K183024 · Implacil DE Bortoli Material Odontologico Ltda · DZE · Feb 13, 2020 · Dental
Device Facts
Record ID
K183024
Device Name
Implacil Implant System
Applicant
Implacil DE Bortoli Material Odontologico Ltda
Product Code
DZE · Dental
Decision Date
Feb 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Implacil Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit and/or multi-unit restorations. When a one-stage surgical approach is applied, the Implacil Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
The Implacil Implant System consists of endosseous dental implants and prosthetic components (abutments, screws, copings) used to support dental restorations. The system includes three interface types: External Hex (HE), Internal Hex (HI), and Morse Taper (CM AR). Implants are surgically placed in the maxillary or mandibular arch by dental professionals. The device provides a foundation for single or multi-unit prosthetics, including crowns, bridges, and overdentures. The system supports both one-stage and two-stage surgical protocols, with immediate loading permitted under conditions of high primary stability. Prosthetic components are available in various designs (straight, angled, cementable, UCLA, ball) to accommodate clinical needs. The system benefits patients by restoring masticatory function and aesthetics in edentulous or partially edentulous sites. Clinical decision-making is guided by the provider's assessment of bone quality and stability.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 11137-1, ISO 17665-1/2), shelf-life/package integrity (ASTM F1980, F1929, F88), biocompatibility (ISO 10993-12), pyrogenicity (AAMI/ANSI ST72), and mechanical performance (dynamic fatigue per ISO 14801, static torsional loading per ISO/TS 13498:2011). No clinical data were included.
Technological Characteristics
Implants made of CPTi (ASTM F67); prosthetic components made of CPTi (ASTM F67) or Ti alloy (ASTM F136). Interfaces: External Hex, Internal Hex, and Morse Taper (CM AR). Surface treatment: grit-blasted and acid-etched. Sterilization: irradiation or steam (for non-sterile components).
Indications for Use
Indicated for patients requiring dental implants in the maxillary or mandibular arch for single or multi-unit restorations. Suitable for immediate loading in one-stage surgical procedures provided good primary stability and appropriate occlusal loading are achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2020
Implacil de Bortoli Material Odontologico Ltda c/o Janine Treter Regulatory Affairs Specialist PR Serviços Regulatórios Administrativos Ltda Rua Alice Além Saadi, 855/ 2402 Ribeirão Preto, São Paulo 14096-570 BRAZIL
Re: K183024
Trade/Device Name: Implacil Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 13, 2020 Received: January 16, 2020
Dear Janine Treter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K183024
Device Name
Implacil Implant System
Indications for Use (Describe)
Implacil Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit and/or multi-unit restorations. When a one-stage surgical is applied, the Implacil Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) | <table><tr><td><span> <span style="font-size:18px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> <span style="font-size:18px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <span> <span style="font-size:18px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### ADMINISTRATIVE INFORMATION
| Sponsor | |
|---------|--|
|---------|--|
| Sponsor | Implacil de Bortoli Material Odontologico Ltda |
|---------|------------------------------------------------|
| | Rua Vicente de Carvalho 178-182 |
| | São Paulo, São Paulo, Brazil 01521020 |
Contact Person and Preparer Janine Treter, PhD Regulatory Affairs Specialist Passarini Regulatory Affairs PR Serviços Regulatórios Administrativos Ltda E-Mail: janine@rapassarini.com.br Telephone +55 (47) 3804 0075
Date Prepared
13/Feb/2020
## DEVICE NAME AND CLASSIFICATION
| Trade/ Proprietary Name | Implacil Implant System |
|-----------------------------------|-----------------------------|
| Common Name | Dental implant and abutment |
| Primary Classification Name | Endosseous dental implant |
| Primary Classification Regulation | 21 CFR 872.3640, Class II |
| Primary Product Code | DZE, NHA |
| Classification Panel | Dental Products Panel |
|----------------------|-----------------------|
| Reviewing Branch | Dental Devices Branch |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device
Reference Devices
K170398 – S.I.N. Dental Implant System – S.I.N. Sistema de Implante Nacional S.A. K062931 – IMPLUS - LEADER Italia S.R.O K072642 - BIOMET 3i Dental Abutments and Restorative Components - Biomet 3i, Inc.
