PERIPATCH SHEET, MODELS 1.5P8, 1.5P16, 4P4, 4P6, 6P8 AND 10P16
K040835 · Pm Devices, Inc. · FTM · Jun 15, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040835
Device Name
PERIPATCH SHEET, MODELS 1.5P8, 1.5P16, 4P4, 4P6, 6P8 AND 10P16
Applicant
Pm Devices, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Jun 15, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PeriPatch Sheet is intended for use as a surgical patch material for: cardiac and vascular reconstruction and repair, soft tissue deficiency repair and reinforcing the suture line during general surgical procedures.
Device Story
PeriPatch™ Sheet is a quadrilateral xenograft patch composed of glutaraldehyde-fixed bovine pericardium. It functions as a surgical patch material for cardiac and vascular reconstruction, soft tissue repair, and suture line reinforcement. The device is provided in various sizes and can be trimmed by the surgeon to fit specific procedural needs. It is intended for use by surgeons in clinical settings. The patch acts as a scaffold for tissue repair, mimicking natural tissue properties. It is supplied in a sterile, liquid alcoholic solution. The device is designed to be integrated into surgical workflows to provide structural support for damaged or deficient tissues.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including physical and performance characteristics such as thickness, weight, melting point, tensile elongation, tensile strength, and burst strength, compared against predicate devices. Sterilization and processing methods were validated for effectiveness.
Technological Characteristics
Material: Glutaraldehyde-fixed bovine pericardium. Form factor: Quadrilateral sheets in various sizes (1.5x8cm to 10x16cm). Sterilization: Liquid alcoholic sterilant (0.2% glutaraldehyde/PBS). Mechanical properties: Tensile strength 46.1±5.4g, Burst strength 8200/159 mmHg. Connectivity: None (standalone).
Indications for Use
Indicated for patients requiring cardiac and vascular reconstruction and repair, soft tissue deficiency repair, or reinforcement of suture lines during general surgical procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
#### K040835 510(K) SUMMARY SECTION 4
1. Submitters Name, Address etc.:
# JUN 1 5 2004
PM Devices Inc. 2135 – 13700 Mayfield Place Richmond, British Columbia V6V 2E4, CANADA Ph: 604.270-4344 Fx: 604.270-4384 www.pmdevices.com Contact: Britta Dombovari Date: March 2004
PeriPatch™ Sheet 2. Name of Devices: Trade Name:
Common Name: Processed bovine pericardial patch
Classification Name: Currently: Intracardiac Patch or Pledget -Class II-Product Code: DXZ Expanded Indications: Mesh, surgical, polymeric –Class Il-Product Code: FTM
- 3. Predicate Devices: Legally marketed devices which PM Devices Inc. claims substantial equivalence:
| Predicate Device | Manufacture | 510(k) # | Class |
|--------------------------------|------------------|----------|-------|
| PeriPatch™ Sheet | PM Devices Inc. | K031948 | II |
| Supple Peri-Guard® Pericardium | BioVascular Inc. | K983162 | II |
| Glycar Pericardial Patch | Glycar Inc. | K963967 | II |
| Vascu-Guard | BioVascular Inc. | K942010 | II |
| Supple Peri-Guard | BioVascular Inc | K921895 | II |
| Hancock Pericardial Patch | Extrcorpeal | K830863 | II |
All of the above previously cleared products are composed of processed bovine pericardium and are all used as a mesh material for surgical repair of pericardial structures and soft tissue deficiencies.
# 4. Device Description:
# 4a. How the Device Works
# PeriPatch™ Sheet
The PeriPatch™ Sheet is a quadrilateral shaped xenograft patch made from a sheet of glutaraldehyde fixed bovine pericardium selected for even thickness. It is designed to repair the body's natural organs and functions like natural tissue. It is intended for intra-cardiac repair procedures. A picture of a patch can be seen in Figure 1 - PeriPatch Sheet, and the engineering drawing can be found in Appendix C, Engineering Drawings. It is identical to other marketed bovine Extensive quality control procedures pericardial patches. assure a consistent, high quality product for clinical use.
