K182946 · Invacare Corporation · ITI · Dec 11, 2018 · Physical Medicine
Device Facts
Record ID
K182946
Device Name
Invacare TDX SP2-HD with Captain's Seat
Applicant
Invacare Corporation
Product Code
ITI · Physical Medicine
Decision Date
Dec 11, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the device is to provide mobility and positioning to persons limited to a sitting position.
Device Story
Invacare TDX SP2-HD with Captain's Seat is a battery-powered, motor-driven, center-wheel drive power wheelchair; intended for adult patients with mobility limitations. Device features rigid, non-folding base; two 12-volt DC batteries; two 4-pole single-stage drive motors; LiNX control system; enhanced SureStep suspension; locking gas cylinder stability mechanism. Operates via user-controlled joystick/interface. Provides mobility and positioning; supports weight capacity up to 450 lbs. Used in clinical or home environments. Healthcare providers/caregivers use device to facilitate patient mobility; patient self-operates via control system. Benefits include increased independence and postural support for individuals unable to ambulate.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including design verification and validation per ANSI/RESNA WC-1, WC-2, and WC-4 standards, and ISO 7176-25 and ISO 14971 risk management.
Technological Characteristics
Center-wheel drive power wheelchair; aluminum frame; 24V DC 4-pole motors; sealed VRLA gel batteries; electro-mechanical friction brake; LiNX electronic control system. Dimensions: 25.5" base width; 31.5"-55.4" length. Complies with ANSI/RESNA WC-1/2/4 standards for stability, braking, energy consumption, and electromagnetic compatibility.
Indications for Use
Indicated for adult patients limited to a sitting position requiring mobility and positioning assistance.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 11, 2018
Invacare Corporation Elijah Wreh Regulatory Affairs Manager One Invacare Way Elyria, Ohio 44035
Re: K182946
Trade/Device Name: Invacare TDX SPV2-HD with Captain's Seat Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: October 22, 2018 Received: October 23, 2018
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182946
Device Name Invacare® TDX® SP2-HD with Captain's Seat
Indications for Use (Describe)
The indication for use of the Invacare® TDX® SP2-HD with Captain's Seat is to provide mobility and positioning to persons limited to a sitting position.
| Type of Use (Select one or both, as applicable) | <table><tr><td> Prescription Use (Part 21 CFR 201 Subpart D) </td></tr><tr><td> Over-The-Counter Use (21 CFR 201 Subpart C) </td></tr></table> | Prescription Use (Part 21 CFR 201 Subpart D) | Over-The-Counter Use (21 CFR 201 Subpart C) |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 201 Subpart D) | | | |
| Over-The-Counter Use (21 CFR 201 Subpart C) | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### SUBMITTER per 21 CFR 807.92(a)(1):
Invacare Corporation One Invacare Way Elyria, OH 44035 Phone: (440) 329-6840 Email: ewreh@invacare.com
CONTACT PERSON: Elijah Wreh Regulatory Affairs Manager
#### MANUFACTURER:
Invacare Corporation 1200 Taylor Street Elyria, OH 44035
#### Date Prepared per 21 CFR 807.92(a)(1):
December 10, 2018
#### DEVICE INFORMATION per 21 CFR 807.92(a)(2)
| Name of Device: | Invacare® TDX® SP2-HD with Captain's Seat | |
|-----------------------|-----------------------------------------------|--|
| Common or Usual Name: | Wheelchair, Powered | |
| Classification Name: | Powered Wheelchair 21 CFR § 890.3860 | |
| Regulatory Class: | 2 | |
| Product Code: | ITI | |
| PREDICATE DEVICE: | Invacare® TDX® SP2 Power Wheelchair (K170507) | |
| Patient Population: | Adult Patient Only | |
{4}------------------------------------------------
#### DEVICE DESCRIPTION per 21 CFR 808.92(a)(4)
The subject device is an update to the existing previously cleared Invacare® TDX® SP2 Power Wheelchair (K170507) with LiNX Electronics and Ultra Low Maxx Seating System. The updated subject version of the Invacare® TDX® SP2 Power Wheelchair has the following changes:
- · Increase in the weight capacity of the previously cleared device from 300 lb to 450 lb.
| Model Number | Description |
|--------------|-------------------------------------------|
| TDXSP2V-HD | Invacare® TDX® SP2-HD with Captain's Seat |
Invacare® TDX® SP2 Power Wheelchair is a battery-powered, motor-driven powered wheelchair, controlled by the LiNX® control system with enhanced suspension and additional back, arm and leg rest types. The subject device is a rigid or "non-folding" type power wheelchair base with center-wheel drive capability, two casters in the rear and two casters in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors. No new accessories were added to the new device.
