K181728 · Shenzhen Kentro Medical Electronics Co., Ltd. · NGX · Oct 17, 2018 · Physical Medicine
Device Facts
Record ID
K181728
Device Name
Muscle Trainer
Applicant
Shenzhen Kentro Medical Electronics Co., Ltd.
Product Code
NGX · Physical Medicine
Decision Date
Oct 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
Muscle Trainer is a powered muscle stimulator for OTC use; delivers electrical pulses to user's skin via electrode patches. Device consists of a host unit and electrode patches; powered by CR2032 button batteries. User selects from 12 intensity levels and 3 operation modes. Device generates biphasic, symmetric, square-wave pulses. Used on abdomen, shoulder, arm, thigh, or calf depending on model. Provides muscle conditioning; benefits include improved muscle performance. Healthcare provider/user interaction involves applying electrodes to skin and adjusting intensity via control buttons. Device includes automatic shut-off and no-load trip features for safety.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1, IEC 60601-1-11), electromagnetic compatibility (IEC 60601-1-2), and performance testing (IEC 60601-2-10). Waveform verification performed under load conditions (0.5kΩ, 2kΩ, 10kΩ). Electrode patch current dispersion and shelf-life testing conducted per ASTM F 1980-07.
Technological Characteristics
Powered muscle stimulator; ABS housing; CR2032 (DC3V) battery. Electrode patches use EVA foam, carbon film, conductive hydrogel, and PET. Output: biphasic, symmetric, square-wave pulses. 12 intensity levels; 3 modes. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Sterilization not specified; electrodes are cleared components (K171381).
Indications for Use
Indicated for healthy individuals to stimulate muscles to improve and facilitate muscle performance. No specific age or gender restrictions; contraindications not explicitly listed.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 17, 2018
Shenzhen Kentro Medical Electronics Co., Ltd % Rain Yip Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Rm. 3005, Area B, Bldg.1, Southward Ruifeng Business Center, Guimiao Road Shenzhen, Cn 518000 Guangdong
Re: K181728
Trade/Device Name: Muscle Trainer Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: July 14, 2018 Received: July 17, 2018
Dear Rain Yip:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181728
Device Name Muscle Trainer
Indications for Use (Describe)
To be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Date: 2018-06-04
#### I. Submitter
SHENZHEN KENTRO MEDICAL ELECTRONICS CO., LTD
No.3, Xihu Industry Zone, Xikeng Village, Henggang Town, Longgang District, Shenzhen City, Guangdong Province, China
Tel.: +86 755 3382 5998 Tax: +86 755 3382 5996
Zewu Zhang (Management representative) Tel: +86 134 1043 0929 Email: kentro@kentro.com.cn
# II. Device
Type of 510(k): Traditional Common Name: Powered muscle stimulator Trade Name: Muscle Trainer Models: KTR-230 series, KTR-231, KTR-232, KTR-233, KTR-234 Classification Name: Stimulator, Muscle, Powered, For muscle conditioning Review Panel: Physical Medicine Regulatory Class: II Product Code: NGX Regulation Number: 21 CFR 890.5850
# III. Predicate Device
| Applicant | | Predicate Device | | 510(k) Number | Approval Date |
|----------------------------|--------|---------------------------------------------------|--|---------------|---------------|
| Well<br>International Ltd. | Brain | (Primary):<br>GYMFORM®ABS &<br>CORE/VDPGYCSET0042 | | K142055 | May 11, 2015 |
| HIVOX<br>INC. | BIOTEK | HIVOX OTC Electrical<br>Stimulator/SEM44 | | K171803 | Nov.29, 2017 |
# IV. Device Description
Muscle Trainer is a product that adopts modern electronic science and technology to delivers electric pulses generated to the user's skin through the electrodes.
Muscle Trainer is mainly composed of the host and electrode patches, as well as it is powered by CR2023 battery. Of which, the electrode patch is cleared in K171381.
{4}------------------------------------------------
Muscle Trainer includes KTR-230 series (KTR-230B, KTR-230W, KTR-230P), KTR-231, KTR-232, KTR-233 and KTR-234 model. All models have three operation modes and one channel, which can give certain electrical pulse through electrode patches on the skin. Their technical parameters are slightly different, but they share the basically same characteristics:
- Electric pulse combination, 0~12 levels can be adjusted and chosen according to personal preference.
- CR2032 Button batteries (DC3V) power supply, easy to use and safe.
