MOBIO Total Knee System Posterior Stabilized (PS and PS+) Tibial Inserts, MOBIO Total Knee System Patellar Component
Applicant
B-One Ortho, Corp.
Product Code
JWH · Orthopedic
Decision Date
Jan 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The b-ONE™ MOBIO™ Total Knee System is intended for total knee arthroplasty due to the following conditions: Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post -traumatic arthritis. Post-traumatic loss of knee joint configuration and function. Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. Revision of previous unsuccessful knee replacement or other procedure. Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-management techniques. Additional indications for Posterior Stabilized (PS) and Posterior Stabilized Plus (PS+) components: Ligamentous instability requiring implant bearing surface geometries with increased constraint. Absent or nonfunctioning posterior cruciate ligament. Severe anteroposterior instability of the knee joint. The b-ONE™ MOBIO™ Total Knee System is intended for implantation with bone cement only. b-ONE™ MOBIO™ Total Knee System components are not intended for use with other knee systems.
Device Story
Modular artificial knee replacement system; replaces diseased knee joint to restore function. Components include symmetric cemented femoral components, tibial baseplate, tibial inserts with locking wires, and patellar resurfacing button. PS and PS+ variants provide increased constraint for ligamentous instability. Implantation requires bone cement; system is closed (no interchangeability with other manufacturers). Intended for use by orthopedic surgeons in clinical settings. Benefits include restoration of joint function and stability for patients with degenerative or traumatic knee conditions. Subject device adds UHMWPE crosslinked with 0.1% tocopherol (Vitamin E) to existing system line.
Clinical Evidence
Bench testing only. Provided characterization of Vitamin E UHMWPE insert material, wear testing, and biocompatibility studies to support substantial equivalence.
Technological Characteristics
Modular knee prosthesis; materials include metal and UHMWPE crosslinked with 0.1% tocopherol. Features symmetric femoral components, tibial baseplate, and patellar button. Posterior stabilized (PS/PS+) design. Cemented fixation only. Supplied sterile, single-use.
Indications for Use
Indicated for patients requiring total knee arthroplasty due to painful, disabling joint disease (osteoarthritis, traumatic arthritis, avascular necrosis, rheumatoid arthritis), post-traumatic loss of joint function, moderate varus/valgus/flexion deformity, revision of failed knee replacement, or distal femur/proximal tibia fractures. PS and PS+ components indicated for ligamentous instability, absent/nonfunctioning posterior cruciate ligament, or severe anteroposterior instability.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Zimmer Persona The Personalized Knee System (K121771)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 28, 2019
b-One Ortho Corp. Allison Gecik Regulatory Affairs Manager 3 Wing Drive Suite #259 Cedar Knolls, New Jersey 07927
Re: K183025
Trade/Device Name: MOBIO Total Knee System Posterior Stabilized (PS and PS+) Tibial Inserts, MOBIO Total Knee System Patellar Component Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Codes: JWH, OIY Dated: January 17, 2019 Received: January 18, 2019
Dear Allison Gecik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is by Peter G. Allen -S. The date of the signature is 2019.01.28. The time of the signature is 13:46:53 -05'00'.
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K183025
Device Name b-ONE MOBIO Total Knee System
Indications for Use (Describe)
The b-ONE MOBIO Total Knee System is intended for total knee arthroplasty due to the following conditions:
· Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post -traumatic arthritis.
· Post-traumatic loss of knee joint configuration and function.
· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Revision of previous unsuccessful knee replacement or other procedure.
· Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture- management techniques.
Additional indications for Posterior Stabilized (PS) and Posterior Stabilized Plus (PS+) components:
· Ligamentous instability requiring implant bearing surface geometries with increased constraint.
- · Absent or nonfunctioning posterior cruciate ligament.
· Severe anteroposterior instability of the knee joint.
