Neodent Implant System – GM Line

K173902 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · NHA · Apr 20, 2018 · Dental

Device Facts

Record IDK173902
Device NameNeodent Implant System – GM Line
ApplicantJjgc Industria E Comercio DE Materiais Dentarios S.A.
Product CodeNHA · Dental
Decision DateApr 20, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.

Device Story

Neodent Implant System - GM Line components; includes GM Exact Co-Cr Abutment for Crown and GM Equator Attachment. Abutments interface with previously cleared GM implants via Morse taper connection. Exact Co-Cr Abutment: Co-Cr alloy base with polyoxymethylene (POM) cylinder for cast-to prosthesis fabrication; single-unit restorations; cement or screw-retained. GM Equator Attachment: Titanium alloy ball-type abutment for partial/full overdentures. Used in dental clinics by clinicians. Components provide mechanical support for prosthetic restorations; restore chewing function; improve patient quality of life through dental rehabilitation.

Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical bench testing: biocompatibility (ISO 10993-5, -12, -18), sterilization validation (ISO 17665-1, -2, ISO 11135, ISO 10993-7), sterile barrier shelf life (seal strength, helium leak, sterility), and bacterial endotoxin testing (AAMI/ANSI ST72).

Technological Characteristics

Materials: Co-Cr-Mo alloy (ASTM F1537) for Exact Co-Cr Abutment; Ti-6Al-4V alloy (ASTM F136) with TiN coating for Equator Attachment. Interface: Morse taper with internal hex indexing. Sterilization: Ethylene oxide (Equator); non-sterile (Exact Co-Cr).

