K182430 · Kuraray Noritake Dental, Inc. · EBF · May 3, 2019 · Dental
Device Facts
Record ID
K182430
Device Name
CLEARFIL MAJESTY IC
Applicant
Kuraray Noritake Dental, Inc.
Product Code
EBF · Dental
Decision Date
May 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
Direct restorations for all cavity classes in anterior and posterior teeth Direct veneers Correction of tooth position and tooth shape (e.g. diastema closure, dwarfed tooth, etc.) Intraoral repairs of fractured restorations
Device Story
CLEARFIL MAJESTY IC is a light-cured, radiopaque dental restorative composite. Delivered via syringe and replaceable tips for direct intraoral injection. Used by dentists for anterior/posterior restorations, veneers, tooth shape/position correction, and repair of fractured restorations. Material properties comply with ISO 4049 (Type 1, Class 2, Group 1). Provides color matching and polishability. Clinical benefit includes restoration of tooth structure and aesthetics.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via comparative bench testing against predicate/reference devices for physical/mechanical properties (ISO 4049:2009) and biocompatibility (ISO 10993-1:2009).
Technological Characteristics
Light-cured, polymer-based restorative composite. Radiopacity equivalent to 1mm aluminum. Classified as ISO 4049 Type 1, Class 2, Group 1. Materials equivalent to predicate devices. Shelf life 48 months at 2-25°C. No releasable agents like fluoride or nitrate ions.
Indications for Use
Indicated for patients requiring direct dental restorations for all cavity classes in anterior and posterior teeth, direct veneers, correction of tooth position and shape, and intraoral repair of fractured restorations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2019
Kuraray Noritake Dental Inc. Yasujiro Ohara Manager Ote Center Bldg. 7F, 1-1-3, Otemachi Chiyoda-ku, Tokyo 100-0004 JAPAN
Re: K182430
Trade/Device Name: CLEARFIL MAJESTY IC Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: January 31, 2019 Received: February 4, 2019
Dear Yasujiro Ohara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K182430
Device Name: CLEARFIL MAJESTY IC
Indications for Use:
[1] Direct restorations for all cavity classes in anterior and posterior teeth
[2] Direct veneers
[3] Correction of tooth position and tooth shape (e.g. diastema closure, dwarfed tooth, etc.)
[4] Intraoral repairs of fractured restorations
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
### 510(k) Summary
Kuraray Noritake Dental Inc.
CLEARFIL MAJESTY IC
Dental light-cured restorative composite
21 CFR 872.3690
EBF
#### 3-1. 510(k) owner (submitter)
- 1) Name
- 2) Address
- 3) Contact person
4) Contact person in US
1621 Sakazu, Kurashiki, Okayama 710-0801, Japan Yasujiro Ohara Quality Assurance Department Manabu Suzuki KURARAY AMERICA, INC. 33 Maiden Lane, 6th Floor, New York, NY 10038 Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676 Fax: (212)-867-3543
#### 3-2. Name of Device
- 1) Trade / Proprietary Name:
- 2) Regulation Number:
- 3) Product Code:
- 4) Common Name
## 3-3. Primary Predicate Device
| CLEARFIL MAJESTY ES-2 | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR:<br>Applicant: | K121583<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. |
|----------------------------------|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| CLEARFIL MAJESTY ES Flow | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR:<br>Applicant: | K130371<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. |
| ESTENIA C&B | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR:<br>Applicant: | K042929<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. |
| Filtek One Bulk Fill Restorative | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR:<br>Applicant: | K163207<br>Tooth shade resin material<br>EBF<br>872.3690<br>3M Company |
| CLEARFIL MAJESTY Posterior PLT | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR:<br>Applicant: | K092281<br>Tooth shade resin material<br>EBF<br>872.3690 |
{4}------------------------------------------------
| | Applicant: | Kuraray Noritake Dental Inc. |
|----------|-----------------|------------------------------|
| EPRICORD | 510(k) Number: | K033267 |
| | Classification: | Tooth shade resin material |
| | Product Code: | EBF |
| | 21 CFR: | 872.3690 |
| | Applicant: | Kuraray Noritake Dental Inc. |
#### 3-4. Device Description
CLEARFIL MAJESTY IC is an intra-oral, light-cured restorative composite that is available direct injection method into preparations with its syringe and replaceable tips. It is the radiopaque equivalent of 1mm aluminium and classified as a Type 1 and Class 2 (Group 1) material by ISO 4049. It is indicated for both anterior and posterior restorations and provides accurate color matching, high polishability and excellent physical properties.
#### 3-5. Indications for Use Statement
The subject device is indicated for the following uses:
- [1] Direct restorations for all cavity classes in anterior and posterior teeth
- [2] Direct veneers
- [3] Correction of tooth position and tooth shape (e.g. diastema closure, dwarfed tooth, etc.)
- [4] Intraoral repairs of fractured restorations
#### 3-6. Substantial Equivalence Discussion
#### 1) Indications for Use
The Indications for Use of the subject device is identical to the primary predicate device. Therefore, the Indications for Use of the subject device is substantially equivalent to that of the primary predicate device.
