K182617 · Alphatec Spine, Inc. · PDQ · May 9, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K182617
Device Name
ATEC IOM Accessory Instruments
Applicant
Alphatec Spine, Inc.
Product Code
PDQ · Ear, Nose, Throat
Decision Date
May 9, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
Device Story
ATEC IOM Accessory Instruments are surgical tools (drills, taps, awls, probes, screwdrivers, dilators, guidewires, needles) used in spine surgery; provide electrical stimulation to body tissues to locate/identify nerves; compatible with standard neuromonitoring consoles via clips or probes; provide surgeon with nerve location, proximity, and integrity data; used in open or percutaneous minimally invasive procedures; assists surgeon in targeting, bone preparation, and implant placement; benefits patient by reducing risk of nerve injury during spinal instrumentation.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including functional performance, electrical safety, biocompatibility, and reprocessing validation.
Technological Characteristics
Instruments include drills, taps, awls, probes, screwdrivers, dilators, guidewires, and needles. Electrical stimulation principle; compatible with external neuromonitoring consoles. Biocompatible materials; limited patient contact (≤ 24 hours). Sterilization via validated steam (ANSI/AAMI/ISO 17665-1). Electrical safety per IEC 60601-1.
Indications for Use
Indicated for use in spine surgical procedures to assist in nerve location during or after preparation and placement of intervertebral fusion cages and pedicle screw fixation devices in open and percutaneous minimally invasive approaches.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
May 9, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Alphatec Spine, Inc Jeremy Markovich Senior Manager, Regulatory and Clinical Affairs 5818 El Camino Real Carlsbad, California 92008
## Re: K182617
Trade/Device Name: ATEC IOM Accessory Instruments Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ, ETN Dated: April 5, 2019 Received: April 8, 2019
## Dear Jeremy Markovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Carlos L. Peña, PhD, MS Director OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182617
Device Name ATEC IOM Accessory Instruments
### Indications for Use (Describe)
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white letter "a" inside of it. To the right of the square is the word "tec" in white font. The background of the logo is a dark blue color.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>5818 El Camino Real<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286<br>Fax: (760) 431-0289 |
|----|------------------------|-------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Jeremy Markovich<br>Senior Manager, Regulatory and Clinical Affairs<br>Contact Phone: (760) 431-9286 |
| | Date Summary Prepared: | May 8, 2019 |
#### II. DEVICE
| Trade or Proprietary Name: | ATEC IOM Accessory Instruments |
|----------------------------|------------------------------------|
| Common Name: | Surgical nerve stimulator/locator. |
| Classification Name: | Neurosurgical Nerve Locator |
| Regulation Number: | 21 CFR 874.1820 |
| Classification: | Class II |
| Product Code: | PDQ, ETN |
#### III. LEGALLY MARKETED PREDICATE DEVICES
| Predicate Device(s): | |
|----------------------|--|
| | |
| 510(k) | Product Name | Clearance Date |
|---------|---------------------------------------------------|-------------------|
| K171807 | Stryker ES2 Neuromonitoring Accessory Instruments | July 18, 2017 |
| K142438 | Pioneer Surgical Nerve Monitoring Cable System | March 5, 2015 |
| K182542 | The EPAD 2 System | February 22, 2019 |
## Reference Device(s):
| 510(k) | Product Name | Clearance Date |
|---------|--------------------------------------|-------------------|
| K183705 | IdentiTi Porous Ti Interbody System | March 1, 2019 |
| K161363 | Arsenal Posterior Fixation System | June 10, 2016 |
| K133221 | Arsenal Posterior Fixation System | March 13, 2014 |
| K123623 | Illico MIS Posterior Fixation System | February 14, 2013 |
| K110170 | Raptor Facet Fixation System | November 28, 2011 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white curved line inside, followed by the word "atec" in white. The "c" in "atec" has a trademark symbol next to it. The background of the logo is dark blue.