The main differences among them are the followings but not affect its intended use:
- Appearance
- The number of control button: KTR-230 series, KTR-231 and KTR-232 have two buttons to control: KTR-233 and KTR-234 have three buttons to control.
- The number of electrode patch used: KTR-230 series, KTR-233 and KTR-234 are equipped with one electrode patch; KTR-231 is equipped with six electrode patches as well as KTR-232 is equipped with two electrode patches.
- Applied parts: KTR-230 series, KTR-233 and KTR-234 apply to abdomen, shoulder, arm, thigh and calf; KTR-231 applies to abdomen in the place where between ribs and hip bones; KTR-232 applies to waist, shoulder, arm, thigh and calf.
# V. Indications for Use
To be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
# VI. Comparison of Technological Characteristics With the Predicate Devices
The Muscle Trainer is substantially equivalent to the predicated device based on intended use, design, specifications and performance.
The Muscle Trainer does not raise different questions of safety and effectiveness as compared to the predicate devices.
Information for predicate device was obtained from publicly available sources, including the 510(k) Summary and device instruction manual. A technical comparison to the predicate is provided below:
{5}------------------------------------------------
| Comparison<br>Elements | Subject Device | Primary Predicate Device<br>K142055 | Predicate Device 1<br>K171803 | | |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------|
| 510(k) number | K181728 | K142055 | K171803 | / | |
| Trade name | Muscle<br>Trainer/KTR-230series,<br>KTR-231, KTR-232, KTR-233,<br>KTR-234 | GYMFORM®<br>VDPGYCSET0042 | HIVOX OTC Electrical<br>Stimulator/SEM44 | / | |
| Product code | NGX | NGX | NUH<br>NGX | / | |
| Indication<br>for<br>use/Intended use | To be used to stimulate healthy<br>muscles in order to improve and<br>facilitate muscle performance. | GYMFORM®ABS& CORE is<br>intended to stimulate healthy<br>muscles in order to improve or<br>facilitate muscle performance. The<br>ABS& CORE may be considered a<br>technique or method for muscle<br>training. 2-area belt is intended for<br>use on the muscles in abdomen or<br>lower back separately. Mini belt is<br>intended for use on the muscles in<br>arms, legs, thighs or buttocks areas<br>separately | "HIVOX OTC Electrical Stimulator,<br>SEM44 –<br>TENS: The device is designed to be<br>used for temporary relief of pain<br>associated with sore and aching<br>muscles in the shoulder, waist, back,<br>neck, upper extremities (arm), lower<br>extremities (leg), abdomen and<br>bottom due to strain from exercise o<br>normal household work activities.<br>EMS: The device is designed to be<br>used for stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance" | SE<br>NOTE 4 | |
| Location for use | OTC | OTC | OTC | SE | |
| BASIC UNIT SPECIFICATIOSN | | | | | |
| Power supply | 2032 Button battery (DC3V) | 2 ×1.5V AAA batteries | 4.5V (batteries, 3×1.5V AAA) | SE | |
| Comparison<br>Elements | Subject Device | Primary Predicate Device<br>K142055 | Predicate Device 1<br>K171803 | | |
| | | | | NOTE 1 | |
| Number of output<br>modes | | 3 | 6 | TENS: 15<br>EMS: 35 | SE<br>NOTE 1 |
| | Channel<br>Number | 1 | 2 | 2 | SE<br>NOTE 1 |
| Number<br>of output<br>channels | Synchro<br>nous or<br>Alternat<br>ing? | N/A | Alternating | Synchronous | SE |
