1. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset. 2. The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment. 3. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (< 70 cc by CTA or MRA, < 25 cc by MR-DWI). Endovascular therapy with the device should start within 6-16 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
Device Story
Neurovascular mechanical thrombectomy device; restores blood flow in large intracranial vessels (ICA, MCA, basilar, vertebral arteries) by removing thrombus. Distal nitinol stent retriever facilitates clot capture; radiopaque Platinum/Iridium markers aid visualization. Used in clinical/hospital settings by trained physicians. Device deployed via catheter to occlusion site; captures thrombus; retrieved to restore perfusion. Output is restored blood flow; clinical benefit is reduced post-stroke disability. Expanded indication supports use in 6-16 hour 'late-window' patients with salvageable tissue.
Clinical Evidence
Supported by sub-analysis of DEFUSE 3 trial (multicenter, randomized, open-label). Analysis cohort (mITT) included 32 Solitaire-treated vs 85 control subjects. Primary endpoint: 90-day mRS score shift (p=0.014). Secondary: functional independence (mRS 0-2) 31.2% (Solitaire) vs 15.3% (control). Safety: sICH 2.6% (Solitaire) vs 4.4% (control); mortality 10.5% (Solitaire) vs 25.6% (control).
Indicated for adult patients (18-90 years) with acute ischemic stroke due to persistent, proximal anterior circulation large vessel occlusion (ICA or MCA-M1). Includes patients ineligible for or failing IV t-PA. Contraindications include specific imaging findings (e.g., large infarct core, hemorrhage), severe systemic illness, or conditions precluding endovascular intervention.
Regulatory Classification
Identification
A neurovascular mechanical thrombectomy device for acute ischemic stroke treatment is a prescription device used in the treatment of acute ischemic stroke to improve clinical outcomes. The device is delivered into the neurovasculature with an endovascular approach, mechanically removes thrombus from the body, and restores blood flow in the neurovasculature.
Special Controls
In combination with the general controls of the FD&C Act, the Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including:
(i) Mechanical testing to demonstrate the device can withstand anticipated tensile, torsional, and compressive forces.
(ii) Mechanical testing to evaluate the radial forces exerted by the device.
(iii) Non-clinical testing to verify the dimensions of the device.
(iv) Non-clinical testing must demonstrate the device can be delivered to the target location in the neurovasculature and retrieve simulated thrombus under simulated use conditions.
(v) Non-clinical testing must demonstrate the device is radiopaque and can be visualized.
(vi) Non-clinical testing must evaluate the coating integrity and particulates under simulated use conditions.
(vii) Animal testing must evaluate the safety of the device, including damage to the vessels or tissue under anticipated use conditions.
(3) Performance data must support the sterility and pyrogenicity of the patient contacting components of the device.
(4) Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the specified shelf-life.
(5) Clinical performance testing of the device must demonstrate the device performs as intended for use in the treatment of acute ischemic stroke and must capture any adverse events associated with the device and procedure.
(6) The labeling must include:
(i) Information on the specific patient population for which the device is intended for use in the treatment of acute ischemic stroke, including but not limited to, specifying time from symptom onset, vessels or location of the neurovasculature that can be accessed for treatment, and limitations on core infarct size.
(ii) Detailed instructions on proper device preparation and use for thrombus retrieval from the neurovasculature.
(iii) A summary of the clinical testing results, including a detailed summary of the device- and procedure-related complications and adverse events.
(iv) A shelf life.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
March 6. 2019
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Helen Chow, PhD, RAC Senior Specialist, Regulatory Affairs 9775 Toledo Way Irvine, California 92618
Re: K181807
Trade/Device Name: Solitaire™ 2 Revascularization Device, Solitaire™ Platinum Revascularization Device (Solitaire™ Revascularization Device) Regulation Number: 21 CFR 882.5600 Regulation Name: Neurovascular Mechanical Thrombectomy Device for Acute Ischemic Stroke Treatment Regulatory Class: Class II Product Code: POL, NRY Dated: February 1, 2019 Received: February 4, 2019
Dear Helen Chow, PhD, RAC:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
S
Digitally signed by John John Marler -Marler -S Date: 2019.03.06 19:37:30 -05'00'
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181807
#### Device Name
Solitaire™ 2 Revascularization Device, Solitaire™ Platinum Revascularization Device (Solitaire™ Revascularization Device)
#### Indications for Use (Describe)
1. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ische to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
2. The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
3. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ische to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (< 70 cc by CTA or MRA, < 25 cc by MR-DWI). Endovascular therapy with the device should start within 6-16 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) Summary - K181807
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration No. 2029214 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer Correa<br>Program Manager, Regulatory Affairs<br>Telephone: (949) 297-5494<br>E-mail: jennifer.l.correa@medtronic.com |
| Alternate<br>Contact: | Helen Chow<br>Sr. Specialist, Regulatory Affairs<br>Telephone: (949) 297-5474<br>E-mail: helen.h.chow@medtronic.com |
| Date Summary<br>Prepared: | March 5, 2019 |
| Trade Name of<br>Device: | Solitaire™ 2 Revascularization Device,<br>Solitaire™ Platinum Revascularization Device<br>(Solitaire™ Revascularization Device) |
| Common Name of<br>Device: | Neurovascular Mechanical Thrombectomy Device for Acute<br>Ischemic Stroke Treatment |
| Classification of<br>Device: | Class II, 21 CFR 882.5600, 21 CFR 870.1250 |
| Product Code: | POL, NRY |
| Predicate Devices: | Solitaire™ 2 Revascularization Device (K162539), |
| Reference Devices: | Solitaire™ Platinum Revascularization Device (K153071,<br>K160641, K161879) |
#### Device Description:
The Solitaire™ Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Revascularization Device facilitates clot retrieval and has Platinum/Iridium radiopaque markers on the proximal and distal ends. The Solitaire™ Platinum Revascularization Device also features radiopaque markers along the circumference of the working length of the devices are supplied sterile and intended for single-use only.
