Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx
Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx
Applicant
Invacare Corporation
Product Code
ITI · Physical Medicine
Decision Date
Jul 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.
Device Story
Battery-powered, center-wheel drive power wheelchair for mobility and positioning. Input: user commands via joystick, alternative driver controls (e.g., sip-n-puff, head array), or attendant controls. LiNX control system processes inputs to vary power to two 24V DC motors. Features include Bluetooth LE for mouse emulation on PCs/tablets and remote diagnostics via iOS. Actuator control module uses an integrated accelerometer for angular feedback to limit driving capability during seating adjustments. Used in clinics or home environments by patients with limited mobility. Healthcare providers use the device to provide postural support and pressure relief. Benefits include increased range of motion, accessibility, and independent repositioning.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 7176 series (stability, braking, energy consumption, dimensions, speed, fatigue, climatic, obstacle climbing, power/control systems), electromagnetic compatibility (ISO 7176-21, ANSI/RESNA WC-2), flammability (CAL117, EN 1021), and wireless coexistence (ANSI C63.27).
Technological Characteristics
Rigid, non-folding center-wheel drive base; aluminum frame; 24V DC motors; 22NF or GP24 VRLA gel batteries. LiNX control system (microprocessor-controlled, CAN bus). Seating: Ultra-Low Maxx system with tilt/recline/elevate actuators. Connectivity: Bluetooth LE. Biocompatibility per ISO 10993-5/10. Software: Moderate level of concern.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility and positioning assistance, including those without adequate upper-body stability who require powered repositioning (tilt/recline/elevate) to maintain comfort, redistribute pressure, or improve accessibility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other to form a single, abstract figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 12, 2017
Invacare Corporation Elijah Wreh Regulatory Affairs Manager (pre-market) One Invacare Way Elyria, Ohio 44035
Re: K170507
Trade/Device Name: Invacare® TDX® SP2 Power Wheelchair TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: June 9, 2017 Received: June 12, 2017
Dear Elijah Wreh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170507
#### Device Name
Invacare® TDX® SP2 Power Wheelchair
TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx
#### Indications for Use (Describe)
The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| | 510(k) Summary (K170507) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | Invacare Corporation<br>One Invacare Way<br>Elyria, OH 44035 |
| | Phone: (440) 329-6840<br>Email: ewreh@invacare.com |
| CONTACT PERSON: | Elijah N. Wreh<br>Regulatory Affairs Manager (Pre-Market) |
| MANUFACTURER: | Invacare Corporation<br>1200 Taylor St<br>Elyria, OH 44035 |
| Date Prepared: | July 12, 2017 |
| DEVICE INFORMATION | |
| Name of Devices: | Invacare® TDX® SP2 Power Wheelchair<br>TDXSP2 TDX SP2 with LiNX Electronics<br>TDXSP2-CG TDX SP2 Base for Single Act. System w/<br>Ultra Low Maxx,<br>TDXSP2-MCG TDX SP2 Base for Multiple Actuator<br>System w/ Ultra Low Maxx |
| Common or Usual Name: | Wheelchair, Powered |
| Classification Name: | Powered Wheelchair 21 CFR § 890.3860 |
| Regulatory Class: | II |
| Product Code: | ITI |
| PRIMARY PREDICATE DEVICE: | Invacare® TDX® SP2 Power Wheelchair (K141783) |
| SECONDARY PREDICATE DEVICE: | Modular Power Positioning System (K150574) |
| REFERENCE DEVICES: | F3 Powered Wheelchair (K143180)<br>F5 Powered Wheelchair (K143014)<br>Quantum Q6 Edge (K143383)<br>Modular Power Positioning System (K150574) |
{4}------------------------------------------------
#### DEVICE DESCRIPTION
The subject device is an update to the existing previously cleared Invacare® TDX® SP2 Power Wheelchair (K141783) with the MK6i™ control system and Formula™ CG Seating System. The updated subject version of the Invacare® TDX® SP2 Power Wheelchair has the following changes:
- . The LiNX® control system, which incorporates updated software, wireless technology and a touch user interface,
- The Ultra-Low Maxx™ Seating System, and ●
- Minor changes to mechanical components such as rim inserts and new colours. .
