Mag Vita TMS Therapy System w/Theta Burst Stimulation
K173620 · Tonica Elektronik A/S · OBP · Aug 14, 2018 · Neurology
Device Facts
Record ID
K173620
Device Name
Mag Vita TMS Therapy System w/Theta Burst Stimulation
Applicant
Tonica Elektronik A/S
Product Code
OBP · Neurology
Decision Date
Aug 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5805
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MagVita TMS Therapy System w/ Theta Burst Stimulation is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Device Story
Computerized electromechanical system delivering non-invasive magnetic fields to cerebral cortex; utilizes Faraday's law via capacitor-driven stimulating coil to induce electrical currents in neural tissue. System includes MagPro stimulator, trolley, Cool-B70 coil with cooler, marking apparatus, and patient fixation (chair, vacuum pillow). Operated by clinicians in clinical settings. Delivers intermittent Theta Burst Stimulation (iTBS): 3-pulse bursts at 50 Hz, repeated at 5 Hz. Output intensity maintained constant across stimuli. Compared to standard 10 Hz rTMS, iTBS provides equivalent excitatory effects with reduced treatment time (approx. 3 minutes vs 19-37 minutes). Output affects cortical excitability to treat MDD. Benefits include enhanced ease of use and shorter treatment sessions for patients and operators.
Clinical Evidence
Effectiveness, safety, and tolerability demonstrated in a large randomized, multicenter trial. iTBS showed similar effectiveness and side effect profile compared to standard-of-care 10 Hz rTMS. Non-clinical bench testing verified magnetic spatial distribution, biocompatibility, and electrical safety (IEC60601-1, IEC60601-1-2).
Technological Characteristics
Electromechanical TMS system; figure-of-eight air-core coil; liquid cooling. Energy source: capacitor-based electrical discharge. Connectivity: standalone system. Standards: ISO 13485:2012, IEC60601-1, IEC60601-1-2. Software: embedded control for stimulation parameters and pulse intensity maintenance.
Indications for Use
Indicated for treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 14, 2018
Tonica Elektronik A/S Sanne Jessen Medical Advisor Lucernemarken 15 Farum, 3520 Dk
Re: K173620
Trade/Device Name: Mag Vita TMS Therapy System w/Theta Burst Stimulation Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulation System Regulatory Class: Class II Product Code: OBP Dated: July 10, 2018 Received: July 13, 2018
Dear Sanne Jessen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173620
#### Device Name
MagVita TMS Therapy System w/Theta Burst Stimulation
Indications for Use (Describe)
The MagVita TMS Therapy System w/ Theta Burst Stimulation is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# MagVita TMS Therapy System w/Theta Burst Stimulation
## 510(k) Summary
# Submitter's Information
| Name of 510(k) owner: | Tonica Elektronik A/S<br>Lucernemarken 15<br>DK-3520 Farum, Denmark | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Phone: | +45 4499 8444 | |
| Fax: | +45 4499 1544 | |
| Contact person: | Sanne Barsballe Jessen<br>Medical Advisor<br>Email: sj@magventure.com<br>Phone: +45 44390418<br><br>Other:<br>Jan Kjoeller<br>Email: jk@tonica.dk | |
| Preparation date: | August 10, 2018 | |
| Trade name: | MagVita TMS Therapy System w/ Theta Burst Stimulation | |
| Common name: | Transcranial Magnetic Stimulator | |
| Classification name: | Repetitive Transcranial Magnetic Stimulator for treatment of<br>Major Depressive Disorder [21 CFR 882.5805, Product Code<br>OBP] | |
| Classification: | Class II Medical Device | |
| Predicate Devices: | MagVita TMS Therapy System (K150641, K171481,<br>K171967 and K172667)<br>21 CFR 882.5805, Repetitive Transcranial Magnetic<br>Stimulation<br>Product code: OBP<br>Device Class: II | |
{4}------------------------------------------------
## 510(k)
# MagVita TMS Therapy System w/Theta Burst Stimulation
### Device description
The MagVita TMS Therapy System w/Theta Burst Stimulation is a computerized. electromechanical medical device that produces and delivers non-invasive, magnetic fields to induce electrical currents targeting specific regions of the cerebral cortex. The MagVita TMS Therapy System w/ Theta Burst Stimulation is indicated for the treatment of Major Depressive Disorder (MDD) in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
