FOUNDATION KNEE SYSTEM WITH TINBN COATING FOUNDATION PS KNEE SYSTEM WITH TINBN COATING 3DKNEE SYSTEM WITH TINBN COATING
K122239 · Encore Medical L.P. · JWH · Apr 11, 2013 · Orthopedic
Device Facts
Record ID
K122239
Device Name
FOUNDATION KNEE SYSTEM WITH TINBN COATING FOUNDATION PS KNEE SYSTEM WITH TINBN COATING 3DKNEE SYSTEM WITH TINBN COATING
Applicant
Encore Medical L.P.
Product Code
JWH · Orthopedic
Decision Date
Apr 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Joint replacement is indicated for patients suffering from disability due to: degenerative, post-traumatic or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; moderate valgus, varus or flexion deformities; treatment of fractures that are unmanageable using other techniques. This device may also be indicated in the salvage of previously failed surgical attempts. The device is intended for cemented applications.
Device Story
Foundation Knee System, Foundation PS Knee System, and 3DKnee System are semi-constrained cemented knee prostheses. Modification involves adding a Titanium Nitride (TiNbN) coating to the entire surface of femoral components and removing porous coating from Foundation Knee Femoral Component. Device used by orthopedic surgeons in clinical settings for total knee arthroplasty. Prosthesis replaces damaged joint surfaces to restore function and reduce pain in patients with arthritis, necrosis, or deformity. Clinical benefit derived from improved wear resistance and surface properties provided by TiNbN coating while maintaining established design and surgical technique.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including acute system toxicity, bone/muscle implantation studies, cytotoxicity, GC/MS fingerprinting, irritation, sensitization, and mechanical coating characterization (composition, thickness, hardness, adhesion, roughness, wear resistance).
Technological Characteristics
Semi-constrained cemented knee prosthesis. Femoral components feature Titanium Nitride (TiNbN) surface coating. Porous coating removed from Foundation Knee Femoral Component. Materials, design, and sterilization methods remain consistent with predicate systems.
Indications for Use
Indicated for patients with disability due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (including patellofemoral erosion/dysfunction or prior patellectomy); moderate valgus, varus, or flexion deformities; or fractures unmanageable by other techniques. Also indicated for salvage of failed surgical attempts. Intended for cemented use.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
# 510(k) Summary
Date: April 9, 2013
Manufacturer:
DJO Surgical (Legal Name: Encore Medical, L.P.) 9800 Metric Blvd Austin, TX 78758
Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (760) 597-3466 Email: teffany.hutto@djoglobal.com
| Product | Original 510(k) Number, Clearance Date/ Classification | Product Code | Product Code | Regulation and Classification Name |
|----------------------------|--------------------------------------------------------|--------------|--------------|------------------------------------------------------------------------------------------------------------------|
| Foundation® Knee System | K923277, February 9, 1993 / Class II | JWH | JWH | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis per 21<br>CFR 888.3560 |
| Foundation® PS Knee System | K933539, November 8, 1994 / Class II | JWH | | |
| 3DKnee™ System | K020114, July 12, 2002 / Class II | JWH | | |
Description: The change that is the subject of this 510(k) is to add a coating of Titanium Nitride (TiNbN) to the entire surface of the above listed femoral components. Additionally, the porous coating has been removed from the Foundation Knee Femoral Component. There is no change to the fundamental scientific technology of the referenced knee systems with the modifications in this 510(k) submission. This includes no changes to materials, design, sterilization, packaging, or method of manufacture.
#### Indications for Use:
Joint replacement is indicated for patients suffering from disability due to:
- degenerative, post-traumatic or rheumatoid arthritis; .
- avascular necrosis of the femoral condyle; .
- post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, . dysfunction or prior patellectomy;
- moderate valgus, varus or flexion deformities; .
- treatment of fractures that are unmanageable using other techniques. .
This device may also be indicated in the salvage of previously failed surgical attempts. The device is intended for cemented applications.
### Predicate Devices:
- DJO Surgical, Foundation® Knee System, K923277 .
- DJO Surgical, Foundation® PS Knee System, K933593 .
- DJO Surgical, 3DKnee™ System, K020114 .
- . DJO Surgical, Movation Knee System, K100900
- DJO Surgical, Foundation® Plasma Coated Femur, K964008 .
Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same indications, dimensions, materials, packaging, sterilization, surgical implantation technique, and intended use.
Non-Clinical Testing: Accute System Toxicity Study, Bone Implantation Study, Cytotoxicity Study, GC/MS Fingerprint Study, Irritation Study, 28 Day Muscle Implantation Study, 90 Day Muscle Implantation Study, Sensitisation Study, Coating Chemical Composition, Coating Thickness, Coating Hardness, Adhesion Strength, Roughness, Wear Resistance
Clinical Testing: None provided.
APR 1 1 2013
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Letter dated: April 11, 2013
Encore Medical, L.P. % Ms. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Boulevard Austin, Texas 78758
Re: K122239
Trade/Device Name: Foundation®, Foundation® PS, and 3DKnee™ Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Codes: JWH Dated: March 7, 2013 Received: March 8, 2013
Dear Ms. Hutto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
## Page 2 -- Ms. Teffany Hutto
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark Nighterson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K122239 510(k) Number (if known):
Device Name: Foundation®, Foundation® PS, and 3DKnee™
Indications for Use:
# Foundation®, Foundation® PS, and 3DKnee™ Indications for Use
Joint replacement is indicated for patients suffering from disability due to:
- degenerative, post-traumatic or rheumatoid arthritis; .
- avascular necrosis of the femoral condyle; .
- post-traumatic loss of joint configuration, particularly when there is patellofemoral . erosion, dysfunction or prior patellectomy;
- . moderate valgus, varus or flexion deformities;
- treatment of fractures that are unmanageable using other techniques. .
This device may also be indicated in the salvage of previously failed surgical attempts.
The device is intended for cemented applications.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley Ph.D.
Division of Orthopaedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.