| Description |
|------------------------------------------------------------------------|
| TDXSP2 TDX® SP2 with LiNX Electronics |
| TDXSP2-CG TDX® SP2 Base for Single Actuator System w/ Ultra Low Maxx |
| TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx |
The TDX® SP2 Power Wheelchair is a battery-powered, motor-driven powered wheelchair, controlled by the LiNX® control system with enhanced suspension and additional back, arm and leg rest types. The subject device is a rigid or "non-folding" type power wheelchair base with centre-wheel drive capability, two casters in the rear and two casters in the front. It is powered by two 12-volt DC batteries and two 4-pole single stage drive motors.
There are three (3) wireless functions available in the LiNX System. These functions are Mouse Mover, Remote Diagnostics of the Power Wheelchair and Configuration/Programming of the LiNX Control System. The Mouse Mover and Remote Diagnostics of the Power Wheelchair functions are built into all the LiNX Remotes and use a Bluetooth LE wireless connection.
The Mouse Mover function allows the end user of the subject Invacare® TDX® SP2 Power Wheelchair device to operate a standard wireless PC mouse where the joystick or Alternative Driver Control can be used to navigate the cursor on the screen on a PC, tablet or other similar device.
The Remote Diagnostics of the Power Wheelchair function send specific diagnostic information relevant to the status of the wheelchair electronics to an Apple iOS device.
{5}------------------------------------------------
The Configuration/Programming of the LiNX Control System requires a separate programming access key (LiNX Access Key: LAK) that plugs into the LiNX Control System via one of its remotes (Joysticks) and communicates over a Bluetooth connection. This is used for remote programming of the LiNX Control system.
Each accessory connects to the LiNX Control system either directly by connecting to the LiNX communication bus (direct access) or indirectly by connecting to an Input module (indirect access).
The associated models and accessories and capabilities include:
- . Invacare® TDX SP2™ with Motion Concepts Ultra Low Maxx Power Positioning
- Primary Driver Controls (Bluetooth LE Wireless Capable: Mouse Mover and Diagnostics)
- o Export data to application
- Attendant Driver Controls (No Wireless)
- o Provide a simple interface for the wheelchair attendant using a LiNX System
- Control of actuators through the joystick o
- Alternative Driver Controls (Bluetooth LE Wireless Capable: Mouse Mover) ●
- Has a compact, low-profile design, and is a secondary remote module designed to o provide a simple interface for the wheelchair occupant using a LiNX Control System
- o Low force joystick option to improve control for those with low muscle strength
- Controls mouse movement о
- Display Modules (Bluetooth LE Wireless Capable: Mouse Mover and Diagnostics) ●
- o Provides a relocatable user interface for use for alternative driver controls
- Provides the same level of functionality as the enhanced rehabilitation primary о display except for the joystick
- Actuator Control (No Wireless) ●
- The Actuator Control Module has an in-built accelerometer used to provide O angular feedback of the seating position which is in turn used to limit driving capability
- 2-Channel Actuator Module o
- o 4-Channel Actuator Module
- Power Modules (No Wireless) ●
- Process the user commands issued via the joystick or alternative drive control and о control the motion of the wheelchair by varying the amount of power supplied to the left and right wheelchair motors
- Stability Control (No Wireless)
- o G-Trac / Gyro
- Input Module with Integrated Sip-n-Puff (No Wireless) ●
- Sip and puff nozzle o
- o Jack input (stereo)
{6}------------------------------------------------
- 2 x LiNX bus connectors o
- o Activity indicator displaying input demands
- USB Charger Module (No Wireless) ●
- The USB charger module includes protection mechanisms for short circuit and o over-current
- 4-way expansion block (No Wireless)
- o harness for expandable systems
- Atom Proton Head Array (Bluetooth LE Wireless: Mouse Mover) ●
- o Direct access
- MEC mini joystick (Bluetooth LE Wireless: Mouse Mover)
- o Direct access
- Extremity control (Bluetooth LE Wireless: Mouse Mover) ●
- o Direct access
- Compact, Single Switch (Bluetooth LE Wireless: Mouse Mover) o Indirect Access (via input module)
- Compact (No Wireless)
- o Indirect Access (via input module)
- . 4-Switch Proximity (Bluetooth LE Wireless: Mouse Mover)
- o Indirect Access (via input module)
- . Sip-n-Puff Head Array (No Wireless)
- o Indirect Access (via input module)
- . Wireless double switch receiver (Bluetooth LE Wireless: Mouse Mover)
- o Indirect Access (via input module)
- Wireless triple switch receiver (Bluetooth LE Wireless: Mouse Mover)
- Wireless mouse emulator (Bluetooth LE Wireless: Mouse Mover) ●
- LiNX Access Key (LAK: Bluetooth Wireless: Remote Programming) ●
# INDICATIONS FOR USE
The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.