There have been no changes to the design of the Solitaire™ Revascularization Device from the currently cleared device (K153071, K160641, K161879, K162539) to support the proposed additional indication. The currently cleared Solitaire™ 2 Revascularization Device is used as a predicate device and the Solitaire™ Platinum Revascularization Devices are used as a reference device for previously completed bench, animal, and clinical data.
{4}------------------------------------------------
#### Indications for Use:
- 1. The Solitaire™M Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset.
- 2. The Solitaire™ Revascularization Device is indicated to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for IV t-PA or who fail IV t-PA therapy are candidates for treatment.
- 3. The Solitaire™M Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion of the internal carotid artery (ICA) or middle cerebral artery (MCA)-M1 segments with smaller core infarcts (< 70 cc by CTA or MRA, < 25 cc by MR-DWI). Endovascular therapy with the device should start within 6-16 hours of time last seen well in patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy.
#### Device Comparison:
A comparison of the technological characteristics of the subject Solitaire™ Revascularization Device and the predicate Solitaire™ 2 Revascularization Device (K162539) is provided in Table 1. The subject Solitaire™ Revascularization device is identical to the predicate Solitaire™ 2 Revascularization Device and reference Solitaire™ Platinum Revascularization Device, apart from the expanded indications for use.
| Table 1: Device Comparison | | | |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Predicate Solitaire™ 2 Revascularization Device (K162539) | Subject Solitaire™ Revascularization Device | Rationale for Difference (if applicable) |
| Indication for Use | 1. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset. | 1. The Solitaire™ Revascularization Device is indicated for use to restore blood flow in the neurovasculature by removing thrombus for the treatment of acute ischemic stroke to reduce disability in patients with a persistent, proximal anterior circulation, large vessel occlusion, and smaller core infarcts who have first received intravenous tissue plasminogen activator (IV t-PA). Endovascular therapy with the device should be started within 6 hours of symptom onset. | The DEFUSE 3 study data analysis has demonstrated that the expanded indications for use does not raise any new or different questions of safety or effectiveness. |
| | Predicate Solitaire™<br>2 Revascularization Device<br>(K162539) | Subject Solitaire™<br>Revascularization Device | Rationale for<br>Difference<br>(if applicable) |
| | 2. The Solitaire™<br>Revascularization Device is<br>indicated to restore blood<br>flow by removing thrombus<br>from a large intracranial<br>vessel in patients<br>experiencing ischemic<br>stroke within 8 hours of<br>symptom onset. Patients<br>who are ineligible for IV t-PA or who fail IV t-PA<br>therapy are candidates for<br>treatment. | 2. The Solitaire™<br>Revascularization Device is<br>indicated to restore blood<br>flow by removing thrombus<br>from a large intracranial<br>vessel in patients experiencing<br>ischemic stroke within 8<br>hours of symptom onset.<br>Patients who are ineligible for<br>IV t-PA or who fail IV t-PA<br>therapy are candidates for<br>treatment.<br>3. The Solitaire™<br>Revascularization Device is<br>indicated for use to restore<br>blood flow in the<br>neurovasculature by removing<br>thrombus for the treatment of<br>acute ischemic stroke to<br>reduce disability in patients<br>with a persistent, proximal<br>anterior circulation, large<br>vessel occlusion of the<br>internal carotid artery (ICA)<br>or middle cerebral artery<br>(MCA)-M1 segments with<br>smaller core infarcts (< 70 cc<br>by CTA or MRA, < 25 cc by<br>MR-DWI). Endovascular<br>therapy with the device<br>should start within 6-16 hours<br>of time last seen well in<br>patients who are ineligible for<br>intravenous tissue<br>plasminogen activator (IV t-<br>PA) or who fail IV t-PA<br>therapy. | |
| Principles<br>of<br>Operation | The device is used in the<br>neurovasculature to restore<br>blood flow for treatment of<br>acute ischemic stroke | Same | N/A |
| Dimensions and Materials | | | |
| Device<br>Size(s) | 4-15 mm<br>4-20 mm<br>4-40 mm<br>6-20 mm<br>6-30 mm | 4-15 mm<br>4-20 mm<br>4-40 mm<br>6-20 mm<br>6-30 mm<br>4-20-05 mm<br>4-20-10 mm<br>4-40-10 mm<br>6-20-10 mm | The subject device<br>includes sizes<br>cleared in both the<br>predicate Solitaire™<br>2 and reference<br>Solitaire™ Platinum<br>Revascularization<br>Device |
| Table 1: Device Comparison | | | |
| | Predicate Solitaire™<br>2 Revascularization Device<br>(K162539) | Subject Solitaire™<br>Revascularization Device | Rationale for<br>Difference<br>(if applicable) |
