The AB Dental Devices implants are intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients. 17 Integral implant, 15 Conical implant, P15 Temporary abutment, P12-T,L Temporary flat connection abutment, and P16 Straight adaptor are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
AB Dental implants and accessories are endosseous dental devices used to restore esthetics and chewing function in edentulous or partially edentulous patients. The system includes various screw-type and conical implants, abutments, healing caps, and adaptors. Implants are surgically placed into the maxillary or mandibular bone. The device functions as a mechanical support structure for prosthetic restorations (crowns, bridges, overdentures). Some components support immediate loading protocols when primary stability is achieved. The devices are used in clinical dental settings by trained dental professionals. The system utilizes titanium alloy (Ti-6Al-4V-ELI) and features sandblasted and acid-etched surface treatments to facilitate osseointegration. The output is a stable foundation for dental prosthetics, which benefits the patient by restoring oral function and appearance.
Clinical Evidence
No clinical testing was performed. Safety and effectiveness were established through non-clinical bench testing, including dynamic fatigue testing per ISO 14801 and pull-out strength testing. Surface characteristics were verified via SEM analysis.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V-ELI (Grade 5). Surface: Sandblasted and acid-etched. Design: Endosseous dental implants, abutments, and accessories. Mechanical: Self-tapping threads. Sterilization: Not specified. Connectivity: None (mechanical device).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring surgical placement of dental implants in the maxillary or mandibular arch to support crowns, bridges, or overdentures. Specific components are indicated for immediate loading given sufficient primary stability and appropriate occlusal loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
K112440
# AB DENTAL SUMMARY
APR 1 8 2012
## Company
AB Dental Devices Ltd.
3 Habosem St, Ashdod
lsrael
Telephone: (+972) 8-853-1388
Fax: (+972) 8-8522262
FDA Establishment Registration # 3005663340
## Contact
Dr. Charles Hurwitz Regulatory Consultant MedicSense Ltd POB 367 Ramat Hasharon 47103 lsrael Phone: +972 3 9233666 Fax: +972 3 9231274 Email: charles@medicsense.com
## Classification Name
Per 21 CFR 872.3640
- Endosseous Dental Implants and Accessories Product Code . DZE
- Abutment, Implant, Dental Endosseous. Product Code NHA .
# Device Description
AB Dental implants are designed to support prosthetic devices in edentulous or partially edentulous patients to restore esthetics and chewing functions.
2
{1}------------------------------------------------
The purpose of this submission is to add new implants, abutments and other items and to receive approval for a new surface treatment to previously-approved implants.
## Indications for Use
The AB Dental Devices implants are intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
17 Integral implant, 15 Conical implant, P15 Temporary abutment, P12-T,L Temporary flat connection abutment, and P16 Straight adaptor are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
{2}------------------------------------------------
# Identification of the Legally Marketed Predicate Devices Used to Claim Substantial Equivalence
| AB Dental Product | Predicate and K number | Predicate and K number |
|------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------|
| 12 Screw Type | AB Dental 12, screw type,<br>K051719 | M.I.S Implant Technologies, Bio-<br>Com, K040807 |
| I5 Conical | AB Dental, I5 Conical, K051719 | M.I.S Implant Technologies,<br>Lance, K040807 |
| I7 Integral | AB Dental, I7 One-piece,<br>K051719 | M.I.S Implant Technologies, Uno<br>one piece screw type, K080162 |
| I10 Trapeze | Nobel Biocare, Nobel Active,<br>K102436 | ADIN, Touareg WP5mmD,<br>K081751 |
| P1-Non Hexed Abutment | AB Dental, P1-Non Hexed<br>Abutment, K051719 | Biohorizon, Titanium Temporary<br>Abutments - Non-hexed,<br>K103691 |
