K132342 · Apteryx, Inc. · LLZ · Dec 20, 2013 · Radiology
Device Facts
Record ID
K132342
Device Name
XVWEB
Applicant
Apteryx, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 20, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
XVWeb is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser. A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations. Acquisition can be included via integration with a DICOM-compatible Imaging and/or PACS system configured to forward images to the XVWeb database. XVWeb is compatible with programs that run on standard "off-the-shelf personal computers, business computers, and servers running standard operating systems. The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase or decrease brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes. XVWeb is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
Device Story
XVWeb is a web-based dental PACS for accessing, manipulating, and enhancing radiological images. It receives images from DICOM-compatible imaging systems or modalities. Users access the system via standard web browsers (e.g., Internet Explorer, Chrome) on off-the-shelf computers. Features include image viewing, rotation, flipping, zooming, brightness/contrast adjustment, gamma correction, and annotation (labels, measurements, lines, arrows). The system supports local or wide-area network access. It is intended for use in dental practices by trained professionals and technicians to assist in diagnostic workflows. The device does not perform automated diagnosis; it provides tools for clinicians to view and process images to support their clinical decision-making.
Clinical Evidence
Clinical trials performed side-by-side with current systems to ensure proper software performance. No specific quantitative metrics (sensitivity/specificity) reported. Non-clinical testing included device hazard analysis, module testing, and system verification/validation testing in a simulated environment.
Technological Characteristics
Web-based software application; operates on standard TCP/IP network infrastructure; compatible with off-the-shelf PCs and servers. Supports DICOM-compliant image acquisition. Features include image manipulation (rotate, flip, zoom), enhancement (brightness, contrast, gamma), and annotation tools. Secure data transmission via HTTPS. Database management and storage included.
Indications for Use
Indicated for use by trained dental professionals and technicians in medium to large dental facilities to query, access, manipulate, and enhance digitally stored hard and soft tissue intraoral radiological images for diagnostic purposes.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Centricity PACS Web Diagnostic (Web DX) PACS System (K083018)
Submission Summary (Full Text)
{0}------------------------------------------------
**APTERYX**
custom software
K132342
XVWeb 510(K) Premarket Notification
# DEC 2 0 2013
## F. 510(k) Summary
- l. Applicant Information
| Submitter Information: | Apteryx, Inc. |
|------------------------|----------------------------|
| | 313 S. High St., Suite 200 |
| | Akron, OH 44308 |
| Contact: | Kevin Crucs, President |
| Telephone Number: | (330) 376-0889 |
| Date Prepared: | February 27, 2013 |
#### 11. Device Name and Classification
| Proprietary Name: | XVWeb |
|-----------------------|----------------------------------------------------|
| Classification Name: | Picture Archiving and Communications System (PACS) |
| Classification: | Class II |
| Regulation Number: | 892.2050 |
| Product Code: | LLZ |
| Classification Panel: | Radiology Devices |
- lll. Predicate Devices
XVWeb is substantially equivalent to the following FDA-approved predicate devices. It is substantially equivalent with regard to indications for use, performance, and technological characteristics.
| 510(K) Number: | K112974 |
|-----------------------|-----------------------------------------------------|
| Trade Name: | Curve Image 2.0 |
| Manufacturer: | Curve Dental, Inc. |
| Classification Name: | Picture Archiving and Communications System (PACS) |
| Classification: | Class II |
| Regulation Number: | 892.2050 |
| Product Code: | LLZ |
| Classification Panel: | Radiology Devices |
| 510(K) Number: | K083018 |
| Trade Name: | Centricity PACS Web Diagnostic (Web DX) PACS System |
| Manufacturer: | GE Healthcare Dynamic Imaging Solutions |
| Classification Name: | Picture Archiving and Communications System (PACS) |
| Classification: | Class II |
| Regulation Number: | 892.2050 |
| Product Code: | LLZ |
{1}------------------------------------------------
XVWeb 510(K) Premarket Notification
#### IV. Device Description
XVWeb is a web-based dental picture archiving and communications system that provides a dental facility the ability to access their patient radiological image library over the internet. It is designed for deployment in medium to large dental facilities with multiple viewing stations and/or across remote sites using standard TCP/IP network infrastructures. The system uses a web-based interface that allows users to display patient images and studies in commercially available web browsers such as Internet Explorer, Mozilla, Firefox, or Google Chrome.
