K161450 · Hamilton Medical AG · CBK · Apr 27, 2017 · Anesthesiology
Device Facts
Record ID
K161450
Device Name
HAMILTON-C3
Applicant
Hamilton Medical AG
Product Code
CBK · Anesthesiology
Decision Date
Apr 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The HAMILTON-C3 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates. Intended areas of use: In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room During transfer of ventilated patients within the hospital The HAMILTON-C3 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
Device Story
HAMILTON-C3 is a continuous ventilator providing invasive/noninvasive positive pressure support; inputs include patient respiratory signals and optional SpO2 monitoring; device uses internal turbine for air and external compressed O2; operates via qualified personnel in clinical settings (ICU, ER, recovery, transport); provides automated ventilation (ASV) and manual modes; output includes real-time ventilation parameters and alarms; assists clinical decision-making by maintaining respiratory support; benefits patients by providing tailored ventilation support.
Clinical Evidence
Bench testing only. Software verification and validation performed per FDA guidance. Waveform performance testing conducted for new ventilation modes. Compliance with IEC 60601-1, ISO 80601-2-12, ISO 80601-2-55, and ISO 80601-2-61 confirmed. No new biocompatibility testing required as materials are unchanged from previous clearances.
Technological Characteristics
Continuous ventilator; internal turbine for air, compressed source for O2; active pneumatic exhalation valve; SpO2 and CO2 monitoring options; embedded system EM10A; IEC 60601-1 (2005) compliant; software level of concern: major.
Indications for Use
Indicated for positive pressure ventilatory support in adults, pediatrics, and optionally infants and neonates. Used in intensive care, intermediate care, emergency wards, long-term acute care, recovery rooms, and during in-hospital patient transfer. Contraindications: Not for use in presence of flammable anesthetic agents or MRI equipment.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 27, 2017
Hamilton Medical AG Annemarie Höft Quality Engineer Via Crusch 8 Bonaduz, 7402 SWITZERLAND
Re: K161450
Trade/Device Name: Hamilton-C3 Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: CBK, DQA Dated: March 20, 2017 Received: March 23, 2017
Dear Ms. Höft:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image contains a signature and a name. The signature is a complex series of loops and lines. The name next to the signature reads "Tina Kiang-S".
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161450
Device Name HAMILTON-C3
Indications for Use (Describe)
The HAMILTON-C3 ventilator is intended to provide positive pressure ventilatory support to adults and optionally infants and neonates.
Intended areas of use:
· In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
· During transfer of ventilated patients within the hospital
The HAMILTON-C3 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Hamilton Medical. The logo is blue and consists of the word "HAMILTON" in large, bold letters, with a stylized "H" symbol to the right. Below "HAMILTON" is the word "MEDICAL" in a similar font but slightly smaller. Underneath the two words is the phrase "Intelligent Ventilation since 1983" in a smaller, lighter font.
Image /page/3/Picture/1 description: The image shows the logo for HAMILTON-C3. The logo is written in a sans-serif font and is blue. The letters are rounded and have a soft appearance. The word "HAMILTON" is followed by a dot and then "C3".
# 510(k) SUMMARY
l. SUBMITTER
Hamilton Medical AG Via Crusch 8 Bonaduz, 7402 Switzerland
Phone: +41 58 610 1225 +41 58 610 0020 Fax:
Contact Person: Annemarie Höft, Quality Engineer Frederike Brühschwein, Senior Manager Regulatory Affairs Date Prepared: 2017-04-24
- II. DEVICE
Name of Devices: HAMILTON-C3 Common or Usual Name: Continuous ventilator Regulation Number and Name: Ventilator, Continuous (21 CFR 868.5895) Device Classification: 2 Product Code: CBK (subsequent: DQA)
- III. PREDICATE DEVICES HAMILTON-C3 (K123637) HAMILTON-G5 (K131774)
#### IV. DEVICE DESCRIPTION
The HAMILTON-C3 is designed for adult, paediatric, infant, and neonatal patients requiring invasive or noninvasive ventilation support. It covers a full range of clinical requirements, including invasive ventilation, automated ventilation with Adaptive Support Ventilation (ASV), and noninvasive ventilation.
The 510(k) submission intends to add the following new features to the previously cleared ventilator HAMILTON-C3:
- The following two new ventilation modes for adult / pediatric patient group: ● (S)CMV and SIMV
- SpO2 monitoring option
#### V. INDICATIONS FOR USE
The HAMILTON-C3 ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates. Intended areas of use:
- In the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room
- During transfer of ventilated patients within the hospital
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the words "HAMILTON MEDICAL" stacked on top of each other. Below the logo is the text "Intelligent Ventilation since 1983".
