K182815 · Apteryx, Inc. · LLZ · Nov 30, 2018 · Radiology
Device Facts
Record ID
K182815
Device Name
XVWeb 3D
Applicant
Apteryx, Inc.
Product Code
LLZ · Radiology
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
XVWeb 3D is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser. A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations. Acquisition can be included via integration with a DICOM-compatible Imaging and/or PACS system configured to forward images to the XVWeb 3D database. XVWeb 3D is compatible with programs that run on standard "off-the-shelf" personal computers, business computers, and servers running standard operating systems. The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes. XVWeb 3D is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
Device Story
XVWeb 3D is a web-based PACS add-on module for the existing XVWeb platform; enables viewing/manipulation of 3D cone beam CT (CBCT) images. Input: DICOM-formatted raw CT frames (axial slices) from CBCT systems. Processing: 3D server computes frame spacing, min/max pixel values, normalizes frames, and downscales/merges data for web optimization; stores as JPEG files. Output: Orthogonal multi-planar reconstruction (MPR) slices (axial, sagittal, coronal), 3D volume renderings, and curved MPR slices. Used in dental clinics; operated by trained dental professionals/technicians. Users access via standard web browsers to perform diagnostic tasks (measurement, enhancement, annotation). Benefits: enables remote access and 3D visualization of craniofacial anatomy to assist in treatment planning.
Clinical Evidence
No clinical testing was performed. Evidence consists of software verification and validation testing conducted in accordance with FDA guidance for software in medical devices.
Technological Characteristics
Web-based PACS module; utilizes DICOM standard for image processing. Compatible with standard off-the-shelf PCs, servers, and web browsers. Software-based image processing includes normalization, downscaling, and rendering of 3D volumes and MPR slices from axial CT frames. No hardware components; operates as an add-on to existing XVWeb infrastructure.
Indications for Use
Indicated for trained dental professionals and technicians in medium to large dental practices to access, manipulate, and enhance dental cone beam CT images of teeth, jaw, and lower skull for diagnostic purposes. Patient population: individuals who have undergone a dental cone beam CT examination.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
November 30, 2018
Apteryx, Inc. Mark Gniewek Regulatory Manager 313 S. High St., Suite 200 AKRON, OH 44308
Re: K182815
Trade/Device Name: XVWeb 3D Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: September 26, 2018 Received: October 3, 2018
Dear Mark Gniewek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D.'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182815
Device Name XVWeb 3D
Indications for Use (Describe)
XVWeb 3D is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser.
A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations.
Acquisition can be included via integration with a DICOM-compatible Imaging and/or PACS system configured to forward images to the XVWeb 3D database. XVWeb 3D is compatible with programs that run on standard "offthe-shelf" personal computers, business computers, and servers running standard operating systems.
The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes.
XVWeb 3D is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the word "APTERYX" in bold, blue letters, with the words "custom software" in a smaller font size underneath. There is a black line separating the word "APTERYX" from the alphanumeric string "K182815", which is also in blue. The image also contains a blue graphic to the right of the word "APTERYX".
# 7. 510(K) SUMMARY
## 1. Submitter Information
Contact: Mark Gniewek, Requlatory Manager Apteryx, Inc. 313 S. High St., Suite 200 Akron, OH 44308 (330) 376-0889
Date Prepared: May 31, 2018
## 2. Device Name and Classification
Trade/proprietary name: XVWeb 3D Common/usual name: Classification name: Regulatory Class: .
3D imaging, analysis, and diagnostic software PACS (21 CFR 892.2050, Product Code LLZ)
## 3. Manufacturer Address and Registration Number
Apteryx, Inc. 313 S. High St., Suite 200 Akron, OH 44308
FDA Reqistration Number: 3007350723
#### 4. Predicate Device Information
The predicate device is XVWeb, K132342, date of concurrence: December 20, 2013.
## 5. Labeling and Intended Use
#### Labeling
No changes to the XVWeb labeling have occurred. For those who wish to add XVWeb 3D to their existing XVWeb workflow, Apteryx simply enables the 3D imaging module on the customer's existing XVWeb site when the purchasing process is complete.
The instructions for use can be found in the product's online documentation.
#### Intended Use
XVWeb 3D is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser. This is the same intended use as previously cleared for XVWeb, K132342.
{4}------------------------------------------------
## 6. Device Description
#### What it is
XVWeb 3D is an add-on software module for XVWeb (K132342) that enables dental practices to query and access digitally stored, 3D hard- and soft-tissue cone beam CT images using an internet/web browser. XVWeb 3D uses the web-based interface provided by XVWeb to display patient 3D images and studies.
XVWeb 3D allows users to manipulate 3D images an infinite number of ways, allowing the dental anatomy to be viewed from every possible perspective, providing the information needed to plan efficient treatments. XVWeb 3D also allows users to enhance images (e.q., increase or decrease brightness/contrast, gamma correction), and add measurement labels to 3D patient images/studies for diagnostic purposes, and view the image's DICOM information entities.