#### INTENDED USE per 21 CFR 807.92(A)(5)
The intended use of the device is to provide mobility and positioning to persons limited to a sitting position.
#### INDICATIONS FOR USE per FORM FDA 3881
The indication for use of the Invacare® TDX® SP2-HD with Captain's Seat is to provide mobility and positioning to persons limited to a sitting position.
#### INDICATIONS FOR USE (IFU) COMPARISON
The Indications for Use statement for the subject device is the same as the previously cleared predicate device. The determination was made based on Section 513(i)(1)(E)(i) of the FD&C Act.
{5}------------------------------------------------
#### COMPARISON of TECHNOLOGICAL CHARACTERISTICS with the PREDICATE DEVICE per 21 CFR 807.92(a)(6)
The device comparison showed that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared predicate device in regard to provide mobility and positioning to persons limited to a sitting position.
The subject device and predicate device are battery powered, motor driven, rigid, non-folding type power wheelchairs with a micro-processor controlled electronic control systems, and nonpowered and powered seating.
The subject device has fixed non-powered and elevate seating systems that are similar to the previously cleared Invacare® TDX® SP2 Power Wheelchair (K170507).
#### BASIS of SUBSTANTIAL EQUIVALENCE per 21 CFR 807.100(b)(2)(ii)(A)
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)/' and the technological characteristics which include materials, design, and other device related features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).
The subject device components are safe and effective as the predicate device and do not raise different questions of safety and effectiveness. The design verification and validation testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate device in regard to the following:
The data generated from the subject device design verification and validation test reports support a finding of substantial equivalence regarding device comparison, wireless capability and coexistence, dimensional analysis, device specifications, and design characteristics.
{6}------------------------------------------------
# Indications for Use (IFU) Comparison Table
| Description | Subject Device Invacare® TDX® SP2-HD with<br>Captain's Seat | Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending Submission | K170507 |
| Indications for<br>Use | The indication for use of the Invacare® TDX® SP2 Power<br>Wheelchair is to provide mobility and positioning to<br>persons limited to a sitting position. | The indication for use of the Invacare® TDX® SP2 Power Wheelchair<br>is to provide mobility and positioning to persons limited to a sitting<br>position. |
# Design Characteristics Comparison – Finished Medical Device
| Description | Subject Device Invacare® TDX® SP2-HD with<br>Captain's Seat | Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair |
|---------------------------------|-------------------------------------------------------------|---------------------------------------------------------------|
| Base Configuration | Centre Wheel Drive | Centre Wheel Drive |
| Suspension | Enhanced SureStep® Suspension | Enhanced SureStep® Suspension |
| Stability Lock | Locking gas cylinder | Locking gas cylinder |
| Speed | 5mph | 5mph or 5.8mph |
| Braking System | Electro-mechanical Friction Brake | Electro-mechanical Friction Brake |
| Range | GP24 Batteries > 20.7 miles | 22NF Batteries > 13.7 miles OR<br>GP24 Batteries > 20.7 miles |
| Weight Capacity | 450lb. | 300lb. |
| Length (without leg rests) | 31.5" to 45.3"<br>(depending on seat configuration) | 31.5" to 45.3"<br>(depending on seat configuration) |
| Base Width | 25.5" | 24" or 25.5"<br>(depending on battery type) |
| Incline Capability | 9° | 9° |
| Total Weight<br>(seat and base) | 291-450lbs.<br>(depending on seat configuration) | 291-450lbs.<br>(depending on seat configuration) |
| Obstacle Climbing | Forward 2.76" AND<br>Reverse 0.79" | Forward 2.95" AND<br>Reverse 0.95" |
| Drive Wheel Frame Material | Aluminum | Aluminum |
| Transport Option | Brackets bolt on frame | Brackets bolt on frame |
{7}------------------------------------------------
## Design Characteristics Comparison – Seating System
| Description | Subject Device | Predicate Device |
