K142438 · Pioneer Surgical Technology, Inc. (Rti Surgical, Inc.) · PDQ · Mar 5, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K142438
Device Name
Nerve Monitoring Cable System
Applicant
Pioneer Surgical Technology, Inc. (Rti Surgical, Inc.)
Product Code
PDQ · Ear, Nose, Throat
Decision Date
Mar 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The Nerve Monitoring Cable in conjunction with dilators, pedicle probes, taps, awls or screw drivers, are intended for tissue dilation/dissection and stimulation of peripheral nerves including spinal nerve roots for location and identification during spinal surgery. The dilators are also intended for use in surgical procedures to provide surgical access by dilating the soft tissue to the intended surgical site to allow passage of current from a point on the proximal end to an uninsulated portion of the distal tip. The purpose of this is to allow controlled monitoring of neural elements near and around the point of access. The dilators are offered sterile/single use.
Device Story
Nerve Monitoring Cable System facilitates tissue dilation and peripheral nerve stimulation during spinal surgery; used in open or percutaneous/minimally invasive procedures. System consists of Nerve Monitoring Cables (1.5 mm female DIN connector) and stimulation accessories (pedicle probes, taps, awls, screw drivers, and sterile/single-use dilators). Dilators are hollow tubes (Aluminum 6061-T6) with Parylene C insulation; distal tip remains uninsulated to deliver electrical current. Surgeon connects cable to an FDA-cleared neuromonitoring console (Type BF or CF rating) and stimulation accessory. Current passes from proximal end to distal tip to monitor neural elements near access site. Provides proximity information to surgeon; assists in nerve identification; helps avoid neural injury during surgical access and instrumentation.
Clinical Evidence
No human clinical data. Evidence includes bench testing (electrical resistance, insulation/Hipot testing, charge/current/power density calculations) and a porcine animal study to assess performance, functionality, and safety. Biocompatibility testing per ISO 10993-1 confirmed non-cytotoxic, non-irritant, non-sensitizing, and non-systemically toxic status. Sterilization, packaging, and shelf-life validated per ISO 11137-2, ISO 11607, and ASTM D4169.
Technological Characteristics
Materials: Aluminum 6061-T6 (ASTM B221) with Parylene C insulation. Form factor: Hollow tubes for tissue dilation. Energy: Electrical stimulation via external neuromonitoring console. Connectivity: Wired connection via 1.5 mm female DIN connector to console; clip connection to accessories. Sterilization: Sterile, single-use. Electrical safety: Compliant with IEC 60601-1.
Indications for Use
Indicated for patients undergoing spinal surgery requiring tissue dilation/dissection and stimulation of peripheral nerves, including spinal nerve roots, for location and identification.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, ribbon-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 5, 2015
Pioneer Surgical Technology, Inc. Sarah Pleaugh Regulatory Affairs Specialist 375 River Park Circle Marquette, MI 49855 USA
Re: K142438
Trade/Device Name: Nerve Monitoring Cable System Regulation Number: 21 CFR 874.1820 Regulation Name: Neurosurgical Nerve Locator Regulatory Class: Class II Product Code: PDQ, ETN, GXZ Dated: January 30, 2015 Receive: February 2, 2015
Dear Ms. Pleaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, Ph.D., M.S. for Carlos L Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142438
Device Name Nerve Monitoring Cable System
#### Indications for Use (Describe)
The Nerve Monitoring Cable in conjunction with dilators, pedicle probes, taps, awls or screw drivers, are intended for tissue dilation/dissection and stimulation of peripheral nerves including spinal nerve roots for location and identification during spinal surgery.