Image /page/0/Picture/15 description: In the image, a translucent rectangular object is being held by a pair of tweezers. The object appears to be thin and flexible, possibly made of plastic or a similar material. The tweezers are holding the object near one of its ends. The background is a dark, textured surface, which contrasts with the lighter color of the object and the tweezers.
Figure 1 - PeriPatch Sheet
The PeriPatch Sheet is available in 6 sizes (Table 4.1 – page 4-2), but can also be trimmed to specific size depending on the procedure.
4
{1}------------------------------------------------
$$\sim / \Psi$$
# Table 4.1 PeriPatch Sheet Model Numbers & Dimensions
| Comments of Concession Commend Commend Comments of Concession Come of Concession Come of Concession Come of Concession Come of Concession of Concession of Concession of Conce | 1.5P8 | 1 | A . A . A | A | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 10P16 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | | 1.5P16 | 4P4 | 4P6 | e68 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| September 1998 - 1998 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 199<br>Size (cm) | 1.5 × 8 | 1.5 x 16<br>------------------------------------------------------------------------------------------------ | and and and and in the commend of the commend of the commend of the comment of<br>4×4<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Status and Address & Address<br>4 × 6 | 6×8 | 10 x 16<br>. |
# 4b. Scientific Concepts
#### PeriPatch Sheet
The PeriPatch Sheet is manufactured from glutaraldehyde fixed bovine pericardium, this is the The I Chi aten Oneono mananate devices. Physical configuration of the PeriPatch sheet is similar to predicate patches, see Table 4.2 below:
Table 4.2: Comparison of PeriPatch Sheet sizes available to current predicate devices.
| PeriPatch Sheet<br>Models (DXZ/FTM) * | 1.5P8 | 1.5P16 | 4P4 | 4P6 | 6P8 | 10P16 |
|-------------------------------------------|-----------|------------|---------|---------|---------|-----------|
| PeriPatch Sheet<br>Sizes (DXZ/FTM) | 1.5cmx8cm | 1.5cmx16cm | 4cmx4cm | 4cmx6cm | 6cmx8cm | 10cmx16cm |
| Bio-Vascular Supple<br>Peri-guard Patch** | N/A | N/A | 4cmx4cm | N/A | 6cmx8cm | 10cmx16cm |
| Bio-Vascular Vascu-<br>Guard*** | 1.5cmx8cm | N/A | N/A | N/A | N/A | N/A |
* PeriPatch Sheet (DXZ) – K031948
** Bio-Vascular Supple Peri-guard Patch - K981895
*** Bio-Vascular Vascu-Guard – K942010
The treatment and processing (relevant for the PeriPatch Sheet) for cross-linking bovine pericardial tissue with glutaraldehyde is well described in the literature (Appendix A, Literature), and similar to those used in predicate devices, and has been validated (Appendix D, Validations). Sterilization is performed using a liquid alcoholic sterilant which is similar to the predicates and validated to be effective (Appendix D, Validations). The finished devices are packaged and labeled in a similar manner as the predicates (Section 6, Proposed Labeling; Appendix E, Predicate Device Labeling).
The PeriPatch Sheet is considered to be similar to the PeriPatch Sheet and Bio-Vascular predicates because:
- Same raw material Bovine Pericardium +
- Same intended medical use .
- Operates using the same fundamental scientific technology
- Similar shapes & sizes .
- Similar method of processing .
- Similar method of sterilization .
- Similar packaging and labelling
# 4c. Physical & Performance Characteristics
The PeriPatch Sheet is designed to repair the body's natural organs and function like natural tissue.
## 4d. Safety & Effectiveness
The devices are designed and manufactured in such a way that, when used under the conditions and the purposes intended, they will not compromise the clinical condition or the
{2}------------------------------------------------
safety of patients, or the safety and health of users or, where applicable, other persons, sancity of patients, or the carety as associated with their use constitute acceptable risks provided that any hiske we the patient and are compatible with a high level of protection of health and safety.
