This system is indicated to be used by a trained physician to provide illumination and visualization of an interior cavity of the body through a natural or surgical opening in diagnostic and operative arthroscopic and endoscopic procedures. Examples of surgical use include but are not limited to procedures on the knee, shoulder, ankle, elbow, wrist, temporomandibular joint (TMJ), spinal, ophthalmic, ENT, and the cervix.
Device Story
NeedleView CH is a single-use, rigid fiberoptic micro-endoscope designed for diagnostic and operative arthroscopic and endoscopic procedures. Input consists of light transmitted through 30um light fibers; output is a visual image captured via a 17,000-pixel coherent silica fiber bundle. The device features a stainless steel shaft with a working channel for instrument passage. It is used by trained physicians in clinical or surgical settings. The scope couples with a non-disposable handpiece containing a light source and camera. The kit includes a cannula, dilator, Tuohy needles, scalpel, guidewire, and hemostasis valve adapter to facilitate access and visualization. By providing direct visualization of interior body cavities, the device assists physicians in performing minimally invasive procedures, potentially reducing trauma compared to open surgery.
Clinical Evidence
No clinical data. Substantial equivalence is established through bench testing, including performance testing of the scope, needles, cannula, HVA, and guidewire, alongside biocompatibility, packaging, and sterilization validation per ISO and ASTM standards.
Technological Characteristics
Rigid fiberoptic scope; 17,000-pixel coherent silica fiber bundle; 30um light fibers; stainless steel shaft and working channel. Dimensions: 3.4mm OD, 1.85mm working channel, 160mm working length. Field of view: 80° (air), 62° (water). Direction of view: 0°. Sterilization: Ethylene oxide (ANSI/AAMI/ISO 11135). Biocompatibility: ISO 10993-1. Packaging: ISO 11607-2, ASTM F2096, ASTM F88, ASTM F1886.
Indications for Use
Indicated for use by trained physicians for illumination and visualization of body cavities during diagnostic and operative arthroscopic and endoscopic procedures. Applicable to patients undergoing procedures on the knee, shoulder, ankle, elbow, wrist, TMJ, spine, ophthalmic, ENT, and cervix.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
BioVision SurgView™ Integrated Visualization System (K082293)
Richard Wolf Medical Instruments Corp. Endoscopic Spine System Set (YESS) (K973405)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2014
BioVision Technologies, LLC Mr. David Sanso President 221 Corporate Circle, Unit H Golden, Colorado 80401
Re: K141326
Trade/Device Name: NeedleView CH™ Scope Kits Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, GCJ Dated: August 22, 2014 Received: September 2, 2014
Dear Mr. Sanso:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Binita S. Ashar -S 2014.10.08 12:36:21 -04'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known): _ K141326
Device Name: NeedleView CH™ Scope Kits
Indications for Use: This system is indicated to be used by a trained physician to provide illumination and visualization of an interior cavity of the body through a natural or surgical opening in diagnostic and operative arthroscopic and endoscopic procedures. Examples of surgical use include but are not limited to procedures on the knee, shoulder, ankle, elbow, wrist, temporomandibular joint (TMJ), spinal, ophthalmic, ENT, and the cervix.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use N/A (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
NeedleView CH™ Scope Kits Traditional 510(k) Submission
## 510(k) SUMMARY
#### Submission Information
| Date: | October 7, 2014 |
|----------------------|-------------------------------------------------------|
| Applicant: | BioVision Technologies, LLC. |
| Address: | 221 Corporate Circle Unit H<br>Golden, Colorado 80401 |
| Telephone Number: | 303-237-9608 |
| Submitter's Contact: | David Sanso, President |
#### Device Information
| Proprietary Name: | NeedleView CH Scope Kit |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 888.1100 |
| Regulation Name: | Arthroscope / Endoscope |
| Regulatory Class: | II |
| Product Codes(s): | HRX, GCJ |
| Predicate Devices | BioVision SurgView™ Integrated Visualization System (K082293)<br>Joimax THESSYS Multiscope (K051827)<br>Hoogland Spine Products maxMorespine Endoscope (K083552)<br>Blazejewski MEDI-Tech Spinal Foraminoscope (K082841)<br>Richard Wolf Medical Instruments Corp. Endoscopic Spine System Set<br>(YESS) (K973405) |
#### Device Description
Name of Device: NeedleView CH™ Scope Kit
The NeedleView CH is a rigid Fiberoptic scope with a working channel designed for single use. The product consists of a 17,000 pixel coherent silica fiber bundle, 30um diameter light fibers and a stainless steel channel encased in a stainless steel shaft. The effective field of view is 80° in air and 62° in water. The direction of view is 0°. The scope has retaining groves to incorporate a drape that will cover the camera and cord that connects to the Surg View System. The NeedleView CH Scope Kit consists of a fiber-optic based micro-endoscope (3.4mm Outer Diameter, 1.85mm Working Channel, 160mm Working Length), 12F Cannula (4mm Inner Diameter), 12F Dilator(4mm Diameter), 14Gx127mm Tuohy Needle, 18Gx152mm Tuohy
{4}------------------------------------------------
Needle, #11 Scalpel, Guidewire (70cm Length), Hemostasis Valve Adapter (HVA), and a sterile drape. All of these items are disposable and for single-use only. The micro-endoscope is made to couple with a non-disposable system hand piece that contains a light source and camera, which is available separately.
