Electrical Systems Used In the Home Healthcare Environment
- . IEC 60601-1-2 Edition 4.0 2014-02 – Medical Electrical Equipment – Part 1-2: General Requirements For Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
- IEC 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment Part 1-6: General • Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- . IEC 60601-2-57 Edition 1.0 2011-01 – Medical Electrical Equipment - Part 2-57: Particular Requirements For The Basic Safety And Essential Performance Of Non-Laser Light Source Equipment Intended For Therapeutic, Diagnostic, Monitoring And Cosmetic/Aesthetic Use
- . The photobiological safety of lamps and lamp systems was evaluated according to IEC 62471 first edition 2006-07. The safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications was tested in accordance with IEC 62133-1:2012. The applied part is made from a common medical grade plastic commonly used in the OLP product code amongst other medical devices. The applied part components were assessed & evaluated for biocompatibility risk in accordance with ISO 10993-1.
A Human Factors (HF) evaluation was conducted to support the safe and effective development of the Light Therapy Acne Mask+ in accordance with "Applying Human Factors and Usability Engineering to Medical Devices" (April 2016).
# Conclusion
Based upon the analysis of the performance data, the characteristics, and the intended use of the subject and predicate device, the Light Therapy Acne Mask+ does not have any differences in the safety or efficacy as the predicate device (K170111), and therefore the subject and predicate devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.