The safety and effectiveness of bovine pericardial patches used for reconstruction and repair r rie salely and encoments of sture (Appendix A, Literature). They have been in use for are well one of and have proven to be effective in achieving the desired result and well tolerated by host tissue.
## 5. Intended use of the Device
Below is a list of the diseases or conditions that the device will treat, prevent, cure or mitigate and a description:
The PeriPatch Sheet has been approved for the following intended use:
- Surgical patch material for cardiac and vascular reconstruction and repair.
The intended expanded indications for use would include:
- ... soft tissue deficiency repair and reinforcing the suture line during general surgical procedures.
## 6. Technological Characteristics
The devices have the same technological characteristics as the predicate devices identified in Section 4-3. A comparison of the PeriPatch Sheet to the predicate devices can be found in Table 4.3 on the next page. As shown in the table, the applicant device is substantially equivalent to the predicates technological characteristics.
{3}------------------------------------------------
PM DEVICES INC. – PERIPATCH™ SHEET 510(K) PREMARKET NOTIFICATION
able 4.3 - Similarities of the PeriPatch Sheet to the predicate devices
| Number | Use | PM Devices Inc.<br>PeriPatch (DXZ & FTM) | PM Devices Inc.<br>PeriPatch (DXZ) K031948 | BioVascular<br>K983162 | Glycar Inc.<br>K963967 | BioVascular<br>K942010 | BioVascular<br>K921895 | Extracorporeal<br>K830863 | Significance |
|------------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|----------------------------------------------------------------------------------|
| | | The PeriPatch Sheet is<br>intended for use as a surgical<br>patch material for: cardiac<br>and vascular reconstruction<br>and repair, soft tissue<br>deficiency repair and<br>reinforcing the suture line<br>during general surgical<br>procedures. | The PeriPatch Sheet is intended<br>for use as a surgical patch<br>material for: cardiac and<br>vascular reconstruction and<br>repair. | Repair of pericardial<br>structures, soft tissue<br>deficiencies, defects<br>of the abdominal and<br>thoracic wall, gastric<br>banding, muscle flap<br>reinforcement, rectal<br>and vaginal pro-<br>lapse, reconstruction<br>of the pelvic floor, and<br>hernias. | Cardiac re-<br>construction &<br>repair, vascular<br>patching,<br>pericardial closure | Peripheral<br>vascular repair/<br>re-construction | For closure of the<br>patient's pericardium<br>& Peri-Strip sleeve<br>configuration - reinforce<br>staple lines during<br>segmental resections | Pericardial closure | PeriPatch<br>indications are<br>same as<br>(combined)<br>K031948 and<br>K983162. |
| | Bovine Pericardium | Same | Same | Same | Same | Same | Same | Same | SE |
| | Glutaraldehyde cross-linking | Same | Same | Same | Same | Same | Same | Same | SE |
| | Flat, square & Rectangular | Same | Same | Same | Not determined at<br>time of submission | Same | Rectangular & Peri-Guard<br>Sleeve, thin strips of<br>bovine pericardium<br>sutured to a polyethylene<br>backing. | Same | SE |
| Size (cm) | | 1.5x8<br>1.5x16<br>4x4<br>4x6<br>6x8<br>10x16 | Same | 4x4<br>6x8<br>8x14<br>10x16 | Not determined at<br>time of submission | 1.5x8<br>1.5x9<br>1.5x10<br>2x8<br>2x9<br>2x10<br>2.5x8<br>2.5x9<br>2.5x10 | 4x4<br>6x8<br>8x14<br>10x16<br>Peri-Guard Sleeves are<br>available in sizes to fit<br>common staplers | 5x12 | SE |
| | | | | | | | | | |
| | Sealed, sterile container | Same | Same | Same | Same | Same | Same | Same | SE |
| Thickness (mm) | | 0.58±0.17 | Same | 0.25 | Unknown | 0.5±0.25 | 0.25 | 0.35 | SE |
| Weight (g/m) | | 1172±311 g/m2 | Same | Unknown | Unknown | 1080±330 g/m2 | Unknown | 4360±1600 g | SE |
| Melting Point (°C) | | 86.2°C | Same | Unknown | Unknown | Unknown | Unknown | 83.6°C | SE |
| Tensile Elongation (g) | | 969±114 | Same | Unknown | Unknown | 1121±102 | Unknown | 1280±108 | SE |
| Tensile Strength (g) | | 46.1±5.4 | Same | Unknown | Unknown | Unknown | Unknown | 31.2±6.0 | SE |
