The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician. Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).
Device Story
Electronically controlled pneumatic ventilator; provides positive pressure ventilatory support. Inputs: user-defined parameters via touchscreen; proximal flow sensor readings. Operation: turbine-based pneumatic system delivers controlled gas mixture to patient. Output: respiratory gas delivery; ventilation monitoring. Used in hospitals, sub-acute facilities, and intra-hospital transport by trained personnel. Healthcare providers use output to manage patient ventilation; supports respiratory function in adult, pediatric, infant, and neonatal populations.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, comparison of ventilation modes, operating ranges, and waveforms against predicate and reference devices.
Technological Characteristics
Turbine-based pneumatic ventilation system; AC/DC/internal battery power. Features proximal flow sensor. Added SpO2 and EtCO2 monitoring capabilities. Connectivity includes touchscreen interface. System is electronically controlled.
Indications for Use
Indicated for positive pressure ventilatory support for adult, pediatric, infant, and neonatal patients. Used by qualified, trained personnel under physician direction in hospitals, sub-acute care facilities, and for intra-hospital transfer. Neonatal use restricted to the NICU.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2019
Imtmedical AG Colleen Watson Director, Regulatory Affairs 26125 N. Riverwoods Blvd. Mettawa, IL 60045 USA
Re: K183364
Trade/Device Name: bellavista 1000 and bellavista 1000e Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK Dated: August 13, 2019 Received: August 15, 2019
Dear Colleen Watson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James Lee Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K183364
Device Name
bellavista 1000 and bellavista 1000e
#### Indications for Use (Describe)
The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician.
Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer
When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vyaire Medical. The word "Vyaire" is in a dark blue color, with the "V" and "y" connected. The word "MEDICAL" is in a light blue color and is located below the word "Vyaire".
### 510(k) Summary
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92
The assigned 510(k) number is:
| Submitted by: | imtmedical AG<br>Gewerbestrasse 8<br>9470 Buchs SG Switzerland |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3004553423 |
| Contact Person: | Colleen Watson<br>Director, Regulatory Affairs<br>Phone: (872) 757-0070<br>e-mail: colleen.okeeffe@vyaire.com |
| Date Summary Prepared: | August 12, 2019 |
| Reason for Premarket<br>Notification: | New Device |
| Trade Name: | bellavista 1000 and bellavista 1000e |
| Common/Usual Name: | Ventilator, continuous, facility use |
| Product Code: | CBK |
| Regulatory Class: | Class II |
| Predicate Device: | bellavista 1000 (K163127) |
vyaire.com
Vyaire Medical, Inc.
26125 N. Riverwoods Blvd. Mettawa, IL 60045 USA
Vyaire Medical, Inc. 22745 Savi Ranch Pkwy Yorba Linda, CA 92887 USA
Vyaire Medical Oy Kuortaneenkatu 2 FI-00510 Helsinki Finland
Vyaire Medical GmbH Leibnizstrasse 7 97204 Hoechberg Germany
{4}------------------------------------------------
#### Reference devices:
Multiple reference devices have been used for the modifications to support substantial equivalence.
| Manufacturer / Tradename | 510(k) Number | Product Code |
|-------------------------------------------------|---------------|--------------|
| Hamilton C3 | K161450 | CBK, DQA |
| Dräger V500 | K093633 | CBK |
| CareFusion Avea | K103211 | CBK |
| Pulmonetic Systems Palmtop<br>Ventilator PTV-10 | K070594 | CBK |
| Vyaire SiPAP | K031745 | CBK |
| Maquet Servo-u | K151814 | CBK |
| Puritan Bennett 840 | K151252 | CBK |
#### Device Description:
bellavista is an electronically controlled pneumatic ventilation system. It is powered by AC or DC and also provided with an internal battery. The bellavista pneumatic system supplies respiratory gas whilst the electrical systems control the pneumatics and provides the power supply.
The user can enter values or parameters in the bellavista via the touch screen. These inputs entail instructions for bellavista's pneumatic system to ventilate the patient with a precisely controlled gas mixture. bellavista gathers readings from the proximal flow sensor within the ventilator.