# INDICATIONS FOR USE (IFU) COMPARISON
The Indications for Use statement for the subject device is the same as the secondary predicate device, the Modular Power Positioning System cleared under K150574 Indications for Use statement, which is to provide "positioning to individuals without adequate upper-body stability." The determination was made based Section 513(i)(1)(E)(i) of the FD&C Act. The seating system for the subject Invacare® TDX® SP2 Power Wheelchair device is Modular Power Positioning System (K150574) which is indicated for positioning functions that are offered on the subject device and is designed to help patients without adequate upper-body
{7}------------------------------------------------
stability to be tilted and allow gravity to hold them in position. The subject device positioning also help the patients to reach higher elevations in a seated position, increasing their range of motion and accessibility.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PRIMARY and SECONDARY PREDICATE DEVICES and REFERENCE DEVICES
The device comparison showed that the subject device is substantially equivalent in intended use, design, materials, and operational principles to the previously cleared primary and secondary predicate devices and/or specified reference devices below in regards to provide mobility and positioning to persons limited to a sitting position.
- Invacare® TDX® SP2 Power Wheelchair (K141783) .
- F3 Powered Wheelchair (K143180)
- F5 Powered Wheelchair (K143014) ●
- Quantum Q6 Edge (K143383)
- Modular Power Positioning System (K150574)
The subject device, predicate device and all reference devices excluding the secondary predicate device Motion Concepts Modular Power Positioning System are battery powered, motor driven, rigid, non-folding type power wheelchairs with a micro-processor controlled electronic control systems, non-powered and powered seating and a selection of alternative drive controls.
The subject device has the same seating system as previously cleared secondary predicate Module Power Positioning System device.
The secondary predicate device and the reference devices were chosen to demonstrate substantial equivalence for the following components and features in the subject device:
- . LiNX® Control System
- Power Wheelchair Base .
- . Power Wheelchair Seating
{8}------------------------------------------------
### BASIS OF SUBSTANTIAL EQUIVALENCE
The substantial equivalence of the subject device was determined as per the FDA guidance document. "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, energy source, and other device features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A).
The subject device components are as safe and effective as the predicate and reference devices and do not raise different questions of safety and effectiveness. The differences between the subject device features and the previously cleared predicate and reference devices do not raise new issues of safety or effectiveness.