The dilators are also intended for use in surgical procedures to provide surgical access by dilating the soft tissue to the intended surgical site to allow passage of current from a point on the proximal end to an uninsulated portion of the distal tip. The purpose of this is to allow controlled monitoring of neural elements near and around the point of access. The dilators are offered sterile/single use.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Pursuant to 21 CFR 807.92
K142438
| 510(k) Owner: | Pioneer Surgical Technology, Inc. (RTI Surgical, Inc.)<br>375 River Park Circle<br>Marquette, MI 49855 USA<br>Ph.: (906) 225-5861<br>Fax: (906) 226-4459<br>Contact: Sarah (McIntyre) Pleaugh<br>Prepared: January 30, 2015 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name: | Nerve Monitoring Cable System |
| Common name: | Neurosurgical Nerve Locator |
| Classification: | Class II;<br>21 CFR 874.1820, PDQ, Neurosurgical nerve locator and ETN, Nerve stimulator<br>21 CFR 882.1350, GXZ, Needle electrode |
| Predicate<br>Devices: | K073229 Pioneer Surgical Technology, Inc. - Nerve Monitoring Cable System<br>K132373 Biomet Spine - Biomet Probes/Guidewires and Dilators<br>K140400 Stryker Spine - ES2® Neuromonitoring Accessory Instruments |
| Device<br>Description: | The Nerve Monitoring Cable System, which consists of Nerve Monitoring Cables<br>in conjunction with various stimulation accessories, allow for tissue dilation and<br>stimulation of peripheral nerve roots for location and identification during open and<br>percutaneous / minimally invasive spinal surgery. The Nerve Monitoring Cable<br>includes a 1.5 mm female DIN connector for use with a stimulating console with a<br>Type BF or CF rating. The other end of the Nerve Monitoring Cable includes a clip<br>for connection to the stimulation accessories of this system.<br><br>The stimulation accessories include general manual orthopedic surgical instruments<br>previously cleared via K073227 (i.e. pedicle probes, taps, awls and screw drivers)<br>and also include sterile/ single-use dilators which are subject of this submission.<br>The subject dilators are a series of hollow tubes that fit over one another for the<br>purpose of dilating tissue in preparation for spinal surgery. Similar to the predicate<br>instrument accessories, they are made from a conducting metal (Aluminum 6061-<br>T6) with an insulated outer coating (Parylene C) to allow passage of current from a<br>point on the proximal end where it connects to the Nerve Monitoring Cable to an<br>un-insulated portion of the distal tip. The purpose of this is to allow controlled<br>monitoring of neural elements near and around the point of access. The subject<br>dilators are offered sterile/single-use. |
| Intended Use: | The Nerve Monitoring Cable in conjunction with dilators, pedicle probes, taps,<br>awls or screw drivers, are intended for tissue dilation/dissection and stimulation of<br>peripheral nerves including spinal nerve roots for location and identification during<br>spinal surgery.<br><br>The dilators are also intended for use in surgical procedures to provide surgical<br>access by dilating the soft tissue to the intended surgical site to allow passage of<br>current from a point on the proximal end to an uninsulated portion of the distal tip.<br>The purpose of this is to allow controlled monitoring of neural elements near and |
{4}------------------------------------------------
| around the point of access. The dilators are offered sterile/single use. | | | | | | | | | | | | | | | | |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Technological<br>Characteristics: | The subject dilators are substantially equivalent to the predicate devices in terms of<br>design, function, principles of operation, technological characteristics, intended use,<br>and performance.<br><br>As compared to the Biomet predicate, the subject dilators are similar in that they<br>are a series of hollow tubes intended for tissue dilation and stimulation of nerves<br>for neuromonitoring applications. They are made from metal with an insulated<br>outer coating. Similar to the Biomet and Stryker predicates, the electrical signal is<br>supplied via FDA cleared neuromonitoring systems attached to a clip or probe that<br>is attached to the neuromonitoring contact zone on the proximal end. The electrical<br>signal is then transferred to a point on the distal end which is uninsulated<br>(minimum exposed surface area is SE to the predicate devices). | | | | | | | | | | | | | | | |