Image /page/4/Picture/1 description: The image shows the word "HAMILTON-C3" in a stylized, sans-serif font. The letters are all in blue. The letters are rounded and have a playful appearance.
The HAMILTON-C3 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.
#### COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICES VI.
A comparative summary of the technological characteristics of the HAMILTON-C3 with the primary predicate and reference devices is presented below.
| Parameters | Application device:<br>HAMILTON-C3 | Predicate device:<br>Currently<br>marketed<br>HAMILTON-C3 | Predicate device:<br>Currently<br>marketed<br>HAMILTON-G5 | Comparison |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Intended use | The HAMILTON-C3<br>ventilator is<br>intended to<br>provide positive<br>pressure<br>ventilatory support<br>to adults and<br>pediatrics and<br>optionally infants<br>and neonates.<br>Intended areas of<br>use:<br>In the<br>intensive care<br>ward,<br>intermediate<br>care ward,<br>emergency<br>ward, long<br>term acute<br>care hospital<br>or in the<br>recovery<br>roomDuring<br>transfer of<br>ventilated<br>patients<br>within the<br>hospitalThe HAMILTON-C3<br>ventilator is a<br>medical device<br>intended for use by<br>qualified, trained<br>personnel under<br>the direction of a<br>physician and<br>within the limits of<br>its stated technical | The HAMILTON-C3<br>ventilator is<br>intended to<br>provide positive<br>pressure<br>ventilatory support<br>to adults,<br>paediatrics,<br>infants, and<br>neonates.<br>Intended areas of<br>use:<br>In the<br>intensive care<br>ward or in the<br>recovery<br>room.During<br>transfer of<br>ventilated<br>patients<br>within the<br>hospital.The HAMILTON-C3<br>ventilator is a<br>medical device<br>intended for use by<br>qualified, trained<br>personnel under<br>the direction of a<br>physician and<br>within the limits of<br>its stated technical<br>specifications. | The HAMILTON-G5<br>ventilator is<br>designed for<br>intensive care<br>ventilation of adult<br>and paediatric<br>patients, and<br>optionally infant<br>and neonatal<br>patients. The<br>device is intended<br>for use in the<br>hospital and<br>institutional<br>environment<br>where healthcare<br>professionals<br>provide patient<br>care. The<br>HAMILTON-G5<br>ventilator is<br>intended for use by<br>properly trained<br>personnel under<br>the direct<br>supervision of a<br>licensed physician.<br>The HAMILTON-G5<br>ventilator may be<br>used for transport<br>within a hospital or<br>hospital-type<br>facility -- provided<br>compressed gas is<br>supplied.<br>The device is not to<br>be used in the<br>presence of<br>flammable<br>anesthetic agents | Substantially<br>equivalent |
| | specifications. | | or other ignition<br>sources. The<br>ventilator is not to<br>be used in an<br>environment with<br>magnetic<br>resonance imaging<br>(MRI) equipment.<br>The device is not<br>intended for<br>transportation<br>outside the<br>hospital or for use<br>in the home<br>environment. | |
| Product<br>classification<br>code | CBK (subsequent:<br>DQA) | CBK | CBK (subsequent:<br>DQA) | Substantially<br>Equivalent |
| CFR citation | 21 CFR 868.5895 | 21 CFR 868.5895 | 21 CFR 868.5895 | Substantially<br>Equivalent |
| Principal<br>operator | Qualified, trained<br>personnel under<br>the direction of a<br>physician | Qualified, trained<br>personnel under<br>the direction of a<br>physician | Qualified, trained<br>personnel under<br>the direction of a<br>physician | Substantially<br>Equivalent<br>Same principal<br>operator as on<br>HAMILTON-C3 and<br>HAMILTON-G5 |
| Environment<br>of use | Intended areas of<br>use:<br>●<br>In the<br>intensive care<br>ward,<br>intermediate<br>care ward,<br>emergency<br>ward, long<br>term acute<br>care hospital<br>or in the<br>recovery<br>room<br>●<br>During<br>transfer of<br>ventilated<br>patients<br>within the<br>hospital | Intended areas of<br>use:<br>●<br>In the<br>intensive care<br>ward or in the<br>recovery<br>room.<br>●<br>During<br>transfer of<br>ventilated<br>patients<br>within the<br>hospital. | Intended areas of<br>use:<br>●<br>In the<br>intensive care<br>ward,<br>intermediate<br>care ward,<br>emergency<br>ward, long<br>term acute<br>care hospital<br>or in the<br>recovery<br>room<br>●<br>During<br>transfer of<br>ventilated<br>patients<br>within the<br>hospital | Substantially<br>Equivalent<br>Same environment<br>of us as on<br>HAMILTON-C3 and<br>HAMILTON-G5 |
