K163207 · 3M Company · EBF · Nov 28, 2016 · Dental
Device Facts
Record ID
K163207
Device Name
Filtek One Bulk Fill Restorative
Applicant
3M Company
Product Code
EBF · Dental
Decision Date
Nov 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
Direct anterior and posterior restorations (including occlusal surfaces) · Base/liner under direct restorations · Core build-ups · Splinting · Indirect restorations including inlays, onlays and veneers · Restorations of deciduous teeth · Extended fissure sealing in molars and premolars · Repair of defects in porcelain restorations, enamel, and temporaries
Device Story
3M Filtek One Bulk Fill Restorative is a visible-light activated, radiopaque composite resin used by dental professionals in clinical settings for restorative procedures. The material consists of a stress-relieving resin system (AUDMA, AFM, diurethane-DMA, 1,12-dodecane-DMA) and inorganic fillers (silica, zirconia, ytterbium trifluoride). It is applied to tooth structure using a methacrylate-based dental adhesive. Upon irradiation with a dental curing light, the resin undergoes polymerization to form a hard, durable composite. The device is designed for bulk placement up to 5mm deep. By improving esthetic properties over previous bulk fill materials, it allows for broader application in both anterior and posterior teeth. Clinical benefits include reduced procedural time, high strength, low wear, and improved esthetics for the patient.
Clinical Evidence
Bench testing only. Performance evaluated via physical property testing including compressive strength, diametral tensile strength, flexural strength, flexural modulus, surface hardness, radiopacity, water sorption, water solubility, volumetric shrinkage, wear, depth of cure, cusp deflection, and polish retention. Biocompatibility assessed per ISO 10993 standards.
Technological Characteristics
Visible-light activated composite resin. Fillers: 20nm silica, 4-11nm zirconia, aggregated zirconia/silica clusters, and ytterbium trifluoride (76.5% wt/58.5% vol). Resin matrix: AUDMA, AFM, diurethane-DMA, 1,12-dodecane-DMA. Complies with ISO 4049:2009 (depth of cure) and ISO 10993 series (biocompatibility). Dispensed via single-use capsules or reusable syringes.
Indications for Use
Indicated for direct anterior and posterior restorations, base/liner placement, core build-ups, splinting, indirect restorations (inlays, onlays, veneers), deciduous teeth restorations, extended fissure sealing, and repair of porcelain, enamel, or temporary defects.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human figures connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 28,2016
3M ESPE Dental Products c/o Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K163207
Trade/Device Name: 3M™ Filtek™ One Bulk Fill Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: November 13, 2016 Received: November 15, 2016
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runno DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used to represent the 3M company.
#### Indications for Use Statement ব
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
| 510(k) Number (if known) | |
|--------------------------|--|
|--------------------------|--|
Device Name
3MTM FiltekTM One Bulk Fill Restorative
Indications for Use (Describe)
- · Direct anterior and posterior restorations (including occlusal surfaces)
- · Base/liner under direct restorations
- · Core build-ups
- · Splinting
- · Indirect restorations including inlays, onlays and veneers
- · Restorations of deciduous teeth
- · Extended fissure sealing in molars and premolars
- · Repair of defects in porcelain restorations, enamel, and temporaries
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| | FOR FDA USE ONLY | | | |
|--|------------------|--|--|--|
|--|------------------|--|--|--|
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# Appendix A Revised Section 3 - 510(k) Summary
#### 510(k) Summary 3.
ЗМ ESPE Dental Products 2510 Conway Avenue St. Paul, MN 55144-1000
Image /page/3/Picture/4 description: The image shows the 3M logo. The logo is red and consists of the number "3" followed by the letter "M". The number "3" is larger than the letter "M". The logo is simple and recognizable.