| Device<br>Materials | Stent: Nitinol<br>Pushwire: Nitinol<br>Markers: 90% Platinum/ 10%<br>Iridium<br>Push-wire shrink Tubing: PTFE<br>Introducer Sheath:<br>PTFE/Grilamid | 6-40-10 mm<br>Same | N/A |
| Sterilization and Packaging | | | |
| Packaging<br>Materials | Stored within dispenser coil,<br>Tyvek pouch, and shipping<br>carton. | Same | N/A |
| Sterilization<br>Method | Ethylene Oxide | Same | N/A |
| How<br>Supplied | Sterile, Single Use | Same | N/A |
{5}------------------------------------------------
{6}------------------------------------------------
# Leveraged Non-Clinical Data:
There are no changes to the subject Solitaire™ Revascularization Device in design, manufacturing process, sterilization, material formulation, principles of operation, or fundamental scientific technology. The only change is the expanded Indications for Use. Therefore, no new or additional non-clinical testing was required or performed.
## Performance Testing - Clinical:
Clinical data from the Endovascular Therapy Following Imaging Evaluation for Ischemic Stroke (DEFUSE 3) study was used to support the expanded indication for the subject Solitaire™ Revascularization Device. The DEFUSE 3 study allowed the use of various cleared endovascular therapy devices. To support substantial equivalence of the subject Solitaire™ Revascularization Device, a sub-analysis of the DEFUSE 3 study data was performed and was limited to subjects treated with the Solitaire™ Revascularization Device in the endovascular plus standard medical therapy (endovascular therapy group) vs. those treated with standard medical therapy alone (control/medicaltherapy group).
## Study Design
The DEFUSE 3 study was a multicenter, randomized, open-label trial, with blinded outcome assessment, of thrombectomy in subjects 6 to 16 hours after they were last known to be well and who had remaining ischemic brain tissue that was not yet infarcted. Subjects with proximal middlecerebral-artery or internal-carotid-artery occlusion, an initial infarct size of less than 70 ml, and a ratio of the volume of ischemic tissue on perfusion imaging to infarct volume of 1.8 or more were randomly assigned to endovascular therapy (thrombectomy) plus standard medical therapy (endovasculartherapy group) or standard medical therapy alone (control/medical-therapy group). The primary outcome was the ordinal score on
<sup>1</sup> Albers GW, Marks MP, Kemp S, et al. Thrombectomy for Stroke at 6 to 16 Hours with Selection by Perfusion Imaging. The New England journal of medicine. Feb 22 2018;378(8):708-718.
{7}------------------------------------------------
the modified Rankin scale (range, 0 to 6, with higher scores indicating greater disability) at day 90.
#### Sample Size
The DEFUSE 3 Study was designed to assess the safety and effectiveness of endovascular therapy using FDA cleared mechanical thrombectomy devices, and was not designed to assess a specific neurothrombectomy device. The DEFUSE 3 study included a total of 182 subjects (92 in the endovascular therapy group and 90 in the control group). Solitaire was the first device used amongst all mechanical thrombectomy devices for 38 subjects in the endovascular therapy group.
#### Statistical Analysis
A subgroup analysis was conducted including all subjects randomized to endovascular intervention and in whom Solitaire was the first device used amongst all mechanical thrombectomy devices, and all subjects randomized to the control group. Results between the two groups thus defined were summarized and compared, including formal statistical hypothesis testing on the primary efficacy endpoint.
All analyses were carried out consistent with the principle of intention-to-treat in that the randomized assignments in DEFUSE 3 as a whole were preserved in making group assignments, and in whom Solitaire was the first device used amongst all mechanical thrombectomy devices (that is, intention-to-treat with Solitaire) constitute the Analysis Cohort. The intention-to-treat (ITT) Analysis Cohort is thus comprised of 90 in the control group and 38 in the Solitaire group.
The DEFUSE 3 study allowed IV t-PA use beyond 3 hours, although IV t-PA is not approved in the United States beyond 3 hours. A total of 11 subjects (5 from the control group and 6 from the Solitaire group) were excluded from the primary and secondary efficacy endpoint analyses due to receiving IV t-PA beyond 3 hours and/or carotid stenting. Therefore, the primary and secondary efficacy endpoint analyses consist of 117 subjects (Analysis Cohort - mITT). In the mITT Analysis Cohort, subjects with multiple interventions (n=8) were treated as failures. All analyses presented for the Analysis Cohort (ITT and mITT) are post-hoc analyses. As such, there may be uncertainty in the interpretation of the clinical study results and any statistically meaningful conclusions due to the post-hoc nature of the analyses, limitations in sample size, and lack of pre-specified hypothesis testing.
