PrimeSight™ UNITY 9000 Video Processor, PrimeSight™ UNITY 9100 Video Processor
K181292 · Cogentix Medical, Inc. · EOX · Nov 15, 2018 · Ear, Nose, Throat
Device Facts
Record ID
K181292
Device Name
PrimeSight™ UNITY 9000 Video Processor, PrimeSight™ UNITY 9100 Video Processor
Applicant
Cogentix Medical, Inc.
Product Code
EOX · Ear, Nose, Throat
Decision Date
Nov 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4710
Device Class
Class 2
Indications for Use
The PrimeSight UNITY 9000 Series Video Processors, when used in conjunction with a Cogentix Medical flexible videoscope, are indicated for the display and management of video and images during trans-nasal esophagoscopy, cystoscopy, bronchoscopy and nasopharyngoscopy procedures.
Device Story
Video processing unit for Cogentix Medical flexible videoscopes; inputs video signals from endoscope; transforms signals for display on integrated 19-inch touchscreen; outputs include live video and captured images; used in clinical settings by physicians; supports patient-centric file storage; enables zoom, white balance, and basic image functions; provides DICOM-formatted storage for EMR integration; replaces legacy DPU-5000 processor to address component obsolescence; includes air pump and stroboscopy interface in Airway Configuration (UVP-9100).
Clinical Evidence
Bench testing only. No clinical data used. Testing included functional performance verification, mechanical requirements, usability evaluation (IEC 62366-1), image quality (ISO 12233), environmental/transport testing (ISO 9022, ISTA 3A), and electrical safety/EMC (IEC 60601 series).
Technological Characteristics
Two-module system: All-in-one PC with 19-inch touchscreen and Camera Interface Module (CIM). Connectivity: DICOM support for EMR. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 14971, ISO 12233. Features: Integrated air pump and stroboscopy interface (UVP-9100 model).
Indications for Use
Indicated for patients undergoing trans-nasal esophagoscopy, cystoscopy, bronchoscopy, or nasopharyngoscopy. Used for display and management of video/images during endoscopic examination of the larynx, esophagus, gastro-esophageal junction, lower urinary tract (bladder), airway/lungs, and upper airway/vocal cords/nasal passages. No contraindications.
Regulatory Classification
Identification
An esophagoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the esophagoscope and is intended to examine or treat esophageal malfunction symptoms, esophageal or mediastinal disease, or to remove foreign bodies from the esophagus. When inserted, the device extends from the area of the hypopharynx to the stomach. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Flexible Trans-Nasal Video Esophagoscope with Digital Video Processor and Disposable EndoSheath Systems (K072088)
Vision-Sciences CST-5000 Video Cystoscope with EndoSheath System (K072180)
Vision Sciences BRS-5000 Video Bronchoscope with EndoSheath System, DPU-5000/DPU-5050 Video Processor and Accessories (K091768)
Vision-Sciences ENT-5000 Video ENT Scope with EndoSheath Technology, Vision-Sciences DPU-5000/DPU-5050 Video Processor (K102733)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2018
Cogentix Medical, Inc. Christa Blaisdell Regulatory Affairs Specialist 40 Ramland Road South Orangeburg, NY 10962
Re: K181292
Trade/Device Name: PrimeSight™ UNITY 9000 Video Processor, PrimeSight™ UNITY 9100 Video Processor Regulation Number: 21 CFR 874.4710 Regulation Name: Esophagoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOX, FAJ, EOB, OCS, EOQ Dated: October 15, 2018 Received: October 16, 2018
Dear Christa Blaisdell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181292
Device Name
PrimeSight™ UNITY 9000 Video Processor, PrimeSight™ UNITY 9100 Video Processor
Indications for Use (Describe)
The PrimeSight™ UNITY 9000 Series Video Processors, when used in conjunction with a Cogentix Medical flexible videoscope, are indicated for the display and management of video and images during trans-nasal esophagoscopy, cystoscopy, bronchoscopy and nasopharyngoscopy procedures.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### General Information
Date Prepared:
Submitter Information:
Registration Number: Contact:
September 11, 2018
Cogentix Medical, Inc. 40 Ramland Road South Orangeburg, New York, 10962
#### 2242464
Christa Blaisdell Regulatory Affairs Specialist Cogentix Medical, Inc. Phone: 952-426-6161 Fax: 866-255-4522 Email: christa.blaisdell@cogentixmedical.com
### Alternate Contact:
Nicole Boser Director, Quality and Regulatory Affairs Cogentix Medical, Inc. Phone: 952-426-6141 Fax: 866-255-4522 Email: nicole.boser@cogentixmedical.com
### Device Information
Trade Name:
PrimeSight™ UNITY 9000 Video Processor, PrimeSight™ UNITY 9100 Video Processor
