K190300 · Biokinemetrics, Inc. · MQB · Mar 13, 2019 · Radiology
Device Facts
Record ID
K190300
Device Name
BIOK7600W Wireless Flat Panel Detector
Applicant
Biokinemetrics, Inc.
Product Code
MQB · Radiology
Decision Date
Mar 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BIOK 7600W is intended for use by a qualified doctor or technologist on both adult and pediativ for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK 7600W provides digital image capture and is intended to replace radiographic film/screen. The x-ray generator, x-ray tube and associated equipment are not provided with the proposed sensor. Prescription use only. The BIOK7600W is not intended for mammography.
Device Story
Wireless digital flat panel detector; captures X-ray images from independent X-ray sources; digitizes images for transfer to PC workstation. Operated by qualified doctor or technologist in clinical settings. ChiroSight software on PC controls detector via Sensor Driver application; initiates exposure; automatically optimizes image quality; stores images in database; allows image manipulation/annotation; supports DICOM transfer. Device uses automatic exposure detection (AED) to trigger acquisition without direct generator connection. Provides digital diagnostic images to assist clinicians in patient diagnosis; replaces traditional film/screen systems.
Clinical Evidence
Clinical concurrence study provided; test images from BIOK7600W compared against images from predicate device to demonstrate equivalent image quality and characteristics.
Technological Characteristics
Wireless digital flat panel detector; solid-state X-ray imager. Compatible with X-ray generators (55-120 kVp, min 30 KW). Conforms to ISO 13485 quality systems. Software-based image processing and optimization.
Indications for Use
Indicated for adult and pediatric patients for diagnostic radiographic exposures of all body parts. Not intended for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
March 13, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" on the first line and "U.S. FOOD & DRUG ADMINISTRATION" on the second line.
Biokinemetrics Inc. % Steven Kraus President 211 East 4th Street CARROLL IOWA 51401
Re: K190300
Trade/Device Name: BIOK7600W Wireless Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: February 9, 2019 Received: February 12, 2019
Dear Steven Kraus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination. product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190300
Device Name BIOK7600W Wireless Flat Panel Detector
## Indications for Use (Describe)
The BIOK 7600W is intended for use by a qualified doctor or technologist on both adult and pediativ for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK 7600W provides digital image capture and is intended to replace radiographic film/screen. The x-ray generator, x-ray tube and associated equipment are not provided with the proposed sensor. Prescription use only.
The BIOK7600W is not intended for mammography.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Biokinemetrics. The logo is in blue and green, with the word "BIOKINE" in blue and the "ME" in green. Below the logo is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in a smaller font.
#### TRADITIONAL 510K - SUMMARY (K190300)
- 1. Device type and name: BIOK7600W Wireless Flat Panel Detector
- 2. Submitter Biokinemetrics Inc. 211 East 4th Street Carroll lowa 51401
- Dr. Steven Kraus Contact person: President Tel: (712) 210-4750 Fax: (888) 800-2149 e-mail: skraus@biokinemetrics.com
Date prepared: March 7, 2019
- 3. Device name: BIOK7600W Wireless Flat Panel Detector BIOK7600W Common name:
- Class II, 21 CFR 892.1680, Solid-State X-ray Imager (Flat panel/ 4. Device classification: Digital Imager)
- 5. Product Code: 90MQB
- 6. Requlation Name: Stationary x-ray system
- 7. Basis for the submission: Addition of a sensor to the list of medical devices compatible with the Biokinemetrics devices
- 8. Predicate device: Wireless Digital Flat Panel Detector (K182551) with iRay software BIOK4600 Digital Radiography Sensor (K092307) Reference: Reference: Aquarius 8600 Flat Panel Detector (K181565) *same classification & code: Class II, 21 CFR 892.1680, 90MQB
- 9. Device description: The BIOK7600W sensor is a wireless digital radiography sensor which automatically collects x-ray images from an x-ray source. The BIOK7600W sensor (flat panel type) collects x-rays and digitizes the images for their transfer and display to a computer. The sensor does not have an x-ray source, which is provided by independent manufacturers. The sensor includes a flat panel for x-ray acquisition and a computer (including proprietary processing software) for processing, annotating and storing x-ray images.
An X-ray generator with a minimum voltage output range between 55 to 120 kVp and a minimum power of 30 KW can be used with the BIOK7600. The BIOK7600 does not connect to the generator. It automatically detects x-rays and enables the image acquisition (AEC). No changes to the generator hardware or software are required to operate the BIOK7600.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Biokinemetrics. The logo is blue and green, with the word "BIOKINETRICS" in large, bold letters. Below the word is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in smaller letters. The logo is simple and modern, and it conveys the company's focus on chiropractic and technology.
The ChiroSight software runs on a PC workstation and provides the user interface to control the BIOK7600 flat panel. ChiroSight initiates the taking of X-ray images from the BIOK7600. through the Sensor Driver application. Once an X-ray is taken. Sensor Driver reads it from the flat panel and transfers it to ChiroSight. After it receives an image from the BIOK7600, ChiroSight:
Automatically optimizes the image quality Permanently stores the image to its database Displays the image to the user Allows the user to perform image manipulations and to add annotations Allows the user store and transfer the images in DICOM format.
The software level of concern for the BIOK7600 has been determined to be moderate based on the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
The cybersecurity risks of the BIOK7600 have been addressed to assure that no new or increased cybersecurity risks were introduced as a part of device risk analysis. These risks are defined as sequence of events leading to a hazardous situation, and the controls for these risks were treated and implemented as proposed in the risk analysis (e.g., requirements, verification).
- The BIOK7600W is intended for use by a qualified doctor or 10. Indications for use: technologist on both adult and paediatric patients for taking diagnostic radiographic exposures of all body parts of the patients. The BIOK7600W is not intended for mammography.
- 11. Comparison with predicate devices: The BIOK7600W sensor is composed of a flat panel detector very similar to the technology used in the Wireless Digital Flat Panel Detector (K182551, predicate device hardware) and the BIOK4600 (K092307) software on the PC. The hardware and firmware of the BIOK7600W sensor are considered to be substantially equivalent to the predicate device hardware, as demonstrated in the non-clinical considerations document, based on the SSXI Guidance from FDA. The BIOK7600W software functionality is equivalent to the original BIOK4600 reference device, except for the Imaray interface application which reads the images from the new BIOK7600W sensor instead of the BIOK 4600 sensor. The BIOK7600W produces images of similar quality and characteristics that are equivalent to those of the predicate device, as demonstrated in the clinical considerations - concurrence study provided in this submission.
- a. Non-clinical considerations: the non-clinical considerations document demonstrates that the BIOK7600W sensor offers equivalent performance, biocompatibility, effectiveness, thermal, electrical and mechanical safety and is equivalent to the predicate device. The design, development and production of the sensor conforms to 892.1680 and ISO 13485 quality systems.
- b. Clinical considerations concurrence study: A set of test images have been submitted along with the equivalent images from the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Biokinemetrics. The logo is in blue and green, with the word "BIOKINE" in blue and "METRICS" in green. Below the logo is the tagline "ALIGNING CHIROPRACTIC AND TECHNOLOGY" in a smaller font.
- c. Conclusion: the device was evaluated against the predicate device (Wireless Digital Flat Panel Detector, K182551) and was found to be equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.