|----------------------|-------------------------------------------|---------------------------------------------------------------------------------------|
| | Invacare® TDX® SP2-HD with Captain's Seat | Invacare® TDX® SP2 Power Wheelchair (K170507) |
| Seat Types | Fixed, Elevate | Fixed, Tilt/Recline/Elevate, Tilt/Recline, Recline, Elevate, Tilt/Elevate, Tilt Only. |
| Seat Widths | 22" to 24" | 16" to 24" |
| Seat Depths | 18" to 22" | 16" to 23" |
| Back Heights | 18.5" | 18" to 25" (tilt only) OR<br>20" to 27" (tilt and recline) |
| Upholstery | Vinyl | Meshtex, Startex, Spacetex, O-Vinyl, Polyester |
| Elevating Seat Range | 10" | 12" |
| Tilt Range | N/A | 50° |
| Recline Range | 117° | 168° |
# POWER BASE SUBSTANTIAL EQUIVALENCE TABLE
| Description | Subject Device<br>Invacare® TDX® SP2-HD with Captain's Seat<br>(Pending) | Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair<br>(K170507) |
|---------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------------|
| Base Configuration | Centre Wheel Drive | Centre Wheel Drive |
| Suspension | Enhanced SureStep® Suspension | Enhanced SureStep® Suspension |
| Stability Lock | Locking gas cylinder | Locking gas cylinder |
| Stability Lock Mechanism | Internal locking valve mechanism in gas cylinder | Internal locking valve mechanism in gas cylinder |
| Stability Lock Side Dependency | Both sides lock/unlock together | Both sides lock/unlock together |
| Motors | 4-Pole SSD | 4-Pole SSD |
| Motor Voltage | 24V nominal | 24V nominal |
| Motor Power | 24V DC/324W (@13.5Amps) | 24V DC/324W (@13.5Amps) |
| Speed | 5mph | 5mph, or 5.8mph |
| Number of Batteries | 2 | 2 |
| Battery Types | GP24 | 22NF OR GP24 |
| Battery Chemistry | Sealed VRLA Gel Batteries | Sealed VRLA Gel Batteries |
| Battery Operating Voltage | 24V nominal<br>(2 * 12V) | 24V nominal<br>(2 * 12V) |
| Battery Amp-Hour Rating | GP24 = 63amp-hrs (C5) | 22NF = 43.2amp-hrs (C5)<br>GP24 = 63amp-hrs (C5) |
| Battery Weight | GP24 = 52lbs. | 22NF = 37lbs.<br>GP24 = 52lbs. |
| | Subject Device | Predicate Device |
| Description | Invacare® TDX® SP2-HD with Captain's Seat<br>(Pending) | Invacare® TDX® SP2 Power Wheelchair<br>(K170507) |
| Battery Chargers | 8-amp off board charger<br>(110V) | 8-amp off board charger<br>(110V) |
| Braking System | Electro-mechanical Friction Brake | Electro-mechanical Friction Brake |
| Range | GP24 Batteries > 20.7 miles | 22NF Batteries > 13.7 miles OR<br>GP24 Batteries > 20.7 miles |
| Stopping Distance | 45.7" to 69.3"<br>(depending on chair configuration) | 45.7" to 69.3"<br>(depending on chair configuration) |
| Weight Capacity | 450lbs. | 300lbs. |
| Length (without leg rests) | 31.5" to 55.4" | 31.5" to 55.4" |
| Base Width | 25.5" | 24" OR 25.5"<br>(depending on battery type) |
| Front Forks | Single OR double sided | Single OR double sided |
| Tires | Black gel tires.<br>Foam filled OR pneumatic | Black gel tires.<br>Foam filled OR pneumatic |
| Drive Wheel Size | 14" x 3" | 14" x 3" |
| Castor Size | 6" x 2" | 6" x 2" |
| Front Castor Force | Applied by proportional force gas spring | Applied by proportional force gas spring |
| Damping on Front Swing Arm | Gas spring has damping | Gas spring has damping |
| Incline Capability | 9° | 9° |
| Turning Diameter | 50.4" to 65.4"<br>(depending on seat configuration) | 50.4" to 65.4"<br>(depending on seat configuration) |
| Base Weight<br>(with batteries) | 264lbs (GP24 Batteries) | 264lbs (GP24 Batteries) |
| Total Weight<br>(seat and base) | 291 to 450lbs.<br>(depending on seat configuration) | 291 to 450lbs.<br>(depending on seat configuration) |
| Ground Clearance | > 2.5" | > 2.5" |
| Obstacle Climbing | Forward 2.95" AND<br>Reverse 0.98" | Forward 2.95" AND<br>Reverse 0.98" |
| Motor Gearbox Sound Level | 54dBa | 54dBa |
| Drive Wheel Frame Material | Aluminum | Aluminum |
| Transport Option | Brackets bolt on frame | Brackets bolt on frame |
{8}------------------------------------------------
{9}------------------------------------------------
#### DESIGN VERIFICATION AND VALIDATION TESTING DATA
Design verification and validation testing was performed on the subject device meets the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise any new questions of safety and effectiveness. The acceptance criteria for the full verification of the design and acceptance criteria for each section of the Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard was met.