XVWeb includes features for dental practices to: manipulate, enhance, add lahels, annotations to patient images/studies, and other relevant operations for diagnostic purposes. XVWeb includes image acquisition capabilities via integration with DICOM-compatible imaging systems that facilitate the production of original images. All images available for viewing in XVWeb have been received from DICOM compliant modalities or systems.
#### V. Intended Use
XVWeb is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard soft tissue intraoral /extraoral radiological images using an internet/web browser.
A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations.
Acquisition can be included via integration with a DICOM-compatible imaging application or server configured to forward images to the XVWeb database.
The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase or decrease brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes.
XVWeb is designed for medium-to-large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
#### VI. Summary of Device Characteristics
- Query and view patient images/studies .
- View single images (e.g. intraoral, panoramic and cephalometric)
- View full mouth series ●
- Manipulate images (e.g. Rotate increments of 90°, flip horizontal/vertically, and zoom in or out) ●
- Enhance images (e.g. Increase/decrease brightness, contrast and gamma correction) .
- . Calibrate and Add Measurement labels to Images
- Add labels and annotations on patient images (e.g. Text, lines, arrows, etc.) .
- . View image information
- . Export a copy of an image
- Integration with a DICOM-compatible imaging application. .
- . Comparison of images
{2}------------------------------------------------
Table F.1 below offers a comparison of XVWeb with the aforementioned predicate devices with respect to intended use, environment of use, technical performance and operations characteristics, etc. The intent is to provide more details to demonstrate substantial equivalence with the predicate devices.
APTERYX
| XVWeb | |
|---------------------------------------------------------|--|
| Curve Image (K112974) | |
| Centricity PACS Web Diagnostic (Web DX) PACS (K0803018) | |
| Product Code | |
| I.I.Z. | |
| Regulation Number | |
| 892.2050 | |
| Class | |
| Class II | |
| Intended Use | |
XVWeb is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored intraoral and extraoral radiological images using an internet/web browser.
A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers such as Internet Explorer, Mozilla, Firefox, or Google Chrome.
Patient images/studies can be accessed by any user locally within the system or across a wide-area network at distributed locations.
The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase or decrease brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes.
Acquisition can be included via integration with a DICOM-compatible Imaging application configured to forward images to the
Curve Image is an internet-based, image management software (PACS) that enables dental offices to keep records of hard and soft tissue charts in a form of digital images. The system uses a Web-based interface and includes acquisition, editing and storage of digital images. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distributed locations. Images can be acquired from standard dental imaging devices, or can be uploaded directly from the user's computer.
Images can be edited (e.g., zoomed, contrast, rotated, etc.), as well as exported. The system is designed to provide images for diagnostic use.
Centricity PACS Web Diagnostic (Web DX) PACS System by GE Healthcare Dynamic Imaging Solutions is a device that receives medical images (including mammograms) and data from various imaging sources. Images and data can be stored, communicated, processed and displayed within the system or across computer networks at distrusted locations.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only interpreted using an FDA Approved monitor that offers at least 5 MPixels resolution and meets other technical specifications reviewed and accepted by FDA.
