K182737 · Nova Medical, Inc. · MOS · Apr 12, 2019 · Radiology
Device Facts
Record ID
K182737
Device Name
3T 32 Channel Head Coil
Applicant
Nova Medical, Inc.
Product Code
MOS · Radiology
Decision Date
Apr 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nova Medical 3T 32 Channel Head Coil is to be used in conjunction with a Magnetic Resonance (MR) Scanner to produce images and/or spectra of the head, that when interpreted by a trained physician yield information that may assist in diagnosis. Model # 3832016 is used with GE 3T MRI Scanner Discovery MR750. Model # 38320169 is used with GE 3T MRI Scanners Signa Architect, Discovery MR750W and Signa PET/MR.
Device Story
3T 32 Channel Head Coil is a receive-only MRI accessory for GE 3T MR systems; used for neuroimaging, fMRI, and spectroscopy. Device features 32 RF elements tuned to 127.75 MHz; constructed of two-part rigid plastic housing with patient viewing opening. Includes optional mirror and foam cushions for positioning. Operated by MRI technicians in clinical settings; coil receives RF signals from patient head during scan. Output is raw MR signal processed by the host scanner into images/spectra for physician interpretation to assist in diagnosis. Split design facilitates patient positioning; device is non-software, passive hardware.
Clinical Evidence
Clinical images were obtained per FDA guidance for Magnetic Resonance Diagnostic Devices. Images were evaluated and certified by a Chief Medical Physicist as having acceptable image quality. No formal clinical trial with diagnostic endpoints was required; performance was validated via bench testing and clinical image assessment.
Technological Characteristics
Receive-only 32-channel phased array head coil; 127.75 MHz resonant frequency. Rigid plastic housing; split design. Dimensions: ~17.5" L x 12" H x 15" W. Biocompatibility per ISO 10993-1. Electrical safety per ANSI/AAMI ES 60601-1, IEC 60601-2-33, and IEC 61000-4-2. Passive device; no internal software or microprocessors.
Indications for Use
Indicated for use in patients of all ages and body types requiring MR imaging or spectroscopy of the head.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## 510(k) Summary
## 3T 32 Channel Head Coils
## K182737
#### 1. Submission Sponsor
Nova Medical, Inc.
150 West Street, Suite 201
Wilmington
MA, 01887
USA
Contact: Cheryl Ledden
Title: Vice President
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
Office Phone: (512) 327.9997
Contact: Julie Powell
Title: VP, Quality Assurance
#### 3. Date Prepared
April 9, 2019
#### 4. Device Identification
Trade/Proprietary Name: 3T 32 Channel Head Coils
- Model 3832016 is compatible with GE MRI Scanner Discovery MR750;
{1}------------------------------------------------
- Model 38320169 is compatible with GE MRI Scanners SIGNA Architect, Discovery MR750W and SIGNA PET/MR.
| Common/Usual Name: | MRI Head Coils |
|-----------------------|-------------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device, Accessory |
| Regulation Number: | 892.1000 |
| Product Code: | MOS |
| Device Class: | Class II |
| Classification Panel: | Radiologv |
#### 5. Legally Marketed Predicate Device and Reference Device
Predicate Device: K072931; MC 3003G - 32 R Head Coil; MR Instruments Inc.
Reference Device: K142098; SIGNA PET/MR with 3.0T HNU Coil; GE Medical Systems, LLC
#### 6. Indication for Use Statement
The Nova Medical 3T 32 Channel Head Coil is to be used in conjunction with a Magnetic Resonance (MR) Scanner to produce images and/or spectra of the head, that when interpreted by a trained physician yield information that may assist in diagnosis.
Model # 3832016 is used with GE 3T MRI Scanner Discovery MR750.
Model # 38320169 is used with GE 3T MRI Scanners Signa Architect, Discovery MR750W and Signa PET/MR.
#### 7. Device Description
The Nova Medical 3T 32 Channel Head Coil is receive only MR coil for imaging the human head when used with GE Healthcare 3T MR Systems. The head coil has 32 RF elements that are tuned to 127.75 MHz. The head coil is constructed of a two part rigid plastic housing with an opening for patient viewing and comfort. The 3T 32 Channel Head Coil comprises a split design for easy patient positioning, measures approximately 17.5" long x 12″ high x 15″ wide without the tray and has an approximate weight of 20 lbs. The device comes with optional accessories; Mirror for patient viewing and Foam Cushions for patient comfort and head positioning.