#### Indications for Use:
The bellavista 1000/1000e ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients and optionally infant and neonatal patients by qualified, trained personnel under the direction of a physician.
Environment of use: hospitals, sub-acute care facilities and intra-hospital transfer
When used on neonatal patients: The environment of use is the Neonatal Intensive Care Unit (NICU).
#### Comparison of bellavista 1000/1000e to the predicate:
We present in the following tables a comparison of the subject device compared to the predicate and reference devices.
{5}------------------------------------------------
| | Subject bellavista<br>1000 / bellavista<br>1000e | Predicate bellavista<br>1000 (K163127) | Remarks, Substantial equivalence |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | CBK | CBK | The same classification |
| Intended use | The bellavista<br>1000/1000e ventilator<br>is intended to provide<br>positive pressure<br>ventilatory support to<br>adult and pediatric<br>patients and<br>optionally infant and<br>neonatal patients. | The bellavista 1000<br>ventilator is intended<br>to provide positive<br>pressure ventilator<br>support to adults and<br>pediatrics (IBW<br>greater than 6 kg). | Neonates are included in the<br>intended patient population of<br>the reference device Hamilton C3<br>(K161450) |
| Environment of use | Environment of use:<br>hospitals, sub-acute<br>care facilities and<br>intra-hospital transfer<br><br>When used on<br>neonatal patients: the<br>environment of use is<br>the NICU | Environment of use:<br>hospitals, sub-acute<br>care facilities and<br>intra-hospital transfer | Both devices are<br>substantially equivalent for<br>environment of use for adult and<br>pediatric patient population.<br><br>The environment of use for neonatal<br>and infant patients is substantially<br>equivalent to the environment of use<br>of the reference device, Hamilton C3<br>(K161450). |
| Intended users | It is intended for use<br>by qualified, trained<br>personnel under the<br>direction of a<br>physician. | It is intended for use<br>by qualified, trained<br>personnel under the<br>direction of a<br>physician. | Both devices are<br>substantially equivalent |
| Patient population | Adults, pediatrics,<br>infants and neonates | Adults and pediatrics<br>(IBW greater than 6<br>kg) | Neonates and infants are included in<br>the intended patient population of<br>the reference device Hamilton C3<br>(K161450) |
## Table 1: Comparison of Indications for Use
{6}------------------------------------------------
# Comparison of Technological Characteristics
We have added SpO2 and EtCO2 monitoring to the device.
# Discussion of the Differences:
## Indications for Use
Indications for use are to provide positive pressure ventilator support to adults, pediatrics and neonates.
# Technology
The technology of a turbine based portable ventilator is identical to the predicate.
# Environment of Use
The environment of use – hospital and intra-hospital transport is equal to the predicate.
# Patient Population
- The patient population of adults and pediatrics is equal to the predicate. .
- . The patient population of neonates is similar to the reference device Hamilton C3 (K161450)
# Ventilation Modes
# Adult/Pediatric Ventilation Modes
Ventilation modes that are equivalent and not changed between subject and predicate device bellavista 1000 are:
The following ventilation modes have been cleared in the predicate device (K163127):
- СРАР ●
- PCV
- P-A/C
- PC-SIMV
- bsv
- ഗ
- S/T ●
- T
- beLevel
- APRV VCV
- V-A/C
- VC-SIMV
The following modes were not available in the predicate device bellavista 1000 (K163127) but are part of the subject device. We have utilized reference devices where applicable.
{7}------------------------------------------------
### Neonatal ventilation modes (modes in K163127 but not for neonates)
As neonatal ventilation had not been cleared in predicate device bellavista 1000 (K163127), all neonatal modes are compared to reference devices.
#### Summary of Ventilation Mode Comparison
Ventilation modes are either equivalent to the predicate – bellavista 1000 (K163127) or are similar to one of the reference devices as identified. We compared the ventilation modes which had similar features and their operating ranges as well as waveforms and they were found to be substantially equivalent.
### Substantial Equivalence Conclusion
We have compared the subject device bellavista 1000/1000e to the predicate device bellavista 1000 and where needed provided reference devices. The testing performed has demonstrated that the bellavista can be found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.