## Study Endpoints
The primary efficacy outcome was the ordinal score on the modified Rankin scale (range, 0 [no symptoms] to 6 [death]) at day 90. The primary safety outcomes were death within 90 days and the occurrence of symptomatic intracranial hemorrhage (sICH) within 36 hours, defined as an increase of at least 4 points in the NIHSS score that was associated with brain hemorrhage on imaging within 36 hours after symptom onset.
Imaging outcomes were:
- . infarct volume measured at 24 hours (with a window of ± 6 hours) after randomization;
- lesion growth (increase in volume of the infarct) between baseline imaging and 24 hours; ●
{8}------------------------------------------------
- reperfusion, defined as a greater than 90% reduction in the region of perfusion delay ● (Tmax of > 6 seconds) between baseline and 24 hours; and
- complete recanalization of the primary arterial occlusive lesion at 24 hours on CTA or ● MRA.
The technical efficacy of the endoyascular procedure in establishing reperfusion was defined in the endovascular-therapy group by a modified Thrombolysis in Cerebral Infarction (mTICI) score of 2b (50 to 99% reperfusion) or 3 (complete reperfusion).
#### Inclusion Criteria
Clinical inclusion criteria:
- 1. Signs and symptoms consistent with the diagnosis of an acute anterior circulation ischemic stroke
- 2. Age 18-90 years
- 3. Baseline NIHSS is ≥ 6 and remains ≥ 6 immediately prior to randomization
- 4. Endovascular treatment can be initiated (femoral puncture) between 6 and 16 hours of stroke onset. Stroke onset is defined as the time the patient was last known to be at their neurologic baseline (wake-up strokes are eligible if they meet the above time limits).
- 5. Modified Rankin Scale (mRS) less than or equal to 2 prior to qualifying stroke (functionally independent for all ADLs)
- 6. Patient/Legally Authorized Representative has signed the Informed Consent form.
Neuroimaging inclusion criteria:
- ICA or MCA-M1 occlusion (carotid occlusions can be cervical or intracranial; with or ● without tandem MCA lesions) by MRA or CTA
AND
- . Target Mismatch Profile on CT perfusion or MRI (ischemic core volume is < 70 ml. mismatch ratio is > 1.8 and mismatch volume* is > 15 ml)
*Notes: The mismatch volume is determined by the RAPID software in real time based on the difference between the ischemic core lesion volume and the Tmax > 6s lesion volume. If both a CT perfusion and a multimodal MRI scan are performed prior to enrollment, the later of the 2 scans is assessed to determine eligibility. Only an intracranial MRA is required for patients screened with MRA; cervical MRA is not required. Cervical and intracranial CTA are typically obtained simultaneously in patients screened with CTA, but only the intracranial CTA is required for enrollment.
Alternative neuroimaging inclusion criteria (if perfusion imaging or CTA/MRA is technically inadequate):
- A. If CTA (or MRA) is technically inadequate:
Tmax>6s perfusion deficit consistent with an ICA or MCA-M1 occlusion AND
Target Mismatch Profile (ischemic core volume is < 70 ml, mismatch ratio is >1.8 and mismatch volume is >15 ml as determined by RAPID software)
- B. If MRP is technically inadequate:
{9}------------------------------------------------
ICA or MCA-M1 occlusion (carotid occlusions can be cervical or intracranial; with or without tandem MCA lesions) by MRA (or CTA, if MRA is technically inadequate and a CTA was performed within 60 minutes prior to the MRI)
AND
MR Diffusion Weighted Imaging (DWI) lesion volume < 25 ml
- C. If CTP is technically inadequate: Patient can be screened with MRI and randomized if neuroimaging criteria are met.
# Exclusion Criteria
Clinical exclusion criteria:
- Other serious, advanced, or terminal illness (investigator judgment) or life expectance is ● less than 6 months.
- . Pre-existing medical, neurological or psychiatric disease that would confound the neurological or functional evaluations
- Pregnant ●
- Unable to undergo a contrast brain perfusion scan with either MRI or CT
- Known allergy to iodine that precludes an endovascular procedure ●
- Treated with t-PA > 4.5 hours after time last known well
- Treated with t-PA 3-4.5 hours after last known well AND any of the following: age > 80, current anticoagulant use, history of diabetes AND prior stroke, NIHSS >25
- . Known hereditary or acquired hemorrhagic diathesis, coagulation factor deficiency; recent oral anticoagulant therapy with INR > 3 (recent use of the new oral anticoagulants is not an exclusion if estimated GFR > 30 ml/min).