| P3 Abutment Anti-rotation | AB Dental, P3 Abutment Anti-<br>rotation, K051719 | Biohorizon, Internal Laser-Lok<br>Abutment, K103691 |
| P3N-Narrow Anti-rotation<br>Abutment | AB Dental, P3 Narrow Abutment<br>Anti-rotation, K051719 | MIS, MN-MAC10, Narrow<br>platform cementing post, NP,<br>K040807 |
| P3W - Wide anti-rotation<br>abutment | MIS, MD-WMAC1 Wide<br>cementing post, SP, K040807 | AlfaBio tec, TLAW, K063364 |
| P3W - Wide anti-rotation<br>abutment | MIS, MD-WMAC1 Wide<br>cementing post, SP, K040807 | AlfaBio tec, TLAW, K063364 |
| P9HG-3.75,11 Gold Composed<br>Hex Abutment | AlfaBio tec, TLABG, K063364 | MIS, MG-GPC10 Direct gold<br>plastic cylinder with hex. Internal<br>hex, K063364 |
| P12-3.75 - Flat Connection<br>Abutment | Adin, FC-6013,6018, K081751 | |
| P12-3.75-T - Temporary Flat<br>Connection Abutment | Adin, FC-6013,6018, K081751 | |
| P15-3x - Temporary Abutment<br>for Immediate Loading | Alpha-Bio Tec, Straight Titanium<br>Abutments ETLA, K063364 | |
| PO-P6H for Aesthetic Abutment<br>with Hex | Alpha Bio, HCT, K042654 | |
| PO-Titanium Healing Cap | Alpha Bio, HS, K042654 | Adin, WP-<br>0046,0047,0048,0049,0050,<br>K081751 |
| P16-Straight adaptor | MIS, Pilier Multi Unit, K040807 | |
The following predicate devices are cited:
. ﻓ
ﺮ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ
{3}------------------------------------------------
| | Current | Predicate | Predicate |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------|
| IMPLANTS | AB Dental | AB Dental | M.I.S Implant Technologies |
| Product<br>Name | I2 Screw Type | I2 Screw Type | Bio-Com |
| K Number | K051719 | K051719 | K040807 |
| Placement<br>Method | Dual-step<br>surgery. Later<br>exposure<br>required. | Dual-step surgery.<br>Later exposure<br>required. | Dual-step surgery. Later<br>exposure required. |
| Indication<br>for Use | For use in dense<br>bone | For use in dense<br>bone | For use in dense bone |
| Length(s) | 8,10,11.5,13,16,<br>18,20 | 8,10,11.5,13,16,18<br>,20 | 8,10,11.5,13,16 |
| Available<br>diameters<br>(mm) | 3.25,3.75,4.2,4.5<br>,5,6 | 3.25,3.75,4.2,4.5,5<br>,6 | 3.3,3.75,4.2,4.7 |
| Integrated<br>Abutment | No | No | No |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium alloy. Grade-4. |
| External<br>Surface | Sandblasted and<br>acid etched<br>surface . (HA) | Sandblasted and<br>acid etched<br>surface (SLA). | Sandblasted and acid etched<br>surface (SLA). |
| Self<br>Tapping | Yes | Yes | Yes |
| | AB Dental | AB Dental | M.I.S Implant Technologies |
| Product<br>Name | I5 Conical | I5 Conical | Lance |
| K Number | K051719 | K051719 | K040807 |
| Placement<br>Method | Dual-step<br>surgery. Later<br>exposure<br>required. | Dual-step surgery.<br>Later exposure<br>required. | Dual-step surgery. Later<br>exposure required. |
| Indication<br>for Use | Used in soft<br>bone and<br>designed to<br>enable the<br>change of<br>direction during<br>implantation. | Used in soft bone<br>and designed to<br>enable the change<br>of direction during<br>implantation. | Used in soft bone |
| Length | 8,10,11.5,13,16 | 8,10,11.5,13,16 | 10,11.5, 13,16 |
| Available | 3.2,3.75,4.2,4. | 3.2,3.75,4.2,4.5,5, | 3.75,4.2,5 |
| diameters<br>(mm) | 5. | 6 | |
| Integrated<br>Abutment | No | No | No |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium alloy. Grade-4. |
| External<br>Surface | Sandblasted and<br>acid etched<br>surface. (HA) | Sandblasted and<br>acid etched<br>surface (SLA). | Sandblasted and acid etched<br>surface (SLA). |
| Self<br>Tapping | Yes | Yes | Yes |
| | AB Dental | AB Dental | M.I.S Implant Technologies |
| Product<br>Name | 17 Integral | One-piece 17 | UNO - ONE PIECE SCREW-<br>TYPE |
| K Number | | K051719 | K080162 |
| Placement<br>Method | Single-step<br>surgery. No<br>exposure<br>required. | Single-step<br>surgery. No<br>exposure required. | Single-step surgery. No exposure<br>required. |
| Indication<br>for Use | For immediate<br>loading. | For immediate<br>loading. | For immediate loading. |
| Length | 10,11.5,13,16 | 10,11.5, 13,16 | 10,11.5, 13,16 |
| Available<br>diameters<br>(mm) | 10,11.5,13,16 | 10,11.5, 13,16 | 3,3.5 |