K090452 - Neoss Implant System 03,25 - Neoss Limited
K101207 - Neodent Implant System -JJGC Industria e Comercio de Materiais Dentarios S.A.
K101945- Neodent Implant System - JJGC Industria e Comercio de Materiais Dentarios S.A.
{4}------------------------------------------------
K133510 - Neodent Implant System - JJGC Industria e Comercio de Materiais Dentarios S.A.
K140440 - Noris Medical Dental Implants System - Noris Medical, Ltd.
K163060 -Deep Conical (DC) Implants and Accessories - Southern Implants (Pty) Ltd
K163634 - External Hex Implants - Southern Implants (Pty) Ltd
K170392 - S.I.N. Dental Implant System - S.I.N. Sistema de Implante Nacional S.A.
K170608 - UF(II) Implant System - DIO Corporation
K173819 – MyPlant II Implant System Hager & Meisinger GmbH
K173902 - Neodent Implant System - GM Line - JJGC Industria e Comercio de Materiais Dentarios S.A.
K180282 - MIS Internal Hex Dental Implant System -MIS Implants Technologies Ltd.
## INDICATIONS FOR USE
Implacil Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit and/or multi-unit restorations. When a one-stage surgical approach is applied, the Implacil Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
## SUBJECT DEVICE DESCRIPTION
Implacil Implant System is composed of three implant lines that are divided according to the implant-toabutment interface: External Hex (HE), Internal Hex (HI) and Morse Taper AR Due Cone (CM AR). HE and HI lines are composed of tissue-level implants while CM AR line of bone-level implants. Each implant line is composed of implants and related prosthetic components available in multiple designs (temporary, screwed, cementable, angled, straight, UCLA, ball).
HE implant line implants are available in two root-forms designs: conical (tapered) and cylindrical. Conical implants are available in three diameters and platforms (3.5, 4.0 and 5.0 mm) and five lengths (7.0, 9.0, 11.0, 13.0 and 15.0 mm). Cylindrical implants are provided in four diameters (3.3, 3.75, 4.0 and 4.75 mm), three platforms (3.5, 4.0 and 5.0 mm) and five lengths (8.0, 10.0, 11.5, 13.0 and 15.0 mm). HE cylindrical implants of diameters 3.75 and 4.0 share the same platform of 4.0 mm.
HI implant line implants are available in two root-forms designs: conical (tapered) and cylindrical. Conical implants are available in three diameters and platforms (3.5, 4.0 and 5.0 mm). For diameters 3.5 and 4.0 are available in the lengths 8.0, 9.0, 11.0, 13.0 and 15.0 mm, and for diameters 5.0 are available in the lengths 7.0, 9.0, 11.0, 13.0 and 15.0 mm. Cylindrical implants are provided in four diameters (3.3, 3.75, 4.3 and 4.75 mm) and three platforms (3.5, 4.0 and 5.0 mm). For diameter 3.3 are available in the lengths 8.0, 9.0, 11.0, 13.0 and 15.0 mm, and for diameters 3.75 and 4.3 are available in the lengths 7.0, 9.0, 11.0, 13.0 and 15.0 mm. HI cylindrical implants of diameters 3.75 and 4.3 share the same platform of 4.0 mm.
{5}------------------------------------------------
CM AR implant line implants are available in conical root-form design only, in four diameters (3.5, 4.0, 4.5 and 5.0 mm) and five lengths (7.0, 9.0, 11.0, 13.0 and 15.0 mm).
lmplacil implants are made of commercially pure titanium (ASTM F67). Implacil prosthetic components are made of commercially pure titanium (ASTM F67) or titanium alloy (ASTM F136). Implant System screws (abutment screw, UCLA screws and coping screws) are made of titanium alloy (ASTM F136).
The subject device abutments components mate exclusively with the subject implants of the same line (HJ, HE, CM AR).
# TECHNOLOGICAL CHARACTERISTICS
The subject device and the predicate devices have the same intended use as and technological characteristics as shown in the tables below. Differences in the design features between the subject devices and the primary predicate device K170398 are addressed by comparison to the reference devices.