The performance testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate and reference devices in regards to the following:
- . Static Stability
- Dynamic Stability of Electric Wheelchairs ●
- Effectiveness of Brakes ●
- Energy Consumption of Electric Wheelchairs and Scooters for Determination of Theoretical Distance Range
- Dimensions Mass and Manoeuvring Space ●
- Maximum Speed Acceleration and Deceleration of Electric Wheelchairs
- Methods for Static Impact and Fatigue Strengths
- Climatic Tests for Electric Wheelchairs
- Climbing Ability of Electrically Powered Wheelchairs
- Power and Control Systems for Electrically Powered Wheelchairs and Scooters ●
- Information Disclosure Documentation and Labeling
- Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Scooters and Battery Chargers
- Software Life Cycle
- Additional Requirements for Wheelchairs (including scooters) with Electrical Systems
- Electromagnetic Compatibility of Electrically Powered Wheelchairs and Motorized Scooters
- Flammability Testing ●
- Assessment of the Ignitability of Upholstered Furniture Ignition Source Smouldering ● Cigarette
- Assessment of the Ignitability of upholstered Furniture Ignition Source Match Flame . Equivalent
- Wireless Coexistence ●
{9}------------------------------------------------
The data generated from the subject Invacare® TDX® SP2 Power Wheelchair Design Verification test report supports a finding of substantial equivalence in regards to the device comparison, dimensional analysis, device specifications, design characteristics and to provide mobility and positioning to persons limited to a sitting position.
{10}------------------------------------------------
# Indications for Use (IFU) Comparison Table
| Description | Subject Device Invacare® TDX® SP2 Power<br>Wheelchair | Primary Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair | Secondary Predicate Device<br>Modular Power Positioning System |
|---------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Pending Submission | K141783 | K150574 |
| Indications for Use | To provide mobility and positioning to persons limited<br>to a sitting position | To provide mobility to<br>persons limited to a sitting position. | The Modular Power Positioning system is appropriate for use by any individual<br>who drives a power wheelchair and who<br>desires or requires a change of position without having to utilize the services of an<br>attendant. Needs for position changes<br>include:<br>- All positioning benefits associated with the tilt/recline product:<br>Comfort: As with any individual, able-bodied or disabled, changes in position are<br>necessary to maintain a state of<br>comfort.<br>Positioning: Individuals without adequate upper-body stability can be tilted to<br>allow gravity to hold them in position.<br>Pressure Relief or Reduction: Individuals who wish, from time to time, to<br>redistribute pressure from one area of the body<br>to another, can do so by tilting and/or reclining. By changing the individual's<br>orientation in space, pressures caused by<br>gravity will shift.<br>- Positioning/Versatility: Individuals are able to reach higher elevations in a<br>seated position, increasing their range of<br>motion and accessibility.<br>Motion Concepts makes no claims as to the therapeutic effectiveness of the<br>products. Our only claims relate to the ability<br>of the products to provide safe and reliable powered repositioning on the<br>equipment onto which they are installed. |
{11}------------------------------------------------
