The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments. Faros Mobile provides the detection and reporting features appropriate for the indications below - Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias. - Evaluation of patients for ST segment changes - Evaluation patients with pacemaker - Evaluating patient rest and stress ECG - Reporting heart rate variability analysis - Wireless transmission of patient ECG data and arrhythmia events for further analysis. Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements. Faros Mobile is contraindicated for - Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias - Pediatric patients weighting less than 10 kg
Device Story
Faros Mobile is a wearable, ambulatory ECG recorder/transmitter; captures heart electrical activity via snap electrodes; stores data on internal memory; transmits data via USB or Bluetooth to companion devices/systems (Cardiac Navigator/Explorer). Used in clinical/home environments by patients; managed by healthcare professionals via Faros Manager application. Device provides computer-generated analysis of cardiac abnormalities; requires physician confirmation. Enables long-term monitoring, arrhythmia detection, and remote data transmission; supports clinical decision-making by providing diagnostic-quality ECG data for rhythm and ST-segment evaluation.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by bench testing, including electrical/mechanical safety, electromagnetic compatibility, usability, and algorithm performance verification in accordance with FDA guidance for arrhythmia detectors.
Technological Characteristics
Ambulatory ECG recorder; 1- or 3-channel configuration; Li-Ion battery; USB and Bluetooth connectivity. Materials: commercially available snap ECG electrodes. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-47, ISO 10993-1, IEC 62304, IEC 62366-1. Software: Cardiac Navigator/Explorer for analysis; Faros Manager for configuration.
Indications for Use
Indicated for adult and pediatric patients (≥10 kg) requiring long-term ambulatory ECG monitoring in or outside clinical settings. Used for evaluating rhythm disturbances, arrhythmias, ST segment changes, pacemaker function, rest/stress ECG, and heart rate variability. Contraindicated for patients requiring attended in-hospital monitoring for life-threatening arrhythmias or pediatric patients <10 kg.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 8, 2018
Bittium Biosignals Ltd. Taneli Vaaraniemi Quality Manager Pioneerinkatu 6 Kuopio. Finland FI-70800
Re: K182030
Trade/Device Name: Faros Mobile Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MLO, DSI, DXH Dated: October 5, 2018 Received: October 9, 2018
Dear Taneli Vaaraniemi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica E. Paulsen -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182030
Device Name Faros Mobile
Indications for Use (Describe)
The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments.
Faros Mobile provides the detection and reporting features appropriate for the indications below
- · Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias.
- · Evaluation of patients for ST segment changes
- · Evaluation patients with pacemaker
- · Evaluating patient rest and stress ECG
- · Reporting heart rate variability analysis
- · Wireless transmission of patient ECG data and arrhythmia events for further analysis.
Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements.
Faros Mobile is contraindicated for
- · Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias
- · Pediatric patients weighting less than 10 kg
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 005 - Abbreviated 510k Summary
## 1. Submitter
| Bittium Biosignals Ltd. | Bittium Biosignals Ltd.<br>Pioneerinkatu 6<br>Kuopio, Finland FI-70800 |
|-------------------------|------------------------------------------------------------------------|
| Phone | +358 17 581 7700 |
| Fax: | +358 17 580 0978 |
| Registration number | 1000425315 |
| Contact | Taneli Vääräniemi (taneli.vaaraniemi@bittium.com)<br>Quality Manager |
## 2. Date
November 7, 2018
# 3. Device identification
| Device name | Faros Mobile |
|------------------------------|-----------------------------------------------------------------|
| Trade/Proprietary Name: | Faros ambulatory ECG recorder, telemetric |
| Common/Usual Name: | Digital Ambulatory ECG recorder, transmitter |
| Classification Product Code: | MLO - Electrocardiograph, Ambulatory, with Analysis Algorithm |
| Regulation Classification | 870.2800 |
| Subsequent Product Code | DSI - Arrhythmia detector and alarm |
| Regulation Classification | 870.1025 |
| Subsequent Product Code | DXH - Transmitters and Receivers, Electrocardiograph, Telephone |
| Regulation Classification | 870.2920 |
| Class of Device: | Class II |
| Panel: | Cardiovascular |
| Device Classification EU: | Class II |
| Medical Device Directive: | 93/42EEC as an amended by 2007/47/EC |
| | Certificate issued by VTT Expert services Ov. NB0537 |
{4}------------------------------------------------
Abbreviated 510(k) Premarket Submission
#### 4. Substantial equivalence, predicate device
K140847, CARDIO SPY ECG Holter Systems by LABTECH KFT.