- Seizures at stroke onset if it precludes obtaining an accurate baseline NIHSS
- Baseline blood glucose of < 50 mg/dL (2.78 mmol) or > 400mg/dL (22.20 mmol) ●
- Baseline platelet count < 50,000/uL ●
- Severe, sustained hypertension (Systolic Blood Pressure >185 mmHg or Diastolic Blood Pressure > 110 mmHg)
- Current participation in another investigational drug or device study ●
- Presumed septic embolus; suspicion of bacterial endocarditis
- Clot retrieval attempted using a neurothrombectomy device prior to 6 hours from symptom onset
- . Any other condition that, in the opinion of the investigator, precludes an endovascular procedure or poses a significant hazard to the subject if an endovascular procedure was performed
Neuroimaging exclusion criteria:
- ASPECT score < 6 on non-contrast CT (if patient is enrolled based on CT perfusion ● criteria)
- . Evidence of intracranial tumor (except small meningioma) acute intracranial hemorrhage, neoplasm, or arteriovenous malformation
- Significant mass effect with midline shift
- Evidence of internal carotid artery dissection that is flow limiting or aortic dissection
- . Intracranial stent implanted in the same vascular territory that precludes the safe deployment/removal of the neurothrombectomy device
{10}------------------------------------------------
- Acute symptomatic arterial occlusions in more than one vascular territory confirmed on ● CTA/MRA (e.g., bilateral MCA occlusions, or an MCA and a basilar artery occlusion).
#### Demographic and other Baseline Characteristics (Analysis Cohort - ITT)
Subject demographics and baseline characteristics are presented in Table 2. Subject age, gender, race, ethnicity, baseline NIHSS, systolic blood pressure and diastolic blood pressure did not differ significantly between the Solitaire and control groups. Median age was 71 years in each treatment group, while gender was similarly balanced with a total of 54% (69/128) females. The majority of the subject population was non-Hispanic and white. Baseline median NIHSS scores were 17 in the Solitaire group and 16 in control, indicating a moderate to severe stroke population.
| Table 2: Demographics and Baseline Characteristics (Analysis Cohort - ITT) | | | | |
|----------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------|---------|------------------------------------------------------|
| | Solitaire<br>Mean ± SD (N)<br>[Median] (IQR)<br>or % (n/N) | Control<br>Mean ± SD (N)<br>[Median] (IQR)<br>or % (n/N) | p-value | All<br>Mean ± SD (N)<br>[Median] (IQR)<br>or % (n/N) |
| Outcome | | | | |
| Age (years) | 68.2 ± 13.0 (38)<br>[71.0] (58.3,76.8) | 68.9 ± 13.4 (90)<br>[71.0] (59.3,80.0) | 0.797 | 68.7 ± 13.2 (128)<br>[71.0] (59.0,79.0) |
| Gender | | | 0.329 | |
| Male | 39.5% (15/38) | 48.9% (44/90) | | 46.1% (59/128) |
| Female | 60.5% (23/38) | 51.1% (46/90) | | 53.9% (69/128) |
| Race | | | 0.765 | |
| American Indian or Alaska<br>Native | 0.0% (0/38) | 1.1% (1/90) | | 0.8% (1/128) |
| Asian | 2.6% (1/38) | 3.3% (3/90) | | 3.1% (4/128) |
| Black or African American | 10.5% (4/38) | 5.6% (5/90) | | 7.0% (9/128) |
| White | 86.8% (33/38) | 88.9% (80/90) | | 88.3% (113/128) |
| Unknown | 0.0% (0/38) | 1.1% (1/90) | | 0.8% (1/128) |
| Ethnicity | | | 0.266 | |
| Hispanic or Latino | 18.4% (7/38) | 11.1% (10/90) | | 13.3% (17/128) |
| Not Hispanic or Latino | 81.6% (31/38) | 88.9% (80/90) | | 86.7% (111/128) |
| NIHSS at baseline | 16.5 ± 5.4 (38)<br>[17.0] (14.3,20.0) | 16.9 ± 6.4 (90)<br>[16.0] (12.0,21.0) | 0.749 | 16.7 ± 6.1 (128)<br>[17.0] (12.0,21.0) |
| Systolic blood pressure (mmHg) | 140.8 ± 23.7 (38)<br>[139.5]<br>(123.3,158.0) | 145.5 ± 19.5 (90)<br>[147.0]<br>(130.5,159.0) | 0.255 | 144.1 ± 20.9 (128)<br>[146.0]<br>(130.0,159.3) |
| Diastolic blood pressure (mmHg) | 74.8 ± 12.9 (38)<br>[74.0] (68.0,82.0) | 79.3 ± 15.9 (90)<br>[81.0] (67.0,88.8) | 0.122 | 78.0 ± 15.1 (128)<br>[77.0] (67.0,87.0) |
#### Procedural Characteristics (Analysis Cohort - ITT)
Procedural characteristics are summarized in Table 3. Among both treatment groups, a majority of subjects were transferred to the endovascular enrolling hospital from an external facility. In the Solitaire group, IV t-PA was administered for 18.4% (7/38) of subjects and general anesthesia was administered in 36.8% (14/38) of subjects. The median time from arrival at the enrolling hospital to arterial puncture was 118.5 minutes, with a median of 65.5 minutes from imaging to arterial puncture. The median time from symptom onset to emergency department
{11}------------------------------------------------
(ED) arrival was 530.0 minutes. Workflow intervals between stroke onset and randomization, ED arrival, arterial puncture and reperfusion times are also provided.