{3}------------------------------------------------
| | | XVWeb 510(K) Premarket Noti |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| XVWeb database.<br>XVWeb is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only. | | |
| Environment of Use | | |
| Dental Offices | Dental Offices | Dental Offices and other Modalities |
| Performance Characteristics & Principles of Operation | | |
| <b>Basic Technology:</b> | | |
| Web-based software application | Web-based software application | Web-based software application |
| <b>Features:</b> | | |
| Browsing images by Date and/or<br>Source | Browsing images by Date and/or<br>Source | Browsing images by Date and/or<br>Source |
| Viewing an Image | Viewing an Image | Viewing an Image |
| Uploading an image file (via<br>integration with a DICOM-<br>compatible imaging application or<br>PACS). | Uploading an Image File | Uploading an Image File |
| Acquiring an Image from a Web<br>Camera (via integration with a<br>DICOM-compatible imaging<br>application. | Acquiring an Image from a Web<br>Camera | Acquiring an Image from a Web<br>Camera |
| Acquiring an Image from a TWAIN<br>device (via integration with a<br>DICOM-compatible imaging<br>application). | Acquiring an Image from a TWAIN<br>Device | Acquiring an Image from a TWAIN<br>Device |
| Acquiring an Image from a TWAIN<br>device (via integration with a<br>DICOM-compatible imaging<br>application). | Acquiring image from standard dental<br>imaging devices | NA |
| Copy an image to the Local<br>Computer | Copying an Image to the Local<br>Computer | Copying an Image to the Local<br>Computer |
| Add labels and annotations on<br>patient images (Text, lines, arrows,<br>etc.) | Annotating an image | NA |
| Saving an image that has been<br>modified with labels/overlays. | Saving a modified image | Saving a modified image |
| Annotating an Image | Annotating an Image | Annotating an Image |
| Zooming In on an Image | Zooming In on an Image | Zooming In on an Image |
| NA | Cropping an Image | NA |
| NA | Inverting colors of an image | Inverting colors of an image |
| Rotating an Image (increments of<br>90°) | Rotating an Image (increments of 90°) | Rotating an Image (increments of 90°) |
| Flipping an image horizontally or<br>vertically | Flipping an image horizontally or<br>vertically | Flipping an image horizontally or<br>vertically |
| Increasing/decreasing image<br>brightness | Increasing/decreasing image brightness | Increasing/decreasing image<br>brightness |
| Adjusting Gamma function | Adjusting Gamma function | Adjusting Gamma function |
| Image measurements | Image measurements | Image measurements |
| DICOM Compliance | NA | DICOM compliance |
| NA | NA | 3D Image view |
| Lossless image compression | Lossless image compression | JPEG and wavelet lossy image compression |
| Zoom depth: Fit and relative. | Zoom depth: absolute x, fit, and relative | Zoom depth: absolute 1x, 2x, 4x, 8x and relative. |
| <b>Technical Characteristics</b> | | |
| Web-based application and interface | Web-based application and interface | Web-based application and interface |
| Secure data transmission (HTTPS) | Secure data transmission (HTTPS) | Secure data transmission (HTTPS) |
| Database Management and storage | Database Management and storage | Database Management and storage |
| Secure Server infrastructure | Secure Server infrastructure | Secure Server infrastructure |
| <b>Administration</b> | | |
| User Login | User login | User login |
| Account Management | Account Management | Account Management |
| Security administration privileges for Administrator and user. | Security administration privileges for Administrator, basic user, clinical user, and practice management user | Security administration privileges for administrator, basic user, and clinical user. |
| <b>Patient Management</b> | | |
| NA | Expanded feature: Patient profile and data management | Patient profile and data management |
| <b>Scheduling</b> | | |
| NA | Expanded feature: Appointment management functions with calendar | NA |
| <b>Billing</b> | | |
| NA | Expanded feature: Setting up billing and insurance information | NA |
| <b>Insurance Management</b> | | |
| NA | Expanded feature : Retrieve insurance payments | NA |
| <b>Claims Management</b> | | |
| NA | Expanded feature: Create and manage insurance claims | NA |
| <b>Dental Charting</b> | | |
| NA | Expanded feature: Dental Chart layout and treatment plans management | Dental Chart layout and treatment plans management |
| <b>Perio Charting</b> | | |
| NA | Expanded feature: Perio Chart, perio exam, and Perio graph management | NA |
| <b>Patient Notes and History</b> | | |
| NA | Expanded feature: Patient text notes, | Keynotes on patient data |
| Prescriptions | | |
| NA | Expanded feature: Prescription notepad<br>writer | NA |
| Patient Education | | |
| NA | Expanded feature: Patient education<br>movies interface | NA |
.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "APTERYX" in bold, uppercase letters. Below the word "APTERYX" is the phrase "custom software" in a smaller font size. To the right of the text is a black graphic that resembles a stylized leaf or wing.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Apteryx. The logo consists of the word "APTERYX" in bold, sans-serif font, with the words "custom software" in a smaller font size underneath. To the right of the text is a stylized image of a kiwi bird, which is the namesake of the company.
Table F.1
#### VII. Comparison Summary
The differences between the subject and the predicate devices, the reason they do not alter the indications for use, or impact the safe and effective use of XVWeb are listed below:
- Cropping an Image Does not raise new types of Safety and Effectiveness questions. . Although this feature may improve image presentation and data size, it does not improve image quality or increase diagnosis accuracy.