| Integrated<br>Abutment | Yes | Yes | Yes |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-6Al-4V-ELI.<br>Grade-5. |
| External<br>Surface | Sandblasted and<br>acid etched<br>surface. (HA) | Sandblasted and<br>acid etched<br>surface (SLA). | Sandblasted and acid etched<br>surface (SLA). |
| Self<br>Tapping | Yes | Yes | Yes |
| Pull out<br>strength<br>test<br>(Newtons) | 162.9N | | 135.8N |
| | AB Dental | AB Dental | |
الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع المو
୍ର ପ୍ରଶାସ
्र
{4}------------------------------------------------
다. 공장
| 波
ी करने के बाद में बाद में कि में बाद में कि में कि में कि में कि में कि में कि में कि में कि कहानी की कहानी की कहानी की कहानी की कहानी की कहानी की किसी है।
संस्कृत
मिली
{5}------------------------------------------------
| Product<br>Name | 110 Trapeze | Nobel Active | Touareg WP5mmD |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| K Number | | K102436 | K081751 |
| Placement<br>Method | Single-step<br>surgery. No<br>exposure<br>required. | Single-step<br>surgery. No<br>exposure required. | Single-step surgery. No exposure<br>required. |
| Indication<br>for Use | Is suitable for all<br>indications and<br>offers<br>exceptional<br>primary stability.<br>Designed to<br>enable the<br>change of<br>direction during<br>implantation. | Is suitable for all<br>indications and<br>offers exceptional<br>primary stability. | Is suitable for all indications. |
| Length | 8,10,11.5,13,16 | 8.5,10,11.5,13,16 | 8.5,10,11.5,13,15 |
| Available<br>diameters<br>(mm) | 3.75, 4.2, 5 | 4.3,5 | 5 |
| Integrated<br>Abutment | No | No | No |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium | Titanium alloy Ti-6Al-4V-ELI. |
| External<br>Surface | Sandblasted and<br>acid etched<br>surface.(HA) | Anodized surface:<br>partial crystalline<br>and phosphate<br>enriched titanium<br>Oxide surface. | OsseoFix, Calcium phosphate<br>treated surface. |
| Self<br>Tapping | Yes | Yes | Yes |
| ABUTMENTS | | | |
| | AB Dental | AB Dental | Biohorizon |
| Product<br>Name | P1-Non Hexed<br>Abutment | P1-Non Hexed<br>Abutment | Titanium Temporary Abutments<br>Non-hexed |
| K Number | | K051719 | K103691 |
| Indication<br>for Use | For connecting<br>two or more<br>implants by<br>using a<br>permanent rod.<br>For permanent | For connecting two<br>or more implants<br>by using a<br>permanent rod.<br>For permanent | Use for multiple-unit, screw-<br>retained, long term temporary<br>restorations (>30 days). |
| | and final<br>restoration. | restoration. | |
| Dimensions | Diameter:3.75<br>Length:9,13 | Diameter:3.75<br>Length:9,13 | Diameter:3.5,4.5,5.7 Length:12 |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium alloy |
| Angle | No | No | No |
| | AB Dental | AB Dental | Biohorizon |
| Product<br>Name | P3 Abutment<br>Anti-rotation | P3 Abutment Anti-<br>rotation | Internal Laser-Lok Abutment |
| K Number | | K051719 | K103691 |
| Indication<br>for Use | When wide<br>spaces are<br>required, for<br>back teeth.<br>When space is<br>limited, between<br>two teeth or<br>implants at the<br>anterior of the<br>mouth. For thick<br>gums when the<br>implant is very<br>deep. | When wide spaces<br>are required, for<br>back teeth. When<br>space is limited,<br>between two teeth<br>or implants at the<br>anterior of the<br>mouth. For thick<br>gums when the<br>implant is very<br>deep. | Designed to be placed at implant<br>surgery or uncover and remain in<br>place through final restoration. |
| Dimensions | Diameter3,3.75,<br>5<br>Length:5,7,9,12,<br>15 | Diameter:3.75<br>Length:5,7,9,12,15 | Diameter:3.5,4.5,5.7Length:4,5.5 |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium Alloy. |
| Angle | No | No | No |
| | AB Dental | AB Dental | Zimmer |
| Product<br>Name | P3S Anatomic<br>Anti-rotation | P3S Anatomic<br>Anti-rotation | Hex-Lock Contour Abutment |
| K Number | | K051719 | K052600 |
| Indication<br>for Use | To follow the<br>gum line. | To follow the gum<br>line. | For limited interocclusal<br>Space. |
| Dimensions | Diameter:3.75<br>Cuff<br>Height:1,2,3. | Diameter:3.75 Cuff<br>Height:2,3,4. | Diameter:3.5,4.5,5.7 Cuff<br>Height:1,2,3. |
