K142055 · Well Brain International , Ltd. · NGX · May 11, 2015 · Physical Medicine
Device Facts
Record ID
K142055
Device Name
GYMFORM ABS & CORE
Applicant
Well Brain International , Ltd.
Product Code
NGX · Physical Medicine
Decision Date
May 11, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GYMFORM®ABS& CORE is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS& CORE may be considered a technique or method for muscle training. 2-area belt is intended for use on the muscles in abdomen or lower back separately. Mini belt is intended for use on the muscles in arms, legs, thighs or buttocks areas separately.
Device Story
Battery-operated (2x AAA) powered muscle stimulator; console generates electrical signals delivered via replaceable electrode pads on belts. User-controlled intensity and mode via console buttons. Used by patients for muscle training on abdomen, lower back, arms, legs, thighs, or buttocks. Device provides alternating electrical current to targeted muscle groups; output parameters (frequency, pulse width, intensity) adjustable. Healthcare provider/user monitors output via console display. Benefits include improved muscle performance through electrical stimulation.
Indicated for stimulation of healthy muscles in adults to improve or facilitate muscle performance. Intended for use on abdomen, lower back, arms, legs, thighs, or buttocks. Contraindications not specified.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 11, 2015
Well Brain International, Ltd. Victor K. Wai Managing Director Room 03, 14/F, Fook Yip Bldg. 53-57 Kwai Fung Crescent Kwai Chung, N. T. Hong Kong, China
Re: K142055
Trade/Device Name: GYMFORM® ABS&CORE, Model: VDPGYCSET0042 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: April 3, 2015 Received: April 10, 2015
Dear Mr. Wai.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Felipe Aquel -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K142055
Device Name GYMFORM® ABS&CORE, Model: VDPGYCSET0042
Indications for Use (Describe)
GYMFORM®ABS& CORE is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS& CORE may be considered a technique or method for muscle training. 2-area belt is intended for use on the muscles in abdomen or lower back separately. Mini belt is intended for use on the muscles in arms, legs, thighs or buttocks areas separately.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Sponsor: Well Brain International Ltd. Subject Device: GYMFORM®ABS& CORE, Model: VDPGYCSET0042 File No.: 510(k) submission report (V1.0), Chapter 5
### Chapter 5. 510(k) Summary
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Date of the summary prepared: May 8, 2015
#### 2. Submitter's Information
- 510(k) Owner's Name: Well Brain International Ltd. �
- � Establishment Registration Number: 3004950644
- � Address: Room 03, 14/F, Fook Yip Bldg., 53-57 Kwai Fung Crescent, Kwai Chung, N. T. Hong Kong, China
- � Phone: (852) 2619-0833
- � Fax: (852) 2429-0960
- � Contact Person: Victor K Wai
- Email: victor@wellbrain-intl.com �
#### 3. Subject Device Information
- GYMFORM®ABS & CORE, Model: VDPGYCSET0042 � Trade Name:
- Powered muscle stimulator Common Name: �
- Stimulator, Muscle, Powered, For muscle conditioning � Classification name:
- Physical Medicine Review Panel: �
- NGX Product Code: �
- ll Regulation Class: �
- 890:5850 � Regulation Number:
#### 4. Predicate Device Information
| Predicate<br>Device | Predicate Device I | Predicate Device II |
|---------------------|-------------------------------|---------------------|
| Sponsor | Well Brain International Ltd. | Contour Technology |
{4}------------------------------------------------
| Predicate<br>Device | Predicate Device I | Predicate Device II |
|----------------------|-----------------------------------------------|-----------------------------------------|
| Device Name | Gymform® ABS-A-ROUND,<br>model: VDPGYCIND0016 | Contour Technology<br>Muscle Stimulator |
| 510(k) Number | K130074 | K111476 |
| Product Code | NGX | NGX |
| Regulation<br>Number | 21 CFR 890.5850 | 21 CFR 890.5850 |
| Regulation<br>Class | II | II |
### 5. Device Description
GYMFORM® ABS&CORE is a two channels battery operated muscle stimulation system specifically designed to stimulation the muscles.