The indications of the subject device and the primary predicate device are as listed on the following table.
| | Subject device | Primary predicate device |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | CLEARFIL MAJESTY IC | CLEARFIL MAJESTY ES-2 |
| Indications<br>for Use | [1] Direct restorations for all cavity classes in<br>anterior and posterior teeth<br>[2] Direct veneers<br>[3] Correction of tooth position and tooth shape<br>(e.g. diastema closure, dwarfed tooth, etc.)<br>[4] Intraoral repairs of fractured restorations | [1] Direct restorations for all cavity classes in<br>anterior and posterior teeth<br>[2] Direct veneers<br>[3] Correction of tooth position and tooth shape<br>(e.g. diastema closure, dwarfed tooth, etc.)<br>[4] Intraoral repairs of fractured restorations |
#### Table 3-6. Indications for Use of the subject device and the primary predicate device
#### 2) Chemical ingredients
All ingredients in the subject device and its accessories have been used in the primary predicate device and the reference devices.
Regarding the primary predicate device and the reference devices, there have not been any reported problems or recalls according to the post-market adverse event reporting requirements in the US.
The chemical ingredients of the subject device and its accessories are equivalent to those of the primary predicate device and the reference devices.
{5}------------------------------------------------
3) Technological characteristics/ Non-Clinical Testing
- < Technological characteristics >
Physical and mechanical properties of the subject device were evaluated according to ISO 4049: 2009 (Dentistry - Polymer-based restorative and materials). According to ISO 4049: 2009, the subject device is classified into the following:
- Type 1: Polymer-based restorative materials;
- Class 2: materials whose setting is effected by light;
- Group 1: materials whose use requires the energy to be applied intra-orally.
In addition, we examined the subject device concerning the Physical Properties described in the "Dental Composite Resin Devices – Premarket Notification [510(k)] Submissions – Guidance for Industry and FDA Staff".
Therefore, we considered sensitivity to ambient light, Depth of cure, flexural strength, water sorption and solubility, shade, filling and restorative materials, color stability after irradiation and water sorption, radio-opacity, compressive strength, elastic modulus, intensity for curing, wavelength for curing, filler particle size distribution, surface hardness and curing time as physical properties when testing the subject device. The subject device doesn't contain chemical ingredients which are corresponded releasable agents "such as fluoride or nitrate ions". The compared test items and the evaluation results are described as follows.
{6}------------------------------------------------
| Test item | For comparison:<br>Primary Predicate device<br>or<br>Reference device | Result |
|------------------------------------------------------------|-----------------------------------------------------------------------|--------------------------------------------------|
| Sensitivity to ambient light | Primary Predicate device | The result demonstrated substantial equivalence. |
| Depth of cure | Primary Predicate device | The result demonstrated substantial equivalence. |
| Flexural strength | Primary Predicate device | The result demonstrated substantial equivalence. |
| Water sorption and solubility | Primary Predicate device | The result demonstrated substantial equivalence. |
| Shade, filling and restorative materials | Primary Predicate device | The result demonstrated substantial equivalence. |
| Color stability after<br>irradiation and water<br>sorption | Primary Predicate device | The result demonstrated substantial equivalence. |
| Radio-opacity | Primary Predicate device | The result demonstrated substantial equivalence. |
| Compressive strength | Primary Predicate device | The result demonstrated substantial equivalence. |
| Elastic modulus | Primary Predicate device<br>Reference device | The result demonstrated substantial equivalence. |
| Intensity for curing | Primary Predicate device | The result demonstrated substantial equivalence. |
| Wavelength for curing | Primary Predicate device | The result demonstrated substantial equivalence. |
| Filler particle size<br>distribution | Primary Predicate device | The result demonstrated substantial equivalence. |
| Surface hardness | Primary Predicate device | The result demonstrated substantial equivalence. |
| Curing time | Primary Predicate device | The result demonstrated substantial equivalence. |
So we conducted the comparative tests for above test items. And all of tests results demonstrated that each characteristic of the subject device is substantially equivalent to that of the primary predicate device or the reference devices.
{7}------------------------------------------------
< Non-Clinical Testing >
The subject device is categorized into the external communicating device (dentin) and permanent contact device.
All the chemical ingredients of the subject device are equivalent to those of the primary predicate device and the reference devices.
Contacting information about the accessories of the subject device is as follows;
| Item | Human contacting information | The reference device |
|--------------------|------------------------------|----------------------------------------------------------|
| Syringe | Indirect | CLEARFIL MAJESTY ES Flow |
| Dispensing tip | Direct | 510(k) Number; K130371 |
| Dispensing tip cap | Indirect | CLEARFIL MAJESTY Posterior PLT<br>510(k) Number; K092281 |
All the chemical ingredients of accessories about the subject device are equivalent to those of the reference device.
And biological evaluation was conducted according to ISO10993-1:2009.
Concerning the track records of the ingredients, Indications for Use, manufacturing process, contact part and contact duration, all of these are substantially equivalent to those of the primary predicate device and the reference devices.
Therefore, the subject device and its accessories are substantially equivalent in biocompatibility to the primary predicate device and the reference devices.
Concerning Shelf Life, the subject device was evaluated by using temperature accelerated samples. And the properties of them which are equivalent to storing at 25°C/ 77 °F more than 48 months were complied with ISO 4049: 2009. So the Shelf Life of the subject device was determined to be 48 months at 2-25 °C/ 36-77 °F.
#### 3-7. Conclusion
The comparison for Indications for Use, chemical ingredients and performance data shows that the subject device is substantially equivalent to the primary predicate device and the reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.