# Design Characteristics Comparison – Finished Medical Device
| Description | Subject Device<br>TDX® SP2 | Primary Predicate Device<br>TDX® SP2 |
|---------------------------------|---------------------------------------------------------------|-----------------------------------------------------------|
| Base Configuration | Centre Wheel Drive | Centre Wheel Drive |
| Suspension | Enhanced SureStep® Suspension | Enhanced SureStep® Suspension |
| Stability Lock | Locking gas cylinder | Locking gas cylinder |
| Speed | 5mph or 5.8mph | 5.8mph |
| Braking System | Electro-mechanical Friction Brake | Electro-mechanical Friction Brake |
| Range | 22NF Batteries > 13.7 miles OR<br>GP24 Batteries > 20.7 miles | 22NF Batteries > 12 miles OR<br>GP24 Batteries > 15 miles |
| Weight Capacity | 300lbs. | 300lbs. (Adult seating)<br>165lbs. (Junior seating) |
| Length (without leg rests) | 31.5" to 45.3"<br>(depending on seat configuration) | 35.5" |
| Base Width | 24" OR 25.5"<br>(depending on battery type) | 24" OR 25.5"<br>(depending on battery type) |
| Incline Capability | 9° | 9° |
| Total Weight<br>(seat and base) | 291-450lbs.<br>(depending on seat configuration) | 291-425lbs.<br>(depending on seat configuration) |
| Obstacle Climbing | Forward 2.95" AND<br>Reverse 0.95" | Forward 2.76" AND<br>Reverse 0.98" |
| Drive Wheel Frame Material | Aluminum | Aluminum |
| Transport Option | Brackets bolt on frame | Brackets bolt on frame |
# Secondary and Reference Predicate Devices
| Subject Device | Component or Sub-Assembly | Secondary Predicate Device and Reference Devices | 510(k) Clearance Code/s | Clearance Date/s |
|---------------------------------------|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------|
| Invacare TDX® SP2 Power<br>Wheelchair | TDX® SP2 Power Base | Invacare® TDX® SP2 Power Wheelchair | K141783 | 1st September 2015 |
| | Ultra-Low Maxx Seating<br>(Secondary Predicate Device) | Motion Concepts Modular Power Positioning System | K150574 | 30th June 2015 |
| | LiNX® Control System | Invacare TDX® SP2<br>Mk6i<br>Permobil F3<br>R-Net<br>Permobil F5<br>R-Net<br>Quantum Q6 Edge EM<br>Q-Logic2 | K141783<br>K143180<br>K143014<br>K143383 | 1st September 2015<br>27th March 2015<br>25th March 2015<br>23rd July 2015 |
{12}------------------------------------------------
# Design Characteristics Comparison – LiNX® Electronics
| Description | Subject Device<br>Invacare® TDX® SP2 Power<br>Wheelchair | Primary Predicate Device<br>Invacare® TDX® SP2 Power<br>Wheelchair (K141783) | Reference Devices<br>F3 Powered Wheelchair & F5<br>Powered Wheelchair | Quantum Q6 Edge |
|------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| System Name | LiNX | MK6i | R-Net | Q-Logic2 |
| Cables | Variable cable lengths<br>A range of standard cable lengths available | Primarily fixed length flying leads, connecting at a single "star point" | Variable cable lengths<br>A range of standard cable lengths available | Variable cable lengths<br>A range of standard cable lengths available |
| System Architecture | Microprocessor Controlled | Microprocessor Controlled | Microprocessor Controlled | Microprocessor Controlled |
| Non-Expandable Options | Yes | Yes | Yes | Yes |
| Expandable Options | Yes | Yes | Yes | Yes |
| Wireless Devices | Bluetooth | None | Bluetooth<br>Infra-Red | Bluetooth<br>Infra-Red |
| Power Source | 24V nominal | 24V nominal | 24V nominal | 24V nominal |
| Bus Interface | CAN | Mixed<br>CAN & 2-Wire Serial | CAN | CAN |
# Design Characteristics Comparison – Seating
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | |
