K181982 · Intravu, Inc. · HRX · Aug 21, 2018 · Orthopedic
Device Facts
Record ID
K181982
Device Name
MIDAScope and Introducer Kit, and MIDASystem
Applicant
Intravu, Inc.
Product Code
HRX · Orthopedic
Decision Date
Aug 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.
Device Story
MIDAScope is a miniaturized, sterile, single-use disposable scope; features distal tip image sensor; LED illumination via fibers. Connects to reusable MIDASystem console via electrical cable. MIDASystem includes integrated LCD display and software for image processing/display. Introducer Kit (cannula, trocar, stopcock) facilitates joint space access. Used in clinical settings for diagnostic/operative arthroscopy/endoscopy. Physician operates device to visualize interior joints; output displayed on console for real-time clinical decision-making during procedures. Benefits include minimally invasive visualization of joint spaces.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included structural/functional evaluation, software validation, electrical/EMC testing, biocompatibility (ISO 10993-1), packaging validation (ISO 11607, ASTM D4169/F2096/F88), and sterilization validation (ISO 11135). All tests met pre-determined acceptance criteria.
Technological Characteristics
Single-use disposable scope; reusable console with LCD. Distal tip image sensor; LED illumination. Sterilization: Ethylene oxide (SAL 10^-6). Standards: ISO 8600-1/3/4 (scope performance), IEC 60601-1/1-2 (electrical/EMC), ISO 10993-1 (biocompatibility), ISO 11135 (sterilization), ASTM F1980 (aging), ASTM D4332 (environmental).
Indications for Use
Indicated for diagnostic and operative arthroscopic and endoscopic procedures requiring illumination and visualization of interior joint cavities and other body cavities via natural or surgical openings.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Neo-ArthroscopeTM - Single Use Digital Video Arthroscopic System (K152511)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter shade of blue.
August 21, 2018
IntraVu. Inc. % Mr. Dave Yungvirt Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, New Jersey 07041
Re: K181982
Trade/Device Name: MIDAScope and Introducer Kit, and MIDASystem Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, GCJ Dated: July 23, 2018 Received: July 25, 2018
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer R. Stevenson
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181982
Device Name
MIDAScope™ and Introducer Kit, and MIDASystem™
Indications for Use (Describe)
The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# 510(k) Applicant
IntraVu, Inc. 610 Price Avenue Redwood City, CA 94063 Phone: (650) 395-6067
#### Contact Person
Plamena Entcheva-Dimitrov, PhD Preferred Regulatory Consulting Inc. Phone: (650) 533-2253 Email: entcheva@preferredregulatoryconsulting.com
## Date of Summary
20 Aug 2018
## Device Overview
| Device Proprietary Name | MIDAScope™ and Introducer Kit, and MIDASystem™ |
|-------------------------|------------------------------------------------|
| Product Class | II |
| Regulation Name | Arthroscope |
| Regulation Number | 888.1100 |
| Product Code | HRX |
| | GCJ |
| Panel | Orthopedic |
#### Predicate Device
| 510(k) Number | Product Name | Manufacturer |
|---------------|----------------------------------------------------------------------------------------|-----------------------------|
| K162475 | mi-eye 2 and mi-eye 2 Monitor (primary predicate) | Trice Medical, Inc. |
| K141326 | NeedleView CHTM Scope Kits | BioVition Technologies, LLC |
| K152511 | Neo-ArthroscopeTM - Single Use Digital Video<br>Arthroscopic System (reference device) | Prosurg, Inc. |
{4}------------------------------------------------
#### Device Description
The MIDAScope™ is a miniaturized, sterile, single-use disposable scope. The image is acquired by an image sensor located at the distal tip. Illumination is provided by an LED and fibers when the disposable scope is connected to the reusable MIDASystem™ via an electrical cable. The scope is available in different lengths to accommodate different treatment locations. The scope is introduced into the joint space using the IntraVu disposable Introducer Kit, consisting of a cannula, trocar and stopcock, which are provided sterile in the same package as the scope.
The MIDASystem™ is a reusable device consisting of an integrated LCD display, and software that receives, processes and displays the images collected by the scope.
#### Intended Use
The device is intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
#### Indications for Use
The MIDAScope™ and Introducer Kit, and MIDASystem™ are indicated for use in diagnostic and operative arthroscopic and endoscopic procedures to provide illumination and visualization of interior cavity joints and other body cavities through a natural or surgical opening.
## Comparison to Predicate Device
The MIDASystem and Introducer Kit and MIDAScope and predicate devices have the same intended use and thus fulfill the requirements of 513(i) of the FD&C Act (21 U.S.C. \$ 360c(i)). Subject device and predicates are similar in design, principle of operation, components, users and use environment. The subject scope and its predicates are sterilized with ethylene oxide to yield SAL106. The subject system is reusable as are the predicate systems. The packaging of the MIDAScope (with the Introducer Kit), and MIDASystem and its predicates is similar and passed testing following the same internationally recognized standards. The subject device and the predicate devices were tested according to the same internationally recognized standards for electromagnetic compatibility, for electrical safety, and for biocompatibility and were shown to be safe and passed the standards' requirements.
Risks associated with the subject device and the predicates are similar. There are no significant technological differences between MIDASystem and MIDAScope (with the Introducer Kit) and the predicate devices.
{5}------------------------------------------------
Therefore, in conclusion, the subject devices MIDAScope (with the Introducer Kit), and MIDASystem are substantially equivalent to the predicates.
#### Performance Testing
Performance testing of the MIDAScope and Introducer Kit, and MIDASystem included structural and functional performance, software validation, electrical and EMC, biocompatibility, packaging and sterilization studies. Devices were subjected to environmental conditioning and simulated distribution to assure devices will perform as intended during marketing distribution. All tests met the pre-determined acceptance criteria at baseline and after aging. The following FDA-recognized standards were followed and the acceptance criteria set forth in them were met:
| Scope performance: | ISO8600-1, ISO8600-3, and ISO8600-4 |
|-----------------------------|------------------------------------------------------------------|
| Biocompatibility: | ISO10993-1 |
| Packaging validation: | ISO11607-1, ISO11607-2, ASTM D4169, ASTM<br>F2096, ASTM F88/F88M |
| Sterilization validation: | ISO11135, ISO10993-7, ISO11737-1 |
| Aging: | ASTM F1980 |
| Environmental conditions: | ASTM D4332 |
| Electrical and EMC testing: | IEC60601-1 and IEC60601-1-2 |
#### Safety and Effectiveness and Conclusion
Based on the evaluation of intended use, performance characteristics and meeting FDArecognized standards for safety and performance, the MIDAScope™ and Introducer Kit, and the MIDASystem™ are substantially equivalent to the predicate device. The labeling of the MIDAScope and Introducer Kit, and MIDASystem contains instructions for use and any necessary cautions and warnings to assure safe and effective use of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.