| Table 3: Procedural Characteristics (Analysis Cohort - ITT) | | |
|-------------------------------------------------------------|-----------------------|-----------------------|
| | Solitaire | Control* |
| | Mean ± SD (N) | Mean ± SD (N) |
| | [Median] (IQR) | [Median] (IQR) |
| Outcome | or % (n/N) | or % (n/N) |
| Transfer Subject | 68.4% (26/38) | 70.0% (63/90) |
| IV t-PA Administration | 18.4% (7/38) | 8.9% (8/90) |
| General Anesthesia | 36.8% (14/38) | - |
| Onset to randomization (min) | 631.2 ± 162.2 (38) | 653.0 ± 153.5 (90) |
| | [644.5] (474.5,742.0) | [644.5] (522.8,780.8) |
| Onset to arterial puncture (min) | 667.9 ± 160.1 (38) | - |
| | [674.5] (527.8,778.0) | - |
| Time of onset to ED arrival (min) | 520.7 ± 162.6 (38) | - |
| | [530.0] (405.5,637.8) | - |
| ED arrival to arterial puncture (min) | 147.2 ± 120.8 (38) | - |
| | [118.5] (80.8,156.8) | - |
| Imaging to arterial puncture (min) | 69.9 ± 31.3 (38) | - |
| | [65.5] (46.0,86.8) | - |
| Onset to reperfusion (min) | 719.0 ± 167.5 (34) | - |
| | [708.0] (569.8,855.2) | - |
| *Note: Data provided for control group where available | | |
## Primary Efficacy Outcomes
In the DEFUSE 3 endovascular-therapy group, treatment was associated with a favorable shift in the distribution of mRS scores at 90 days compared to the control group. Similarly, in the modified ITT Analysis Cohort, the Solitaire cohort shows a favorable shift in the distribution of mRS scores at 90 days compared to the control (p-value=0.014), as presented in Table 4. The shift toward better outcomes was consistent in direction across all levels of the mRS (Figure 1).
| Table 4: Primary Effectiveness Endpoint (Analysis Cohort – mITT) | | | |
|------------------------------------------------------------------|-----------------|-----------------|---------|
| | Solitaire | Control | |
| | Mean ± SD (N) | Mean ± SD (N) | |
| | [Median] (IQR) | [Median] (IQR) | |
| mRS Score at 90 days | or % (n/N) | or % (n/N) | p-value |
| - | 3.2 ± 1.7 (32) | 4.0 ± 1.8 (85) | 0.014 |
| | [3.0] (2.0,4.0) | [4.0] (3.0,6.0) | |
| 0 | 6.2% (2/32) | 8.2% (7/85) | |
| 1 | 12.5% (4/32) | 3.5% (3/85) | |
| 2 | 12.5% (4/32) | 3.5% (3/85) | |
| 3 | 25.0% (8/32) | 15.3% (13/85) | |
| 4 | 21.9% (7/32) | 27.1% (23/85) | |
| 5 | 9.4% (3/32) | 15.3% (13/85) | |
| 6 | 12.5% (4/32) | 27.1% (23/85) | - |
{12}------------------------------------------------
Image /page/12/Figure/0 description: This image is a bar graph comparing the number of people in the Solitaire group (n=32) and the Control group (n=85) across different categories labeled 0 through 6. In the Solitaire group, the numbers are 6, 13, 13, 25, 22, 9, and 13 for categories 0 to 6, respectively. For the Control group, the numbers are 8, 4, 4, 15, 27, 15, and 27 for categories 0 to 6, respectively.
Image /page/12/Figure/1 description: This image contains the title "Patients (%)" and the text "Figure 1: mRS at 90 days (Analysis Cohort - mITT)". The text indicates that the figure likely presents data related to the modified Rankin Scale (mRS) scores of patients at 90 days. The analysis is based on the modified intent-to-treat (mITT) cohort. The figure likely shows the percentage of patients achieving different mRS scores at the specified time point.
## Secondary Efficacy Outcome
The proportion of subjects functionally independent at 90 days (mRS 0-2) favors the Solitaire group over control (31.2% vs 15.3%). Results are presented in Table 5.
| Table 5: Secondary Efficacy Endpoint, mRS 0-2 at 90 days (Analysis Cohort – mITT) | | |
|-----------------------------------------------------------------------------------|----------------------|--------------------|
| Outcome | Solitaire<br>% (n/N) | Control<br>% (n/N) |
| Functional Independence (mRS 0-2) at 90 days | 31.2% (10/32) | 15.3% (13/85) |
## Additional Efficacy, Imaging and Technical Efficacy Outcomes
Early neurological improvement was defined as a > 8-point improvement in NIHSS from baseline to the 24-hour visit, or an NIHSS score of 0-1 attained at the 24-hour visit irrespective of the baseline value of NIHSS. Results are presented in Table 6.