| Materials: | The subject dilators are made from Aluminum 6061-T6 conforming to ASTM B221<br>with an insulated outer coating (Parylene C). | | | | | | | | | | | | | | | |
| Biocompatibility: | Biocompatibility testing was conducted on the subject dilators in accordance with<br>the testing recommendations in ISO 10993-1 (Biological Evaluation of Medical<br>Devices Part 1: Evaluation and Testing). The test results are summarized in the<br>table below:<br>Test Results Conclusion Cytotoxicity No evidence of cytotoxicity. Non-cytotoxic Irritation No evidence of irritation. Non-irritant Sensitization No evidence of sensitization. Non-sensitizing Systemic Toxicity No evidence of toxicity. Not systemically toxic | | | | | | | | | | | | | | | |
| | As shown in the table, biocompatibility testing found the subject dilators to be non-<br>cytotoxic, non-irritant, non-sensitizing, and not systemically toxic. | | | | | | | | | | | | | | | |
| Summary of<br>Performance<br>Data: | This 510(k) premarket notification is seeking clearance of subject dilators as<br>accessory instruments during neuromonitoring applications. Performance testing<br>and engineering calculations were performed to demonstrate that the subject<br>devices are substantially equivalent to the identified predicates in terms of design,<br>performance and intended use.<br><br>Engineering analysis confirmed the maximum charge, current, and power densities<br>of the subject devices fall within the range of predicate devices. IEC 60601-1<br>testing for electrical safety on applicable devices of the subject system was<br>supplied. A porcine animal study was conducted to assess the performance,<br>functionality, and safety or the subject dilators as neuromonitoring accessory<br>instruments. Electrical/resistance bench testing confirmed the subject dilators are<br>capable of transferring an electrical current with little or no resistance in a similar<br>manner to predicate devices. This bench testing, along with in-process Hipot<br>testing, also confirmed the coating of each device provides adequate electrical<br>insulation. Sterilization, packaging, and shelf-life validations were also completed<br>per applicable test standards (ISO 11137-2, ISO 11607, and ASTM D4169<br>respectively). | | | | | | | | | | | | | | | |
{5}------------------------------------------------
| The performance data confirmed the substantial equivalence of the subject devices |
|-------------------------------------------------------------------------------------|
| and demonstrates the subject devices are as safe, as effective, and perform as well |
| as, or better than, the predicate devices. |
### Substantial Equivalence Summary - Comparison of Subject and Predicate Systems
| Characteristic | Subject<br>K142438 Nerve<br>Monitoring Cable<br>System | Predicate<br>K073229 Nerve<br>Monitoring Cable<br>and Reusable<br>Accessory<br>Instruments | Predicate<br>K132373<br>Biomet Spine<br>Probes/<br>Guidewires and<br>Dilators | Predicate K140400<br>ES2® Neuro-<br>monitoring<br>Accessory<br>Instruments | Substantial<br>Equivalence |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Neuromonitoring<br>Accessories<br>Instruments | Dilators | General, manual<br>orthopedic reusable<br>surgical<br>instruments (e.g.<br>probes, awls, taps,<br>inserters) | Probes,<br>guidewires and<br>dilators | Awls, Taps and<br>Screwdriver to be<br>used with Dilators or<br>Tap Sleeves | Yes |
| Use of Dilators | Dilators | No dilators in<br>K073229 | Dilators | Dilators or Tap<br>Sleeve | Yes |
| Compatible with<br>Common<br>Neuromonitoring<br>Consoles &<br>Software | Compatible with<br>FDA cleared<br>neuromonitoring<br>systems | Compatible with<br>FDA cleared<br>neuromonitoring<br>systems | Compatible with<br>FDA cleared<br>neuromonitoring<br>systems | Compatible with<br>FDA cleared<br>neuromonitoring<br>systems | Yes |
| Connection to<br>Neuromonitoring<br>Unit | Clip | Clip | Clip or Probe | Clip or Probe (based<br>on Neuromonitoring<br>System used) | Yes |