#### DEVICE DESCRIPTION IV.
The ATEC IOM Accessory Instruments are surgical instruments that provide electrical stimulation to the body to locate and identify nerves in either open, minimally invasive, or percutaneous procedures. These surgical instruments are compatible with common FDA cleared neuromonitoring platforms as they are connected via a compatible clip or probe depending on the system. The neuromonitoring capability provides the surgeon with spinal nerve location, proximity, and integrity information. This information assists the surgeon during targeting, bone preparation, and placement of orthopedic implants such as intervertebral fusion devices (e.g., interbodies) and bone screws (e.g., pedicle screws).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the "atec" logo. The logo consists of a green square with a white stylized "a" inside it, followed by the word "tec" in white. There is a trademark symbol after the word "tec". The background is dark blue.
| | Predicate Device<br>K171807 | Predicate Device<br>K142438 | Predicate Device<br>K182542 | Subject Device |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Stryker ES2<br>Neuromonitoring Accessory<br>Instruments | Pioneer Surgical Nerve<br>Monitoring Cable System | The EPAD 2 System | ATEC<br>IOM Instruments |
| Indications for<br>Use | The ES2 Neuromonitoring<br>instruments (Awls, Taps,<br>Screwdriver and LITe Y-<br>NEEDLE 200, 300 and 400) can<br>be used by the surgeon to assist<br>in location of the spinal nerves<br>by providing proximity<br>information before, during or<br>after bone preparation and<br>placement of bone screws in<br>open and percutaneous<br>minimally invasive posterior<br>surgical approaches of the non-<br>cervical spine. | The Nerve Monitoring Cable in<br>conjunction with dilators,<br>pedicle probes, taps, awls or<br>screw drivers, are intended for<br>tissue dilation/dissection and<br>stimulation of peripheral nerves<br>including spinal nerve roots for<br>location and identification<br>during spinal surgery.<br>The dilators are also intended for<br>use in surgical procedures to<br>provide surgical access by<br>dilating the soft tissue to the<br>intended surgical site to allow<br>passage of current from a point<br>on the proximal end to an<br>uninsulated portion of the distal<br>tip. The purpose of this is to<br>allow controlled monitoring of<br>neural elements near and around<br>the point of access. The<br>dilators are offered sterile/single<br>use. | The EPAD 2 system is intended<br>for use in monitoring<br>neurological status by recording<br>somatosensory evoked potentials<br>(SSEP), electromyography<br>(EMG), or assessing the<br>neuromuscular junction (NMJ).<br>Neuromonitoring procedures<br>include intracranial,<br>extracranial, intratemporal,<br>extratemporal, neck dissections,<br>upper and lower extremities,<br>spinal degenerative treatments,<br>pedicle screw fixation,<br>intervertebral fusion cages,<br>rhizotomy, orthopedic surgery,<br>open or percutaneous, lumbar,<br>thoracic, and cervical<br>surgical procedures. | The ATEC IOM Accessory<br>Instruments are utilized in spine<br>surgical procedures to assist in<br>location of the nerves during or<br>after preparation and placement<br>of implants (intervertebral fusion<br>cages and pedicle screw fixation<br>devices) in open and<br>percutaneous minimally invasive<br>approaches. |
| Regulation,<br>Product Code, &<br>Class | 21 CFR 874.1820<br>PDQ<br>Class II | 21 CFR 874.1820<br>21 CFR 882.1350<br>PDQ, ETN, GXZ<br>Class II | 21 CFR 882.1870<br>21 CFR 874.1820<br>21 CFR 882.1350<br>21 CFR 882.1320<br>GWF, ETN, PDQ, GXY, GXZ<br>Class II | 21 CFR 874.1820<br>PDQ, ETN<br>Class II |
| IEC 60601<br>Compliant | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) |
| Instrument Type<br>(Description) | Awls, Taps, Screwdrivers,<br>Guidewires (K-wires), and 200,<br>300, 400 LITE Y-Needles | Dilators, Pedicle Probes, Taps,<br>Awls and Screwdrivers | Probes, Dilators, Electrodes | Drills, Taps, Awls, Probes,<br>Screwdrivers, Dilators (Sleeves),<br>Guidewires, and Needles |
