K183513 · LeMaitre Vascular, Inc. · GXQ · Jun 13, 2019 · Neurology
Device Facts
Record ID
K183513
Device Name
XenoSure Dura Biologic Patch
Applicant
LeMaitre Vascular, Inc.
Product Code
GXQ · Neurology
Decision Date
Jun 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.
Device Story
DuraSure Biologic Patch is a bovine pericardial tissue patch used as a dura substitute during neurosurgery; tissue is treated with glutaraldehyde to crosslink collagen and minimize antigenicity; device is liquid chemical sterilized and stored in a glutaraldehyde solution; surgeon applies patch to close dural defects; provides a physical barrier to support healing and prevent cerebrospinal fluid leakage; benefits include biocompatibility and mechanical strength equivalent to predicate devices.
Clinical Evidence
No human clinical data. Evidence consists of bench testing and a prospective animal study. Bench testing (tensile strength, elongation, burst strength, suture retention, thickness) confirmed performance meets specifications. Animal study (n=49 rabbits) evaluated local tissue response at 1, 8, and 26 weeks post-implantation. Histopathological evaluation of brain tissue, dural integrity, and lymph nodes showed the device is locally non-toxic and performs equivalently to the predicate control.
Technological Characteristics
Material: Bovine pericardium. Sterilization: Liquid chemical (ISO 14160:2011). Storage: 0.2% glutaraldehyde solution. Form factor: Patch. Mechanical properties: Tensile strength ≥ 2 MPa, burst strength ≥ 12 PSI, suture retention ≥ 300 gf, thickness 0.35-0.75 mm.
Indications for Use
Indicated for use as a surgical patch material to close dura mater during neurosurgery in patients requiring dural repair.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 13, 2019
LeMaitre Vascular Inc. Xiang Zhang Vice President of Regulatory Affairs 63 Second Ave Burlington, Massachusetts 01803
Re: K183513
Trade/Device Name: DuraSure Biologic Patch Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: May 8, 2019 Received: May 14, 2019
Dear Xiang Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger, M.S.E Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183513
Device Name DuraSure Biologic Patch
Indications for Use (Describe)
The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510k Summary K183513
| Submitter's Information | |
|--------------------------------|--------------------------------------------------------------------------------------------------------|
| Name: | LeMaitre Vascular, Inc. |
| Address: | 63 Second Avenue, Burlington, MA 01803 |
| Phone: | 781-425-1706 |
| Fax: | 781-425-5049 |
| Contact Person: | Xiang (Vic) Zhang<br>VP of Regulatory Affairs<br>LeMaitre Vascular, Inc.<br>Email: xzhang@lemaitre.com |
| Date Prepared: | May 8, 2019 |
| Device Name: | DuraSure Biologic Patch |
| Trade Name: | DuraSure Biologic Patch |
| Common Name: | Dura Substitute |
| Regulation Number: | 21CFR §882.5910 |
| Classification Panel: | Neurology |
| Class: | II (2) |
| Product Code: | GXQ |
| Establishment<br>Registration: | 1220948 |
| Establishment: | 63 Second Avenue<br>Burlington, MA 01803 |
| Predicate Device: | Dura-Guard Dura Repair Patch K973706 |
| Reference Device: | XenoSure Biologic Patch K040835 |
{4}------------------------------------------------
| Device Description: | The DuraSure consists of one piece of bovine pericardial tissue that has been selected for minimal tissue blemishes. The tissue is treated with a glutaraldehyde process which crosslinks the collagen fibers and minimizes antigenicity. The DuraSure is liquid chemical sterilized and packaged in a plastic jar containing sterile glutaraldehyde storage solution. The DuraSure is designed to repair the body's natural organs. | | |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for Use: | The DuraSure Biologic Patch is intended for use as a surgical patch material to close dura mater during neurosurgery. | | |
| Summary of Technological<br>Characteristics: | Comparisons of the DuraSure Biologic Patch with the predicate Dura-Guard Patch show that technological characteristics such as materials, biocompatibility, performance, and sterilization of the proposed device are substantially equivalent to the predicate device. The difference between the proposed device and the predicate device is: The DuraSure biologic patch is stored in glutaraldehyde solution and the Dura-Guard is stored in propylene oxide solution. This difference does not raise additional risk concern because DuraSure biologic patch has satisfactory biocompatibility results and in-vivo animal study results. | | |
| | Proposed Device | Predicate Device | Comparison |
| | Product Name:<br>DuraSure Biologic Patch | Product Name:<br>Dura-Guard Patch | |
| Manufacturer | LeMaitre Vascular Inc. | Synovis/Baxter | |
| Clearance | This submission | K973706 | |
| Indications for<br>Use | The XenoSure Dura Biologic<br>Patch is intended for use as a<br>surgical patch material to<br>close dura mater during<br>neurosurgery. | For use as a dura<br>substitute for the<br>closure of dura mater<br>during neurosurgery. | Same |
| Common Name | Dura Substitute | Dura Substitute | Same |
| Classification | II (2) | II (2) | Same |
| Regulation<br>Number | 882.5910 | 882.5910 | Same |
| Product Code | GXQ | GXQ | Same |
| Materials | Bovine pericardium | Bovine pericardium | Same |
| Sterility | Chemical sterilization with 106 SAL | Chemical sterilization with 10-6 SAL | Same |
| Single Use | Yes | Yes | Same |
| Medical<br>Specialists | Neurology | Neurology | Same |
| Packaging | Jar/Lid | Jar/Lid | Same |
| Storage solution | 0.2% glutaraldehyde | 1% propylene oxide | Different |
| Functional/Safety Testing: | The verification activities conducted indicate that DuraSure biologic patch meets the product performance requirements of the device specifications and does not raise any additional safety issues. | | |