| | Method<br>of<br>Channel<br>Isolation | N/A | Press MODE button for 3 seconds | By electrical circuit and software | SE |
| Regulated Current<br>or Voltage? | | Regulated Voltage | Regulated Voltage | Regulated Voltage | SE |
| Software/Firmware<br>/Microprocessor<br>Control? | | Yes | Yes | Yes | SE |
| Automatic overload<br>trip voltage level? | | No | No | Yes | SE |
| Automatic no-load<br>trip? | | Yes | Yes | Yes | SE |
| Automatic shut off | | Yes | Yes | Yes | SE |
| Comparison<br>Elements | Subject Device | Primary Predicate Device<br>K142055 | Predicate Device 1<br>K171803 | | |
| Patient override<br>control? | Yes | Yes | Yes | SE | |
| Indicator<br>Display | On/Off<br>Status? | Yes | Yes | Yes | SE |
| | Low<br>Battery? | Yes | Yes | Yes | SE |
| | Voltage/<br>Current<br>Level? | N/A | Yes | Yes | SE |
| Timer range | | Default 15-minute | Default 10-minute | 5-100 minutes | SE |
| Compliance<br>with<br>voluntary standards | | ■ IEC60601-1-2<br>■ IEC60601-1<br>■ IEC60601-1-11<br>■ IEC60601-2-10 | ■ IEC60601-1-2<br>■ IEC60601-1<br>■ IEC60601-2-10 | ■ IEC60601-1-2<br>■ IEC60601-1<br>■ IEC60601-2-10 | SE |
| Compliance<br>with<br>21CFR 898 | | Yes | Yes | Yes | SE |
| Dimensions<br>(L*W*H) | | (Host)<br>KTR-230 series: φ51×13.5mm<br>KTR-231: φ51×13.5mm<br>KTR-232: φ51×13.5mm<br>KRT-233: 53.4×47.5×11.8mm<br>KTR-234: 49×49×12.6mm | 100mm×68mm×24.5mm | 132mm×63mm×29.5mm (including<br>belt clip) | SE<br>NOTE 2 |
| Weight | | (Host)<br>KTR-230 series: 0.6oz<br>KTR-231: 0.6oz<br>KTR-232: 0.6oz<br>KTR-233: 0.6oz | 50g (Without batteries) | 89g (including belt clip, without<br>batteries)<br>123g (including belt clip and<br>batteries) | SE<br>NOTE 2 |
| Comparison<br>Elements | Subject Device | Primary Predicate Device<br>K142055 | Predicate Device 1<br>K171803 | | |
| | KTR-234: 0.56oz | | | | |
| Housing<br>material<br>and construction | ABS | ABS | ABS | SE | |
| OUTPUT SPECIFICATIOSNS | | | | | |
| Waveform | Mode 1: Pulsed symmetric, biphasic,<br>square wave<br>Mode 2: Pulsed symmetric, biphasic,<br>square wave<br>Mode 3: Pulsed symmetric, biphasic,<br>square wave | Symmetrical, rectangular | Biphasic, square | SE | |
| Maximum<br>output<br>voltage | Mode 1:<br>(±10%)Vp<br>36.5V @500Ω<br>58.5V @2kΩ<br>93 V @10kΩ<br>Mode 2:<br>(±10%)Vp<br>36.5V @500Ω<br>58.5V @2kΩ<br>93 V @10kΩ<br>Mode 3:<br>(±10%)Vp<br>36.5V @500Ω<br>58.5V @2kΩ<br>93 V @10kΩ | (±10%)Vp<br>66V @ 500 Ω<br>69V @ 2k Ω<br>70V @ 10k Ω | (±10%)Vp<br>50V @ 500 Ω<br>60V @ 2k Ω<br>125V @ 10k Ω | SE<br>NOTE 3 | |
| Maximum<br>output | Mode 1:<br>(±10%) Ip | (±10%)Ip<br>132mA @ 500 Ω | (±10%)Ip<br>100mA @ 500 Ω | SE<br>NOTE 3 | |
| Comparison<br>Elements | Subject Device | Primary Predicate Device<br>K142055 | Predicate Device 1<br>K171803 | | |
| current | 73mA @ 500 Ω<br>29.25mA @ 2k Ω<br>9.3mA @ 10k Ω<br><br>Mode 2:<br>(±10%) Ip<br>73mA @ 500 Ω<br>29.25mA @ 2k Ω<br>9.3mA @ 10k Ω<br><br>Mode 3:<br>(±10%) Ip<br>73mA @ 500 Ω<br>29.25mA @ 2k Ω<br>9.3mA @ 10k Ω | 34.5mA @ 2k Ω<br>7mA @ 10k Ω | 45mA @ 2k Ω<br>12.5mA @ 10k Ω | | |
| Net charge (per<br>pulse) | @500Ω<br>Mode 1: 0<br>Mode 2: 0<br>Mode 3: 0 | 19.2μC @ 500 Ω | 0.001μC @ 500 Ω | SE | |
| Maximum phase<br>charge @500Ω | Mode 1: 6.85μC@500 Ω<br>Mode 2: 10.11µC@500 Ω<br>Mode 3: 10.11µC@500 Ω | 16.4µC @ 500 Ω | 0.045μC @ 500 Ω | SE<br>NOTE 3 | |
| Maximum average<br>current @500Ω | Mode 1: 3.6mA@500 Ω<br>Mode 2: 1.1mA@500 Ω<br>Mode 3: 5.4mA@500 Ω…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.