#### Risk Management
Risk Management has been conducted in accordance with ISO 14971:2007 - Medical Devices -Application of Risk Management to Medical Devices.
#### Non-Clinical Test per 21 CFR 807.92(b)(1)
Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing was performed to demonstrate that the subject Invacare® TDX® SP2-HD with Captain's Seat meets the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise any new questions of safety and effectiveness.
The following testing was performed according to FDA recognized consensus standards:
- ANSI/RESNA WC-1:2009 Section 1: Determination of Static Stability ●
- . ANSI/RESNA WC-1:2009 Section 2: Determination of Dynamic Stability of electrically Powered Wheelchairs
- ANSI/RESNA WC-1:2009 Section 3: Determination of Effectiveness of Brakes .
- ANSI/RESNA WC-2:2009 Section 4: Energy Consumption of Electrically Powered ● Wheelchairs and Scooters for Determination of Theoretical Distance Range
- ANSI/RESNA WC-1:2009 Section 5: Determination of Dimensions, Mass . and Maneuvering Space
- ANSI/RESNA WC-2:2009 Section 6: Determination of Maximum Speed, . Acceleration and Deceleration of Electrically Powered Wheelchairs
- ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of Seating and . Wheel Dimensions
- · ANSI/RESNA WC-1:2009 Section 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
- ANSI/RESNA WC-2:2009 Section 9: Climatic Tests for Electrically Powered Wheelchairs
{10}------------------------------------------------
- ANSI/RESNA WC-2:2009 Section 10: Determination of Obstacle-Climbing Ability . of Electrically Powered Wheelchairs
- ANSI/RESNA WC-1:2009 Section 11: Test Dummies .
- ANSI/RESNA WC-1:2009 Section 13: Determination of Coefficient of ● Friction of Test Surfaces
- . ANSI/RESNA WC-2:2009 Section 14: Power and Control Systems for Electrically Powered Wheelchairs
- ANSI/RESNA WC-1 :2009 Section 15: Requirements for Information . Disclosure, Documentation and Labeling
- ANSI/RESNA WC-1:2009 Section 16: Resistance to Ignition of Upholstered . Parts
- . ANSI/RESNA WC-4: Section 19: Wheelchairs Used as Seats in Motor Vehicles
- · ANSI/RESNA WC-2: 2009 Section 21: Requirements and Test Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Motorized Scooters
- ISO 7176-25 First Edition 2013 Wheelchairs Part 25: Batteries and Chargers for . Powered Wheelchairs
- . ISO 14971:2012 - Medical Devices - Application of Risk Management to Medical Devices
Verification testing demonstrated that the subject Invacare® TDX® SP2-HD with Captain's Seat is substantially equivalent to the marketed predicate device.
#### Animal Study
Animal testing was not required for this submission.
#### Clinical Testing
Clinical testing was not required for this submission.
#### CONCLUSIONS per 21 CFR 807.92(b)(3)
The subject device has the same intended use and technological characteristics as the predicate device. The design verification and validation data support the safety and performance of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions. Therefore, the subject Invacare® TDX® SP2-HD with Captain's Seat accessories are compatible with the previously cleared Invacare® TDX® SP2 Power Wheelchair (K170507) device and demonstrate a basis of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.