SGS® Dental Implants System is intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. SGS® Dental Implants System may be immediate loading when good primary stability is achieved and with appropriate occlusal loading. Narrow diameter implants (3.2) are intended for maxillary lateral incisors and mandibular lateral and central incisors. Two Stage Implants: P7N, P1D, P5D, P7D. PEEK Temporary Abutments are not to exceed 30 days.
Device Story
SGS Dental Implants System consists of endosseous root-form dental implants and various abutments; used for supporting prosthetic restorations (crowns, bridges, overdentures). Implants are surgically placed in maxillary or mandibular bone by dental professionals. System includes internal hex and internal conical connection designs; various diameters and lengths available. Abutments include healing, straight, anatomic, angular, ball attachment, and PEEK temporary types. Device functions by osseointegration of the implant into the jawbone, providing a stable base for prosthetic attachment. Clinical benefit includes restoration of chewing function and aesthetics for edentulous or partially edentulous patients. System is provided sterile; some components are CNC machined from titanium alloy (Ti 6Al-4V ELI).
Clinical Evidence
No clinical data included. Evidence consists of bench testing: fatigue testing per ISO 14801, sterilization validation (gamma and steam) per ISO 11137 and ISO 17665, shelf-life validation per ASTM F-1980, and biocompatibility assessment leveraged from the primary predicate.
Technological Characteristics
Materials: Ti 6Al-4V ELI (ASTM F-136/ISO 5832-3) and PEEK-CLASSIX. Design: Endosseous root-form, internal hex or internal conical connections, tapered or straight thread designs. Surface treatment: Calcium Phosphate. Sterilization: Gamma radiation or moist heat (steam).
Indications for Use
Indicated for surgical placement in maxillary and/or mandibular arches to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. Narrow diameter implants (3.2mm) indicated for maxillary lateral incisors and mandibular lateral and central incisors. May be used for immediate loading given good primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 29, 2019
SGS International Ltd % Daniela Levy Regulatory Consultant Sterling Medical Registration 22815 Ventura Blvd Woodland Hills, California 91364
Re: K182219
Trade/Device Name: SGS® Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 10, 2019 Received: July 31, 2019
Dear Daniela Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for SGS Dental. The logo features an orange square with rounded corners on the left side, containing a stylized white graphic that resembles a series of curved lines. To the right of the square, the letters "SGS" are written in a bold, black sans-serif font. Below "SGS", the word "Dental" is written in a smaller, orange sans-serif font.
#### Indication for Use Statement
510(k) Number (if known):
#### K182219
Device Name:
SGS® Dental Implants System
Indications for Use (Describe)
#### Indications for Use:
SGS® Dental Implants System is intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. SGS® Dental Implants System may be immediate loading when good primary stability is achieved and with appropriate occlusal loading. Narrow diameter implants (3.2) are intended for maxillary lateral incisors and mandibular lateral and central incisors.
Two Stage Implants: P7N, P1D, P5D, P7D
PEEK Temporary Abutments are not to exceed 30 days.
Type of Use (Select one or both, as applicable)
区Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use(21 CFR 801 Subpart
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
T"D
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for SGS Dental. The logo features an orange square with rounded corners on the left side, containing a white, stylized image of a spiral or helix. To the right of the square, the text "SGS" is written in a bold, sans-serif font, with the word "Dental" appearing below it in a smaller, orange font.
# 510(k) Summary (21 CFR 807.92)
# 510(k) Number K182219
- Type of Submission Premarket Notification (21 CFR 807.90(e)) 510(k) Traditional 1
| 2 | Submission Owner | SGS International Ltd | | | |
|---|------------------------------------------------------|--------------------------------------|-----|--------------------------------|------------|
| | | Michaeli Shabtai – CEO | | | |
| | | Karolyi Istvan Street 1-3. | | | |
| | | Budapest, Hungary | | | |
| | | H 1047, Hungary | | | |
| | | Telephone Number +36-309611579 | | | |
| | | Facsimile (Fax) Number +36-309611579 | | | |
| 3 | Official Correspondent<br>Contact Person | Sterling Medical Registration | | | |
| | | Daniela Levy - Regulatory Consultant | | | |
| | | 22815 Ventura blvd. | | | |
| | | Woodland Hills, CA 91364 | | | |
| | | Phone: 1-213-787-3027 | | | |
| | | Email: sterlingmedical2017@gmail.com | | | |
| 4 | Prepared Date | August 28, 2019 | | | |
| 5 | Device Trade Name | SGS® Dental Implants System | | | |
| 6 | Regulation Description | Endosseous Dental Implants Abutment | | | |
| 7 | Classification | Product Code: | DZE | | |
| | | Device Name | : | Implant, endosseous, root-form | |
| | | Regulation No | : | 872.3640 | |
| | | Class | : | II | |
| | | Panel | : | Dental | |
| | | Secondary Product Code: | | NHA | |
| 8 | Reason for the Premarket Notification Submission | | | : | New Device |
| 9 | Identification of Legally Marketed Predicate Devices | | | : | |
- Primary Predicate:
SGS International Ltd K133362;
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for SGS Dental. The logo features an orange square with rounded corners on the left side, containing a white abstract design resembling a series of curved lines. To the right of the square, the text "SGS" is displayed in bold, black letters, with the word "Dental" underneath in a smaller, orange font. The overall design is clean and professional, suggesting a company in the dental industry.