{6}------------------------------------------------
| Table 5.1: Comparison on indication for use statement | | |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT |
| SUBJECT DEVICE | K183024 – Implacil Implant System<br>Implacil de Bortoli Material Odontologico<br>Ltda | Implacil Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit and/or multi-unit<br>restorations. When a one-stage surgical approach is applied, the Implacil Implant System is intended for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. |
| PRIMARY PREDICATE<br>DEVICE | K170398 - S.I.N. Dental Implant System<br>S.I.N. Sistema de Implante Nacional S.A | S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit<br>restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. Revolution Compact with a 6 mm length is intended for delayed loading<br>only. |
| REFERENCE DEVICES | K062931 – IMPLUS<br>LEADER Italia S.R.O | IMPLUS implant fixtures are intended to be surgically placed in the bone of the mandibular and/or maxillary dental arches in order to provide<br>support for fixed and/or removal prosthetics in order to restore original features and masticatory functions.<br>Implus implant fixtures are indicated for permanent use.<br>Implus implant fixtures are disposable and are for one-time use only. These fixtures are not to be recleaned or re-sterilized. |
| | K072642 – BIOMET 3i Dental Abutments and<br>Restorative Components<br>Biomet 3i, Inc. | BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implant to support a prosthetic device in a partially or<br>completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or<br>maxilla. The prosthesis can be screw retained or cement retained.<br>Restorative Components<br>* Temporary Healing Abutments are intended for use to shape and maintain the soft tissue opening during healing.<br>* Castable restorative components are intended for use as accessories to endosseous dental implants to aid in the fabrication of dental<br>prosthetics.<br>* Screw components are intended for use as accessories to endosseous dental implants for retention of screw retained abutments to the<br>dental implant. |
| | K090452 – Neoss Implant System 03,25<br>Neoss Ltd | The Neoss Implant System 03,25 is for single-stage and two-stage surgical procedures and cement or screw retained restorations.<br>The Neoss Implant System 03,25 are intended for immediate loading on single tooth and /or multiple tooth applications recognizing sufficient<br>bone stability and appropriate occlusal loading, to restore chewing function. Multiple tooth applications may be splinted with a bar.<br>The Neoss Implant 03,25 abutments are designed to be connected to the Neoss implants and intended for use as an aid in prosthetic<br>rehabilitation. |
| | K101207 - Neodent Implant System<br>JJGC Industria e Comercio de Materiais<br>Dentarios S.A. | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic<br>devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple<br>unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple<br>tooth applications may be rigidly splinted. |
| | KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT |
| REFERENCE DEVICES | K101945- Neodent Implant System<br>JJGC Industria e Comercio de Materiais<br>Dentarios S.A. | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic<br>devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple<br>unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple<br>tooth applications may be rigidly splinted. |