#### 8. Substantial Equivalence Discussion
The following table compares the Nova Medical, Inc. 3T 32 Channel Head Coil to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{2}------------------------------------------------
| Manufacturer | Nova Medical | MR Instruments |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | 3T 32 Channel Head Coil | MC 3003G-32R Head Coil |
| | (3T 32 CH Head Coil) | |
| 510(k) Number | K182737 | K072931 |
| Attribute | | |
| Product Classification | Class II | Class II |
| Product Code | MOS | MOS |
| Regulation Number | 892.100 | 892.100 |
| Regulation Name | Magnetic Resonance diagnostic<br>device | Magnetic Resonance diagnostic<br>device |
| Indications for Use | The Nova Medical 3T 32 Channel<br>Head Coil is to be used in conjunction<br>with a Magnetic Resonance (MR)<br>Scanner to produce images and/or<br>spectra of the head, that when<br>interpreted by a trained physician<br>yield information that may assist in<br>diagnosis.<br>Model # 3832016 is used with GE 3T<br>MRI Scanner Discovery MR750.<br>Model # 38320169 is used with GE 3T<br>MRI Scanners Signa Architect,<br>Discovery MR750W and Signa<br>PET/MR. | The MR Instrument MC 3003G-32R<br>Head Coil is to be used in<br>conjunction with a MR Scanner to<br>produce images and /or spectra of<br>the head, that when interpreted by<br>a trained physician yield<br>information that may assist in<br>diagnosis. |
| Channel Array / Coil Type | 32 Channel Whole-Brain Receive-only<br>Array | 32 Channel Whole-Brain Receive-<br>only Array |
| Application | High Sensitivity Neuroimaging/<br>fMRI/spectroscopy | High Sensitivity Neuroimaging/<br>fMRI/spectroscopy |
| Conditions of Use | Model # 3832016 is used with GE 3T<br>MRI Scanner Discovery MR750.<br>Model # 38320169 is used with GE 3T<br>MRI Scanners Signa Architect,<br>Discovery MR750W and Signa<br>PET/MR. | Use with GE Discovery MR750<br>Scanner |
| Maximum Static Field | 3T | 3T |
| Manufacturer | Nova Medical | MR Instruments |
| Trade Name | 3T 32 Channel Head Coil<br>(3T 32 CH Head Coil) | MC 3003G-32R Head Coil |
| RF (nominal) | 127.75 MHz | 127.8 MHz |
| Resolution | Scanner Sequence Dependent | Scanner Sequence Dependent |
| Size | Head Coil dimensions:<br>• Outside Housing width 25.7 cm<br>• Inside Housing width 18 cm<br>• Outside Housing height 27.6 cm<br>• Inside Housing height 22 cm<br>Head Coil with Tray dimensions:<br>• Coil Length 38.5 cm<br>• Coil Width 44 cm<br>• Coil Height 29.25 cm<br>Weight - 20 lbs. including the Tray | • Inner Dimensions – 23.5cm<br>• Coil Size – 29cm W x 30cm L<br>• Coil Base – 41cm W x 48cm L<br>• Weight – 6.8kg with base |
| Shape | Rigid plastic,<br>cylindrical case, with an opening for<br>patient viewing and tray | Rigid plastic, cylindrical case with<br>an opening for patient viewing |
| Anatomical Location | Head | Head |
| Patient population (including age,<br>anatomy) | All ages and body types | All ages and body types |
| Housing Materials | Plastic | Plastic |
| Duration of Body Contact | Generally between<br>20 min. – 1 hour | Generally between<br>20 min. – 1 hour |
| Software Microprocessor | No | No |
| Complies with ISO 10993-1 | Yes | Unknown |
| Electrical Safety Testing | ANSI/AAMI ES 60601-1<br>IEC 60601-2-33<br>IEC 61000-4-2 per IEC 60601-1-2 | Unknown |
### Table 5A – Comparison of Characteristics
{3}------------------------------------------------
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of 3T 32 Channel Head Coils and in showing substantial equivalence to the predicate device, Nova Medical, Inc. completed a number of non-clinical performance tests. The 3T 32 Channel Head Coils meet all the requirements for overall design, performance, biocompatibility, and electrical safety, confirming that the design output meets the design inputs and specifications for the device.
Reference Device, K142098 (SIGNA PET/MR with 3.0T HNU Coil; GE Medical Systems, LLC), was a device used along with the Nova Medical 3T 32 Channel Head Coil for a study performed on GE SIGNA PET/MR, which is part of the 'Performance Testing' referenced below.
{4}------------------------------------------------
The 3T 32 Channel Head Coils passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- . Biocompatibility testing per ISO 10993-1
- . Electrical safety testing per IEC 60601-1
- Electrostatic discharge testing per IEC 61000-4-2 (per IEC 60601-1-2).
- Safety & essential performance of magnetic resonance equipment per IEC 60601-2-33
- . Magnetic Resonance Diagnostic Devices Testing per NEMA MS 6
- . Transport Testing per ISTA Procedure 2B
- . Performance Testing: A number of performance tests were performed by Nova Medical, Inc. as part of the design and development process; results were deemed acceptable.
#### 10. Clinical Performance Data
Clinical images were obtained in accordance with the FDA Guidance (November 18, 2016) for Submission of Premarket Notification for Magnetic Resonance Diagnostic Devices. The clinical images were deemed to be of acceptable image quality and were certified by a Chief Medical Physicist.
#### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
The Nova Medical, Inc. 3T 32 Channel Head Coils, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nova Medical, Inc. % Ms. Julie Powell Vice President, Quality Assurance Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 AUSTIN TX 78746
April 12, 2019
Re: K182737
Trade/Device Name: 3T 32 Channel Head Coils Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: March 22, 2019 Received: March 25, 2019
Dear Ms. Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{6}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.