eference Devices: UNIQA Dental Itd K180598; A.B.Dental Devices K112440; Osstem lmplant Co Ltd K182091;
in terms of intended use, indication for use, raw material, technological characteristics and performance. The primary predicate and referenced devices are Class II medical devices.
10 Device Description:
SGS® Dental Implants System is consist of endosseous form dental implants, internal hex implants, tapered design; cover screws, healing caps and abutment systems; Dental Implants:
P7N Narrow Conical Implant, Internal Hex available as follow:
Diameter 3.2 mm Length 8 mm
P1D Conical Groovy Double Connection Implant, Internal hex Conical Connection available as follow:
Diameter 3.5, 3.75, 4.2, 5, 6 mm Length 8, 10, 11.5, 13, 16.
P5D Conical Groovy Double Connection Implant, Internal hex Conical Connection available as follow:
Diameter 3.5, 3.75, 4.2, 4.5, 5, 6 mm Length 8, 10, 11.5, 13, 16.
P7D Conical Groovy Double Connection Implant, Internal hex Conical Connection available as follow:
Diameter 3.5, 3.75, 4.2, 4.5, 5, 6 mm Length 8, 10, 11.5, 13, 16.
Dental Abutments consist of:
Healing Abutments, Anatomic Straight Abutments, Straight Abutments, Straight Abutments Narrow/Wide, PEEK Temporary Abutments for bars, Angular Anatomic Abutments Ti, Angular Abutments Slim/Narrow, Multi-Base Angular Abutment, Anqular Ball Attachment, S-lock Straight/Anqular Abutment, Easy-Fix Angular Abutment, Straight/One-piece angular Multi-unit Abutment, Ball Attachments, Abutment for temporary restoration, Easy Fix Abutments Straight/Angular, Smart Lock abutments Straight/Angular, Flat connection abutments for casting, overdenture attachments. Titanium abutments are CNC machined with no anodized surface.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for SGS Dental. The logo consists of an orange square with rounded corners on the left side, containing a white abstract design of curved lines. To the right of the square are the letters "SGS" in black, with the word "Dental" in orange underneath. The logo is simple and modern, with a clean design.
11 Indication for Use:
> SGS® Dental Implants System is intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. SGS® Dental Implants System may be immediate loading when good primary stability is achieved and with appropriate occlusal loading. Narrow diameter implants (3.2) are intended for maxillary lateral incisors and mandibular lateral and central incisors.
Two Stage Implants: P7N, P1D, P5D, P7D.
PEEK Temporary Abutments are not to exceed 30 days.
- 12 Performance Standards or Special Controls
- ISO 7405 Second edition 2008-12-15 Dentistry Evaluation of biocompatibility of . medical devices used in dentistry.
- ISO 5832-3:1996 Implants for surgery -- Metallic materials -- Part 3: Wrought . titanium 6-aluminium 4-vanadium alloy.
- . ISO 14801 Second edition 2007-11-15 Dentistry-Implants-Dynamic fatigue test for endosseous dental implants.
- . ISO 11137-1:2006 Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- . ISO 11737-2: 2009 Sterilization of medical devices -- Microbiological methods --Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- . ASTM F-1980 - 16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ISO 17665-1:2006 Sterilization of health care products Moist heat Part 1: . Requirements for the development, validation and routine control of a sterilization process for medical devices
- . ISO 17665-2:2009 Sterilization of health care products -- Moist heat -- Part 2: Guidance on the application of ISO 17665-1
- ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in . vitro cytotoxicity
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for SGS Dental. The logo features an orange square with rounded corners on the left side, containing a series of white curved lines resembling a stylized globe or tooth shape. To the right of the graphic is the text "SGS" in bold black letters, with the word "Dental" underneath in a smaller, orange font.
- FDA guidance document: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff.