#### Statement of Intended Use
The system is indicated to be used by a trained physician to provide illumination and visualization of an interior cavity of the body through a natural or surgical opening in diagnostic and operative arthroscopic and endoscopic procedures. Examples of surgical use include but are not limited to procedures on the knee, shoulder, ankle, elbow, wrist, temporomandibular joint (TMJ), spinal, ophthalmic, ENT, and the cervix.
### Comparison to Predicate Devices
- The table below shows relevant similarities and difference between the NeedleView CH Table 1: and its predicate devices.
| Feature | BioVision<br>SurgView™<br>Integrated<br>Visualization<br>System | Joimax<br>THESSYS<br>Multiscope | Hoogland<br>Spine Products<br>msxMorespine<br>Endoscope &<br>System | Blazejewski<br>MEDI-Tech<br>Spinal<br>Foraminoscope | Richard Wolf<br>Medical<br>Instruments Corp.<br>Endoscopic Spine<br>System Set (YESS) |
|-----------------------------------|-----------------------------------------------------------------|---------------------------------|---------------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------------------------------------|
| Similarities | | | | | |
| Intended<br>Use | X | X | X | X | X |
| Scope Shaft<br>Material | X | X | X | X | X |
| Labeling (Single<br>Use/Reusable) | X | X | X | X | |
| Differences | | | | | |
| Labeling (Single<br>Use/Resuable) | | X | X | X | X |
| Outer Diameter | X | | | X | |
| Number of<br>Channels | | X | | X | X |
### Summary of Technologies
The technological characteristics of the BioVision NeedleView CH are similar to those of the predicate devices listed.
{5}------------------------------------------------
# Performance Testing
The following performance tests were completed on the BioVision NeedleView CH Kit.
# NeedleView CH Scope Performance Testing
- NeedleView CH Performance Testing ●
- NeedleView CH Accelerated Aging Performance Testing .
# Tuohy Needle Performance Testing
- . Flexural Test for 14G Tuohy Needle
- Flexural Test for 18G Tuohy Needle ●
- 18G Tuohy Needle Buckling Test .
- . Tuohy Needle Puncture Test Summary
# Cannula Performance Testing
- Cannula Insertion and Withdrawal Test including the 12F dilator ●
- Cannula Leak Test ●
- Cannula Bending Test .
- Cannula Crushing Test ●
- . Cannula Tensile Test
# HVA
- . HVA Fit and Leak Test
# Guidewire Test
- Guidewire Tensile Test ●
- Guidewire Compression Test ●
- . Guidewire Flexural Test
Biocompatibility testing was completed utilizing the following standard:
- ISO 10993-1:2009, Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process.
Packaging, sterilization, distribution and accelerated aging testing were completed utilizing the following standards:
- ISO 11607-2:2006 Packaging for terminally sterilized medical devices Part 2: . Validation requirements for forming , sealing and assembly processes
- ASTM F2096:2011, Standard Test Method for Detecting Gross Leaks in Medical . Packaging by Internal Pressurization (Bubble Test)
- ASTM F88:2009, Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM F1886:2009, Standard Test Method for Determining Integrity of Seals for Medical Packaging by Visual Inspection
- ANSI/AAMI/ISO 11135:2007 Part 1, Sterilization of health care products Ethylene . oxide - Part 1 : Requirements for development, validation and routine control of a sterilization process for medical devices.
{6}------------------------------------------------
- EN 556:2006, Sterilization of medical devices Requirements for terminally-sterilized ● medical devices to be labeled "Sterile"
- AAMI TIR28:2009, Product adoption and process equivalency for ethylene oxide ● sterilization
- ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide ● sterilization residuals
- ANSI/AAMI/ISO 11135-2:2008, Guidance on the application of 11135-1 ●
- ISO 11737-1:2006. Determination of a population of microorganisms on products
- ISO 11137-3:2004. Guidance on evaluation and interpretation of bioburden data
- ISTA 2A:2011, Partial-Simulation Performance Test Procedure: Packaged-Products ● 150lb (68 kg) or Less
- . ASTM D4169:2009. Standard Practice for Performance Testing of Shipping Containers and Systems
- . ASTM F 1980:2007. Guide to Accelerated Aging of Sterile Medical Device Packages
## Clinical Testing
Clinical testing was not used to prove substantial equivalence. Establishment of equivalence is based on similarities of intended use, design, physical characteristics and geometry between the BioVision NeedleView CH and predicate devices.
## Conclusion
Based on the evaluation of the performance characteristics, construction, manufacturing processes, and indications of use, Bio Vision Technologies has concluded that the BioVision NeedleView CH is substantially equivalent to the predicate devices listed. In all cases, the characteristics of this device are identical or similar to the predicate devices as they relate to the intended uses or application. The indications and contraindications of this device and those of the predicates listed in this submission are in congruence between devices. BioVision Technologies has also determined that the differences in characteristics indicated in this evaluation do not adversely affect the safety and effectiveness of the BioVision NeedleView CH.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.