| Biocompatibility<br>Patient Contact<br>Duration | Limited patient duration contact<br>(≤ 24 hours) | Limited patient duration contact<br>(≤ 24 hours) | Limited patient duration contact<br>(≤ 24 hours) | Limited patient duration contact<br>(≤ 24 hours) |
| Biocompatible | Yes<br>(Characterizatio | Yes | Yes | Yes |
| Surgical<br>Approach | Open or Percutaneous/<br>Minimally Invasive | Open or Percutaneous/<br>Minimally Invasive | Open or Percutaneous/<br>Minimally Invasive | Open or Percutaneous/<br>Minimally Invasive |
| Sterility | Sterile and Non-sterile<br>Non-sterile devices are provided<br>with validated steam sterilization<br>parameters to assure an SAL of<br>$10^{-6}$ | Sterile and Non-sterile<br>Non-sterile devices are provided<br>with validated steam sterilization<br>parameters to assure an SAL of<br>$10^{-6}$ | Sterile and Non-sterile | Non-sterile<br>Non-sterile devices are provided<br>with validated steam sterilization<br>parameters to assure an SAL of<br>$10^{-6}$ |
| Reusable/<br>Single Use | 200, 300, and 400 LITE Y<br>Needles – Single Use<br>Awls, Taps, and Screwdrivers -<br>Reusable | Dilators – Single Use<br>Pedicle probes, taps, awls and<br>screw drivers - Reusable | Single Use | Guidewires, Targeting Needles<br>and Dilators – Single Use<br>Awls, Drills/Taps, Probes,<br>Dilators (Sleeves), and<br>Screwdrivers - Reusable |
| Compatible with<br>Common<br>Neuromonitoring<br>Consoles &<br>Software | Yes | Yes | Yes | Yes |
| Connection to<br>Neuromonitoring<br>Unit | Clip or probe<br>(based on Neuromonitoring<br>system) | Clip or probe<br>(based on Neuromonitoring<br>system) | Clip or probe<br>(based on Neuromonitoring<br>system) | Clip or probe<br>(based on Neuromonitoring<br>system) |
| Minimum<br>exposed surface<br>area during<br>tissue<br>stimulation | 0.53 mm² | 12.9032 mm²<br>(0.02 in²) | Unknown | 8.6 mm² |
# Table 1: Comparison for Substantial Equivalence
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for a company called "atec". The logo consists of a green square with a white "a" inside of it, followed by the word "tec" in white. The background of the logo is a dark blue color. There is a trademark symbol after the word "tec".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Atec. The logo features a green square with a white letter "a" inside of it. To the right of the square is the word "tec" in white font with a trademark symbol.
| Table 2: Comparison for Substantial Equivalence for Reference Device(s) | | | | | |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Reference Device<br>K183705 | Reference Device<br>K110170 | Reference Device<br>K123623 | Reference Device<br>K133221<br>K161363 | Subject Device |
| Regulation,<br>Product Code, &<br>Class | IdentiTi Porous Ti<br>Interbody System<br>21 CFR 888.3080<br>MAX<br>Class II | Raptor Facet Fixation<br>System<br>21 CFR 888.4540<br>LXH<br>Class I | Illico MIS Posterior<br>Fixation System<br>21 CFR 888.4540<br>LXH<br>Class I | Arsenal Posterior<br>Fixation System<br>21 CFR 888.4540<br>LXH<br>Class I | ATEC<br>IOM Instruments<br>21 CFR 874.1820<br>PDQ, ETN<br>Class II |
| Instrument Type<br>(Description) | Inserter | Screwdrivers, Guidewires | Guidewires, Dilators<br>(Sleeves) | Screwdrivers | Drills, Taps, Awls, Probes,<br>Screwdrivers, Dilators<br>(Sleeves), Guidewires, and<br>Targeting Needles |
| Biocompatibility<br>Patient Contact<br>Duration | Limited patient duration<br>contact<br>(≤ 24 hours) | Limited patient duration<br>contact<br>(≤ 24 hours) | Limited patient duration<br>contact<br>(≤ 24 hours) | Limited patient duration<br>contact<br>(≤ 24 hours) | Limited patient duration<br>contact<br>(≤ 24 hours) |
| Biocompatible | Yes (Data provided to<br>characterize materials)<br>Non-sterile | Yes (Data provided to<br>characterize materials)<br>Non-sterile | Yes (Data provided to<br>characterize materials)<br>Non-sterile | Yes (Data provided to<br>characterize materials)<br>Non-sterile | Yes<br>Non-sterile |
| Sterility | Non-sterile devices are<br>provided with validated<br>steam sterilization<br>parameters to assure an<br>SAL of 10-6 | Non-sterile devices are<br>provided with validated<br>steam sterilization<br>parameters to assure an<br>SAL of 10-6 | Non-sterile devices are<br>provided with validated<br>steam sterilization<br>parameters to assure an<br>SAL of 10-6 | Non-sterile devices are<br>provided with validated<br>steam sterilization<br>parameters to assure an<br>SAL of 10-6 | Non-sterile devices are<br>provided with validated<br>steam sterilization<br>parameters to assure an<br>SAL of 10-6 |