Indications for Use

Indicated for patients requiring dental implants in the upper or lower jaw to support prosthetic devices (artificial teeth) and restore chewing function. Suitable for single or multiple unit restorations in single-stage or two-stage procedures; immediate loading permitted given good primary stability and appropriate occlusal loading.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. JJGC Industria e Comercio de Materiais Dentarios S.A. Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130 April 20, 2018 Re: K173902 Trade/Device Name: Neodent Implant System - GM Line Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 22, 2018 Received: March 22, 2018 # Dear Kevin Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Mary S. Runner -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173902 Device Name Neodent Implant System - GM Line Indications for Use (Describe) The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ # 510(k) Summary # K173902 ### Neodent Implant System - GM Line ## JJGC Indústria e Comércio de Materiais Dentários SA ### April 18, 2018 #### ADMINISTRATIVE INFORMATION | Manufacturer Name | JJGC Indústria e Comércio de Materiais Dentários SA<br>Av. Juscelino Kubitschek de Oliveira, 3291 - CIC<br>Curitiba, Paraná, 81270-200, Brazil | | | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--------------------| | | Telephone: | +55 41 2169 1000 | | | | Fax: | +55 41 2169 4061 | | | Official Contact | Julianne de Oliveira Capucho Lechechem<br>Regulatory Affairs Manager | | | | Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | | | | Telephone: | +1-858-792-1235 | | | | Fax: | +1-858-792-1236 | | | | Email: | kthomas@paxmed.com | flarson@paxmed.com | #### DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name | Neodent Implant System - GM Line | |---------------------------|------------------------------------| | Common Name | Dental implant abutment | | Classification Name | Endosseous dental implant abutment | | Classification Regulation | 21 CFR 872.3630, Class II | | Product Code | NHA | | Classification Panel | Dental Products Panel | | Reviewing Branch | Dental Devices Branch | #### PREDICATE DEVICE INFORMATION The primary predicate device is K163194. The reference predicate devices are K101207, K121843, and K133696. #### INDICATIONS FOR USE The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. {4}------------------------------------------------ # SUBJECT DEVICE DESCRIPTION The purpose of this submission is to expand the Neodent Implant System - GM Line components cleared under K163194, which included dental implants with a Morse taper abutment interface (called Grand Morse, or GM), mating abutment screws, and other associated components. This submission includes the GM Exact Co-Cr Abutment for Crown and the GM Equator Attachment. The subject device components mate exclusively with the GM implants cleared in K163194. The GM Exact Co-Cr Abutment for Crown is provided in three sizes to fit the prosthetic diameter sizes of the GM implants cleared in K163194. The subject device abutments consist of a Co-Cr alloy component that interfaces directly to the implant and a cvlinder of polyoxymethylene (POM) for fabrication of a cast prosthesis by a burn-out technique. The subject device abutments are provided in three (3) sizes to match the various platforms of the mating GM implants: 3.5/3.75 mm platform diameter, 4.0/4.3 mm platform diameter, and 5.0/6.0 mm platform diameter. The subject device abutments are for single-unit prosthetic restorations only. The final prosthetic restoration may be cement-retained or screw-retained. The GM Equator Attachment abutments are straight, ball-type abutments for the attachment of full or partial overdentures. GM Equator Attachments are provided in five gingival heights (1.5, 2.5, 3.5, 4.5, and 5.5 mm), and may be used with all GM implant diameter sizes cleared in K163194. The subject device GM Exact Co-Cr abutments are made of Co-Cr alloy conforming to ASTM F1537, with mating abutment screws made of titanium alloy conforming to ASTM F136. The GM Equator Attachments made of titanium alloy conforming to ASTM F136 with a TiN coating identical to the TiN coating cleared in K133696. ### PERFORMANCE DATA Non-clinical data submitted to demonstrate substantial equivalence include biocompatibility testing of the subject device GM Exact Co-Cr Abutment for Crown according to ISO 10993-5, ISO 10993-12 and ISO 10993-18. Non-clinical data relied upon to demonstrate substantial equivalence with regard to biocompatibility of titanium alloy components were incorporated by reference to K163194 and K133696. Sterilization validation testing of subject devices was performed according to ISO 17665-1. ISO TS 17665-2, ISO 11135, and ISO 10993-7. Sterile barrier shelf life testing was performed on real time aged samples (referenced from K163194) and included seal strength testing, helium leak detection, and sterility testing of the package contents. Bacterial endotoxin testing was performed using methods described in AAMI / ANSI ST72. Clinical data were not submitted in this premarket notification. ### EQUIVALENCE TO MARKETED DEVICES JJGC Indústria e Comércio de Materiais Dentários S.A. submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices: K163194, Neodent Implant System - GM Line, JJGC Indústria e Comércio de Materiais Dentários S.A.; K101207, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A.; K121843, NP-Cast Abutment System, OSSTEM Implant Co., Ltd.; and K133696, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários S.A. A comparison of the Indications for Use and the technological characteristics of the subject device and the primary predicate device K163194 is provided in the following table. {5}------------------------------------------------ | | Subject Device | Primary Predicate Device | |---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison | Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | K163194<br>Neodent Implant System - GM Line<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | | Indications for Use<br>Statement | The Neodent Implant System is intended to be surgically placed<br>in the bone of the upper or lower jaw to provide support for<br>prosthetic devices such as artificial teeth, to restore chewing<br>function. It may be used with single-stage or two-stage<br>procedures, for single or multiple unit restorations, and may be<br>loaded immediately when good primary stability is achieved and<br>with appropriate occlusal loading. | Indications for Use for GM implants and conventional<br>abutments:<br>The Neodent Implant System is intended to be surgically placed<br>in the bone of the upper or lower jaw to provide support for<br>prosthetic devices such as artificial teeth, to restore chewing<br>function. It may be used with single-stage or two-stage<br>procedures, for single or multiple unit restorations, and may be<br>loaded immediately when good primary stability is achieved<br>and with appropriate occlusal loading.