iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2). The arrhythmia detection and ST Segment analysis are intended for adult only. The monitor is additionally intended for use during patient transport inside and outside of the hospital environment. The monitor is not intended for airplane, helicopter transport, home use and MRI environments.
Device Story
iM20 is a portable patient monitor for continuous physiological monitoring (ECG, RESP, NIBP, SpO2, PR, TEMP, IBP, CO2). It functions as an independent monitor or as a module for V-series monitors (elite V5, V6, V8). Inputs include patient-connected sensors for electrical, pressure, and gas signals. The device processes these signals to display trends, store data, and trigger alarms for clinicians. Used in hospital settings and during patient transport; operated by trained healthcare professionals. When docked to V-series monitors, the host system controls alarms and power. Output allows clinicians to assess patient status and manage care. Benefits include continuous monitoring and portability for patient transport.
Clinical Evidence
Clinical validation of NIBP performed per ISO 81060-2:2013. Non-clinical evidence includes biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing against various IEC/ISO standards (e.g., IEC 60601-2-27 for ECG, ISO 80601-2-61 for SpO2). Software verification and validation conducted for 'major' level of concern.
Technological Characteristics
Portable patient monitor; 5-inch LCD (800x480); connectivity via Wi-Fi (802.11b/g/n). Materials comply with ISO 10993-1. Power: 11.1V-19.8V DC or AC-DC adapter. Standards: IEC 60601-1, IEC 60601-2-25/27/34/49, ISO 80601-2-55/56/61. Enclosure IP44. Arrhythmia/ST analysis algorithms for adults.
Indications for Use
Indicated for monitoring, storing, reviewing, and alarming of physiological parameters (ECG, RESP, TEMP, SpO2, PR, NIBP, IBP, CO2) in adult, pediatric, and neonatal patients by trained healthcare professionals in hospital and transport environments. Arrhythmia and ST-segment analysis indicated for adults only. Contraindicated for airplane, helicopter, home, and MRI use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2016
Edan Instruments. Inc. % Doug Worth Sr. Dir. US RA/QA Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94086
Re: K152552
Trade/Device Name: Patient Monitor, model iM20 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, GXY, DPS, DRG Dated: March 22, 2016 Received: March 28, 2016
Dear Doug Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152552
Device Name Patient Monitor, model iM20
#### Indications for Use (Describe)
iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
The monitored physiological parameters include: ECG. respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).
The arrhythmia detection and ST Segment analysis are intended for adult only.
The monitor is additionally intended for use during patient transport inside of the hospital environment.
The monitor is not intended for airplane, helicopter transport, home use and MRI environments.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:20px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:20px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) Summary
# Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>Shekou, Nanshan Shenzhen, |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 518067 P.R. China |
| | Tel: +86(0755) 26858736 |
| | Fax: +1 (408) 418-4059 |
| Contact person: | Queena Chen |
| Preparing date: | March 22, 2015 |
| 2. Device name and<br>classification: | Device Name: Patient Monitor |
| | Model: iM20 |
| | Classification Name/ Product code: |
| | 870.1025 monitor, physiological, patient(with arrhythmia<br>detection or alarms)/ MHX |
| | 870.1025 Detector and Alarm, Arrhythmia/ DSI |
| | 870.1025 Monitor, ST Segment with Alarm/ MLD |
| | 870.2300 Cardiac monitor (including cardiotachometer and rate<br>alarm)/ DRT |
| | 870.1130 Non-Invasive blood pressure/ DXN |
| | 870.1110 Blood pressure computer/ DSK |
| | 880.2910 Clinical Electronic Thermometers-Temperature<br>Monitor with Probe/ FLL |
| | 870.2700 Oximeter, Pulse/ DQA |
| | 868.1400 Carbon Dioxide Gas Analyzer/ CCK |
| | 882.1320 Cutaneous electrode/GXY |
| | 870.2340 Electrocardiograph/DPS |
| | 870.2910 Radiofrequency physiological signal transmitter and<br>receiver/ DRG |
| | Regulatory Class: Class II |
| 3.Premarket<br>Notification Class III<br>Certification and<br>Summary | Not applicable, the subject device is Class II. |
| 4. Predicate Device(s): | 1) Philips Medizin.Systeme Boeblingen GmbH IntelliVue MP2 cleared under K102562. |
| | 2) Edan Instruments, Inc. iM70/ K131971 |
| 5. Reason for<br>Submission | Introduce new device iM20 |
| 6. Pre-Submission,<br>IDE | Not applicable, there is no prior submission. |