| | K133510 - Neodent Implant System<br>JJGC Industria e Comercio de Materiais<br>Dentarios S.A. | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lowerjaw to provide support for prosthetic<br>devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple<br>unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Multiple<br>tooth applications may be rigidly splinted. |
| | K163060 -Deep Conical (DC) Implants and<br>Accessories<br>Southern Implants (Pty) Ltd | Southern Implants Dental Implants are intended for both one- and two-stage surgical procedures in the following situations and with the<br>following clinical protocols:<br>• replacing single and multiple missing teeth in the mandible and maxilla,<br>• immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge,<br>• immediate loading in all indications, except in single tooth situations on implants shorter than 8mm or in soft bone (type IV) where<br>implant stability may be difficult to obtain and immediate loading may not be appropriate.<br>The intended use for 3.0 Deep Conical implants is limited to replacement of maxillary lateral incisors and mandibular incisors |
| | K163634 - External Hex Implants<br>Southern Implants (Pty) Ltd | Southern Implants' External Hex Implants are intended for surgical placement in the upper or lower<br>jaw to provide a means for prosthetic attachment of crowns, bridges or overdentures utilizing delayed or immediate loading.<br>Southern Implants' External Hex Implants are intended for immediate function when good primary<br>stability with appropriate occlusal loading is achieved. |
| | K170608 - UF(II) Implant System<br>DIO Corporation | The UF(II) Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or<br>multiple units' prosthetic attachment to restore a patient's chewing function. The UF(II) Implant System (Ø3.8 ~ Ø5.5) can be placed with a<br>conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for<br>immediate loading when good primary stability is achieved with appropriate occlusal loading. |
| | K170392 - S.I.N. Dental Implant System<br>S.I.N. Sistema de Implante Nacional S.A | S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit<br>restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. Implants with lengths less than 7 mm are intended for delayed loading<br>only. |
| | K173819 – MyPlant II Implant System<br>Hager & Meisinger GmbH | The MyPlant II implants are surgically placed in the maxilla or mandible to enable prosthetic restorations in edentulous or partially edentulous<br>patients. The implants are to be used exclusively with MyPlant II abutments and prosthetic components. The abutments serve for prosthetic<br>restorations and can include individual crowns, bridges, partial or full prostheses. Abutments can be used for single tooth restorations or for<br>the restoration of several teeth. The implants are intended for delayed loading with two surgical interventions. In case of appropriate primary<br>stability (35 Ncm), immediate temporary restoration with appropriate occlusal load can also be performed. |
| | KNUMBER/ MANUFACTURER | INDICATION FOR USE STATEMENT |
| REFERENCE DEVICES | K173902 - Neodent Implant System - GM<br>Line | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic<br>devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple<br>unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. |
| | JJGC Industria e Comercio de Materiais<br>Dentarios S.A. | |
| | K180282 - MIS Internal Hex Dental Implant<br>System | MIS dental implant systems are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic<br>devices, such as artificial teeth, in order to restore masticatory function. When a one-stage surgical procedure is applied, the implant may be<br>immediately loaded when good primary stability is achieved and the occlusal load is appropriate. |