Common Name: Video Processor Esophagoscope (flexible or rigid) and accessories Classification Name: 21 CFR 874.4710 Regulation Number: Product Classification: Class II Primary Product Code: EOX Secondary Product Code(s): FAJ, OCS, EOQ, EOB Primary Predicate Device Flexible Trans-Nasal Video Esophagoscope with Digital Video Processor and Disposable EndoSheath Systems (K072088)
{4}------------------------------------------------
| Secondary Predicate Device(s) | Vision-Sciences CST-5000 Video Cystoscope with EndoSheath<br>System (K072180) |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| | Vision Sciences BRS-5000 Video Bronchoscope with<br>EndoSheath System, DPU-5000/DPU-5050 Video Processor and<br>Accessories (K091768) |
| | Vision-Sciences ENT-5000 Video ENT Scope with EndoSheath<br>Technology, Vision-Sciences DPU-5000/DPU-5050 Video<br>Processor (K102733) |
# Device Description
The PrimeSight™ UNITY 9000 and PrimeSight™ UNITY 9100 Video Processors (herein referred to as "UNITY 9000", "UNITY 9000 Processor" or "UNITY 9000 Series Video Processors") are video processing units designed to work exclusively with the Cogentix Medical TNE-5000, CST-5000, CST-5000i. BRS-5100 and ENT-5000 models of flexible videoscopes (herein referred to as "5000 Series Videoscopes") to display images and videos during endoscopic procedures. The UNITY 9000 Processor consists of two modules which are physically connected with a hinge mechanism. The all-inone computer (AIO PC) unit is used for the presentation and display of images and includes a large 19'' wide screen display with an integrated touchscreen. The Camera Interface Module (CIM) base unit allows for the connection of a videoscope or camera to the Processor, as well as any endoscopic peripherals being used during the procedure, as well as image processing.
There are two different model configurations available for the UNITY 9000. These models are offered specific to the type of endoscopic procedure being performed. The Base Configuration model (UVP-9000) supports urology and bronchoscopy applications, while the Airway Configuration model (UVP-9100) offers an integrated air pump for trans-nasal esophagoscopy and a stroboscopy interface for use during Ear, Nose, and Throat (ENT) stroboscopy examinations.
# Indications for Use
The PrimeSight UNITY 9000 Series Video Processors, when used in conjunction with a Cogentix Medical flexible videoscope, are indicated for the display and management of video and images during trans-nasal esophagoscopy, cystoscopy, bronchoscopy and nasopharyngoscopy procedures.
## Intended Use
| Cogentix Medical<br>Flexible Videoscope | Flexible Videoscope<br>Model Number | Endoscopy System Indication |
|-----------------------------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Trans-nasal<br>esophagoscope | TNE-5000 | For the display and management of video and<br>images during endoscopic examination of the<br>larynx, esophagus, and gastro-esophageal junction |
| Cystoscope | CST-5000<br>CST-5000i | For the display and management of video and<br>images during endoscopic examination of the lower<br>urinary tract, including the bladder |
The PrimeSight UNITY 9000 Series Video Processors are intended to be used with a Cogentix Medical flexible videoscope to form an endoscopy system utilized for the following indications:
{5}------------------------------------------------
| Cogentix Medical<br>Flexible Videoscope | Flexible Videoscope<br>Model Number | Endoscopy System Indication |
|-----------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bronchoscope | BRS-5100 | For the display and management of video and<br>images during airway management of the trachea<br>and other major passages of the lungs and<br>endoscopic diagnosis of the airways |
| Nasopharyngoscope | ENT-5000 | For the display and management of video and<br>images during endoscopic examination of the upper<br>airway, vocal cords and/or nasal passages |
Additional information regarding the indication and intended use for each compatible flexible videoscope is contained within the User Manual specific to that device.
## Contraindications
There are no patient populations for which the PrimeSight UNITY 9000 Series Video Processors are contraindicated for.
## Comparison to Predicate Device
Cogentix Medical, Inc. (formerly Vision-Sciences, Inc.) holds the 510(k) clearances to market the current DPU-5000 Series Video Processor as part of its video endoscopy systems. These products were cleared for market through multiple 510(k) submissions, which include: the Flexible Trans-Nasal Video Esophagoscope with Digital Video Processor and Disposable EndoSheath Systems (K072088); the CST-5000 Video Cystoscope with EndoSheath System (K072180); the BRS-5000 Video Bronchoscope with EndoSheath System, DPU-5000/DPU-5050 Video Processor and Accessories (K091768); and, the ENT-5000 Video ENT Scope with EndoSheath Technology and DPU-5000/DPU-5050 Video Processor (K102733).