| Material | Titanium alloy Ti-<br>Titanium Alloy. | | |
| Indication<br>for Use | To be used<br>when space is<br>limited, between<br>two teeth or<br>implants, usually<br>at the anterior of<br>the mouth. | To be used when<br>space is limited,<br>between two teeth<br>or implants,<br>usually at the<br>anterior of the<br>mouth. | To be used when space is<br>limited, between two teeth or<br>implants, usually at the anterior of<br>the mouth. |
| Dimensions | Diameter:3.75<br>Length:5,7,9. | Diameter:3.75<br>Length:5,7,9. | Diameter:3.75 Length:11 |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | Titanium alloy |
| Angle | No | No | No |
| | AB Dental | MIS | AlfaBio tec |
| Product<br>Name | P3W - Wide<br>anti-rotation<br>abutment | MD-WMAC1<br>Wide cementing<br>post, SP | TLAW |
| K Number | | K040807 | K063364 |
| Indication<br>for Use | To be used<br>when there is a<br>wide space<br>between two<br>teeth or<br>implants, usually<br>at the posterior<br>part of the<br>mouth | To be used when<br>there is a wide<br>space between<br>two teeth or<br>implants, usually<br>at the posterior<br>part of the mouth | To be used when there is a wide<br>space between two teeth or<br>implants, usually at the posterior<br>part of the mouth |
| Dimensions | Diameter: 3.75<br>Length: 9, 12. | Diameter: 5.5<br>Length:11 | Diameter: 4.5<br>Length:9 |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium | Titanium |
| Angle | 0 | 0 | 0 |
| | Ti-6Al-4V-ELI.<br>Grade-5. | 6Al-4V-ELI.<br>Grade-5. | |
| Angle | No | No | No |
| Product<br>Name | AB Dental<br>P3N-Narrow<br>Anti-rotation<br>Abutment | AB Dental<br>P3N-Narrow Anti-<br>rotation Abutment | MIS<br>MN-MAC10<br>Narrow platform<br>cementino post. NP |
| | Product Name | | |
| | P12-3.75 - Flat<br>Connection<br>Abutment | FC-6013,6018 | |
| K Number | | K081751 | |
| Indication<br>for Use | Abutment with a<br>flat connection<br>for screw<br>retained<br>restoration on<br>tilted implants | Abutment with a<br>flat connection for<br>screw retained<br>restoration on<br>tilted implants | |
| Dimensions | Diameter:4.5<br>Length:5 | Diameter:4<br>Length:5 | |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | |
| Angle | 0 | 0 | |
| | AB Dental | Adin | |
| Product Name | P12-3.75-T -<br>Temporary Flat<br>Connection<br>Abutment | FC-6013,6018 | |
| K Number | | K081751 | |
| Indication<br>for Use | Temporary<br>abutment with a<br>flat connection<br>for screw<br>retained<br>restoration on<br>tilted implants | Abutment with a<br>flat connection for<br>screw retained<br>restoration on<br>tilted implants | |
| Dimensions | Diameter:4.5<br>Length:15 | Diameter:4<br>Length:5 | |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5. | |
| Angle | 0 | 0 | |
| | AB Dental | Alpha-Bio Tec | |
| Product Name | P15-3x -<br>Temporary<br>Abutment for<br>Immediate<br>Loading | Straight Titanium<br>Abutments<br>ETLA | |
| K Number | | K063364 | |
| Indication<br>for Use | Used for | Used for | |
| for Use | temporary<br>restoration. | temporary<br>restoration. | |
| Dimension<br>s | Diameter:3, 3.75<br>Length: 1,2,3,4,5<br>,7 | Diameter:3.9,4.5<br>Length: 7 | |
| Material | Titanium alloy<br>Ti-6Al-4V-ELI.<br>Grade-5+plastic. | Titanium alloy Ti-<br>6Al-4V-ELI.<br>+plastic. | |
| Angle | 0 | 0 | |
| HEALING CAPS | | | |
| | AB Dental | Alpha Bio | |
| Product<br>Name | PO-P6H for<br>Aesthetic<br>Abutment with<br>Hex | HCT | |
| K<br>Number | | K042654 | |
| Indication<br>for Use | Used to allow the<br>gingiva to heal<br>around P6H<br>abutment | Used to allow the<br>gingiva to heal<br>around P6H<br>abutment | |
| Dimensio<br>ns | Diameter:4.7<br>Length:5 | Diameter:4.7<br>Length:5 | |
| Material | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5 | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5 | |
| | AB Dental | Alpha Bio | Adin |
| Product<br>Name | PO-Titanium<br>Healing Cap | HS | WP-0046,0047,0048,0049,0050 |
| K<br>Number | | K042654 | K081751 |
| Indication<br>for Use | Used to allow the<br>gingiva to heal<br>around implants | Used to allow the<br>gingiva to heal<br>around implants | Used to allow the gingiva to heal<br>around implants |