| Materials | ASTM B221 6061<br>Aluminum (UNS<br>A96061) with<br>Parylene C Coating | -ASTM F899 17-4<br>SST with Halar®<br>6014 Coating<br>-Detachable<br>silicone handles or<br>RADEL® handles | ASTM F899<br>Stainless steel<br>with polymeric<br>dielectric<br>coating | -Awl, Taps, &<br>Screwdriver:<br>Surgical Grade<br>Stainless Steel<br>-Dilators & Tap<br>Sleeve: RADEL® | Yes |
| Indications for<br>Use | The Nerve<br>Monitoring Cable in<br>conjunction with<br>dilators, pedicle<br>probes, taps, awls or<br>screw drivers, are<br>intended for tissue<br>dilation/ dissection<br>and stimulation of<br>peripheral nerves<br>including spinal nerve<br>roots for location and<br>identification during<br>spinal surgery.<br>The dilators are<br>also intended for use<br>in surgical procedures<br>to provide surgical<br>access by dilating the<br>soft tissue to the<br>intended surgical site<br>to allow passage of<br>current from a point | The Pioneer Nerve<br>Monitoring Cable,<br>in conjunction with<br>Pioneer pedicle<br>probes, taps, awls<br>or screw drivers, is<br>intended to<br>stimulate<br>peripheral motor<br>nerves during<br>surgery for the<br>purpose of locating<br>and identifying<br>these nerves,<br>including spinal<br>nerve roots during<br>the incision and<br>removal of soft and<br>hard tissue or bone. | The Biomet<br>probes/<br>guidewires and<br>dilators are<br>intended for<br>tissue dilation<br>and stimulation<br>of peripheral<br>nerves including<br>spinal nerve<br>roots for<br>location and<br>identification<br>during spinal<br>surgery. | The ES2® Awl,<br>ES2® Taps, and<br>ES2® Screwdriver<br>can be used to assist<br>in location of the<br>spinal nerves by<br>providing proximity<br>information before,<br>during or after bone<br>preparation and<br>placement of bone<br>screws in open and<br>percutaneous<br>posterior surgical<br>approaches of the<br>non-cervical spine. | Yes – There are<br>no new or<br>different<br>intended uses as<br>compared to the<br>predicates. The<br>subject and<br>predicate<br>instruments are<br>employed in the<br>same manner<br>when used for<br>tissue dilation<br>and as tools to<br>assist the<br>surgeon in<br>locating spinal<br>nerves before,<br>during, or after<br>surgery for<br>open and |
{6}------------------------------------------------
| | on the proximal end<br>to an uninsulated<br>portion of the distal<br>tip. The purpose of<br>this is to allow<br>controlled monitoring<br>of neural elements<br>near and around the<br>point of access. The<br>dilators are offered<br>sterile/single use. | | | | percutaneous<br>surgical<br>approaches. |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Sterilization | Instruments<br>provided as single-<br>use sterile packed<br>devices. | Neuromonitoring<br>Clip provided as<br>single-use sterile<br>packed devices.<br>Accessory<br>instruments provided<br>as reusable non-<br>sterile devices with<br>validated sterilization<br>parameters to assure<br>a SAL of 10-6 | Instruments<br>provided as<br>single-use sterile<br>packed devices. | Instruments provided<br>as reusable non-<br>sterile devices with<br>validated<br>sterilization<br>parameters to assure<br>a SAL of 10-6 | Yes |
| Surgical<br>Approach | Open or<br>Percutaneous/<br>Minimally Invasive | None specified | None specified | Open or<br>Percutaneous/<br>Minimally Invasive | Yes |
| Electromagnetic<br>Compatibility &<br>Electrical Safety | IEC 60601-1 | IEC 60601-1 | ASTM D149:<br>2009 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-3-2<br>IEC 60601-3-3 | Yes - Subject<br>system<br>compliant with<br>applicable<br>sections of IEC<br>60601-1. |
| Min. exposed sur-<br>face area during<br>tissue stimulation | 0.02 in2 | 0.02 in2 | .167 cm2 (.026<br>in2) | 0.53 mm2 (.0008 in2) | Yes |
| Maximum<br>Charge Density<br>( $\mu$ C/cm2) | 0.108 | N/A | 0.097 | 0.306 | Yes |
| Maximum<br>Current Density<br>(mA/cm2) | 1.080 | N/A | 0.970 | 3.057 | Yes |
| Maximum Power<br>Density (500 Ω<br>load) (mW/cm2) | 0.087 | N/A | 0.079 | 0.248 | Yes |
| Maximum Power<br>Density (10kΩ<br>load) (mW/cm2) | 1.750 | N/A | 1.571 | 4.952 | Yes |
| Conclusion: | Based on the information included in this premarket notification, the subject system is substantially equivalent to the predicates. The subject dilators are employed in the same manner, have similar intended uses, principles of operation, technological characteristics, and performance. There are no new issues of safety or efficacy. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.