The b-ONE MOBIO Total Knee System is intended for implantation with bone cement only. b-ONE MOBIO Total Knee System components are not intended for use with other knee systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed announny to and review the collection of information. Send comments regarding this burden estimate or any other spect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K183025 Page 1/3
# TRADITIONAL 510(K) SUMMARY As required by 21 CFR 807.92
### Submitter Information:
| Submitter's Name: | b-ONE Ortho Corp. |
|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Address: | 3 Wing Drive |
| | Suite 259 |
| | Cedar Knolls, NJ 07927 |
| Telephone: | 866-276-4538 |
| Contact Person: | Allison Gecik |
| Telephone: | 973-587-8431 |
| Date Prepared: | October 31, 2018 |
| Proprietary Name: | b-ONE™ MOBIO™ Total Knee System |
| Classification: | Class II |
| Classification Panel: | Orthopedic |
| Common Name: | Total Knee Joint Replacement |
| Product Code(s): | JWH, OIY |
| | Classification Name(s): |
| | Regulation<br>Number |
| | Prosthesis, Knee, Patellofemorotibial, semi-constrained<br>cemented, Polymer/Metal/Polymer |
| | 888.3560 |
| | Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer + additive/metal/polymer + additive |
| | 888.3560 |
| Legally Marketed<br>Predicate Devices to Which<br>Substantial Equivalence is<br>Claimed: | b-ONE Total Knee System; K180446 |
| Legally Marketed Reference<br>Devices Used to Support<br>Substantial Equivalence: | Zimmer Persona The Personalized Knee System;<br>K121771 |
### Intended Use:
The b-ONE™ MOBIO™ Total Knee System is intended for total knee arthroplasty due to the following conditions:
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint ● disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post -traumatic arthritis.
{4}------------------------------------------------
K183025 Page 2/3
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be . returned to adequate function and stability.
- . Revision of previous unsuccessful knee replacement or other procedure
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture-● management techniques.
Additional indications for Posterior Stabilized (PS) and Posterior Stabilized Plus (PS+) components:
- Ligamentous instability requiring implant bearing surface geometries with increased constraint. ●
- Absent or nonfunctioning posterior cruciate ligament. ●
- Severe anteroposterior instability of the knee joint
The b-ONE™ MOBIO™ Total Knee System is intended for implantation with bone cement only.
b-ONE™ MOBIO™ Total Knee System components are not intended for use with other knee
systems.
# Device Description/Technological Characteristics:
The b-ONE™ MOBIO™ Total Knee System is a modular artificial knee replacement system comprised of symmetric cemented femoral components with optional femoral distal pegs, symmetric cemented tibial baseplate, symmetric tibial inserts with locking wires, symmetric patellar resurfacing button, and reusable surgical instruments. The therapeutic effect is replacement of the diseased joint with artificial components to restore joint function. The system is a posterior stabilized device two options for posterior stabilization, a PS and a PS+. The PS+ is designed with additional constraint when more stability is required. Compatibility of the system components is only claimed with the b-ONE Total Knee System. There is no allowed interchangeability with systems manufactured by other companies.
The purpose of this submission is to add a line extension to the existing system to offer the Tibial Inserts and Patella components in UHMWPE Crosslinked with 0.1% tocopherol.
All system components are supplied sterile and are single use devices.
# Comparison of Technological Characteristics
The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The b-ONE™ MOBIO™ Total Knee System and the predicate devices share the following characteristics:
- Materials of construction
- Manufacturing processes
- Sizes offered ●
- Product design for shape and macrostructures
- Sterilization methods .
{5}------------------------------------------------
#### K183025 Page 3/3
# Performance Data
The following performance data were provided in support of the substantial equivalence determination. All performance testing performed for the predicate system was conducted on the subject material and therefore is applicable to the subject devices. Additional performance testing condiucted to support the Vitamin E UHMWPE material is listed below.
Non-Clinical Studies
- Characterization of Vitamin E UHMWPE Insert Material ●
- Wear Testing
- Biocompatibility ●
# Conclusion
The information provided above supports that the b-ONE™ Total Knee System is as safe and effective as the predicate devices with the same intended use. Applicable reference devices have been cited to support the conclusion that the addition of new material does not raise any new questions of safety and effectiveness. The b-ONE™ MOBIO™ Total Knee System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.