K143032 · Mega Electronics , Ltd. · DXH · Mar 29, 2015 · Cardiovascular
Device Facts
Record ID
K143032
Device Name
eMotion Faros ECG Mobile
Applicant
Mega Electronics , Ltd.
Product Code
DXH · Cardiovascular
Decision Date
Mar 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research. The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation. Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters. Indicated for adult patients who require clinical or non-clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment. The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.
Device Story
Mobile telemetry solution for ambulatory monitoring of chronic patients; collects health data (ECG, weight, blood pressure, SpO2) from external plug-in devices via Bluetooth. Data transmitted via mobile network to 'Virtual Clinic' server; healthcare professionals review data via PC-based interface. Device includes eMotion Faros ECG sensors (90, 180, 360 models) which record ECG data to internal memory or transmit in real-time. Used in clinical settings or home environments under physician supervision. Provides no automatic analysis or diagnosis; serves as data acquisition and transmission tool to support clinical decision-making.
Clinical Evidence
No clinical testing performed. Substantial equivalence established via non-clinical bench testing, verification, and validation against established design protocols and international standards.
Technological Characteristics
Wearable ECG sensors (Faros 90, 180, 360) with 14-bit or 24-bit resolution; 3.7V Li-ion battery; Bluetooth connectivity; internal memory (512MB-1GB). Standards: EN 60601-1 (safety), EN 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), EN 60601-2-25 (ECG performance).
Indications for Use
Indicated for adult patients requiring clinical or non-clinical ECG monitoring in healthcare facilities under physician supervision or via physician prescription for home-based data acquisition.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2015
Mega Electronics Ltd. Taneli Vaaraniemi Quality Engineer Pioneerinkatu 6 Kuopio, 70800 FI
Re: K143032
Trade/Device Name: Emotion Faros ECG Mobile Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II Product Code: DXH Dated: February 6, 2015 Received: February 9, 2015
Dear Taneli Vaaraniemi,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143032
Device Name eMotion Faros ECG Mobile
#### Indications for Use (Describe)
The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.
The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation.
Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.
Indicated for adult patients who require clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment.
The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Mega Electronics Ltd. 510(k) Premarket Submission eMotion Faros ECG Mobile
Image /page/3/Picture/1 description: The image shows the logo for Mega. The logo consists of a square with rounded corners that is pink. Inside the square is a white silhouette of an archer aiming an arrow. To the right of the square is the word "mega" in black, lowercase letters.
## 007 – 510(k) Summary
### 1. Submitter
Mega Electronics Ltd Pioneerinkatu 6 KUOPIO FI-70800 Finland Phone: +358 17 581 7700 Fax: +358 17 580 0978 Contact: Taneli Vääräniemi, Quality Engineer Email: taneli.vaaraniemi@megaemg.com
Registration Number: 1000425315
### 2. Date
March 25, 2015
### 3. Device Identification
| Modified ECG Module Names: | a) eMotion Faros 90<br>b) eMotion Faros 180<br>c) eMotion Faros 360 |
|-----------------------------------|---------------------------------------------------------------------|
| Modified Health Gateway Interface | Virtual Clinic |
| Trade/Proprietary Name: | eMotion Faros ECG Mobile |
| Common/Usual Name: | Digital Ambulatory Monitor |
| Classification Name: | Transmitters And Receivers, Electrocardiograph Telephone |
| Classification Regulation: | 21 CFR 870.2920 |
| Product Code: | DXH |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
| Device Classification EU: | Class IIa |
| Medical Device Directive: | 93/42EEC as a amended by 2007/47/EC |
# 4. Substantial Equivalence, Predicate Device
eMotion ECG Mobile, 510(k) Number: K131699
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains a logo with a pink square on the left side and the word "mega" in black on the right side. The pink square contains a white silhouette of an archer aiming an arrow. The word "mega" is written in a bold, sans-serif font.
## 5. Device Description
The eMotion Faros ECG Mobile is a mobile device. PC and Internet based telemetry solution for the ambulant monitoring of the pluq-in device data of chronic patients via a mobile network. Pluq-in devices can be ECG devices, blood pressure monitors, weighing scales, etc. The device reads data from the plug-in device via a Bluetooth connection. The mobile device sends the data to a server (Health Gateway) over mobile networks using the secured connection.
Data can be viewed from the Health Gateway server using the Virtual Clinic interface. Monitoring is performed using PC application that reads data from the server over the internet using the secured connection. ECG data is possible to record on eMotion Faros ECG sensors internal memory. Data from ECG sensors can be reviewed with eMotion EDF Viewer or uploaded via USB connection.
Data acquisition is performed using eMotion Faros ECG sensors and applicable plug-in devices. Faros ECG sensors are modified versions of eMotion ECG sensor which the FDA granted clearance on November 26, 2013 (K131699).