Transcranial magnetic stimulation (TMS) is a non-invasive technique for stimulating brain and neural tissue. The principle of magnetic stimulation is implicit in Faraday's law. The pulses of current are generated with a circuit containing a capacitor connected to the stimulating coil. With the capacitor charged to a certain level, the conducting state will cause the discharging of the capacitor through the coil. A magnetic field is generated proportional to this current. The rapid change in the magnetic field induces a current in conducting materials e.g. the body tissue. If the current induced in the human body is of sufficient amplitude and duration, it will excite neurons. The standard-of-care FDA approved TMS protocol for treatment of MDD uses repetitive transcranial magnetic pulses applied at a frequency of 10 Hz. Such stimulation has been shown to be effective in modulating cortical excitability. The observed and documented increase in cortical excitability after high frequency (10 Hz rTMS) repetitive TMS (rTMS) has been shown to persist beyond the duration of the train of stimulation, and 10 Hz rTMS on the Left-Dorsolateral prefrontal cortex (L-DLPFC) has been shown to be effective and safe in the treatment of MDD. A newer form of the TMS, called intermittent Theta Burst Stimulation (iTBS), mimics the endogenous theta rhythms of the human brain, and has been shown to evoke equally potent excitatory effects. Intermittent TBS provides bursts of three individual stimuli at a rate of 50 Hz (i.e. 20 ms apart), repeated at 5 Hz (i.e. 200 ms between burst) mimicking the theta rhythm of the brain. Thus, the iTBS protocol is more demanding in terms of stimulator output. The clinical performance is based on the fact that the actual individual stimuli are of equal intensity, to ensure that a constant dose of stimuli is delivered during treatment.
MagVita TMS therapy system w/Theta Burst stimulation is applied to the human brain of the left dorsolateral prefrontal cortex (L-DLPFC) using iTBS.
The MagVita TMS Therapy System w/ Theta Burst Stimulation is an integrated system consisting of the following components:
- MagPro Stimulator and Trolley .
- o MagPro family
- Trolley with holding arrangements o
- Coil for MT determination and Depression treatment .
- o Coil Cool-B70 with Coil Cooler Unit
- Marking apparatus for locating treatment area .
- Marking plate for Coil Cool-B70 o
- Pen for marking, Cap, Ruler o
- Patient head fixation .
- Treatment Chair o
- Vacuum Pump and Vacuum pillow o
- Super Flexible Arm mounted on the trolley o
{5}------------------------------------------------
- Isolation Transformer .
### Intended Use/Indication for Use:
The MagVita TMS Therapy System w/Theta Burst Stimulation and the predicate device are indicated for:
Treatment of Major Depressive Disorder in adult patients who have failed to receive satisfactory improvement from prior antidepressant medication in the current episode.
#### Performance Standards:
The MagVita TMS Therapy System w/Theta Burst Stimulation has been tested and complies with the following standards
- ISO 13485:2012 .
- IEC60601-1 .
- . IEC60601-1-2
#### Non-Clinical performance data:
The non-clinical performance testing of the components of the Mag Vita TMS Therapy System w/ Theta Burst Stimulation has been tested as required according to the standards listed above.
To establish substantial equivalency for the new coil and the predicate devices we have performed a comparative testing of the magnetic field distribution for the Cool-B70 coil and the predicate devices, that determines that the magnetic spatial distribution is substantially equivalent.
The MagVita TMS Therapy System w/Theta Burst Stimulation consists of components that are identical to those of the predicate devices. The New Device is identical to the predicate devices in terms of biocompatibility, design elements, such as cable lengths, coil materials, size and coil windings as well as cooling media, isolation design and functionalities. All coils are subject to high-voltage tests and leakage current tests to ensure safety.