The 2-area belt is intended for use on the muscles in abdomen or lower back separately.
Mini belt is intended for use on the muscles in arms, legs, thighs and buttocks areas separately.
It is comprised of a console for signal generation, two belts (2-area belt and mini belt) for fixation, and electrode pads for signal connection to skin. The electrode pads are replaceable.
Power is derived from 2 "AAA" batteries located in a compartment protected by a removable battery cover. There is no current passed from side to side. The user cannot access the wiring or connectors within the belt.
The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the device are controlled by the buttons. Its intensity level can be adjustable by user.
### 6. Intended Use / Indications for Use
GYMFORM®ABS& CORE is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS& CORE may be considered a technique or method for muscle training. 2-area belt is intended for use on the muscles in abdomen or lower back separately. Mini belt is intended for use on the muscles in arms, legs, thighs or buttocks areas separately.
### 7. Test Summary
ABS & CORE has been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1 and IEC 60601-2-10 standards �
{5}------------------------------------------------
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
- � Waveform test report to verify the output specifications of the device according to IEC 60601-2-10 and Guidance for Powered Muscle Stimulator.
- � Dispersion test and Shelf test according to ASTM F 1980-07, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devicesand Guidance: Shelf Life of Medical Device.
- � Adhesion test according to Section 5.4 of AAMI EC 12_2000_(R) 2010 Disposable ECG electrodes.
#### Comparison to predicate device and conclusion 8.
The technological characteristics, features, specifications, materials, mode of operation, and intended use of GYMFORM®ABS & CORE is substantially equivalent to the predicate devices quoted above.
The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements<br>of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
|------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|--------------|
| Basic Unit Characteristics | | | | | |
| Device Name and<br>Model | GYMFORM® ABS &<br>CORE, Model:<br>VDPGYCSET0042 | Gymform®<br>ABS-A-ROUND,<br>model:<br>VDPGYCIND0016 | Contour Technology<br>Muscle Stimulator | -- | |
| 510 (K) Number | Applying | K130074 | K111476 | -- | |
| Intended Use | GYMFORM®ABS&<br>CORE is intended to<br>stimulate healthy<br>muscles in order to<br>improve or facilitate<br>muscle<br>performance. The<br>ABS& CORE may<br>be considered a<br>technique or method<br>for muscle training. | Intended Use /<br>Indications for Use:<br>The<br>GYMFORM®ABS-A<br>-ROUND is<br>intended to<br>stimulate healthy<br>muscles in order to<br>improve or facilitate<br>muscle<br>performance. The<br>ABS-A-ROUND | The Contour<br>Technology Muscle<br>Stimulator is<br>intended to stimulate<br>healthy muscles in<br>order to improve or<br>facilitate muscle<br>performance. The<br>Contour Technology<br>Muscle Stimulator<br>may therefore be<br>considered a | SE | |