{13}------------------------------------------------
| Table 6: Early neurological improvement at 24 Hours (Analysis Cohort – ITT) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|--------------------|
| Outcome | Solitaire<br>% (n/N) | Control<br>% (n/N) |
| Early neurological improvement | 31.6% (12/38) | 5.6% (5/89*) |
| *Note: NIHSS at 24 hours, which is needed to compute early neurological improvement, was<br>available for 89 of 90 subjects in the control group. | | |
In the Solitaire group, arterial reperfusion as measured by the modified TICI score by the central reader, was assessed immediately post-procedure. Results are presented in Table 7.
| Table 7: TICI Immediate Post-Procedure* (Analysis Cohort – mITT) | |
|---------------------------------------------------------------------|----------------------|
| Outcome | Solitaire<br>% (n/N) |
| TICI post-procedure (central reading) | |
| 0 | 21.9% (7/32) |
| 2a | 12.5% (4/32) |
| 2b | 43.8% (14/32) |
| 3 | 21.9% (7/32) |
| TICI≥2b post-procedure | 65.6% (21/32) |
| *TICI was not assessed for Control Group immediately post-procedure | |
Reperfusion was analyzed both as a continuous metric and dichotomized to define successful reperfusion (> 90% reperfused) at 24 hours post-procedure. Complete recanalization of the primary arterial occlusive lesion was assessed on CTA or MRA at 24 hours postprocedure. Results are presented in Table 8.
| Table 8: Reperfusion and Recanalization 24 hours Post-Procedure* (Analysis Cohort – mITT) | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------|
| | Solitaire<br>Mean ± SD (N)<br>[Median] (IQR)<br>or % (n/N) | Control<br>Mean ± SD (N)<br>[Median] (IQR)<br>or % (n/N) |
| Outcome | | |
| Reperfusion rate (%) | 92.6 ± 20.2 (24)<br>[100.0] (99.0,100.0) | 48.7 ± 46.0 (63)<br>[53.8] (24.7,84.7) |
| Successful reperfusion (>90%) | 83.3% (20/24) | 17.5% (11/63) |
| Complete recanalization at 24h | 82.8% (24/29) | 19.2% (14/73) |
| * Results are presented for subjects in the mITT analysis cohort with available data. Reperfusion<br>percentage and successful reperfusion were available on 24 of 32 subjects in the Solitaire group, and<br>63 of 85 subjects in the control group. Complete recanalization was available on 29 of 32 subjects in<br>the Solitaire group and 73 of 85 subjects in the control group. | | |
Additional imaging endpoints at 24 hours post procedure were assessed by the core lab and are presented in Table 9.
{14}------------------------------------------------
| Table 9: Imaging Endpoints (Analysis Cohort – ITT) | | |
|----------------------------------------------------|----------------------------------------------|----------------------------------------------|
| Imaging Outcomes* | Solitaire<br>Mean ± SD (N)<br>[Median] (IQR) | Control**<br>Mean ± SD (N)<br>[Median] (IQR) |
| Infarct volume (ml) at 24h per core lab | 64.5 ± 67.2 (38)<br>[35.0] (18.6,81.2) | 74.3 ± 80.7 (89)<br>[41.0] (25.4,106.2) |
| Infarct growth (ml) at 24h per core lab | 48.6 ± 61.4 (38)<br>[19.9] (3.4,79.9) | 57.6 ± 70.6 (89)<br>[32.8] (18.3,74.8) |
| Tmax6 volume (ml) at 24h per core lab | 9.2 ± 32.5 (30)<br>[0.0] (0.0,0.0) | 59.5 ± 59.6 (68)<br>[46.6] (11.1,78.7) |
*Measured by Rapid Software
**One subject in the control group was missing imaging data 24-hours post-procedure.
#### Primary Safety Outcomes
All-cause mortality at the 90-day follow-up visit was less frequent in the Solitaire group compared to control (10.5% vs 25.6%). The rate of sICH was low in both study groups, with a rate of 2.6% (1/38) with Solitaire and 4.4% (4/90) in control. Results are presented in Table 10.
| Table 10: Primary Safety Outcomes (Analysis Cohort – ITT) | | |
|-----------------------------------------------------------|----------------------|--------------------|
| Outcome | Solitaire<br>% (n/N) | Control<br>% (n/N) |
| All-cause mortality | 10.5% (4/38) | 25.6% (23/90) |
| Symptomatic ICH | 2.6% (1/38) | 4.4% (4/90) |
## Additional Important Safety Results
Procedural complications for the Solitaire group were reported by the clinical site and are presented in Table 11.
| Table 11: Procedural Complications (Analysis Cohort – ITT) | |
|---------------------------------------------------------------------------------|----------------------|
| Outcome | Solitaire<br>% (n/N) |
| Arterial dissection (per site) | 0.0% (0/38) |
| Access site complication requiring surgical repair or<br>transfusion (per site) | 0.0% (0/38) |
| Embolization to previously unaffected territory (per<br>site) | 0.0% (0/38) |
| Vessel perforation (per site) | 2.6% (1/38) |
Adverse events observed during the study were reported and coded via the MedDRA classification system and are presented in Table 12 through Table 14.