- Chemical and SEM analysis performed on implants per Class II Special Controls . Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments - Guidance for Industry and FDA Staff.
- Endotoxin batch testing protocol per FDA Guidance Document Submission and . Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for SGS Dental. The logo consists of an orange graphic on the left, which features a series of curved lines that resemble a spiral or a stylized representation of teeth. To the right of the graphic, the text "SGS" is displayed in a bold, sans-serif font, with the word "Dental" appearing below it in a smaller, slightly different font, also in orange.
# 13 Substantial Equivalence
# SE Table - Indication for Use:
| | Candidate | Primary Predicate | Reference Device |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SGS International | SGS International | UNIQA Dental Itd |
| K Number | K182219 | K133362 | K180598 |
| Indication for<br>use | SGS® Dental Implants<br>System is intended for<br>surgical placement in the<br>maxillary and/or mandibular<br>arch to support crowns,<br>bridges, or overdentures in<br>edentulous or partially<br>edentulous patients. SGS®<br>Dental Implants System may<br>be immediate loading when<br>good primary stability is<br>achieved and with<br>appropriate occlusal loading.<br>Narrow diameter implants<br>(3.2) are intended for<br>maxillary lateral incisors and<br>mandibular lateral and central<br>incisors.<br>Two Stage Implants: P7N,<br>P1D, P5D, P7D. PEEK<br>Temporary Abutments are<br>not to exceed 30 days. | SGS® Dental Implants System<br>is intended for surgical<br>placement in the maxillary<br>and/or mandibular arch to<br>support crowns, bridges, or<br>overdentures in edentulous or<br>partially edentulous patients.<br>SGS® Dental Implants System<br>may be immediate loading<br>when good primary stability is<br>achieved and with appropriate<br>occlusal loading.<br>Two Stage Implants: P1, P7,<br>P7N.<br>One Stage: P7S, P9S.<br>One Stage & One-Piece 3.0<br>mm diameter implants: P7S,<br>P9S are intended for<br>placement at the mandibular<br>central and lateral incisors and<br>maxillary and lateral incisors.<br>Indicated also for denture<br>stabilization using multiple<br>implants.<br>One stage & One-Piece 2.4<br>mm diameter implants for<br>temporary use or long term<br>use:<br>P9S permit immediate splint<br>stability and long term fixation<br>of new or existing crown,<br>bridge and prosthesis, and<br>protection of graft sites.<br>PEEK Temporary Abutments<br>are not to exceed 30 days. | UNIQA Dental ® Dental<br>Implants System is indicated<br>for use in surgical and<br>restorative applications for<br>placement in the bone of the<br>upper or lower jaw to provide<br>support for prosthetic devices,<br>such as artificial teeth, in order<br>to restore the patient's chewing<br>function. UNIQA Dental ®<br>Dental Implants System is<br>indicated also for immediate<br>loading when good primary<br>stability is achieved and with<br>appropriate occlusal loading.<br>UNIQA Conical Connection<br>Implants: #UCI & #UC7 are to<br>be used only with straight<br>abutments.<br>Conical Mini implants<br>(Ø3.3mm) are indicated for use<br>in surgical and restorative<br>applications for placement only<br>in the mandibular central,<br>lateral incisor and maxillary<br>lateral incisor regions of<br>partially edentulous jaws, to<br>provide support for prosthetic<br>devices such as artificial teeth,<br>in order to restore the patient'<br>chewing function. Mandibular<br>central and lateral incisors<br>must be splinted if using two or<br>more narrow implants adjacent<br>to one another. |
| | Reference Device | Reference Device | Reference Device |
| | Osstem Implant Co Ltd | A.B. Dental Devices | |
| K Number | K182091 | K112440 | |
| Indication for<br>use | Osstem Abutment System is<br>intended for use with a dental<br>implant to provide support for<br>prosthetic restorations such<br>as crowns, bridges, or<br>overdentures. | The AB Dental Devices<br>implants are intended for<br>surgical placement in the<br>maxillary and/or mandibular<br>arch to support crowns,<br>bridges, or overdentures in<br>edentulous or partially<br>edentulous patients. I7 Integral<br>implant, I5 Conical implant,<br>P15 Temporary abutment. | |
.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for SGS Dental. The logo consists of an orange square with rounded corners on the right side. Inside the square are white curved lines. To the right of the square is the text "SGS" in black, with the word "Dental" in orange below it.
| | P12-T,L Temporary flat connection abutment, and P16 adaptor are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading. |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a logo for SGS Dental. The logo consists of an orange square with rounded corners on the left side, and the text "SGS Dental" on the right side. The orange square contains a white design that looks like a stylized tooth or a series of curved lines. The text "SGS" is in black, and the text "Dental" is in orange.