## Intended Use
The eMotion Faros ECG sensor is a wearable, externally applied, electrocardiograph recorder and transmitter for the purpose of health monitoring, biofeedback and scientific research.
The eMotion Faros ECG Mobile is intended for use in clinical settings to collect and transmit health parameters to healthcare professionals for monitoring and evaluation.
Health parameters are collected from a variety of commercially available, external plug-in devices such as ECG sensors, Weight Scales, Blood Pressure Meters and Pulse Oximeters.
Indicated for adult patients who require clinical or non-clinical ECG monitoring in healthcare facility environment under supervision of a physician or prescript by the supervising physician to supplement data acquisition in home environment.
The eMotion Faros ECG Mobile does not provide any automatic analysis or diagnosis.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with a pink square on the left and the word "mega" in black on the right. The pink square contains a white silhouette of an archer aiming an arrow. The word "mega" is written in a simple, sans-serif font and is positioned to the right of the square.
## 6. Comparison of Technological Characteristics
The eMotion Faros ECG Mobile is modified device of the predicate device. With respect to intended use, technological characteristics and principles of operation are identical with Predicate device, eMotion ECG Mobile K131699. Comparison to Legally Marketed Device in section 009 provides more detailed information regarding the basis for the determination of substantial equivalence.
The technological characteristics and principles of operation of the modified device are the same as the predicate device.
| | eMotion ECG<br>sensor (K131699) | eMotion Faros 90<br>(modified device) | eMotion Faros 180<br>(modified device) | eMotion Faros 360<br>(modified device) |
|-------------------------|-----------------------------------------------|---------------------------------------|-----------------------------------------------|-----------------------------------------------|
| ECG<br>Channel(s) | Online: 1-channel | Offline:<br>1-channel | Online and offline:<br>1-channel | Online: 1-channel<br>Offline:3-channels |
| Sensor<br>Dimensions | 35 x 35 x 15 (mm) | 48 x 29 x 12 (mm) | 48 x 29 x 12 (mm) | 48 x 29 x 12 (mm) |
| Sensor Weight | 16g | 13g | 13g | 13g |
| Internal Data<br>logger | no | Yes | yes | yes |
| Memory<br>Capacity | no | 512Mb | 1 GB | 1 GB or more |
| ECG Sampling<br>Rate | Selectable: 100,<br>125, 250, 500, 1000<br>Hz | Selectable: 100,<br>125, 250 Hz | Selectable: 100,<br>125, 250, 500, 1000<br>Hz | Selectable: 100,<br>125, 250, 500, 1000<br>Hz |
| Resolution | 14-bit | 14-bit | 14-bit | 24-bit |
| Bluetooth | yes | no | yes | yes |
| Battery | 3,7 V Li-ion battery | 3,7 V Li-ion battery | 3,7 V Li-ion battery | 3,7 V Li-ion battery |
Table 2: Technological Characteristics comparison Between ECG Sensors
| | Web Monitor<br>(Predicate Device) | Virtual Clinic<br>(Subject Device) |
|-------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------|
| Login screen | Yes | Yes |
| Patient list | Yes | Yes |
| Document list | Yes | Yes |
| Doctor info view | Yes | Yes |
| Doctor list | Yes | Yes |
| Measurement views<br>- ECG<br>- ECG Report and listing<br>- Blood pressure<br>- Body weight<br>- SpO2 | Yes | Yes |
| Patient management | Yes | Yes |
| Measurement value limits with<br>notifications | No | Yes |
| Administration Management | Yes | Yes |
Table 3: Comparison Between UI Interfaces features
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Mega Limited. On the left side of the image is a pink square with rounded corners. Inside the square is a white silhouette of an archer holding a bow and arrow. To the right of the square is the word "mega" in black, lowercase letters.
## 7. Performance Data Summary
The eMotion Faros ECG Mobile has been fully verified and validated against written test protocols to demonstrate that the design meets the requirements and performs as intended. The test results including product release information summary is documented as a part of the section 011 Summary of Design Control Activities.
Performance is tested and evaluated according to the applicable sections of the following standards:
- EN 60601-1:2006, Medical Electrical Equipment Part 1 General Requirements for Basic ● Safety and Essential Performance
- EN 60601-1-2: 2007, Medical Electrical Equipment Part 1-2: General Requirements for ● Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- ISO 10993-1: 2003, Biological Evaluation of Medical Devices Part 1: Evaluation and ● Testing
- . EN 60601-2-25:1999 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs
## 8. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing performed supports the substantial equivalence of the device.
## 9. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the modified device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the eMotion Faros ECG Mobile and the predicate devices do not raise any questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the eMotion Faros ECG Mobile is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.