| Reusable/<br>Single Use | Inserter - Reusable | Guidewires - Single Use<br>Screwdrivers - Reusable | Guidewires - Single Use<br>Dilator - Reusable | Screwdrivers - Reusable | Guidewires, Targeting<br>Needles and Dilators -<br>Single Use<br>Awls, Drills/Taps, Probes,<br>Dilators (Sleeves), and<br>Screwdrivers - Reusable |
| Compatible with<br>Common<br>Neuromonitoring<br>Consoles &<br>Software | No | No | No | No | Yes |
| Connection to<br>Neuromonitoring<br>Unit | Not applicable | Not applicable | Not applicable | Not applicable | Clip or probe<br>(based on Neuromonitoring<br>system) |
# Table 2: Comparison for Substantial Equivalence for Reference Device(s)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the "atec" logo. The logo consists of a green square with a white design inside, followed by the word "atec" in white font. The background is a dark blue color.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for a company called "atec". The logo consists of a green square with a white stylized "a" inside of it, followed by the white text "tec" with a trademark symbol. The background of the logo is a dark blue color. The logo is simple and modern, and the colors are eye-catching.
#### V. INDICATIONS FOR USE
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
#### TECHNOLOGICAL COMPARISON TO PREDICATES VI.
The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent.
#### VII. PERFORMANCE DATA
Performance testing demonstrates the subject ATEC IOM Accessory Instruments are appropriate for neuromonitoring applications and do not introduce a new worst case for minimum exposed surface area during tissue stimulation compared to other legally marketed devices cleared by FDA.
The table below summarizes the testing which was performed on the subject devices to show substantial equivalence to the predicate devices.
| Test | Test Method Summary | Results |
|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional<br>performance testing<br>and verification<br>analysis | Insulation Effectiveness Electrical Resistance Current Density | All functional performance testing<br>passed. Substantial equivalence has been<br>shown via analysis and testing. |
| Electrical safety<br>testing and/or<br>evaluation | Evaluation and testing was performed<br>on the subject devices in accordance<br>with IEC 60601-1: 2005. | Subject devices passed electrical safety<br>testing and/or evaluation, demonstrating<br>that the devices meet the requirements. |
| Biocompatibility<br>testing | Biocompatibility testing conducted<br>per ISO 10993. Cytotoxicity Sensitization Irritation/Intracutaneous<br>Reactivity Acute Systemic Toxicity Material Mediated Pyrogenicity Hemocompatibility | A risk analysis was performed taking<br>into account nature of body contact and<br>duration to categorization the use of<br>existing data, end-specific testing, and<br>endpoint assessment to cover the<br>identified test methods. Additionally,<br>data was leveraged by other means (e.g.,<br>authorized use of Master File, predicate<br>and reference devices, well known and<br>characterized materials) to support the<br>biocompatibility of the subject devices. |
| Reprocessing | Cleaning validation study based<br>on acceptance criteria from<br>AAMI TIR30:2011 Steam sterilization validation<br>performed per ANSI/AAMI/ISO<br>17665-1:2006/(R)2013. | The subject devices can be adequately<br>cleaned and steam sterilized prior to use. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the "atec" logo. The logo consists of a green square with a white stylized "a" inside, followed by the white text "tec" with a trademark symbol. The background is a dark blue color.
## Clinical Information
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
#### CONCLUSION VIII.
Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regards to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.