{15}------------------------------------------------
| (Analysis Cohort – ITT) | | |
|---------------------------------------------------------|-------------------------------|-----------------------------|
| System Organ Class | Solitaire<br>% (n/N) [events] | Control<br>% (n/N) [events] |
| TOTAL | 28.9% (11/38) [14] | 4.4% (4/90) [4] |
| Cardiac disorders | 2.6% (1/38) [1] | 1.1% (1/90) [1] |
| Injury, poisoning and<br>procedural complications | 2.6% (1/38) [1] | - |
| Nervous system disorders | 23.7% (9/38) [10] | 1.1% (1/90) [1] |
| Vascular disorders | 5.3% (2/38) [2] | - |
| General disorders and<br>administration site conditions | - | 1.1% (1/90) [1] |
| Investigations | - | 1.1% (1/90) [1] |
# Table 12: Procedure Related Adverse Events by MedDRA Classification
| Table 13: All Adverse Events by MedDRA Classification | | |
|-------------------------------------------------------|--------------------|---------------------|
| (Analysis Cohort – ITT) | | |
| | Solitaire | Control |
| System Organ Class | % (n/N) [events] | % (n/N) [events] |
| TOTAL | 78.9% (30/38) [97] | 86.7% (78/90) [229] |
| Blood and lymphatic system disorders | 5.3% (2/38) [2] | 6.7% (6/90) [7] |
| Cardiac disorders | 21.1% (8/38) [10] | 16.7% (15/90) [19] |
| Gastrointestinal disorders | 21.1% (8/38) [11] | 12.2% (11/90) [11] |
| General disorders and administration site conditions | 13.2% (5/38) [5] | 12.2% (11/90) [12] |
| Infections and infestations | 10.5% (4/38) [4] | 24.4% (22/90) [28] |
| Injury, poisoning and procedural complications | 7.9% (3/38) [3] | 4.4% (4/90) [4] |
| Investigations | 5.3% (2/38) [2] | 5.6% (5/90) [5] |
| Metabolism and nutrition disorders | 13.2% (5/38) [7] | 14.4% (13/90) [17] |
| Musculoskeletal and connective tissue disorders | 7.9% (3/38) [3] | 4.4% (4/90) [5] |
| Nervous system disorders | 42.1% (16/38) [22] | 58.9% (53/90) [64] |
| Psychiatric disorders | 13.2% (5/38) [5] | 4.4% (4/90) [4] |
| Renal and urinary disorders | 10.5% (4/38) [4] | 4.4% (4/90) [5] |
| Respiratory, thoracic and mediastinal disorders | 13.2% (5/38) [5] | 25.6% (23/90) [29] |
| Skin and subcutaneous tissue disorders | 2.6% (1/38) [1] | - |
| Vascular disorders | 26.3% (10/38) [13] | 16.7% (15/90) [16] |
| Congenital, familial and genetic disorders | - | 1.1% (1/90) [1] |
| Eye disorders | - | 1.1% (1/90) [1] |
| Product issues | - | 1.1% (1/90) [1] |
{16}------------------------------------------------
| Table 14: Serious Adverse Events by MedDRA Classification | | |
|-----------------------------------------------------------|--------------------|--------------------|
| (Analysis Cohort -ITT) | | |
| | Solitaire | Control |
| System Organ Class | % (n/N) [events] | % (n/N) [events] |
| TOTAL | 47.4% (18/38) [26] | 53.3% (48/90) [68] |
| Cardiac disorders | 5.3% (2/38) [2] | 4.4% (4/90) [5] |
| Gastrointestinal disorders | 7.9% (3/38) [4] | 3.3% (3/90) [3] |
| General disorders and<br>administration site conditions | 2.6% (1/38) [1] | 1.1% (1/90) [1] |
| Infections and infestations | 2.6% (1/38) [1] | 7.8% (7/90) [8] |
| Nervous system disorders | 23.7% (9/38) [10] | 31.1% (28/90) [30] |
| Psychiatric disorders | 2.6% (1/38) [1] | - |
| Renal and urinary disorders | 5.3% (2/38) [2] | - |
| Respiratory, thoracic and<br>mediastinal disorders | 10.5% (4/38) [4] | 14.4% (13/90) [15] |
| Vascular disorders | 2.6% (1/38) [1] | 3.3% (3/90) [3] |
| Injury, poisoning and<br>procedural complications | - | 2.2% (2/90) [2] |
| Product issues | - | 1.1% (1/90) [1] |
#### Conclusion
The DEFUSE 3 study was a multicenter, randomized, open-label trial, with blinded outcome assessment, of thrombectomy in patients 6 to 16 hours after they were last known to be well and who had remaining ischemic brain tissue that w…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.