| Sterilization: | DuraSure biologic patch is chemically sterilized according to ISO14160: 2011, "Sterilization of health care products -- Liquid chemical sterilizing agents for single-use medical devices utilizing animal tissues and their derivatives -- Requirements for characterization, development, validation and routine control of a sterilization process for medical devices". | | |
| Biocompatibility: | The material for DuraSure Biologic Patch is bovine pericardium. It is identical to that in the predicate device and the reference device which have established biocompatibility. This submission is to expand the indication with no change to product design, materials, or manufacturing processes. | | |
| Summary of Product Testing: | The following tests have been completed to evaluate the safety and performance of the DuraSure Biologic Patch when compared with the predicate device:<br>• Tensile test<br>• Burst strength test<br>• Suture retention test<br>• Thickness measurement | | |
| Test | Test method summary | Results | |
| Tensile | Use Instron pull the sample until it fails. Record the ultimate tensile strength in MPa. | All DuraSure patches passed the acceptance criteria of ≥ 2 MPa. The mean of tensile strength of the predicate device was | |
| | | of tensile strength of DuraSure<br>patch was measured as 11.9<br>MPa. | |
| Elongation | Use Instron pull the sample until it<br>fails. Record the elongation at the<br>failure as percent of the original sample<br>length. | All DuraSure patches passed the<br>acceptance criteria of 5~50%<br>elongation. The mean of<br>elongation of the predicate<br>device was measured as 32%.<br>The mean of elongation of<br>DuraSure patch was measured as<br>25%. | |
| Burst strength | Sample is secured in the testing fixture<br>as a membrane between two chambers.<br>One side of the sample is slowly<br>pressurized using water. Record the<br>pressure at the time of burst or leak. | All DuraSure patches passed the<br>acceptance criteria of ≥ 12 PSI.<br>The mean of burst strength of<br>the predicate device was<br>measured as 59 PSI. The mean<br>of burst strength of DuraSure<br>patch was measured as 127 PSI. | |
| Suture retention | Make sutures on the edge of the patch.<br>Pull the suture using Instron until either<br>patch or suture fail. Record the force at<br>the failure. | All DuraSure patches passed the<br>acceptance criteria of ≥ 300 gf.<br>The mean of suture retention of<br>the predicate device was<br>measured as 978 gf. The mean<br>of suture retention of DuraSure<br>patch was measured as 970 gf. | |
| Thickness | Measure the thickness of the patch<br>using a thickness gauge | All DuraSure patches passed<br>acceptance criteria of 0.35~0.75<br>mm. The thickness of the<br>predicate device was measured<br>between 0.32~0.71 mm. | |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
| Summary of Pre-clinical Study: | The purpose of this animal study was to assess the local tissue response to implantation following 1, 8, and 26-week brain dural repair/neural implantation in rabbits compared to a predicate control to satisfy ISO 10993-6, Annex D, Tests for local effects after implantation and US-FDA Guidance Document for Dura Substitutes Devices, Section XI, 2000.<br><br>Forty nine (49) rabbits received single dural defects of approximately 0.6 cm x 0.6 cm. Test or control article were sutured in place to cover each dural defect. The bone flap was replaced, the dermal incisions were surgically closed, and animals were recovered and returned to standard housing. Animals were observed closely and weighed weekly to as one measure of general clinical health. After 1, 8, or 26 weeks post-surgery, animals were humanely euthanized and draining lymph nodes and the whole skull with brain was excised and placed in formalin. The formalin-fixed lymph nodes and implant sites were processed by histopathology techniques, using decalcification (as needed) and paraffin embedding. Resulting lymph node slides were stained with H&E. The dura/brain sections were treated with H&E, Fluoro-Jade, Luxol Fast Blue, anti-GFAP (glial fibrillary acidic protein) and anti-Iba-1 immunostaining on slides. All slides were evaluated by a veterinary pathologist for neuropathological changes in brain tissue, dural integrity, neoduralization and local tissue reaction according to ISO 10993-6 and FDA Guidance.<br><br>The overall conclusion of the control and test article brain dural repair local effect evaluation based on in vivo and terminal evaluations, including an extensive histopathological evaluation of the implantation site, adjoining areas and draining lymph nodes, is that the implantation and dural repair surgery was successful and the test article is locally non-toxic. This applies to evaluations of animals exposed for 1, 8 or 26-weeks. General |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | clinical evaluations/cageside observations, neurophysiological evaluations and body weights provide no evidence supporting an adverse effect of the surgery, control material or test article. These conclusions are supported by macroscopic evaluations and microscopic evaluations with scoring. Resorption of the test article or control article was minimal at the last exposure time evaluated. Adhesions at or near the repair site were noted, especially at one week exposure, and evidence of hydrocephaly was also noted, especially at 1 and 8 week exposure, however neither were associated with any pathology or adverse effects. From this investigation, it appears that the test and control articles performed equivalently in the in-vivo rabbit model of dural repair. LeMaitre Vascular has demonstrated that DuraSure Biologic Patch is substantially equivalent to the predicate devices based on its intended use and fundamental scientific technology. |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.