| Intended<br>patient<br>population | Adults, pediatrics,<br>infants and<br>neonates | Adults, pediatrics,<br>infants and<br>neonates | Adults, pediatrics,<br>infants and<br>neonates | Substantially<br>Equivalent<br>Same Intended<br>patient population<br>as on HAMILTON-<br>C3 and HAMILTON-<br>G5 |
| Patient<br>interface | Delivered<br>invasively (via ET<br>tube) or<br>noninvasively (via<br>mask) | Delivered<br>invasively (via ET<br>tube) or<br>noninvasively (via<br>mask) | Delivered<br>invasively (via ET<br>tube) or<br>noninvasively (via<br>mask) | Substantially<br>Equivalent<br>Same patient<br>interface as on<br>HAMILTON-C3 and<br>HAMILTON-G5 |
| Power source | AC, DC, Battery | AC, DC, Battery | AC, Battery | Substantially<br>Equivalent<br>Power source is<br>unchanged. |
| Operational<br>modes | PCV+ SPONT APVcmv /<br>(S)CMV+ APVsimv /<br>SIMV+ ASV (only for<br>adult/pediatric patients) PSIMV+ DuoPAP APRV NIV NIV-ST nCPAP-PS<br>(only for<br>neonatal<br>patients) (S)CMV (only<br>for<br>adult/pediatric<br>patients) SIMV (only<br>for<br>adult/pediatric<br>patients) | PCV+ SPONT (S)CMV+<br>SIMV+ ASV (only for<br>adult/pediatric<br>patients) PSIMV+ DuoPAP APRV NIV NIV-ST nCPAP-PS<br>(only for<br>neonatal<br>patients) | (S)CMV (only<br>for adult and<br>pediatric<br>patients) P-CMV SIMV (only for<br>adult and<br>pediatric<br>patients) P-SIMV SPONT APVcmv<br>APVsimv ASV (only for<br>adult and<br>pediatric<br>patients) DuoPAP APRV NIV (only for<br>adult and<br>pediatric<br>patients) NIV-ST (only<br>for adult and<br>pediatric<br>patients) nCPAP-PS<br>(only for<br>neonatal<br>patients) VS | Substantially<br>Equivalent |
| Electrical<br>safety | IEC 60601-1: 2005<br>(3rd Edition): all<br>applicable<br>requirements met. | IEC 60601-1:1988<br>(A1:1991 +<br>A2:1995): all<br>applicable<br>requirements met. | IEC 60601-1: 2005<br>(3rd Edition): all<br>applicable<br>requirements met. | Substantially<br>Equivalent |
| Emergency air<br>intake | In case of a power<br>supply, technical,<br>or pneumatics<br>failure the ambient<br>valve allows<br>spontaneous | In case of a power<br>supply, technical,<br>or pneumatics<br>failure the ambient<br>valve allows<br>spontaneous | In case of a power<br>supply, technical,<br>or pneumatics<br>failure the ambient<br>valve allows<br>spontaneous | Substantially<br>Equivalent<br>Same emergency<br>air intake as on<br>HAMILTON-C3 and<br>HAMILTON-G5 |
| Active<br>exhalation<br>valve? | Yes, pneumatic | Yes, pneumatic | Yes, pneumatic | Substantially<br>Equivalent<br>Active exhalation is<br>unchanged. |
| Alarms and<br>monitoring | Yes | Yes | Yes | Substantially<br>Equivalent |
| Supply gas | Oxygen, ambient<br>air | Oxygen, ambient<br>air | Oxygen, Air, Heliox | Substantially<br>Equivalent |
| Method of<br>supply gas<br>pressurization | Internal turbine for<br>air, compressed<br>source for 02 | Internal turbine for<br>air, compressed<br>source for 02 | Compressed<br>source for Air, 02,<br>Heliox | Substantially<br>Equivalent |
| CO2<br>monitoring<br>option | Yes | Yes | Yes | Substantially<br>Equivalent<br>CO2 monitoring is<br>unchanged. |
| SpO2<br>monitoring<br>option | Yes | No | Yes | Substantially<br>Equivalent |
| Differences in<br>technological<br>characteristics | Embedded<br>-<br>system type:<br>EM10A<br>Internal flow<br>sensor with<br>the ability to<br>detect reflow<br>Flashing<br>-<br>alarm lamp | Embedded<br>-<br>system type:<br>EM10A<br>Internal flow<br>sensor<br>without the<br>ability to<br>detect reflow<br>Alarm lamp<br>not flashing,<br>only<br>illuminated | Embedded<br>-<br>system type:<br>EM01A<br>Internal flow<br>valve without<br>the ability to<br>detect reflow<br>Flashing<br>-<br>alarm lamp | Substantially<br>Equivalent |
Table 1: Comparison of HAMILTON-C3 with predicate device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the words "HAMILTON" and "MEDICAL" stacked on top of each other. Below the logo is the text "Intelligent Ventilation since 1983".