|----------------------|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--|
| Description | Invacare® TDX® SP2 Power Wheelchair | Invacare® TDX® SP2 Power Wheelchair<br>(K141783)<br>(Formula CG) | Motion Concepts Modular Power Positioning System<br>(K150574) | |
| Seat Types | Fixed, Tilt/Recline/Elevate, Tilt/Recline,<br>Recline, Elevate, Tilt/Elevate, Tilt Only. | ADJASBA, ADJRECL, Formula TRE,<br>Tilt/Recline/Elevate, Tilt/Recline, Recline, TRE,<br>Elevate, Tilt/ Elevate | Fixed, Tilt/Recline/Elevate, Tilt/Recline, Recline, Elevate,<br>Tilt/Elevate, Tilt Only | |
| Seat Widths | 16" to 24" | 12" to 24" | 15" to 24" | |
| Seat Depths | 16" to 23" | 12" to 22" | 15" to 22" | |
| Back Heights | 18" to 25" (tilt only) OR<br>20" to 27" (tilt and recline) | 12" to 28" | 18" to 30" | |
| Upholstery | Meshtex, Startex, Spacetex, O-Vinyl, Polyester | Cloth & Vinyl | Meshtex, Startex, Spacetex, O-Vinyl, Polyester | |
| Elevating Seat Range | 12" | ידיי | 12" | |
| Tilt Range | 500 | Not Specified | 500 | |
| Recline Range | 1680 | Not Specified | 1680 | |
{13}------------------------------------------------
# POWER BASE SUBSTANTIAL EQUIVALENCE TABLE
| | Subject Device | Primary Predicate Device |
|---------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------------------|
| Description | Invacare® TDX® SP2 Power Wheelchair | Invacare® TDX® SP2 Power Wheelchair |
| | (Pending) | (K141783) |
| Base Configuration | Centre Wheel Drive | Centre Wheel Drive |
| Suspension | Enhanced SureStep® Suspension | Enhanced SureStep® Suspension |
| Stability Lock | Locking gas cylinder | Locking gas cylinder |
| Stability Lock Mechanism | Internal locking valve mechanism in gas cylinder | Internal locking valve mechanism in gas cylinder |
| Stability Lock Side Dependency | Both sides lock/unlock together | Both sides lock/unlock together |
| Motors | 4-Pole SSD | 4-Pole SSD |
| Motor Voltage | 24V nominal | 24V nominal |
| Motor Power | 24V DC/324W (@13.5Amps) | 24V DC/324W (@13.5Amps) |
| Speed | 5mph, or 5.8mph | 5.8mph |
| Number of Batteries | 2 | 2 |
| Battery Types | 22NF OR GP24 | 22NF OR GP24 |
| Battery Chemistry | Sealed VRLA Gel Batteries | Sealed VRLA Gel Batteries |
| | 24V nominal | 24V nominal |
| Battery Operating Voltage | (2 * 12V) | (2 * 12V) |
| | 22NF = 43.2amp-hrs (C5) | 22NF = 43.2amp-hrs |
| Battery Amp-Hour Rating | GP24 = 63amp-hrs (C5) | GP24 = 63amp-hrs |
| | 22NF = 37lbs. | 22NF = 37lbs. |
| Battery Weight | GP24 = 52lbs. | GP24 = 52lbs. |
| | 8-amp off board charger | 8-amp off board charger |
| Battery Chargers | (110V) | (110V or 220V) |
| Braking System | Electro-mechanical Friction Brake | Electro-mechanical Friction Brake |
| | 22NF Batteries > 13.7 miles OR | 22NF Batteries > 12 miles OR |
| Range | GP24 Batteries > 20.7 miles | GP24 Batteries > 15 miles |
| | 45.7" to 69.3" | 74.8" to 102.4" |
| Stopping Distance | (depending on chair configuration) | (depending on chair configuration) |
| | | 300lbs. (Adult seating) |
| Weight Capacity | 300lbs. | 165lbs. (Junior seating) |
| Length (without leg rests) | 31.5" to 55.4" | 35.5" |
| | 24" OR 25.5" | 24" OR 25.5" |
| Base Width | (depending on battery type) | (depending on battery type) |
| Front Forks | Single OR double sided | Single OR double sided |
| | Black gel tires. | Pneumatic OR gel/foam filled flat free tires. |
| Tires | Foam filled OR pneumatic | |
| Drive Wheel Size | 14" x 3" | 14" x 3" |
| Castor Size | 6" x 2" | 6" x 2" |
| Front Castor Force | Applied by proportional force gas spring | Applied by proportional force gas spring |
| Damping on Front Swing Arm | Gas spring has damping | |