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| | intended for use on<br>the muscles in<br>abdomen or lower<br>back separately.<br>Mini belt is intended<br>for use on the<br>muscles in arms,<br>legs, thighs or<br>buttocks areas<br>separately. | may be considered<br>a technique or<br>method for muscle<br>training.<br>The 3-area belt is<br>intended for use on<br>the muscles in<br>abdomen, left waist<br>and right waist<br>alternately.<br>The Mini belt is<br>intended for use on<br>the muscles in<br>arms, legs (lower<br>extremities), thighs<br>and buttocks areas<br>separately. | technique or method<br>for muscle training.<br>The Contour<br>Technology Muscle<br>Stimulator Ab Belt<br>accessory is<br>intended for use on<br>abdominal muscles<br>only for<br>strengthening and<br>toning of abdominal<br>muscles.<br>The Contour<br>Technology Muscle<br>Stimulator BackPad<br>accessory is<br>intended for use on<br>the lower back<br>muscles only. | | |
| Stimulated muscles | Abdomen, lower<br>back, arms, legs,<br>thighs and buttocks | Abdomen, lower<br>back, arms, legs,<br>thighs and buttocks | Abdomen, lower<br>back | SE | |
| Power Sources | 2 x 1.5V AAA<br>batteries | 3 x 1.5V AAA<br>batteries | 4 x 1.5V AAA<br>batteries | SE<br>Note 1 | |
| Method of Line<br>Current Isolation | Battery Supply N/A | Battery Supply N/A | Battery Supply N/A | SE | |
| Patient<br>Leakage<br>Current | Normal<br>Condition | 10μA | 5.8μA | -- | SE<br>Note 1 |
| | Single<br>Fault<br>Condition | 50μA | 8.5μA | -- | SE<br>Note 1 |
| Number of Modes | | 6 | 6 | -- | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Number of Channels | 2 | 3 | -- | SE<br>Note 1 | |
| -Synchronous or<br>Alternating | Alternating | Alternating | -- | SE | |
| -Method of Channel<br>Isolation | Press MODE button<br>for 3 seconds | Press FLR button | -- | SE<br>Note 1 | |
| Number Intensity<br>Level | 31 steps | 99 steps | -- | SE<br>Note 1 | |
| Regulated Current or<br>Regulated Voltage | Regulated Voltage | Regulated Voltage | -- | SE | |
| Software/Firmware/Mi<br>croprocessor control | Yes | Yes | -- | SE | |
| Automatic Overload<br>Trip | No | No | -- | SE | |
| Automatic No-load<br>Trip | Yes. | Yes. | -- | SE | |
| Automatic Shut Off | Yes. | Yes | -- | SE | |
| Patient Override<br>Control | Yes | Yes | -- | SE | |
| Indicator On/Off Status | Yes | Yes | -- | SE | |
| Indicator Low Battery | Yes | Yes | -- | SE | |
| Display Voltage/Current Level | Yes | No | -- | SE | |
| Timer Range | Default time is 10<br>minutes | Default time is 19<br>minutes | -- | SE<br>Note 1 | |
| Console weight | 50g(Without<br>batteries) | 70g (Without<br>batteries) | -- | SE<br>Note 1 | |
| Accessories weight | Big belt: 150g<br>Small belt: 65g<br>Electrode pad: 15g<br>Carry bag: 60g | 3-area belt: 310g<br>Mini belt: 25g<br>Electrode pad (big):<br>40g<br>Electrode pad | -- | SE<br>Note 1 | |
| Elements<br>of | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Comparison | | | | | |
| | | (small): 26g | | | |
| Console dimensions | 100 mm(L) x 68 mm<br>(W) x 24.5 mm (H) | 91.8 mm(L) x 25.5 mm (W) x82 mm (H) | -- | SE<br>Note 1 | |