| | MIS Implants Technologies Ltd. | Narrow implants (Ø3.3mm & UNO) are indicated for use in surgical and restorative applications for placement only in the mandibular central,<br>lateral incisor and maxillary lateral incisor regions of partially edentulous jaws, to provide support for prosthetic devices such as artificial<br>teeth. Mandibular central and lateral incisors must be splinted if using two or more narrow implants adjacent to one another.<br>The long MIS (18 & 20 mm) implants can be used in a tilted manner.<br>MIS short implants are to be used only with straight abutments.<br>M4 short implants are indicated for delayed loading only. |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Page 7 of 22
| Table 5.2: SE comparison on HE implants | | | | | |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SUBJECT DEVICES | PRIMARY PREDICATE DEVICE | | REFERENCE DEVICES | |
| Trade Name | K183024 - Implacil Implant<br>System<br>Implacil de Bortoli Material<br>Odontologico Ltda | K170398 - S.I.N. Dental Implant<br>System<br>S.I.N. Sistema de Implante<br>Nacional S.A. | K101207 – Neodent Implant<br>System<br>JJGC Industria e Comercio de<br>Materiais Dentarios S.A. | K062931 - IMPLUS<br>LEADER Italia S.R.O. | K163634 – External Hex Implant<br>Southern Implants (Pty) Ltd.<br>November 6, 2017 |
| Information | | | | | |
| Implant-to-abutment connection | HE interface | HE interface | HE interface | HE interface | HE interface |
| Raw Material | CPTi | CPTi | CPTi | CPTi | CPTi |
| Surface | Grit-blasted and acid-etched | Acid-etched and HA | Grit-blasted and acid-etched | Sand-blasted and acid-etched | Grit-blasted and acid-etched<br>Machine collar versions<br>available |
| Root type design | Cylindrical | Cylindrical | Cylindrical | Cylindrical | Cylindrical |
| | Tapered | | Tapered | Tapered | Tapered |
| Implant diameter (mm) | Cylindrical:<br>3.3: 8, 10, 11.5, 13, 15<br>3.75: 8, 10, 11.5, 13, 15<br>4.0: 8, 10, 11.5, 13, 15<br>4.75: 8, 10, 11.5, 13, 15<br><br>Tapered:<br>3.5: 7, 9, 11, 13, 15<br>4.0: 7, 9, 11, 13, 15<br>5.0: 7, 9, 11, 13, 15 | Cylindrical:<br>3.25: 8.5, 10, 11.5, 13, 15<br>3.5: 7, 8.5, 10, 11.5, 13, 15<br>3.75: 7, 8.5, 10, 11.5, 13, 15<br>4.0: 6, 7, 8.5, 10, 11.5, 13, 15<br>4.5: 8.5, 10, 11.5, 13, 15<br>5.0: 6, 7, 8.5, 10, 11.5, 13, 15 | Cylindrical:<br>3.3: 9, 11, 13, 15, 17<br>3.75: 9, 11, 13, 15, 17, 19<br>4.0: 9, 11, 13, 15, 17, 19<br>4.5: 9, 11, 13<br>5.0: 7, 9, 11, 13<br><br>Tapered:<br>3.5: 10, 13, 16<br>4.3: 10, 13, 16<br>5.0: 10, 13, 16 | Cylindrical:<br>3.3: 10, 11.5, 13, 15<br>3.75: 8, 10, 11.5, 13, 15, 18, 20<br>5.0: 8, 10, 11.5, 13<br><br>Tapered:<br>4.0: 8, 10, 11.5, 13, 15<br>5.0: 8, 10, 11.5, 13 | Cylindrical:<br>3.75: 7, 8.5, 10, 11.5, 13, 15, 18,<br>20<br>5.0: 6, 7, 8.5, 10, 11.5, 13, 15<br>6.0: 7, 8.5, 10, 11.5, 13, 15<br><br>Tapered:<br>3.25: 8.5, 10, 11.5, 13, 15, 18<br>4.0: 6, 8.5, 10, 11.5, 13, 15<br>4.7: 10, 11.5, 13, 15, 18<br>5.0: 6, 8.5, 10, 11.5, 13, 15<br>5.7: 10, 11.5, 13, 15, 18 |
| Sterilization | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation |
{10}------------------------------------------------
Page 8 of 22
| Table 5.3: SE comparison on HI implants | | | | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SUBJECT DEVICES | PRIMARY PREDICATE DEVICE | REFERENCE DEVICES | |
| Trade Name<br>Information | K183024 – Implacil Implant System | K170398 - S.I.N. Dental Implant System | K062931 - IMPLUS | K180282 - MIS Internal Hex Dental Implant<br>System |
| | Implacil de Bortoli Material Odontologico<br>Ltda | S.I.N. Sistema de Implante Nacional S.A. | LEADER Italia S.R.O. | MIS Implants Technologies Ltd. |