| Incline Capability | 0% | 0% |
| Description | Subject Device<br>Invacare® TDX® SP2 Power Wheelchair<br>(Pending) | Primary Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair<br>(K141783) |
| Turning Diameter | 50.4" to 65.4"<br>(depending on seat configuration) | < 46" |
| Base Weight<br>(with batteries) | 264lbs (GP24 Batteries) | 264lbs (GP24 Batteries) |
| Total Weight<br>(seat and base) | 291 to 450lbs.<br>(depending on seat configuration) | 291 to 425lbs.<br>(depending on seat configuration) |
| Ground Clearance | > 2.5" | 3" |
| Obstacle Climbing | Forward 2.95" AND<br>Reverse 0.98" | Forward 2.76" AND<br>Reverse 0.98" |
| Motor Gearbox Sound Level | 54dBa | 58dBa |
| Drive Wheel Frame Material | Aluminum | Aluminum |
| Transport Option | Brackets bolt on frame | Brackets bolt on frame |
{14}------------------------------------------------
## General Comparison of the Performance Characteristics Associated with the Control System
| | Subject Device | Primary Predicate Device | Reference Devices | |
|---------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Description | Invacare® TDX® SP2<br>Power Wheelchair<br>(Pending) | Invacare® TDX® SP2 Power<br>Wheelchair<br>(K141783) | Permobil F3 & F5<br>(K143180 & K143014) | Quantum Q6 Edge (K143383) |
| Cables | Variable cable lengths<br>A range of standard cable<br>lengths available | Primarily fixed length flying<br>leads, connecting at a single<br>"star point" | Variable cable lengths<br>A range of standard cable<br>lengths available | Variable cable lengths<br>A range of standard cable lengths available |
| System Architecture | Microprocessor Controlled | Microprocessor Controlled | Microprocessor Controlled | Microprocessor Controlled |
| Non-Expandable<br>Options | Yes | Yes | Yes | Yes |
| Expandable Options | Yes | Yes | Yes | Yes |
| Wireless Devices | Bluetooth | None | Bluetooth<br>Infra-Red | Bluetooth<br>Infra-Red |
| Power Source | 24V nominal | 24V nominal | 24V nominal | 24V nominal |
| Bus Interface | CAN | Mixed<br>CAN & 2-Wire Serial | CAN | CAN |
{15}------------------------------------------------
| Description | Subject Device<br>Invacare® TDX® SP2 Power Wheelchair<br>(Pending) | Primary Predicate Device<br>Invacare® TDX® SP2 Power Wheelchair<br>(K141783) |
|---------------------------|--------------------------------------------------------------------|------------------------------------------------------------------------------|
| | REM110 | Mk6i SPJ |
| Mounting | 2 x M5 screws suitable for both tube and plate. | 4 x M5 screws suitable for both tube and plate. |
| Connector Type | Direct | Direct |
| User Display | LED | LED |
| Joystick | Magnetic | Inductive |
| On/Off Button | Yes | Yes |
| Horn | Yes | Yes |
| Speed Selection | Speed dial. | Speed up and speed down buttons. |
| Number of Drive Functions | 1 | 1 |
| Battery Gauge | 5 x LED's | 8 x LED's |
| Charger Port | XLR | XLR |
| Status Indicator | Power button 'flash codes'. | Service LED. |
| Remote Diagnostics | Yes.<br>Built-in Bluetooth connection. | No. |
# Comparison of the Performance Characteristics Associated with the Drive Only Primary Remotes
{16}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Devices | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Description | Invacare® TDX® SP2<br>Power Wheelchair<br>(Pending) | Invacare® TDX® SP2 Power<br>Wheelchair<br>(K141783) | Permobil F3 & F5<br>(K143180 & K143014) | Quantum Q6 Edge<br>(K143383) |
| | REM400 | Mk6i-MPJC | R-Net CJSM2 | Q-Logic2 EX Joystick |