| Dimensio<br>ns | Diameter:3.75<br>Length:2,3,4,5,<br>6,7 | Diameter:4.5<br>Length:2,3,4,5,6,7 | Diameter:4<br>Length:2,3,4,5,6 |
| Material | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5 | Titanium | Titanium alloy Ti-6Al-4V-ELI.<br>Grade-5 |
| ADAPTORS | | | |
| | AB Dental | MIS | |
| Product<br>Name | P16-Straight<br>adaptor | Pilier Multi Unit | |
| K<br>Number | K040807 | | |
| Indication<br>for Use | Used for screw<br>retained<br>restoration | Used for<br>temporary<br>restoration | |
| Dimensions | Diameter:3.0,3.75<br>Length: 1,2,3,4,5. | Diameter:3.75<br>Length: 1,2,3,4 | |
| Material | Titanium alloy Ti-<br>6Al-4V-ELI.<br>Grade-5 | Titanium alloy Ti-<br>6Al-4V-ELI. | |
ళ్ళు
all and the collection of the state of the starter of the start
{6}------------------------------------------------
న్నాయ
{7}------------------------------------------------
9
September 19, 10, 10, 10,
.
100
{8}------------------------------------------------
:
{9}------------------------------------------------
{10}------------------------------------------------
יל איני מיני .
(අබ
Contractor Comments
િત્ત્વ
ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤ ﻟﻠﻘﺎﻧﻮﻥ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ
> ಿಗಳು
> ವಿಧಾರವಾಗಿ ಪ್ರಮ : ﺍﻟﻤﻮﺍﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮ
{11}------------------------------------------------
# Brief Discussion of the Non-Clinical Tests Submitted
As part of demonstrating safety and effectiveness of AB Dental dental implants and in showing substantial equivalence to the predicate devices that are the subject of this 51 0(k) submission, AB Dental submitted a selected number of its dental implants for fatigue testing in accordance with ISO 14801, Dentistry -Implants - Dynamic Fatigue Test for Endosseous Dental Implants. Further, AB Dental dental implants also underwent extensive SEM surface analysis and surface topography studies to prove that the blasting and postblasting surface cleaning process used during the manufacture of these devices produced implants that resulted in a clean, textured, surface. In addition, the 17 implants were subjected to pullout testing.
## Clinical Testing
No clinical testing was performed. Non-clinical testing was used to support the decision of safety and effectiveness.
# Conclusions
We believe that the AB Dental Implants products, which are the subject of this submission are substantially equivalent predicate devices cited. The device constitutes a safe, reliable and effective medical device, meeting all declared requirements of its intended use and the device does not introduce new risks and does not present any adverse health effects or safety risks to patients when used as intended.
{12}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
A.B. Dental Devices, LTD C/O Charles Hurwitz, Ph.D. MedicSense LTD POB 367 Ramat Hasharon ISRAEL 47103
APR 1 8 2012
Re: K112440
Trade/Device Name: AB Dental Devices Implants and Accessories Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 10, 2012 Received: April 13, 2012
Dear Dr. Hurwitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{13}------------------------------------------------
#### Page 2 - Dr. Hurwitz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
· Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{14}------------------------------------------------
### Indications for Use
#### 510 (k) Number K112440
Device Name: AB Dental Dental Devices Implants and Accessories
### Indications For Use :
. ()
The AB Dental Devices implants are intended for surgical placement in the maxillary and/or mandibular arch to support crowns, bridges, or overdentures in edentulous or partially edentulous patients.
17 Integral implant, 15 Conical implant, P15 Temporary abutment, P12-T,L Temporary flat connection abutment, and P16 Straight adaptor are appropriate for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use ( 21CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE(
Page 1 of 1
Susan Quare
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K160446
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.