# 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Submitter................................ | 3M ESPE Dental Products<br>2510 Conway Ave.<br>St. Paul, MN 55144 USA |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact person................................ | Scott Erickson, RAC<br>Senior Regulatory Affairs Specialist<br>Phone: (651) 736-9883<br>Fax: (651) 736-1599<br>sterickson@mmm.com |
| Date Summary was Prepared...... | 19Oct2016 |
| Trade Name................................ | 3MTM FiltekTM One Bulk Fill<br>Restorative |
| Common Name(s)............................ | Tooth shade resin material<br>Restorative |
| Recommended Classification...... | 21 CFR 872.3690<br>Tooth shade resin material<br>Product Code: EBF |
Predicate Device: Filtek™ Bulk Fill Posterior Restorative (K141081)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used on 3M products and marketing materials.
## Description of Device:
3MTM FiltekTM One Bulk Fill Restorative, is a visible-light activated, restorative composite optimized to create fast and easy restorations. This material provides excellent strength and low wear for durability and improved esthetics. The material can be placed and cured up to 5mm deep, enabled by a stress relieving resin system and optimized optical properties. Filtek™ One Bulk Fill Restorative serves to enhance the 3M ESPE lineup of restorative materials by improving the esthetic properties of a bulk fill material to allow for broader use in both posterior and anterior restorations. Filtek™ One Bulk Fill Restorative is offered in a range of tooth-colored shades. All shades are radiopaque.
### Indications for Use:
- · Direct anterior and posterior restorations (including occlusal surfaces)
- · Base/liner under direct restorations
- Core build-ups
- · Splinting
- · Indirect restorations including inlays, onlays and veneers
- · Restorations of deciduous teeth
- · Extended fissure sealing in molars and premolars
- · Repair of defects in porcelain restorations, enamel, and temporaries
### Technological Characteristics:
Filtek™ One Bulk Fill Restorative is a modification of predicate device, Filtek™ Bulk Fill Posterior Restorative, also manufactured by 3M ESPE Dental Products. The formulation was modified to improve the esthetics of the cured restorative material.
The fillers are a combination of a non-agglomerated/non-aggregated 20 nm silica filler, a non-agglomerated/non-aggregated 4 to 11 nm zirconia filler, an aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4 to 11 nm zirconia particles), and a ytterbium trifluoride filler consisting of agglomerate 100 mm particles. The inorganic filler loading is about 76.5% by weight (58.5% by volume). FiltekTM One Bulk Fill Restorative contains AUDMA. AFM. diurethane-DMA. and 1. 12dodecane-DMA. Filtek™ One Bulk Fill Restorative is applied to the tooth following use of a methacrylate-based dental adhesive, such as manufactured by 3M ESPE, which permanently bonds the restoration to the tooth structure.
When irradiated by light. the methacrylate functionalities of the resins and fillers undergo, in conjunction with the photoinitiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the 3M logo. The logo is red and consists of the number "3" followed by the letter "M". The logo is simple and recognizable.