Image /page/5/Picture/1 description: The image shows the logo for HAMILTON-C3. The logo is written in a sans-serif font and is a bright blue color. The letters are bold and easy to read.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Hamilton Medical. The logo has the word "HAMILTON" on the top line with a symbol to the right of the word. The word "MEDICAL" is on the second line, and the phrase "Intelligent Ventilation since 1983" is on the third line. All of the text and the symbol are in blue.
Image /page/6/Picture/1 description: The image shows the word "HAMILTON.C3" in blue font. The letters are bold and rounded. The word is written in all capital letters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the word "HAMILTON" stacked on top of the word "MEDICAL". To the right of the word "HAMILTON" is a blue circle with a white "H" inside. Below the logo is the text "Intelligent Ventilation since 1983".
Image /page/7/Picture/1 description: The image shows the text "HAMILTON-C3" in a bold, sans-serif font. The text is a solid blue color. The letters are rounded and have a slightly cartoonish appearance. The word "HAMILTON" is followed by a dot and then the letter "C" and the number "3".
As can be seen in the table above, the application device HAMILTON-C3 has the same technological characteristics as the predicate HAMILTON-C3, with the exception of its internal flow sensor. The applicant device HAMILTON-C3 add new (S)CMV and SIMV ventilation mode which have same algorithms for delivering breaths as the predicate device HAMILTON-G5.
Altogether, the technological characteristics of the application device HAMILTON-C3 are substantially equivalent to the predicate devices. Thus, the comparison of the HAMILTON-C3 to its predicate devices does not raise new safety and effectiveness concerns.
#### VII. PERFORMANCE DATA
The following performance and nonclinical data are provided in support of the substantial equivalence determination.
The Software Design and Validation process, together with the bench testing of the device, demonstrated that the HAMILTON-C3 operates as intended.
In particular, testing demonstrated that the HAMILTON-C3 is compliant with the following guidelines and standards:
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Hamilton Medical. The logo is in blue and consists of the words "HAMILTON" and "MEDICAL" stacked on top of each other. To the right of the word "HAMILTON" is a circular symbol. Below the logo is the text "Intelligent Ventilation since 1983".
Image /page/8/Picture/1 description: The image shows the logo for HAMILTON-C3. The logo is written in a sans-serif font and is blue. The letters are rounded and have a playful appearance. The word "HAMILTON" is followed by a dot and then the number "C3".
- ANSI/AAMI ES60601-1 (2005/ (R) 2012): Medical electrical equipment General -Requirements for Safety
- -IEC 60601-1-2 (2014): Medical electrical equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- -ISO 80601-2-12 (2011): Medical electrical equipment - Part 2-12: Particular requirements for basic safety and essential performance of critical care ventilators
- -IEC 60601-1-8 (2006 + Am.1: 2012): Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- -IEC 60601-1-6 (2010 + A1 :2013): Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62366 (2008)+A1(2014): Medical devices Application of usability engineering to medical devices
- IEC 62304 (2006): Medical device software Software life-cycle processes -
- -ISO 80601-2-55 (2011): Medical electrical equipment -- Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors
- -ISO 80601-2-61 (2011): Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
Additional software verification and validation testing were conducted and documentation was provided as recommended by the FDA's "Guidance for Industry and FDA Staff: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Testing of the modified HAMILTON-C3, with the new features, was conducted. The new ventilation modes were subjected to waveform performance testing. The data provided from these tests was shown to be substantially equivalent to the legally marketed devices.
Since only materials already used in earlier 510(k)s are introduced with this 510(k), Hamilton Medical did not conduct any additional biocompatibility testing.
#### VIII. CONCLUSION
The results of verification, validation, and testing activities demonstrate that the modified HAMILTON-C3 ventilator is as safe and as effective as the legally marketed devices identified herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.