| Burst Strength (mmHg) | | 8200 / 159 | Same | Unknown | Unknown | 7033 / 136 | Unknown | Unknown | SE |
| Sterilization | | 0.2% Glutaralde-hyde /<br>phosphate buffered solution<br>(PBS) | Same | Propylene oxide in<br>water | Propylene oxide<br>in water | Propylene oxide<br>in water | Propylene oxide in water | Same | SE |
| Rinsing Conditions | | Two 2 min. rinses | Same | One 3 min. Rinse | Unknown | One 3 min. rinse | One 3 min. Rinse | Three 2 min. rinse | SE |
| Sterility Method | | Liquid Alcoholic | Same | Same | Unknown | Same | Same | Same | SE |
|<040835
c/4
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three stylized waves or lines, which may represent the flow of information or resources. The image is black and white.
Public Health Service
JUN 1 5 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Britta Dombovari Regulatory Affairs/Quality Assurance PM Devices, Inc. 2135-13700 Mayfield Place Richmond, British Columbia V6V 2E4 Canada
Re: K040835
Trade/Device Name: PeriPatch™ Sheet Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTM Dated: March 29, 2004 Received: March 31, 2004
Dear Ms. Dombovari:
We have reviewed your Section 510(k) premarket notification of intent to market the device i we nave teviewed your bection of the device is substantially equivalent (for the indications felerenced above and nave determined and and and and one inces marketed in interstate for use stated in the encrosule) to regars actual date of the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Medical Device Americ commerce prior to May 20, 1770, the enastance with the provisions of the Federal Food, Drug, devices that have been recuire approval of a premarket approval application (PMA). and Cosmelle Act (Act) market the device, subject to the general controls provisions of the Act. The r ou may, merciole, market the act include requirements for annual registration, listing of general controls provisions of use ice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (sociation major regulations affecting your device can
may be subject to such additional controls. Existing major regulations of the contract hilly be subject to suell additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Obas neements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised hiat i Dri s issuated on a business with other requirements of the Act that I DA has made a colorimiation administered by other Federal agencies. You must of any Federal statutes and regulations damading, but not limited to: registration and listing (21 comply with an the Ace STequirements, were 801); good manufacturing practice requirements as set and CFR Part 807), adocing (21 OFF Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Britta Dombovari
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin manieting your antial equivalence of your device to a legally premarket notification. THC FDA mixing of substantial equate and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice ion your ac not and subser note the regulation entitled, contact the Office of Complanes as (Set notification" (21CFR Part 807.97). You may obtain Misorallung by release to promance no promance notice ander the Act from the Division of Small other general information on your response at its toll-free number (800) 638-2041 or Manufacturers, International and Soless http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
(Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040835
Device Name: PeriPatchTM Sheet - Processed Bovine Pericardial Pati
Indications For Use:
The Por Use.
The Por Parch sheet is intended for intended for works of a construction The Perforth 5neet 15 Orders and vasular re-constitution.
Patch material for ; cardiac and vasular and soft fissne deficiency repair and repair surgical 1 general the mitche line during re inforcing procedures .
Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please Do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
(Division Sign-Off Division of General, Restorative. and Neurological Devices
510(k) Number_K 040835
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.