K142098 · Ge Medical Systems, LLC · OUO · Nov 24, 2014 · Radiology
Device Facts
Record ID
K142098
Device Name
SIGNA PET/MR
Applicant
Ge Medical Systems, LLC
Product Code
OUO · Radiology
Decision Date
Nov 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, and diagnosis of diseases and disorders. MR is intended to produce transverse, sagittal, coronal and oblique cross-sectional MR images and/ or spectra, and displays the internal struction of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles. PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems.
Device Story
SIGNA PET/MR combines 3.0T MR and PET scanners for simultaneous, isocentric, whole-body imaging; used for oncology, neurology, and cardiology. MR component provides anatomical/spectroscopic data; PET component measures radiopharmaceutical distribution. System uses MR data for PET attenuation correction, replacing CT-based attenuation. Operated by trained healthcare professionals in clinical settings. Output includes fused high-resolution metabolic and anatomic images; aids physicians in disease diagnosis and metabolic/physiologic evaluation. Supports interventional procedures with MR-compatible devices. Benefits include simultaneous acquisition, precise anatomical reference, and reduced radiation exposure by eliminating CT-based attenuation correction.
Clinical Evidence
No clinical trials; bench testing only. Evidence consists of sample images collected from multiple sites to confirm simultaneous diagnostic image quality and verification/validation testing including design reviews, module verification, integration testing, and simulated use testing.
Technological Characteristics
3.0T superconducting magnet; gradient coil; body coil; local surface RF coils. PET detectors modified for MR bore integration. Connectivity: DICOM. Standards: IEC 60601-1, 60601-1-2, 60601-2-33; NEMA MS1-MS5, MS8, MS9, NU2, PS3.1-3.18. Software-based image reconstruction and processing.
Indications for Use
Indicated for use by trained healthcare professionals to aid in detection, localization, and diagnosis of diseases and disorders, including cardiovascular disease, neurological disorders, and cancer, in patients requiring MR and/or PET imaging.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other and are connected by a flowing line.
November 24, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
GE Medical Systems, LLC % Ms. Robin Martin Regulatory Affairs Manager 3000 N. Grandview Blvd. WAUKESHA WI 53188
Re: K142098 Trade/Device Name: SIGNA PET/MR Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: OUO Dated: November 14, 2014 Received: November 17, 2014
Dear Ms. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### k142098
Device Name SIGNA PET/MR
#### Indications for Use (Describe)
The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, and diagnosis of diseases and disorders. MR is intended to produce transverse, sagittal, coronal and oblique cross-sectional MR images and/ or spectra, and displays the internal struction of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles. PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. There are decorative swirls or flourishes around the circle, adding a touch of elegance to the design. The logo is simple, recognizable, and represents the well-known multinational corporation.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | July 31, 2014 |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter (Manufacturer): | GE Medical Systems, LLC.<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Manufacturing Site: | GE Medical Systems, LLC.<br>3000 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contact Person: | Robin Martin<br>Regulatory Affairs Manager<br>GE Medical Systems, LLC<br>Phone: (414) 336-9228<br>Fax: (262) 894-4968 |
| Secondary Contact Person: | Glen Sabin<br>Regulatory Affairs Director<br>GE Medical Systems, LLC<br>Phone: (262) 521-6848<br>Fax: (262) 894-4968 |
| Device Trade Name: | SIGNA PET/MR |
| Common/Usual Name: | Magnetic Resonance Diagnostic Device /<br>Positron Emission Tomography (PET) System |
| Classification Names:<br>Product Code: | Emission Computed Tomography System per 21 CFR 892.1200<br>OUO |
| Predicate Device(s): | K133226 Siemens Biograph mMR<br>Reference Devices:<br>K081496 Discovery XR (Currently marketed as Discovery PET/CT 690)<br>K132376 Discovery MR750w 3.0T |
| Device Description: | The GE SIGNA PET/MR system is a combined Magnetic Resonance Diagnostic<br>Device (MRDD) and Positron Emission Tomography (PET) scanner. The system<br>is designed for whole body oncology, neurology and cardiology examinations.<br>The SIGNA PET/MR system provides simultaneous acquisition of high-<br>resolution metabolic and anatomic information from the two major components<br>of each system (MR and PET). Additional components of the system include: a<br>patient table and both the acquisition and processing workstations with associated<br>software.<br><br>The SIGNA PET/MR includes a 3.0T superconducting magnet, gradient coil,<br>body coil and local surface RF coils based on those of the reference device<br>Discovery MR750w 3.0T. The system includes dual drive capabilities. The<br>SIGNA PET detectors have been modified from those of the reference device, the<br>Discovery PET/CT 690, to allow them to be integrated into the bore of the MR.<br>This allows for simultaneous, precisely aligned whole body MR and PET<br>acquisition. Similar to Discovery PET/CT D690, PET supports Time of Flight |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. There are decorative swirls around the outside of the circle, giving it a classic and recognizable appearance.