| 7. Device Description: | The iM20 Patient Monitor System (hereinafter called iM20) can<br>perform long-time continuous monitoring of multiple<br>physiological parameters, including ECG, respiration (RESP),<br>non-invasive blood pressure (NIBP), oxygen saturation of the<br>blood (SpO2), pulse rate (PR), temperature (TEMP), invasive<br>pressure (IBP), Carbon Dioxide (CO2). The system capabilities<br>include storing, displaying, analyzing and controlling such data<br>afterwards. When necessary, alarms will be produced so that<br>doctors and nurses can manage patient care appropriately. The<br>system is intended to be used during patient transport inside and<br>outside of the hospital environment.<br>iM20 can be utilized in two ways, as an independent monitor<br>and as a module of V series Patient Monitor (including models<br>elite V5, elite V6 and elite V8). When used as an independent<br>monitor, it can simultaneously monitor, store, review several<br>parameters data. And transfer patient data to V series patient<br>monitor only under the transport mode. As a highly portable<br>monitor, its compact design makes it particularly appropriate<br>inside hospital and vehicle ambulance transport environments.<br>When as a multi-measurement module, when the iM20 is<br>directly connected to a V series patient monitor, it can provide<br>the measurements, trends, and patient information.<br>When<br>connected, the V series Patient Monitor controls the connected<br>iM20, including all alarm functionality. So no alarms are<br>available on iM20 in such application and iM20 takes power<br>from the V series Patient Monitor. |
| 8. Intended<br>Use/Indications for<br>Use: | iM20 Patient Monitor (hereinafter called monitor) is intended to<br>be used for monitoring, storing, and reviewing of, and to<br>generate alarms for, multiple physiological parameters of<br>adults, pediatrics and neonates. The monitors are intended for<br>use by trained healthcare professionals in hospital<br>environments.<br><br>The monitored physiological parameters include: ECG,<br>respiration (RESP), temperature (TEMP), oxygen saturation of |
{4}------------------------------------------------
{5}------------------------------------------------
pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).
The arrhythmia detection and ST Segment analysis are intended for adult only.
The monitor is additionally intended for use during patient transport inside and outside of the hospital environment.
The monitor is not intended for airplane, helicopter transport, home use and MRI environments.
## 9. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters improvement, detailed substantial equivalence discussion are included in the following tables.
| Item | Proposed device: iM20 | Predicate device:<br>IntelliVue MP2 | Compariso<br>n Result |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| 510(k)<br>Number | Current Submission | K102562 | |
| | Indications for Use | | |
| Intended Use | iM20 Patient Monitor is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended to be used by trained healthcare professionals in a hospital environment. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), and carbon dioxide (CO2). The arrhythmia detection and ST-segment monitoring are intended for adult only. Neonatal and pediatric | Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics and neonates in hospital environments. The | Different |
| | | | |
| | patients are not clinically validated.<br>The monitors are indicated for use by<br>health care professionals or under their<br>direction whenever there is a need for<br>monitoring the physiological parameters<br>of patients.<br>The monitors are additionally intended<br>for use during patient transport inside<br>and outside of the hospital environment.<br>The monitor is not intended for airplane,<br>helicopter transport, home use or MRI<br>environments. | MP2, X2, MP5,<br>MP5T, MP20,<br>MP30, MP40, and<br>MP50 are<br>additionally<br>intended for use in<br>transport situations<br>within hospital<br>environments. The<br>MP2, X2and MP5<br>are also intended<br>for use during<br>patient transport<br>outside of a<br>hospital<br>environment. | |
| ECG Function | | | |
| HR Calculation | | | |
| Range | ADU: 15 bpm to 300 bpm<br>PED/NEO: 15 bpm to 350 bpm | Adult/pedi: 15 to<br>300 bpm<br>Neo range: 15 to<br>350 bpm | Different |
| Accuracy | $\pm$ 1% or 1 bpm, whichever is greater | $\pm$ 1% of range | Different |
| Resolution | 1 bpm | 1bpm | Same |
| Sensitivity | $\ge$ 300 $\mu$ VPP | $\ge$ 200 $\mu$ Vpeak | Different |
| PVC Rate | | | |
| Range | ADU: 0 to 300 PVCs/ min<br>PED/NEO: 0 to 350 PVCs/ min | 0 to 300 bpm | Different |
| Resolution | 1 PVCs/min | 1 bpm | Different |
| ST Numeric | | | |
| Range | -2.0 mV to +2.0 mV | -20 to +20 mm | Different |
| Accuracy | -0.8 mV to +0.8 mV: $\pm$ 0.02 mV or 10%,<br>which ever one is greater.<br>Beyond this range: not specified. | $\pm$ 0.5mm or 15%,<br>which ever one is<br>greater | Different |
| Resolution | 0.01 mV | 0.1mm | Different |
| Range of Sinus and SV Rhythm | | | |
| Brady | Adult: RR interval for 5 consecutive<br>QRS complex $\ge$ 1.5 s.<br>Pediatric/neonatal: RR interval f…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.