There are no significant differences in the intended use, mechanical and functional performance or the functional scientific technology between the UNITY 9000 Series Processors and the predicate device (DPU-5000). Both systems provide image presentation and capture and storage of pictures; provide the ability for the operator to store files under a patient centric directory; and allow controls on the endoscope to interface with the Processor for the purposes of zoom, white balance or other basic image functions of the system. The UNITY 9000 Processor simply provides an updated mechanical design which deals with parts obsolescence - and a new graphical interface when compared to its predicate. UNITY 9000 also introduces video capability and the ability to store images in a Digital Imaging and Communications in Medicine (DICOM) format for easier transposition to electronic medical records systems.
{6}------------------------------------------------
## Summary of Testing
Bench testing and ISO standard compliance testing, as described below in Table 1.1 – Summary of Testing, was performed on the PrimeSight UNITY 9000 Series Video Processors. The results of this testing confirm that the UNITY 9000 Processor is safe and effective and performs in a manner making it substantially equivalent to its predicate device. Clinical data was not used to demonstrate substantial equivalence as the subject device and predicate device are so similar in design and functionality.
| Category | Evaluation Test Reports | Associated Standards |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional<br>Performance | RPT-175-0020-VER – PrimeSight<br>Product Requirements Verification<br>Report | |
| | RPT-175-0021-VER – PrimeSight<br>Camera Interface Module Verification<br>Report | |
| | RPT-175-0023-VER – PrimeSight<br>Mechanical Requirements Verification<br>Report | EN ISO 14971:2012 – Medical devices.<br>Application of risk management to medical<br>devices |
| | TR 013-18 – Usability Evaluation of<br>UNITY 9000 Video Processor | IEC 62366-1:2015 – Medical devices - Part 1:<br>Application of usability engineering to medical<br>devices |
| | TR 006-18 – PrimeSight UNITY 9000<br>Gap Verification | IEC 62304:2006+A1:2015 – Medical device<br>software – Software life cycle processes |
| | TR 020-18 – Video Processor Image<br>Quality Performance | ISO 12233:2017 – Resolution and spatial<br>frequency responses |
| Software | RPT-175-0022-VER – PrimeSight User<br>Interface Software Requirements<br>Verification Report | |
| | RPT-175-0021-VER – PrimeSight<br>Camera Interface Module Verification<br>Report | |
| Environmental,<br>Shipping and<br>Transportation | | ISO 9022-1:2012 – Optics and photonics –<br>Environmental test methods - Part 1:<br>Definitions, extent of testing |
| | TR 032-17 – Environment and Transport<br>Testing of PrimeSight UNITY 9000 | ISO 9022-2:2015 – Optics and photonics –<br>Environmental test methods - Part 2: Cold,<br>heat and humidity |
| | | ISTA 3A:2008 – General Stimulation<br>Performance Tests |
| Category | Evaluation Test Reports | Associated Standards |
| Electrical Safety and<br>Electromagnetic<br>Compatibility | Report Number 103144900BOX-002 –<br>Cogentix Medical, Inc. Test Report –<br>EMC Testing – UVP-9100<br>Report Number 103480453BOX-001 –<br>Test Report – IEC 60601-1 Testing –<br>Endoscopic Video Processor<br>Report Number 103480453BOX-002 –<br>Test Report – IEC 60601-1-6 Testing –<br>Endoscopic Video Processor<br>Report Number 103480453BOX-003 –<br>Test Report – IEC 62304 Testing –<br>Endoscopic Video Processor<br>Report Number 103480453BOX-004 –<br>Test Report – IEC 60601-2-18 Testing –<br>Endoscopic Video Processor<br>Report Number 103480453BOX-005 –<br>Listing Constructional Data Report | IEC 60601-1:2005 + A1:2012 – Medical<br>electrical equipment. General requirements for<br>basic safety and essential performance<br>IEC 60601-2-18:2009 – Medical electrical<br>equipment – Part 2-18: Particular requirements<br>for the basic safety and essential performance<br>of endoscopic equipment<br>IEC 60601-1-2:2014 – Medical electrical<br>equipment. General requirements for basic<br>safety and essential performance. Collateral<br>Standard. Electromagnetic disturbances.<br>Requirements and tests<br>IEC 60601-1-6:2010 + A1:2013 – Medical<br>electrical equipment – Part 1-6: General<br>requirements for basic safety and essential<br>performance – Collateral standard: Usability<br>IEC 62304:2006+A1:2015 – Medical device<br>software – Software life cycle processes<br>EN ISO 14971:2012 – Medical devices.<br>Application of risk management to medical<br>devices |
## Table 1.1 – Summary of Testing
{7}------------------------------------------------
# Conclusion
Cogentix Medical, Inc. believes that based on the data contained in this 510(k) related to the indications for use, technological characteristics, and performance testing, the PrimeSight UNITY 9000 Series Video Processors have demonstrated substantial equivalence to their predicate device and are considered to be safe and effective.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.