## SE - Dental Implants
| Characteristic | Candidate<br>P7N Narrow Conical Implant | Primary Predicate<br>P7 Conical Groovy Implant |
|-------------------------------|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K133362 |
| Manufacturer | SGS International Ltd. | SGS International Ltd. |
| Product Name | P7N Narrow Conical Implant | P7 Conical Groovy Implant |
| Thread Design | Tapered, Double Thread | Tapered, Double Thread |
| Measurements<br>Ø / Length mm | 3.2Ø: 8 | 3.2Ø: 8, 10, 11.5<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.2Ø: 8, 10, 11.5, 13, 16<br>4.5Ø: 8, 10, 11.5, 13, 16<br>5.0Ø: 8, 10, 11.5, 13, 16<br>6.0Ø: 8, 10, 11.5, 13, 16 |
| Connection Type | Internal Hex | Internal Hex |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI |
| Surface<br>Treatment | Calcium Phosphate | Calcium Phosphate |
| Self tapping | Yes | Yes |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
| Characteristic | Candidate<br>P1D Conical Groovy Double<br>Connection Implant | Primary Predicate<br>P1 Screw Type Groovy Implant |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K133362 |
| Manufacturer | SGS International Ltd. | SGS International Ltd. |
| Product Name | P1D Conical Groovy Double<br>Connection Implant | P1 Screw Type Groovy Implant |
| Thread Design | Straight | Tapered, Double Thread |
| Measurements<br>Ø / Length mm | 3.5Ø: 8, 10, 11.5, 13, 16<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.2Ø: 8, 10, 11.5, 13, 16<br>5Ø: 8, 10, 11.5, 13, 16<br>6Ø: 8, 10, 11.5, 13, 16 | 3.2Ø: 8, 10, 11.5<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.2Ø: 8, 10, 11.5, 13, 16<br>4.5Ø: 8, 10, 11.5, 13, 16<br>5Ø: 8, 10, 11.5, 13, 16<br>6Ø: 8, 10, 11.5, 13, 16 |
| Connection Type | Internal conical | Internal Hex |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI |
| Surface<br>Treatment | Calcium Phosphate | Calcium Phosphate |
| Self tapping | Yes | Yes |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for SGS Dental. The logo features a rounded square shape with an orange background and a series of white, curved lines resembling a spine or vertebrae. To the right of the graphic is the text "SGS" in bold, black letters, with the word "Dental" underneath in a smaller, orange font.
| Characteristic | Candidate<br>P5D Conical Groovy Double<br>Connection Implant | Primary Predicate<br>P7 Conical Groovy Implant |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K133362 |
| Manufacturer | SGS International Ltd. | SGS International Ltd. |
| Product Name | P5D Conical Groovy Double Connection<br>Implant | P7 Conical Groovy Implant |
| Thread Design | Tapered, Double Thread | Tapered, Double Thread |
| Measurements<br>Ø / Length mm | 3.50Ø: 8, 10, 11.5, 13, 16<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.20Ø: 8, 10, 11.5, 13, 16<br>4.50Ø: 8, 10, 11.5, 13, 16<br>5.00Ø: 8, 10, 11.5, 13, 16<br>6.00Ø: 8, 10, 11.5, 13, 16 | 3.20Ø: 8, 10, 11.5<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.20Ø: 8, 10, 11.5, 13, 16<br>4.50Ø: 8, 10, 11.5, 13, 16<br>5.00Ø: 8, 10, 11.5, 13, 16<br>6.00Ø: 8, 10, 11.5, 13, 16 |
| Connection Type | Internal conical | Internal Hex |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI |
| Surface<br>Treatment | Calcium Phosphate | Calcium Phosphate |
| Self tapping | Yes | Yes |
| Sterilization | Gamma Ray | Gamma Ray |
| Packaging | Double packaging | Double packaging |
| Characteristic | Candidate<br>P7D - Dental Implant with Double<br>Connection | Primary Predicate<br>P7 Conical Groovy Implant | |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | TBD | K133362 | |
