K182872 · Waldemar Link GmbH & Co. KG · JWH · Apr 3, 2019 · Orthopedic
Device Facts
Record ID
K182872
Device Name
LINK GEMINI SL Total Knee System
Applicant
Waldemar Link GmbH & Co. KG
Product Code
JWH · Orthopedic
Decision Date
Apr 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LINK® GEMINI® SL® Total Knee System is indicated for patients suffering from disability due to: - Degenerative, post-traumatic or rheumatoid arthritis; - Avascular necrosis of the femoral condyle; - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy: - Moderate valgus, varus or flexion deformities. This device may also be indicated in the salvage of previously failed surgical attempts. The device is indicated for cemented use. Only cementless labeled modular stems are indicated for uncemented use.
Device Story
LINK® GEMINI® SL® Total Knee System is a semi-constrained, patellofemorotibial, cemented knee prosthesis designed to replace three articular portions of the knee joint. System includes Cruciate Retaining (CR) and Posterior Stabilized (PS) designs. Components are manufactured from CoCrMo and feature PorEx® surface modification. Device is used by orthopedic surgeons in clinical settings to restore joint function in patients with degenerative, inflammatory, or post-traumatic knee conditions. Output is a mechanical replacement of the knee joint; clinical decision-making relies on surgeon assessment of patient deformity and joint degradation. Benefits include restoration of joint configuration and mobility.
Clinical Evidence
No clinical performance testing was required for this device. Substantial equivalence is supported by non-clinical bench testing, including tibial baseplate fatigue (ISO 14879, ASTM F1800), tibial bearing wear (ISO 14243-1/2), particle analysis (ISO 17853, ASTM F1877), fretting corrosion (ASTM F1875-98), and tibial post fatigue, constraint, and range of motion analyses.
Technological Characteristics
Semi-constrained, cemented total knee prosthesis. Materials: CoCrMo with PorEx® surface modification. Design: Cruciate Retaining (CR) and Posterior Stabilized (PS) configurations. Modular stems available for cemented/cementless use. Testing standards: ISO 14879, ASTM F1800, ISO 14243-1/2, ISO 17853, ASTM F1877, ASTM F1875-98.
Indications for Use
Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion/dysfunction/prior patellectomy); or moderate valgus, varus, or flexion deformities. Also indicated for salvage of failed surgical attempts.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 3, 2019
Waldemar Link GmbH & Co. KG Andre von Malotki Regulatory Affairs Oststraße 4-10 22844 Norderstedt, Germany
Re: K182872
Trade/Device Name: LINK® GEMINI® SL® Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 1, 2019 Received: March 4, 2019
Dear Andre von Malotki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is for Peter G. Allen -S. The date of the signature is 2019.04.03. The time of the signature is 23:55:12 -04'00'.
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K182872
Device Name
LINK® GEMINI® SL® Total Knee System
Indications for Use (Describe)
The LINK® GEMINI® SL® Total Knee System is indicated for patients suffering from disability due to:
- Degenerative, post-traumatic or rheumatoid arthritis;
- Avascular necrosis of the femoral condyle;
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy:
- Moderate valgus, varus or flexion deformities.
This device may also be indicated in the salvage of previously failed surgical attempts. The device is indicated for cemented use. Only cementless labeled modular stems are indicated for uncemented use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k)<br>Submitter: | Waldemar Link GmbH & Co. KG<br>Barkhausenweg 10<br>22339 Hamburg, Germany<br>Phone: +49-40-539950<br>Facility Registration:3003386935 (Oststraße 4-10)<br>Facility Registration: 3007118403 (Harckesheyde 95) |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact<br>Person: | Waldemar Link GmbH & Co. KG<br>André von Malotki (Regulatory Affairs)<br>Oststraße 4-10<br>22844 Norderstedt, Germany<br>Phone: +49-40 53995-530<br>Fax: +49-40 53995-174<br>E-Mail: a.vonmalotki@linkhh.de |
| Date<br>Prepared: | December 28th, 2018 |
| Trade Name: | LINK® GEMINI® SL® Total Knee System |
| Common<br>Name: | Total Knee Prosthesis |
| Classification<br>Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis; 21 CFR §888.3560, product code<br>JWH |
| Classification<br>and Panel: | Class II, Orthopedic / 87 |
| Predicate<br>Devices: | Foundation® Knee System, manufactured by DJO surgical,<br>K923277, cleared 02/09/1993 |
| | Foundation® Posterior Stabilized Knee, manufactured by DJO<br>surgical, K933539, cleared 11/08/1994 |
| | Foundation® Knee System with TiNbN Coating, manufactured by<br>DJO surgical, K122239, cleared 04/11/2013 |
| Device<br>Description: | The LINK® GEMINI® SL® Total Knee System is a semi-constrained,<br>patellofemorotibial, cemented knee prosthesis. It is intended to<br>replace the three articular portions of the knee joint.<br>The system compromises two different designs: “Cruciate Retaining<br>(CR) |
{4}------------------------------------------------
tibial components are made of CoCrMo and are available with PorEx® (K152431) surface modification.
The LINK® GEMINI® SL® Total Knee System is indicated for Indications for Use: patients suffering from disability due to:
- I Degenerative, post-traumatic or rheumatoid arthritis:
- Avascular necrosis of the femoral condyle: 트
- I Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy:
- I Moderate valgus, varus or flexion deformities.
This device may also be indicated in the salvage of previously failed surgical attempts.
The device is indicated for cemented use. Only cementless labeled modular stems are indicated for uncemented use.
The LINK® GEMINI® SL® Total Knee System is substantially Comparison equivalent to the commercially available device Foundation® Knee to Predicate Device: System, in that both have same intended use and similar indications, design, materials, technological characteristics, and principles of operation. The minor technological differences between the LINK® GEMINI® SL® Total Knee System and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the LINK® GEMINI® SL® Total Knee System is as safe and effective as the Foundation® Knee System. Both devices are intended for cemented use and contain modular stems for cemented and cementless use. The femoral and tibial components are made of CoCrMo and are available with PorEx® (K152431) surface modification. The predicate device is available with the surface modification AmorCoat™ (K122239) produced by the same manufacturer DOT GmbH with same specifications.
> Thus, the LINK® GEMINI® SL® Total Knee System is substantially equivalent.
#### Performance Non-Clinical Performance and Conclusions:
Non-Clinical performance testing was conducted with consideration Data: to Draft Guidance For The Preparation of Premarket Notifications (510(k)s) for cemented, semi-constrained Total Knee Prostheses, April 1993, Guidance Document for Knee Joint patellofemorotibial femorotibial metal/polymer porous-coated uncemented and Prostheses, January 16, 2003
> Non-clinical performance testing included: Tibial Baseplate Component fatigue test per ISO 14879 and ASTM F1800; Tibial Bearing Component wear test per ISO 14243-1 and -2; Particle Analyses test per ISO 17853 and ASTM F1877; Fretting Corrosion Testing per ASTM F1875-98; Tibial Post Fatigue Testing, Constraint
{5}------------------------------------------------
and Range of Motion Analyses.
The results of non-clinical performance testing demonstrate that the device is as safe, as effective, and substantially equivalent to the predicate device.
# Clinical Performance and Conclusions:
There was no clinical performance testing required for this device.
- The subject device LINK® GEMINI® SL® Total Knee System is Conclusion: substantially equivalent to the predicate devices identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.