| Mounting | 2 x M5 screws suitable<br>for both tube and plate. | 4 x M5 screws suitable for both tube and<br>plate. | 2 x M5 screws suitable for<br>both tube and plate. | 4 x 10-32 screws suitable for both tube and plate. |
| Connection | Direct | Flying Lead | Flying Lead | Flying Lead |
| User Display | LCD - colour. | LCD - colour. | LCD - colour. | LDC - colour. |
| Viewable LCD Size | 49 x 74mm | 46 x 35mm | 53 x 71mm | 44 x 57mm |
| Joystick | Magnetic | Inductive | Magnetic | Not Available |
| Text & Graphics | Icons & translations.<br>Customisable. | Icon based with text. | Icon & translations.<br>Multi-language support | Icon and translations.<br>5 languages. |
| Touch Interface | Yes - capacitive | No | No | No |
| On/Off Button | Yes | Yes | Yes | Yes |
| Horn | Yes | No | Yes | Yes |
| Mode/Function | Button or Touch | Button | Toggle | Button |
| Programmable Multi-<br>Function Keys | 2 x Configurable | No | 4 x Configurable | 2 x Configurable |
| Jack Sockets | 2 x Stereo. | 1 x Mono * 1 x Stereo | 2 x Mono | 2 x Mono |
| Speed Selection | Virtual speed dial<br>operated by touch. | Speed dial. | Virtual speed dial operated by<br>toggles. | Speed dial. |
| Speed Indication | Yes | Yes | Yes | Yes |
| Number of Drive Functions | 36 | 4 | 8 | 5 |
| Battery Gauge | Bar on LCD<br>(continuous) | Battery icon with fill level on LCD<br>(6 discrete segments) | Bar on LCD<br>(10 discrete segments) | Battery icon with fill level on LCD<br>(continuous) |
| Seating Control | Up to 8 actuators | Up to 6 actuators | Up to 8 actuators | Up to 6 actuators |
| Lighting Control | Yes | No | Yes | Yes |
| Real Time Clock | Yes | Yes | Yes | Yes |
| Charger Port | XLR | XLR | XLR | XLR |
| Status Indicator | Error codes on display<br>with supporting icons.<br>Additional LED flash<br>codes for faulty LCD. | Error codes on display with brief<br>description. | Error codes with brief<br>description on display with<br>LED for faulty LCD. | Error codes with brief description on display |
| Mouse Mover | Yes.<br>Built-in Bluetooth<br>connection. | No | Yes.<br>Built-in Bluetooth connection. | Yes.<br>Built-in Bluetooth connection. |
| Remote Diagnostics | Yes.<br>Built-in Bluetooth<br>connection. | No | No | No |
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | |
| Description | Invacare® TDX® SP2 Power Wheelchair<br>(Pending) | Invacare® TDX® SP2 Power Wheelchair –<br>(CG Formula)<br>(K141783) | Modular Power Positioning System (K150574) | |
| Seat Types | Fixed, Tilt/Recline/Elevate, Tilt/Recline, Recline,<br>Elevate, Tilt/Elevate, Tilt Only. | ADJASBA, ADJRECL, Formula TRE,<br>Tilt/Recline/Elevate, Tilt/Recline, Recline,<br>TRE, Elevate, Tilt/ Elevate | Fixed, Tilt/Recline/Elevate, Tilt/Recline, Recline, Elevate,<br>Tilt/Elevate, Tilt Only | |
| Seat Widths | 16" to 24" | 12" to 24" | 15" to 24" | |
| Seat Depths | 16" to 23" | 12" to 22" | 15" to 22" | |
| Back Heights | 18" to 25" (tilt only) OR<br>20" to 27" (tilt and recline) | 12" to 28" | 18" to 30" | |
| Upholstery | Meshtex, Startex, Spacetex, O-Vinyl, Polyester | Cloth & Vinyl | Meshtex, Startex, Spacetex, O-Vinyl, Polyester | |
| Elevating Seat Range | 12" | 7" | 12" | |
| Tilt Range | 50° | Not Specified | 50° | |
| Recline Range | 168° | Not Specified | 168° | |
| Seat Cushion<br>Accessories | Matrx Libra, Matrx PS, Matrx Vi, Matrx Flo-tech Lite<br>and Contour | In-TouchTM, TargitTM Cushions, Infinity<br>Cushions, Leg Cutouts, Leg Wedges | Matrx Libra, Matrx PS, Matrx Vi | |
| Back Types | High Back, Matrx PB, Matrx PB Elite, Matrx PB<br>Deep and Matrx PB Elite TR…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.