#### Reference device: technology, design functionality
K143032, eMotion Faros ECG Mobile by Mega Electronics Ltd.
#### 5. Indications for Use Statement
The Faros Mobile system is intended for use in clinical long term ambulatory ECG monitoring, data transfer and analysis. Faros Mobile is indicated for adult and pediatric patients who require ECG monitoring inside or outside hospital or healthcare facility environments.
Faros Mobile provides the detection and reporting features appropriate for the indications below
- . Evaluation of patients with symptoms related to rhythm disturbances or symptoms suggesting arrhythmias.
- Evaluation of patients for ST segment changes
- Evaluation patients with pacemaker
- Evaluating patient rest and stress ECG
- Reporting heart rate variability analysis
- Wireless transmission of patient ECG data and arrhythmia events for further analysis.
Interpretation algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. The Faros Mobile does not provide interpretive statements.
Faros Mobile is contraindicated for
- . Those patients requiring attended, in-hospital monitoring for life threatening arrhythmias
- Pediatric patients weighting less than 10 kg
{5}------------------------------------------------
#### 6. Device Description
Faros Mobile is used for long term registration of a patient's heart electrocardiogram (ECG) and the wireless or wired transmission of the registered electrocardiogram arrhythmia events to a receiver or analysis system running on computer. The device is connected to a patient via electrode leads and the patient wears the device during the recording period and/or while he or she performs normal daily activities. The device is used with commercially available snap ECG electrodes.
Data is transferred to computer via USB. The ECG recording is analyzed at a medical facility using a Cardiac Navigator or Cardiac Explorer software.
In wireless transmission mode, Bluetooth communication protocol enables use of the Faros sensor as a part of a Mobile Cardiac Telemetry (MCT) or Cardiac Event Monitor (CEM) system. Data is transferred from device to companion device. For clarity, all the data is stored on sensor memory when the device operates via Bluetooth. The communication protocol is provided for 30° party integration purposes.
Device measurement configurations are managed via Faros Device Manager. The application operator can manage measurement configurations which include: ECG and motion data sampling frequencies, ECG channel count, heart rate variability, temperature, auto start and measurement auto stop features. Values for cardiac arrhythmia event detections are handled via Faros Manager application.
Faros Mobile system consists of:
- . Faros ECG sensor
- o Bluetooth communication protocol and documentation for implementing interface for data collection from Faros sensor
- channel and 3-channel cable sets o
- general micro-USB cable for recharging device O
- Faros Manager application
- Cardiac Explorer application
- Cardiac Navigator application
{6}------------------------------------------------
# 7. Comparison of technological characteristics
The following table compares the Faros Mobile to the predicate device with respect to indications for use, technological characteristics and principles of operation providing more information regarding the basis for the determination of substantial equivalence.
The technological characteristics and principles of operation of the Faros Mobile are the same as the predicate device.
| Manufacturer | Bittium Biosignals Ltd | LABTECH KFT. | Mega Electronics Ltd |
|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Faros Mobile | CARDIOSPY ECG Holter Systems | eMotion Faros ECG Mobile |
| | Subject device | Predicate device | Reference device |
| 510(k) number | K182030 | K140847 | K143032 |
| Product code and regulation number | DSI - 21 CFR 870.1025 | MLO - 21 CFR 870.2800 | DXH - 21 CFR 870.2920 |
| Subsequent Product Code and regulation number | DXH - 21 CFR 870.2920<br>MLO 21 CFR 870.2800 | n/a | n/a |
| ECG channels | 1- or 3-channels | 1, 2, 3, 12 -channels | 1- or 3-channels |
| ECG performance standard | IEC 60601-2-47 | IEC 60601-2-47 | IEC 60601-2-25 |
| RR Algorithm | Yes | Yes | Yes |
| Cardiac arrhythmia detection | Yes | Yes | No |
| Event marker | Yes | Yes | Yes |
| Data transfer | Yes | Yes | Yes |
| Battery type | Li Ion battery | 1x1.2 V AAA NiMH battery<br>(or 1x1.5 V AAA alkaline battery) | Li Ion battery |
| Software | Cardiac Navigator, Cardiac Explorer | Cardiospy software | LiveECG and Virtual Clinic |
| Simple, user friendly software with multiple functions | Yes | Yes | Yes |
| Full disclosure ECG review tool | Yes | Yes | Yes |
| Precise QRS classification and rhythm analysis | Yes | Yes | No |
| Arrhythmia analysis, arrhythmia overview | Yes | Yes | No |
| Atrial Fibrillation analysis | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.