In addition, the MagVita TMS Therapy System w/Theta Burst Stimulation has been verification tested to ensure that the intensity of the individual stimuli in iTBS is equal and kept constant. The clinical performance is based on the fact that the actual individual stimuli are of equal intensity, to ensure that a constant dose of stimuli is delivered during treatment.
We have also tested the device according to IEC60601 3rd edition and verified that the device complies with the specified permissible sound pressure levels. The device also complies with the permissible thresholds for exposure defined by the Occupational Safety and Health Administration (OSHA).
These tests provide evidence that the MagVita TMS Therapy System w/Theta Burst Stimulation does not pose any risk for potential hearing reduction or loss in either patients or operators.
{6}------------------------------------------------
### 510(k) MagVita TMS Therapy System w/Theta Burst Stimulation
#### Clinical performance data:
The effectiveness, safety and tolerability of iTBS have been demonstrated in a large randomized, multicenter trial. iTBS has similar effectiveness and side effect profile compared to standard-of-care 10 Hz rTMS. Based on performance testing, the MagVita TMS Therapy System w/Theta Burst Stimulation, is as safe and effective and performs as well as the predicate devices with enhanced ease of use and reduced treatment time to the benefit of both patient and operator.
#### Substantial equivalence:
The MagVita TMS Therapy System w/Theta Burst Stimulation is substantially equivalent to the predicate device. Mag Vita TMS Therapy System. The Mag Vita TMS Therapy System w/Theta Burst Stimulation and the predicate device have identical indication for use, and the technological characteristics are very similar such that they can be considered equivalent. The clinical performance and effectiveness, as well as tolerability and side effect profile for the new device and the predicate devices are substantially equivalent.
Design of the MagVita TMS Therapy System w/Theta Burst Stimulation is identical to the predicate device MagVita TMS Therapy System as both systems apply Transcranial Magnetic Stimulation at an intensity of 1.2xMT (120% of MT) as repetitive pulse trains delivered as brief rapidly alternating magnetic fields to induce electrical currents over the prefrontal cortex.
Both the MagVita TMS Therapy System w/Theta Burst Stimulation and the predicate device MagVita TMS Therapy System consist on the same components, that is TMS stimulator with software, electromagnetic coil and an articulated arm for positioning of the treatment coil. The operational procedures including system setup, patient preparations, motor threshold determination and coil positioning are essentially the same. The two systems only differ in predefined treatment stimulation parameters.
| Area | New Device<br>MagVita TMS Therapy<br>System w/<br>Theta Burst Stimulation | Predicate Device<br>MagVita TMS Therapy<br>System<br>Tonica Elektronik A/S<br>(K150641, K171481, K171967 and<br>K172667) |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance | Waveforms: Biphasic.<br>Frequency: 0.1 -30 pulses<br>per second or 0.1 -100<br>pulses, depending on model<br>Preset range of % MT: 0% -<br>140% | Waveforms: Biphasic.<br>Frequency: 0.1 -30 pulses<br>per second or 0.1 -100<br>pulses, depending on model<br>Preset range of % MT: 0% -<br>140% |
#### Characteristics of the Device as Compared to Predicate Devices*
{7}------------------------------------------------
# 510(k)
| Area | New Device<br>MagVita TMS Therapy<br>System w/<br>Theta Burst Stimulation | Predicate Device<br>MagVita TMS Therapy<br>System<br>Tonica Elektronik A/S<br>(K150641, K171481, K171967 and<br>K172667) |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | iTBS treatment protocol:<br>Stimulation Intensity: 120%<br>MT<br>(MT=Motor Threshold<br>intensity)<br>Repetition rate:<br>50 Hz (5 pulses pr sec)<br>Train duration:<br>2 sec<br>Interval between Trains:<br>8 secs<br>Burst pulses: 3<br>Bursts : 200<br>Inter pulse interval:<br>20 msec | Standard treatment:<br>Stimulation Intensity: 120%<br>MT<br>(MT=Motor Threshold<br>intensity)<br>Repetition rate:<br>10 Hz<br>Train duration:<br>4 sec<br>Interval between pulses: 11-<br>26 sec |