| Implant-to-abutment<br>connection | HI interface | HI interface | HI interface | HI interface |
| Raw Material | CPTi | CPTi | CPTi | Ti alloy (TI-6AI-4V ELI) |
| Surface | Grit-blasted and acid-etched | Acid-etched and HA | Sand-blasted and acid-etched | Anodized, sand blasted and acid etched<br>Sand blasted and acid etched |
| Root type design | Cylindrical<br>Tapered | Cylindrical | Cylindrical<br>Tapered | Cylindrical<br>Tapered |
| Diameter: length (mm) | Cylindrical<br>3.3: 8, 9, 11, 13, 15<br>3.75: 7, 9, 11, 13, 15<br>4.3: 7, 9, 11, 13, 15<br>4.75: 7, 9, 11, 13, 15<br>Tapered:<br>3.5: 8, 9, 11, 13, 15<br>4.0: 8, 9, 11, 13, 15<br>5.0: 7, 9, 11, 13, 15 | Cylindrical<br>3.8: 8.5, 10, 11.5, 13, 15<br>4.5: 8.5, 10, 11.5, 13, 15<br>5.0: 8.5, 10, 11.5, 13, 15 | Cylindrical<br>3.3: 8, 10, 11.5, 13, 16<br>3.75: 8, 10, 11.5, 13, 16<br>4.5: 8, 10, 11.5, 13, 16<br>5.5: 8, 10, 11.5, 13<br>Tapered:<br>4.0: 8, 10, 11.5, 13, 16<br>5.0: 8, 10, 11.5, 13, 16<br>6.0: 8, 10, 11.5, 13 | Cylindrical<br>3.3: 10, 11.5, 13, 16<br>3.75: 8, 10, 11.5, 13, 16, 18, 20<br>4.2: 6, 8, 10, 11.5, 13, 16, 18, 20<br>5.0: 6, 8, 10, 11.5, 13, 16<br>6.0: 6, 8, 10, 11.5, 13<br>Tapered:<br>3.3: 10, 11.5, 13, 16<br>3.75: 8, 10, 11.5, 13, 16, 18, 20<br>4.2: 8, 10, 11.5, 13, 16, 18, 20<br>5.0: 8, 10, 11.5, 13, 16<br>6.0: 8, 10, 11.5, 13 |
| Sterilization | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation |
{11}------------------------------------------------
Page 9 of 22
| Table 5.4: SE comparison on CM AR implants | | | | |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SUBJECT DEVICES | REFERENCE DEVICES | | |
| Trade Name<br>Information | K183024 – Implacil Implant System | K101945 – Neodent Implant System | K170608 – UF(II) Implant System | K163060 -Deep Conical (DC) Implants and<br>Accessories |
| | Implacil de Bortoli Material Odontologico<br>Ltda | JJGC Industria e Comercio de Materiais<br>Dentarios S.A. | DIO Corporation | Southern Implants (Pty) Ltd |
| Implant-to-abutment<br>connection | Conical connection (CM AR) | Conical connection (CM) | Conical connection (CM) | Conical connection (Deep Conical) |
| Raw Material | CPTi | CPTi | CPTi | CPTi |
| Surface | Grit-blasted and acid-etched | Grit-blasted and acid-etched | Hybrid Sand-blasted and acid-etched | Roughened surface |
| Root type design | Tapered | Cylindrical<br>Tapered | Tapered | Cylindrical<br>Tapered |
| Implant-to-abutment<br>indexation feature | Dodecagon (double-hexagon) | Hexagon | Hexagon | Dodecagon (double-hexagon) |
| Implant diameter: lenghts (mm) | Tapered:<br>3.5: 7, 9, 11, 13, 15<br>4.0: 7, 9, 11, 13, 15<br>4.5: 7, 9, 11, 13, 15<br>5.0: 7, 9, 11, 13, 15 | Cylindrical:<br>3.5: 7, 8, 9, 11, 13, 15, 17<br>3.75: 7, 8, 9, 11, 13, 15, 17<br>4.0: 7, 8, 9, 11, 13, 15, 17, 19<br>5.0: 7, 8, 9, 11, 13<br>Tapered:<br>3.5: 8, 10, 11.5, 13, 16<br>4.3: 8, 10, 11.5, 13, 16<br>5.0: 8, 10, 11.5, 13, 16 | Tapered:<br>3.8: 8.5, 10, 11.5, 13, 15, 16<br>4.0: 8.5, 10, 11.5, 13, 15, 16<br>4.5: 7, 8.5, 10, 11.5, 13, 15, 16<br>5.0: 7, 8.5, 10, 11.5, 13, 15, 16<br>5.5: 7, 8.5, 10, 11.5, 13, 15, 16 | Cylindrical:<br>3.0: 11, 13, 15<br>3.5: 8, 9, 11, 13, 15<br>4.0: 6, 8, 9, 11, 13, 15<br>5.0: 9, 11, 13, 15<br>Tapered:<br>3.0: 9, 11, 13, 15<br>3.5: 8, 9, 11, 13, 15<br>4.0: 6, 8, 9, 11, 13, 15<br>5.0: 9, 11, 13, 15 |
| Sterilization | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation | Provided sterile by irradiation |
{12}------------------------------------------------
Page 10 of 22
| Table 5.5: SE comparison on Covers | | | | | |
|--------------------------------------------------|-----------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------|----------------------------------|