XVWeb 3D is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental cone beam CT images of the teeth, jaw, and lower skull area for diagnostic purposes only. The patient population are those who have had a dental cone beam CT examination.
XVWeb 3D does not function by itself (i.e., it is not standalone software that can be launched and ran), it is integrated with XVWeb to query/access/view/ patient 3D images. XVWeb 3D is not required for XVWeb, it is an extra-cost option available to those who wish to add 3D-imaging capability to their existing XVWeb workflow. XVWeb 3D does not perform any radiographic image acquisition.
#### How it works
XVWeb 3D uses the DICOM file format standard to process the raw images/CT frames that have rescale type HU (Hounsfield unit) from the cone beam CT system. These files are sent from wherever the cone beam CT examination is taking place (i.e., the doctor's office, or "client") to the XVWeb DICOM server via web service. The DICOM files contain the raw images/CT frames in the form of parallel, horizontal slices of the head, starting from the bottom and proceeding to the top (i.e., the axial plane view).
When all of the CT frames for a given patient/series are received by the XVWeb DICOM server, a separate server, referred to as the 3D server, preprocesses the axial plane CT frames. This preprocessing consists of several steps:
- The spacing between each CT frame is computed ●
- The minimum and maximum pixel value in all CT frames is computed ●
- If necessary, frames are downscaled and/or merged to optimize them for web browser . bandwidth and throughput
- Each CT frame is normalized using the minimum and maximum pixel values computed earlier
- . Each CT frame is stored as a JPEG file on the customer's existing XVWeb site
When the preprocessing is complete, a user can then access the stored CT data using the XVWeb user interface in a web browser. One way the CT data will appear in the browser is as a series thumbnail image in the series bar on the left side of the XVWeb user interface. When the user clicks the CT thumbnail image, the browser begins to request the JPEG files created during the preprocessing steps. When all of the JPEG files are loaded into the browser, XVWeb 3D renders the orthogonal MPR slices (the axial, sagittal, and coronal plane views) and the 3D volume view and displays them in the XVWeb user interface. The user can then manipulate the 3D volume and slices using the controls within the XVWeb interface. XVWeb 3D also renders the
{5}------------------------------------------------
curved MPR slice (including cross-sectional slices) and makes it available for manipulation when the Curved MPR View button is clicked and a curve is drawn in the XVWeb interface. Crosssectional slices related to the user-drawn curve are also rendered, but not until after the curve is drawn.
## 7. Indications for Use
XVWeb 3D is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser.
A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations.
Acquisition can be included via integration with a DICOM-compatible Imaging and/or PACS system configured to forward images to the XVWeb 3D database. XVWeb 3D is compatible with programs that run on standard "off-the-shelf" personal computers, business computers, and servers running standard operating systems.
The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase or decrease brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes.
XVWeb 3D is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
#### 8. Comparison
The only modification that was made to the originally-cleared device is the 3D data recognition. processing, and image manipulation capabilities within the XVWeb interface. These capabilities are contained in the add-on software module.
The XVWeb 3D module extends the cloud capability of XVWeb to include cone beam CT datasets, allowing 3D images of dental structures, soft tissues, nerve paths and bone in the human craniofacial region to be included in the XVWeb query/access/view/diagnose workflow.
Engineering drawings, schematics, etc. are not applicable to the device (figures are not pertinent to describe the differences between subject and predicate devices).
#### Substantial Equivalence
XVWeb 3D has the following similarities to the previously-cleared XVWeb:
- . Same indicated use
- Ability to query and view patient images ●
- Ability to manipulate images (e.g., view from different perspectives, zoom in/out)
- Ability to enhance images (e.g., increase/decrease brightness, contrast, and gamma correction)
- . Ability to add measurement labels
- Ability to view DICOM information entities (e.g., patient, study, series, image)
- Ability to export a copy of images
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Apteryx Custom Software. The logo features the word "APTERYX" in bold, dark blue letters, with the words "custom software" in a smaller font size underneath. To the right of the text is a stylized, dark blue graphic that resembles a feather or a leaf. The overall design is clean and professional.
While XVWeb applied the above device characteristics to 2D images, XVWeb 3D allows these same characteristics to be applied to 3D images.
#### Software Verification and Validation Testing
XVWeb 3D verification and validation testing were conducted and documentation was provided, including pass/fail criteria, as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Apteryx believes the XVWeb 3D Level of Concern to be Moderate, since a latent design flaw could result in an erroneous diagnosis that would likely lead to minor injury.
#### Clinical Studies
No clinical testing was performed.
#### 9. Conclusions
In summary, the XVWeb 3D module described in this submission is, in the opinion of Apteryx, substantially equivalent to the predicate device, XVWeb.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.