| | Number of trains : 20<br>Numbers of pulses/session:<br>600<br>Total duration: 3 min and 9<br>sec | Number of trains: 75<br>Numbers of pulses/ session:<br>3000<br>Total duration: 19-37.5 min. |
| Area | New Device<br>MagVita TMS Therapy System w/<br>Theta Burst Stimulation | Predicate Device<br>MagVita TMS Therapy System<br>Tonica Elektronik A/S<br>(K150641, K171481, K171967 and K172667) |
| | Output Stimulation<br>Parameters:<br>Available Stimulation<br>Amplitude in Standard<br>Motor Threshold (SMT) units<br>Amplitude Range: 0 - 1.7 SMT<br><br>The clinical performance of TBS is dependent on the fact that the stimuli are of equal intensity. At the relevant TBS intensities required in the treatment setting, the individual intensity of the three stimuli is kept constant (i.e. at maximum a 1% drop referred to maximum machine output between the first stimuli and the third stimuli in a burst). | Output Stimulation<br>Parameters:<br>Available Stimulation<br>Amplitude in Standard<br>Motor Threshold (SMT) units<br>Amplitude Range: 0 - 1.7 SMT |
| Coil<br>Configuration | Figure-of-eight coil<br>Air core | Figure-of-eight coil<br>Air core |
| Cooling | Liquid cooling | Liquid cooling |
| Standards met | Company complies with ISO<br>13485:2012. | Company complies with ISO<br>13485:2012. |
| Electrical safety | Complies with IEC60601-1<br>and IEC60601-1-2. | Complies with IEC60601-1<br>and IEC60601-1-2. |
# MagVita TMS Therapy System w/Theta Burst Stimulation
{8}------------------------------------------------
# 510(k)
# MagVita TMS Therapy System w/Theta Burst Stimulation
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
*For a more comprehensive comparison of devices please refer to section 12 Substantial Equivalence Comparison and to section 20 Clinical Performance.
{9}------------------------------------------------
### Conclusion:
The indication for use, the target population, the treatment procedure and the treatment position are all 100 % identical for the MagVita TMS Therapy System w/Theta Burst Stimulation and the predicate device MagVita TMS Therapy System.
The protocol parameters are not equivalent in terms of dosage, repetition rate, number of pulses in a train or the number of trains. Hence, the dosage is much smaller for the New Device, whereas the repetition rate in contrast is higher. The intensity is identical for the Mag Vita TMS Therapy System w/Theta Burst Stimulation and the predicate devices. However, the iTBS protocol is more demanding in terms of stimulator output, and the MagVita TMS Therapy System w/Theta Burst Stimulation delivers stimuli of equal intensity, thereby verifying that the dose is maintained throughout the course of treatment.
The performance and the clinical effectiveness are substantially equivalent to that of the predicate devices, and the iTBS protocol with MagVita TMS Therapy System w/Theta Burst Stimulation in addition offers advantages over the predicate devices in terms of treatment duration to the benefit of the patient as well as the operator.
The transducer design (figure-of-eight) is equivalent, the magnetic properties of the MagVita TMS Therapy System w/Theta Burst Stimulation and the predicate devices, as well as the realized magnetic properties are substantially equivalent for the coils (K172667). Additionally, the new device offers the advantage that the same coil can be used for MT determination and treatment to the benefit of the Operator.
The reliability of the positioning method used by the MagVita TMS Therapy System w/Theta Burst Stimulation is, based on the direct relationship of the underlying cortical brain anatomy to the patient's scalp, identical to the method used for the predicate device, MagVita TMS Therapy System and previously FDA cleared (K150641 and K171481).
The MagVita TMS Therapy System does not introduce any new safety considerations in comparison to the predicate devices based on clinical trial data.
All other identified differences between the two systems are minor and without any impact on safety or efficacy.
The above comparison demonstrates and supports the substantial equivalency of the MagVita TMS Therapy System w/Theta Burst Stimulation to the predicate device, MagVita TMS Therapy System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.