| | SUBJECT DEVICES | REFERENCE DEVICES | | | |
| Trade Name | K183024 - Implacil Implant System | K101207 - Neodent Implant<br>System | K180282 - MIS Internal Hex<br>Dental Implant System | K101945 – Neodent Implant<br>System | K170608 - UF(II) Implant System |
| Information | Implacil de Bortoli Material<br>Odontologico Ltda | JJGC Industria e Comercio de<br>Materiais Dentarios S.A. | MIS Implants Technologies Ltd. | JJGC Industria e Comercio de<br>Materiais Dentarios S.A. | DIO Corporation |
| Implant-to-abutment connection<br>(Implant line) | HE interface<br>HI interface<br>Conical interface (CM AR) | HE interface | HI interface | Conical interface (CM) | Conical interface (CM) |
| Diameter (mm) | HE: 3.5, 4.0, 5.0<br>HI: 3.5, 4.0, 5.0<br>CM AR: 2.5 | HE: 3.3, 4.1, 4.3, 5.0 | HI: 3.3, 3.75, 4.7 | CM: 2.5 | CM: 2.7, 2.794, 3.6, 3.8 |
| Gingival Height (mm) | HE: 0<br>HI: 0<br>CM AR: 2 | HE: 0 | HI: 0 | CM: 0, 2 | CM: 0, 1, 2, 3 |
| Raw material | CPTi | Ti alloy | Ti alloy | Ti alloy | CPTi |
| Surface | Machined | Machined | Machined and Anodized | Machined | Machined and anodized |
| Angulation | 0° | 0° | 0° | 0° | 0° |
| Load | No occlusal load | No occlusal load | No occlusal load | No occlusal load | No occlusal load |
| Sterility | Provided sterile by irradiation. | Provided sterile by ethylene<br>oxide. | Provided sterile by irradiation | Provided sterile by ethylene<br>oxide. | Provided sterile by irradiation. |
{13}------------------------------------------------
Page 11 of 22
| Table 5.6: SE comparison on Healing Abutments | | | | | |
|-----------------------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Trade Name Information | SUBJECT DEVICES | PRIMARY PREDICATE DEVICES | REFERENCE DEVICES | | |
| | K183024 – Implacil Implant System<br>Implacil de Bortoli Material<br>Odontologico Ltda | K170398 – S.I.N. Dental Implant System<br>S.I.N. Sistema de Implante<br>Nacional S.A. | K101207 – Neodent Implant System<br>JJGC Industria e Comercio de<br>Materiais Dentarios S.A. | K180282 – MIS Internal Hex Dental<br>Implant System<br>MIS Implants Technologies Ltd. | K101945 – Neodent Implant System<br>JJGC Industria e Comercio de<br>Materiais Dentarios S.A. |
| Implant-to-abutment connection | HE interface<br>HI interface<br>Conical interface (CM AR) | HE interface<br>HI interface | HE interface | HI interface | Conical interface (CM) |
| Diameter (mm) | HE: 3.5, 4.0, 5.0<br>HI: 3.5, 4.0, 5.0<br>CM AR: 3.5, 4.5 | HE: 3.6, 4.1, 5.0 | HE: 3.3, 4.1, 4.3, 5.0 | HI: 4.0, 4.3, 4.8, 5.0, 5.5, 6.5 | CM: 3.5, 4.0 |
| Gingival Height (mm) | HE: 2, 3, 4, 5, 6, 7<br>HI: 2, 3, 4, 5, 6, 7<br>CM AR: 0.8, 1.5, 2.5, 3.5, 4.5, 5.5 | HE: 1, 2, 4, 6, 8 | HE: 2, 3, 4, 5, 6 | HI: 2, 3, 4, 5, 6, 8 | CM: 0.8, 3.5, 4.5, 5.5, 6.5 |
| Raw material | CPTi | Ti alloy | Ti alloy | Ti alloy | Ti alloy |
| Surface | Machined | Machined | Machined | Machined and Anodized | Machined |
| Angulation | 0° | 0° | 0° | 0° | 0° |
| Load | No occlusal load | No occlusal load | No occlusal load | No occlusal load | No occlusal load |
| Sterility | Provided sterile by irradiation. | Provided sterile by irradiation. | Provided sterile by ethylene<br>oxide. | Provided sterile by irradiation | Provided sterile by ethylene<br>oxide. |
#### Table 5.6: SE comparison on Healing Abutments
{14}------------------------------------------------
## Table 5.7: SE comparison on Healing Abutment Covers
| | SUBJECT DEVICES | PRIMARY PREDICATE DEVICES | REFERENCE DEVICES | |
|--------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Trade Name<br>Information | K18…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.