V9 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp). V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patients. It is not intended for helicopter transport, hospital ambulance or home use.
Device Story
Portable patient monitor; inputs physiological signals via NIBP cuff, temperature sensor, CO2 sensor, and SpO2 sensor; processes signals using integrated modules (BLT, Suntech, Nellcor, or Masimo) to display, store, and alarm on vital signs; used in hospital environments (out-patient departments, wards, NICU) by clinicians; output provides real-time vital sign data to assist clinical decision-making; benefits patient through continuous monitoring of physiological status.
Clinical Evidence
Bench testing only. No clinical studies were performed. Substantial equivalence is supported by the use of previously FDA-cleared physiological monitoring modules and verification of design specifications against international standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 80601-2-30, IEC 80601-2-61, ISO 80601-2-55, ISO 80601-2-56).
Technological Characteristics
Portable patient monitor; AC-powered; modular design supporting various SpO2, NIBP, CO2, and temperature sensors. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-8 (alarms), IEC 80601-2-30 (NIBP), IEC 80601-2-61 (SpO2), ISO 80601-2-55 (CO2), and ISO 80601-2-56 (temperature).
Indications for Use
Indicated for monitoring, displaying, reviewing, storing, and alarming physiological parameters (SpO2, PR, NIBP, CO2, Temp) in adult, pediatric, and neonatal patients within controlled hospital environments.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 5, 2019
Guangdong Biolight Meditech Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 Cn
Re: K181919
Trade/Device Name: Patient Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: February 19, 2019 Received: February 26, 2019
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181919
Device Name Patient Monitor Model: V9
### Indications for Use (Describe)
V9 Patient Monitor is intended to be used for monitoring, displaying, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp).
V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patiented for helicopter transport, hospital ambulance or home use.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K181919
- 1. Date of Preparation: 4/5/2019
- 2. Sponsor Identification
## Guangdong Biolight Meditech Co., Ltd.
No.2 Innovation First Road, Technical Innovation Coast Hi-Tech Zone, Zhuhai, 519085, Guangdong, China.
Establishment Registration Number: 3007305624
Contact Person: Dan Hou Position: Manager Tel: +86-0756-3399971 Fax: +86-0756-3399989 Email: hou d@blt.com.cn
### 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)
# Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850 Fax: 001-3609253199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Patient Monitor Model: V9 Classification Name: Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) Classification: II Product Code: MWI Regulation Number: 21 CFR 870.2300 Review Panel: Cardiovascular
Indications for Use:
V9 Patient Monitor is intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters of patients, including Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp).
V9 Patient Monitor is used to monitor vital signals for patients and is intended to be used in controlled, hospital environments. It is applicable for adult, pediatric and neonatal patients. It is not intended for helicopter transport, hospital ambulance or home use.
## Device Description
The V9 Patient Monitor can provide the monitoring of physiological parameters such as the Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Carbon dioxide (CO2) and Temperature (Temp). The V9 Patient Monitor is a portable device, consists of main unit and accessories, including NIBP cuff, Temperature sensor, CO2 sensor and SpO2 sensor.
It consists of main unit and accessories, including NIBP cuff, Temperature sensor, CO2 sensor and SpO2 sensor.
All modules, boards and accessories used in V9 Patient Monitor has already been cleared by FDA, referring to Table 1. In which, two NIBP and three SpO2 modules are optional for proposed device.
| Module | | 510(k) Number |
|-------------|---------------------|---------------|
| NIBP | BLT NIBP Module | K153135 |
| | Suntech NIBP Module | K152552 |
| SpO2 | Nellcor SpO2 Module | K052186 |
| | MASIMO SpO2 Module | K101896 |
| | BLT SpO2 Module | K153135 |
| Temperature | Temperature Module | K153135 |
| CO2 | CO2 Module | K153135 |
| Table 1. List of 510(k) of Modules | | |
|------------------------------------|--|--|
|------------------------------------|--|--|
{5}------------------------------------------------
#### 5. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- > IEC 60601-1-8: 2012 Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- > IEC 80601-2-30:2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- > IEC 80601-2-61:2011 Medical electrical equipment -- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ISO 80601-2-55:2011 Medical electrical equipment Part 2-55: Particular requirements for the > basic safety and essential performance of respiratory gas monitors
- A ISO 80601-2-56: 2009 Medical electrical equipment - Part 2-56: Particular requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
- > ANSI C63.27-2017 American National Standard For Evaluation Of Wireless Coexistence
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device predicate.