<br>Indications for Use for GM Titanium Base abutments:<br>Titanium Base Abutment is a titanium base placed onto Neodent<br>dental implants to provide support for customized prosthetic<br>restorations. It is used with a coping and crown, or crown alone,<br>and is indicated for cement-retained single or multi-unit<br>restorations, or screw-retained single restorations. All digitally<br>designed copings and/or crowns for use with the Neodent<br>Titanium Base Abutment System are intended to be sent to<br>Straumann for manufacture at a validated milling center.<br>Indications for Use for GM Pro Peek Abutments:<br>The Pro PEEK Abutments are indicated to be used on Neodent<br>implants to provide temporary support for prosthesis structure<br>for up to 6 months. They can be used in one or two stage<br>procedures and also immediate load when there is good primary<br>stability. | | Product Code | NHA | DZE, NHA | | Intended Use | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | | Design | | | | Implant Design | | Threaded root-form implants, 3 thread designs;<br>Internal Morse taper abutment connection | | Implant Diameter | Not applicable | Various, 3.5 mm to 5.0 mm | | Implant Length | | Various, 7 mm to 18 mm | | Abutment Design | GM Exact Co-Cr Abutment for Crown:<br>Component for cast-to fabrication<br>GM Equator Attachment:<br>Ball-type for partial or full overdentures | Healing; temporary; conventional 1-piece,<br>2-piece, CAD-CAM | | Gingival Height | GM Exact Co-Cr Abutment for Crown: 1.0 mm<br>GM Equator Attachment: 1.5, 2.5, 3.5, 4.5, 5.5 mm | 0.8 mm to 5.5 mm | | Implant-Abutment<br>Platform Diameter | Exact Co-Cr Abutment for Crown: To match GM<br>implant diameters 3.5/3.75, 4.0/4.3, and 5.0/6.0<br>GM Equator Attachment: Fits all GM implants | To match implant diameters | | Prosthesis Attachment | Exact Co-Cr Abutment for Crown:<br>Cement-retained, screw-retained<br>GM Equator Attachment:<br>Ball-type for partial or full overdentures | Cement-retained and screw-retained | | Restoration | Exact Co-Cr Abutment for Crown: Single-unit<br>GM Equator Attachment: Multi-unit | Single-unit and multi-unit | | Abutment Angle | None (straight 0° only) | None (straight 0°), 17°, 30° | | Abutment-Implant Interface | Exact Co-Cr Abutment for Crown:<br>Morse taper (GM Line)<br>GM Equator Attachment: Morse taper (GM Line) | Morse taper (GM Line) | | Materials | | | | Abutment | Exact Co-Cr Abutment for Crown:<br>Co-Cr-Mo alloy, ASTM F1537<br>GM Equator Attachment:<br>Ti-6Al-4V alloy, ASTM F136, with TiN coating | Ti-6Al-4V alloy, ASTM F136; PEEK | | Abutment Screw | Exact Co-Cr Abutment for Crown:<br>Ti-6Al-4V alloy, ASTM F136<br>GM Equator Attachment: Not applicable | Ti-6Al-4V alloy, ASTM F136 | | Comparison | Subject Device | Primary Predicate Device | | | Neodent Implant System – GM Line<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | K163194<br>Neodent Implant System – GM Line<br>JJGC Indústria e Comércio de Materiais Dentários S.A. | | How Provided | | | | | Exact Co-Cr Abutment for Crown: Non-sterile | | | Sterility/Sterilization | GM Equator Attachment:<br>Sterile by ethylene oxide exposure | Abutments: Sterile by ethylene oxide exposure | | Usage | Single-patient, single-use | Single-patient, single-use | {6}------------------------------------------------ The primary predicate device K163194 is for substantial equivalence of the subject device abutmentimplant interface design. The subject device abutments have the identical implant interface design and are to mate exclusively with the Neodent GM implants cleared in K163194. The subject device abutments and abutments cleared in K163194 are provided in straight (0°) designs with internal hex prosthetic indexing and titanium alloy abutment screws. The differences between the subject device abutments and the abutments cleared in K163194 are the addition of the cast-to design and the Co-Cr alloy material. The reference predicate device K101207 is for substantial equivalence of the design of the cast-to, UCLA-type abutment. K10107 included cast-to, UCLA-type abutments consisting of a titanium alloy component that interfaces directly to the implant and a POM burn-out sleeve for fabrication of a cast abutment. The differences between the subject device abutments and the abutments cleared in K 101207 are the implant interface and the Co-Cr alloy material. The reference predicate device K121843 is for substantial equivalence of the cast-to, UCLA-type abutment design consisting of a Co-Cr alloy component that interfaces directly to the implant and a POM burn-out sleeve for fabrication of a cast abutment. The subject device abutments and the abutments cleared in K121843 include Morse taper connections with internal hex indexing. The primary difference between the subject device abutments and the abutments cleared in K121843 is the implant interface. The Indications for Use Statements for the subject device and primary predicate device K163194 are similar, with the language that the subject device abutments are for support of prosthetic restorations. The intended use of the subject device abutments is similar to that of the abutments in each of the predicate devices. The slight differences in wording among the Indications for Use Statements for the subject device and the predicate devices reflects the fact that the predicate submissions include components, such as CAD/CAM abutments, that are not included in the subject submission. The differences do not affect the intended use with dental implants for rehabilitation of the edentulous maxilla or mandible. # CONCLUSION The subject device and the predicate devices have the same intended use, have similar technological characteristics, and are made of the same or similar materials. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...