# Substantial Equivalence:
| Technological property | Filtek™ One<br>Bulk Fill<br>Restorative | Filtek™ Bulk<br>Fill Posterior<br>Restorative<br>(K141081) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Photoinitiator system | X | X |
| Methacrylate-based resin matrix | X | X |
| Compatible with methacrylate-based dental adhesives | X | X |
| Inorganic fillers | X | X |
| Oxide fillers are silane treated so that they bond to the<br>resin matrix when the restorative is cured | X | X |
| Single-surface light-cure (up to 3 mm depth of cure) | X1 | - |
| Single-surface light-cure (up to 4 mm depth of cure) | X | X |
| Multi-surface light-cure (5 mm depth of cure, Class II) | X2 | X2 |
| Multi-surface light-cure (5 mm depth of cure, Core<br>Build-Up) | X3 | - |
| When irradiated by light, the methacrylate<br>functionalities of the resins and fillers undergo, in<br>conjunction with the photoinitiator system, a light-<br>induced polymerization to form a hard composite that<br>is bonded to the tooth structure with a permanent<br>dental adhesive. | X | X |
| Dispensing system:<br>single-use capsule (intraoral)4<br>reusable syringe (extraoral)5 | X | X |
| Recommended for load-bearing occlusal surfaces | X | X |
| FDA-recognized standards followed | Risk Management:<br>ISO 14971:2007<br><br>Biocomp stds:6<br>ISO 10993-1:2009<br>ISO 10993-3:2003<br>ISO 10993-3:2014<br>ISO 10993-5:2009<br>ISO 10993-6:2007<br>ISO 10993-10:2010<br>ISO 10993-11:2006<br>ISO 10993-12:2012<br>ISO 7405:2008/<br>Amd1 2013<br><br>Product stds:7<br>ISO 4049:2009 | Risk Management:<br>ISO 14971:2007<br><br>Biocomp stds:6<br>ISO 10993-1:2009<br>ISO 10993-3:2003<br>ISO 10993-5:2009<br>ISO 10993-6:2007<br>ISO 10993-10:2010<br>ISO 10993-11:2006<br>ISO 10993-12:2012<br>ISO 7405:2008<br><br>Product stds: 7<br>ISO 4049:2009 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the 3M logo. The logo is red and features the number 3 followed by the letter M. The logo is simple and recognizable.
- 1. Curing protocol for < 3mm depth, with reduced light-cure time versus 4mm depth, is added for Filtek™ One Bulk Fill Restorative. This difference is not significant since test data shows both Filtek™ One Bulk Fill Restorative and the predicate device pass ISO 4049 3mm depth of cure requirement when using the Filtek™ One Bulk Fill Restorative light-cure recommendations.
- 2. In order to obtain 5 mm depth of cure for Class II restorations, product is lightcured from the occlusal surface and, after the matrix band is removed, light-cured from the buccal and lingual surfaces through tooth structure.
- 3. In order to obtain 5 mm depth of cure for Core Build-Up, product is light-cured from the occlusal. buccal and lingual surfaces. Core Build-Up is less challenging to light-cure than a Class II restoration (2 above), because the curing light is placed in closer proximity to the restorative material for a Core Build-Up. The multi-surface light-cure protocol provides satisfactory light-cure for Filtek™ One Bulk Fill Restorative 5mm Core Build-up and would also be expected to adequately cure predicate device 5mm Core Build-up.
- 4. Restorative material is dispensed from a single-use capsule in the mouth.
- 5. Restorative material is dispensed from a reusable syringe outside the mouth (e.g., onto a pad).
- 6. Newer versions of two biocompatibility standards were applied to the evaluation of Filtek™ One Bulk Fill Restorative, due to time elapsed since the predicate device was evaluated. This difference is not significant because for both Filtek™ One Bulk Fill Restorative and the predicate device, Filtek™ Bulk Fill Posterior Restorative (K141081):
- a. A Diplomate of the American Board of Toxicology assessed the safety of the product.
- b. Standard risk assessment techniques and consideration of internationally recognized guidelines were used in the evaluation.
- c. The conclusion of the assessment is that the device is safe for its intended use.
- 7. Newer version of ISO 6874 was applied to the evaluation of Filtek™ One Bulk Fill Restorative, due to time elapsed since the predicate device was evaluated, however, this difference is not significant, since the relevant physical property requirement (Depth of Cure) did not change.
Test results for the following physical properties were included in this submission: Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus. Surface Hardness. Radiopacity. Water Sorption. Water Solubility. Volumetric Shrinkage, Wear, Depth of Cure, Cusp Deflection and Polish Retention.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" and the letter "M" in a bold, sans-serif font. The logo is simple and recognizable, and it is often used to represent the 3M company.
# Conclusion:
Filtek™ One Bulk Fill Restorative is substantially equivalent to predicate device Filtek™ Bulk Fill Posterior Restorative in terms of intended use, indications for use, formulation, physical properties and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.