- 6. Clinical Test Conclusion
No clinical study is included in this submission, because all modules have been previously cleared by FDA. The K Number of each module is listed in Table 1 of this document.
- 7. Identification of Predicate Device
Predicate Device 510(k) Number: K153135 Product Name: V6 Vital Signs Monitor
{6}------------------------------------------------
### 8. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device<br>Vital Signs Monitor V6, K153135 |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | MWI | MWI |
| Regulation Number | 21 CFR 870.2300 | 21 CFR 870.2300 |
| Class | II | II |
| Indications for Use | V9 Patient Monitor is intended to be<br>used for monitoring, displaying,<br>reviewing, storing and alarming of<br>multiple physiological parameters of<br>patients, including Pulse Oxygen<br>Saturation (SpO2), Pulse Rate (PR),<br>Non-invasive Blood Pressure (NIBP),<br>Carbon dioxide (CO2) and Temperature<br>(Temp).<br>V9 Patient Monitor is used to monitor<br>vital signals for patients and is intended<br>to be used in controlled, hospital<br>environments. It is applicable for adult,<br>pediatric and neonatal patients. It is not<br>intended for helicopter transport,<br>hospital ambulance or home use. | The vital signs monitor is intended to be<br>used for monitoring, displaying, reviewing,<br>storing and alarming of multiple<br>physiological parameters of patients,<br>including Pulse Oxygen Saturation (SpO2),<br>Pulse Rate (PR), Non-invasive Blood<br>Pressure (NIBP), Carbon dioxide (CO2)<br>and Temperature (Temp).<br>This vital signs monitor is used to monitor<br>vital signals for patients and is suitable for<br>use in hospital environments including<br>out-patient department, wards and NICU. It<br>is not intended for helicopter transport,<br>hospital ambulance or home use. And it is<br>applicable for adult, pediatric and neonatal<br>patients. |
| Intended Population | The monitors are intended for adult,<br>pediatric and neonatal. | The monitors are intended for adult,<br>pediatric and neonatal. |
| Intended<br>Environment | The V9 Patient Monitor is used to be<br>used in hospital environments including<br>out-patient outpatient departments,<br>wards and NICU. It is not intended for<br>helicopter transport, hospital ambulance<br>or home use. | The vital signs monitor is used to be used<br>in hospital environments including<br>out-patient outpatient departments, wards<br>and NICU. It is not intended for helicopter<br>transport, hospital ambulance or home use. |
| Power Supply | The monitors can be supplied by AC<br>power. | The monitors can be supplied by AC<br>power. |
| Number of Patient<br>be monitored each<br>time | Single patient each time | Single patient each time |
| Sterile | No | No |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Alarm Function | Comply with IEC60601-1-8 | Comply with IEC60601-1-8 |
| SpO2 Module | BLT, Nellcor and Masimo<br>Comply with IEC 80601-2-61 | BLT and Nellcor<br>Comply with IEC 80601-2-61 |
| | NIBP Module | BLT and Suntech<br>Comply with IEC 80601-2-30 |
| CO2 Module | Comply with ISO 80601-2-55 | Comply with ISO 80601-2-55 |
| Temp Module | Comply with ISO 80601-2-56 | Comply with ISO 80601-2-56 |
| Patient Contact<br>Accessories | SpO2 Sensor, NIBP cuff, CO2 Sensor<br>and Temp Sensor | SpO2 Sensor, NIBP cuff, CO2 Sensor and<br>Temp Sensor |
| | Biocompatibility | All the patient contact accessories were<br>previously cleared by FDA |
# Table 2 Substantially Equivalent Comparison
{7}------------------------------------------------
The SpO2 and NIBP module of proposed device and predicate device is different. Masimo SpO2 and Suntech NIBP are not included in predicate device, however, the Masimo SpO2 and Suntech NIBP have been cleared by FDA under K131898 and K152552. Therefore, this difference is considered not to affect the substantially equivalency (SE) between the proposed and predicate device.
### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.