| Manufacturer | SGS International Ltd. | SGS International Ltd. | |
| Product Name | P7D - Dental Implant with Double<br>Connection | P7 Conical Groovy Implant | |
| Thread Design | Tapered, Double Thread | Tapered, Double Thread | |
| Measurements<br>Ø / Length mm | 3.50 Ø: 8, 10, 11.5, 13, 16<br>3.75Ø: 8, 10, 11.5, 13, 16<br>4.20 Ø: 8, 10, 11.5, 13, 16<br>4.5Ø: 8, 10, 11.5, 13, 16<br>5.00 Ø: 8, 10, 11.5, 13, 16<br>6.00 Ø: 8, 10, 11.5, 13, 16 | 3.20 Ø: 8, 10, 11.5<br>3.75Ø:8, 10, 11.5, 13, 16<br>4.20 Ø: 8, 10, 11.5, 13, 16<br>4.5Ø: 8, 10, 11.5, 13, 16<br>5.00 Ø: 8, 10, 11.5, 13, 16<br>6.00 Ø: 8, 10, 11.5, 13, 16 | |
| Connection Type | Internal conical | Internal Hex | |
| Material | Ti 6Al-4V ELI | Ti 6Al-4V ELI | |
| Surface<br>Treatment | Calcium Phosphate | Calcium Phosphate | |
| Self tapping | Yes | Yes | |
| Sterilization | Gamma Ray | Gamma Ray | |
| Packaging | Double packaging | Double packaging | |
| | Candidate | Primary Predicate | Reference Device |
| | SGS International | SGS International | UNIQA Dental Itd |
| K Number | | K133362 | K180598 |
| Product Name | Healing Abutments<br>(HD,HND,HW7D,HW8D,H<br>WD) | Healing Abutments<br>(HN,HNN,HWN,H,HN, HW) | Healing Cap # UOHX |
| Dimensions | Narrow, Length: wide 2,3,4;<br>Normal, Length: 2,3,4,5,6<br>Wide, Length: 3,5 | Narrow, Length: 2,3,4<br>Normal, Length: 2,3,4,5,6<br>Wide, Length: 3,5 | Diameter: Mini (#UOHM) -<br>Diameter 4.0, 4.5 Heights 3,<br>4, 5,7<br>Regular (#UOHR) -<br>Diameter 4.0, 4.5, 5.0, 6.0,<br>7.0 Heights 3, 4, 5, 7 |
| Material | Titanium alloy Ti-6Al-4V-<br>ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal conical | Internal Hex | Internal conical |
| | | | |
| | Candidate | Primary Predicate | Reference Device |
| | SGS International | SGS International | UNIQA Dental Itd |
| K Number | | K133362 | K180598 |
| Product Name | Anatomic Straight Titanium<br>Abutment S1AD | Anatomic Straight Titanium<br>Abutment S1A, S1AN | Transfer Abutment #UOTX |
| Dimensions | Normal Profile: Height 1, 2,<br>3 mm | Narrow: Straight Anatomic<br>Abutment-Profile Height 1<br>mm<br>Normal: Profile Height 1, 2, 3<br>mm | Mini Diameter: 4.5 mm,<br>Height 1, 2, 3, 4, 5 mm<br>Lengths: 5.5, 7 mm (#UOTM)<br>Regular Diameter: 4.5, 5.0,<br>6.0 mm, Height 1, 2, 3, 4, 5<br>mm, Lengths: 4.0(not for 4.5),<br>5.5, 7 mm(#UOTR)<br>Provided with: Screw Mini<br>Conical (#UOSM-0001);<br>Screw Regular Conical<br>(#UOSR-0001) |
| Material | Titanium alloy Ti-6Al-4V-<br>ELI. | Titanium alloy Ti-6Al-4V-ELI. | Titanium alloy Ti-6Al-4V-ELI. |
| Connection | Internal Conical | Internal Hex | Internal Conical |
| | | | |
| | Candidate | Primary Predicate | Reference Device |
| | SGS International | SGS International | UNIQA Dental Itd |
| K Number | | K133362 | K180598 |
| Product Name | Straight Titanium Abutment<br>S1WN,S1D, S1ND, S1WD,<br>S1W, S1W8D, S1W8 | Straight Titanium Abutment<br>S1N,S1WN, S1,S1N,S1W | Transfer Abutment #UOTX |
| Dimensions | Narrow: Wide - Length:9<br>Normal: Length: 7, 9, 12, 15<br>Narrow: Length: 7, 9<br>Wide: Length:9, 12<br>S1W8D, S1W8 - 8mm wide<br>with 10mm length | Narrow: Narrow, Wide -<br>Length:9<br>Normal: Length: 5, 7, 9, 12, 15<br>Narrow: Length: 7, 9<br>Wide: Length:9 | Mini Diameter: 4.5 mm,<br>Height 1, 2, 3, 4, 5 mm<br>Lengths: 5.5, 7 mm (#UOTM)<br>Regular Diameter: 4.5, 5.0,<br>6.0 mm, Height 1, 2, 3, 4, 5<br>mm, Lengths: 4.0(not for 4.5), |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.