| Electrode<br>pad<br>dimension | 40 mm (L) x 70 mm<br>(W) x 3mm (H) | Small: 33.0 cm²<br>Big: 34.5 cm² | 40.5 cm² | SE<br>Note 1 | |
| Housing<br>Materials<br>and Construction | Console: ABS plastic<br>Belt: Polyester<br>Electrode pads:<br>Glycerine,<br>Polyacrylic acid,<br>Water and Salt | Console: ABS plastic<br>Belt: Polyester<br>Electrode pads:<br>Glycerine,<br>Polyacrylic acid,<br>Water and Salt | -- | SE | |
| Output Specification | | | | | |
| Waveform | Symmetrical | Symmetrical | -- | SE | |
| Shape | Rectangular | Rectangular | -- | SE | |
| Maximum Output<br>Voltage(Vpeak-topeak)<br>(+/- 10%) | 132V @ 500Ω<br>138V @ 2kΩ<br>140V @ 10kΩ | 108V @ 500Ω<br>124V @ 2kΩ<br>126V @ 10kΩ | -- | SE<br>Note 2 | |
| Maximum Current<br>Density(Ipeak-topeak) | 264mA @ 500Ω<br>69mA @ 2kΩ<br>14mA @ 10kΩ | 216mA @ 500Ω<br>62mA @ 2kΩ<br>12.6mA @ 10kΩ | -- | SE<br>Note 2 | |
| Frequency range | 2 Hz, 10 Hz, 50 Hz,<br>90 Hz, 120 Hz | 2 Hz, 10 Hz, 50 Hz,<br>90 Hz, 120 Hz | 1 to 120 Hz | SE<br>Note 2 | |
| Pulse width range | 108μs / 124μs | 100 μs / 120 μs | -- | SE<br>Note 2 | |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Pulse duration(Only<br>changes with the<br>mode) | Mode 1: 500ms;<br>Mode 2: 11.1ms;<br>Mode 3: 8.3ms;<br>Mode 4:<br>Front 90s: 11.1ms;<br>Back 90s: 8.3ms;<br>Mode 5: 100ms;<br>Mode 6: 20ms | Mode 1: 500ms;<br>Mode 2: 11.1ms;<br>Mode 3: 8.33ms;<br>Mode 4:<br>Front 90s: 11.1ms;<br>Back 90s: 8.33ms;<br>Mode 5: 100ms;<br>Mode 6: 20ms | 340 µs | SE<br>Note 2 | |
| Phase duration(Only<br>changes with the<br>mode) | Mode 1: Not<br>applicable(Continuo<br>us pulse)<br>Mode 2: 2s<br>Mode 3: 2s<br>Mode 4: 2s<br>Mode 5: 10s<br>Mode 6: 16s | Mode 1: Not<br>applicable(Continuo<br>us pulse)<br>Mode 2: 2s<br>Mode 3: 2s<br>Mode 4: 2s<br>Mode 5: 10s<br>Mode 6: 16s | -- | SE | |
| Net Charge | 19.2µC@ 500Ω | 15.7µC @ 500Ω | -- | SE<br>Note 2 | |
| Maximum Phase<br>Charge | 16.4µC@ 500Ω | 13.0µC@ 500Ω | -- | SE<br>Note 2 | |
| Maximum Current<br>Density | 0.082 mA/cm²@<br>500Ω | 0.057 mA/cm²@<br>500Ω | 0.55 mA/cm² @<br>500Ω | SE<br>Note 2 | |
| Maximum Power<br>Density | 94.8 μW/cm² @<br>500Ω | 53.8 μW/cm² @<br>500Ω | -- | SE<br>Note 2 | |
| ON time | 0.5s | 1s | -- | SE<br>Note 2 | |
| OFF time | 0.5s | 1s | -- | SE<br>Note 2 | |
| Contraction<br>and<br>Relaxation time | Adjustable, due to<br>different modes. | Adjustable, due to<br>different modes. | -- | SE | |
| Burst Mode | | | | | |
| Pulse per burst | 1~397 | 1~397 | -- | SE | |
| Elements of Comparison | Subject Device | Predicate Device I | Predicate Device II | Remark | |
| Bursts per second | 0.125~1 | 0.125~1 | -- | SE | |
| Burst Duration(s) | 1~8 | 1~8 | -- | SE | |
| Duty Cycle | 0.02%~1.28% | 0.02%~1.19% | -- | SE<br>Note 2 | |
| Additional Features | | | | | |
| Environment for operating | Temperature: 5 ~ 40°C<br>Humidity: 20 ~65% RH | Temperature: 5 ~ 45°C<br>Humidity: 20 ~65% RH | -- | SE | |
| | | | | | |
| | | | | | |
| Environment for storage | Temperature: 0~40° C<br>Humidity: 10 ~90% RH | Temperature: 5 ~ 45°C<br>Humidity: 20 ~65% RH | -- | SE | |
| | | | | | |
| | | | | | |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | SE | |
| | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.