| | (ToF). SIGNA PET/MR software is based on a combination of Discovery<br>MR750w with Discovery PET/CT 690 software. It is used for patient<br>management, data management, scan control, image reconstruction and image<br>archival and evaluation. All images conform to DICOM imaging format<br>requirements.<br>The SIGNA PET/MR system and surface coil suite, the subject of this<br>application, is substantially equivalent to the commercially available devices<br>above with modifications made to integrate the two modalities together into a<br>whole-body system. |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | whole-body system. |
| Intended Use: | The SIGNA PET/MR system combines magnetic resonance diagnostic devices<br>(MRDD) and Positron Emission Tomography (PET) scanners that provide<br>registration and fusion of high resolution physiologic and anatomic information,<br>acquired simultaneously and isocentrically. The combined system maintains<br>independent functionality of the MR and PET devices, allowing for single<br>modality MR and / or PET imaging. |
| | These systems are intended to be utilized by appropriately trained health care<br>professionals to aid in the detection, localization, and diagnosis of diseases and<br>disorders. The MR is intended to produce transverse, sagittal, coronal and<br>oblique cross-sectional MR images, spectroscopic images and/or spectra, and<br>displays the internal structure and/or function of the human body. Other physical<br>parameters derived from the images and/or spectra may also be produced.<br>Depending on the region of interest, approved contrast agents may be used, as<br>described in their labeling. This system may also be used for imaging during<br>interventional procedures when performed with MR compatible devices, such as<br>MR safe biopsy needles. |
| | The PET images and measures the distribution of PET radiopharmaceuticals in<br>humans to aid the physician in determining various metabolic (molecular) and<br>physiologic functions within the human body for evaluation of diseases and<br>disorders such as, but not limited to, cardiovascular disease, neurological<br>disorders and cancer. |
| | The combined system utilizes the MR for radiation-free attenuation correction<br>maps for PET studies. The system provides inherent anatomical reference for the<br>fused PET and MR images due to precisely aligned MR and PET image<br>coordinate systems. |
| Technology: | The proposed SIGNA PET/MR employs much of the same fundamental<br>scientific technology as its respective reference devices for MR and PET<br>technology and has the same intended use as the claimed predicate device. The<br>proposed device combines PET and MR technologies for simultaneous imaging.<br>Subsequently, the PET subsystem requires different technology from the<br>Discovery PET/CT 690 for compatibility with the MR subsystem (detectors,<br>chiller system, etc.). Additionally, MR images are used to correct for PET<br>attenuation compared to the reference PET/CT, which uses CT technology for<br>PET attenuation. |
| Determination of<br>Substantial Equivalence: | Summary of Non-Clinical Tests:<br>The SIGNA PET/MR and Coil Suite comply with the following voluntary |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and the circle is surrounded by a thin blue line. The logo is simple and recognizable, and it is often used to represent the company.
| | standards (3rd Edition):<br>IEC 60601-1 (system and coil suite) IEC 60601-1-2 (system and coil suite) IEC 60601-2-33 (system and coil suite) In addition, this SIGNA PET/MR scanner and surface coils are in compliance<br>with the applicable NEMA standards, including NEMA MS1, MS2, MS3, MS4,<br>MS5, MS8, MS9, NEMA NU2 and NEMA PS3.1-3.18.<br><br>The following quality assurance measures were applied to the development of the<br>system:<br>Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) Summary of Clinical Tests:<br>Sample images from the SIGNA PET/MR were collected from multiple sites to<br>confirm simultaneous diagnostic image quality. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | GE Medical Systems, LLC. considers the SIGNA PET/MR to be as safe, as<br>effective, and performance is substantially equivalent to the predicate device and<br>reference devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.