- Inverting colors of an image Does not raise new types of Safety and Effectiveness . questions. In regards to safety, the feature does not improve the quality of an image, therefore, does not contribute to a reduced number of retakes for a patient. Effectiveness of XVWeb is not impacted by the exclusion of the feature. Color inversion does not significantly reduce diagnostic accuracy. Diagnostic accuracy is dependant on the quality of the raw, source data.
- Patient profile and data management An expanded feature not defined as an intended use . in the predicate device(s).
- . Appointment management functions with calendar - An expanded feature not defined as an intended use in the predicate device(s).
- . Setting up billing and insurance information - An expanded feature not defined as an intended use in the predicate device(s).
- Retrieve insurance payments An expanded feature not defined as an intended use in the . predicate device(s).
- Create and manage insurance claims An expanded feature not defined as an intended use . in the predicate device(s).
- Dental Chart layout and treatment plans management An expanded feature not defined as . an intended use in the predicate device(s).
- Perio Chart, perio exam, and Perio graph management An expanded feature not defined as . an intended use in the predicate device(s).
- Patient text notes, note history An expanded feature not defined as an intended use in the . predicate device(s).
- Prescription notepad writer An expanded feature not defined as an intended use in the o predicate device(s).
- Patient education movies interface An expanded feature not defined as an intended use in . the predicate device(s).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Alteryx custom software. The logo has the word "ALTERYX" in bold, black letters. Below the word "ALTERYX" is the phrase "custom software" in a smaller, black font. To the right of the text is a black and white graphic that appears to be a stylized leaf or abstract shape.
#### Summary of Non-Clinical Testing VIII.
Apteryx has performed thorough non-clinical testing to ensure the safety and effectiveness of XVWeb. Stress testing in a simulated environment has been performed to ensure that the XVWeb system can consistently operate in a manner similar to the indications for use of the predicate devices listed. Quality assurance test activities included:
- Device Hazard Analysis
- . Module Testing
- . System Verification and Validation Testing
#### IX. Summary of Clinical Testing
Clinical trials have been performed side-by-side with current systems to ensure proper performance of the software.
#### X. Conclusions
The subject (XVWeb) is substantially equivalent to the predicated devices (Curve Image 2.0 and Centricity PACS Web Diagnostic [Web DX] PACS System] when comparing the intended use and technological characteristics to these predicate devices. XVWeb combines the basic features, functionality, and web-based technology of Curve Image (the stand-alone product) and Centricity PACS. The differences between the subject and the predicate devices do not alter the indications for use or raise new issues that impact the safe and effective use of XVWeb.
{7}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2013
Apteryx, Inc. % Mr. Kevin Crucs President 313 South Hight Street, Suite 200 AKRON OH 44308
Rc: K132342
Trade/Device Name: XV Web Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 31, 2013 Received: November 4, 2013
## Dear Mr. Crucs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{8}------------------------------------------------
Page 2-Mr. Crucs
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use Form F.
510(k) Number: K132342
Device Name: XVWeb
# Indications for Use:
XVWeb is a Picture Archiving and Communications System (PACS) that enables dental facilities to query and access digitally stored hard and soft tissue intraoral radiological images using an internet/web browser.
A web-based interface provides users the needed functionality to display patient images and studies in commercially available web browsers. Patient images/studies can be accessed by users locally within the system or across a wide-area network at distributed locations.
Acquisition can be included via integration with a DICOM-compatible Imaging and/or PACS system configured to forward images to the XVWeb database. XVWeb is compatible with programs that run on standard "off-the-shelf personal computers, business computers, and servers running standard operating systems.
The system allows users to: manipulate (e.g. rotate, flip, etc.); enhance (e.g. increase or decrease brightness/contrast, gamma correction); add labels (e.g. measurements, lines, arrows, etc.), annotations to patient images/studies and other relevant operations for diagnostic purposes.
XVWeb is designed for medium to large dental practices and is intended for trained dental professionals and technicians to access, manipulate, and enhance dental images for diagnostic purposes only.
Prescription Use: Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over the Counter Use: No (21 CFR PART 801 Subpart C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)

Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510K__________________________________________________________________________________________________________________________________________________________________________ K132342
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.