VIDEO SCIENCES BRS-5000 VIDEO BRONCHOSCOPE WITH ENDOSHEATH SYSTEM, DPU-5000/DPU-5050 VIDEO PROCESSOR & ACCESSORIES
K091768 · Vision-Sciences, Inc. · EOQ · Aug 3, 2009 · Ear, Nose, Throat
Device Facts
Record ID
K091768
Device Name
VIDEO SCIENCES BRS-5000 VIDEO BRONCHOSCOPE WITH ENDOSHEATH SYSTEM, DPU-5000/DPU-5050 VIDEO PROCESSOR & ACCESSORIES
Applicant
Vision-Sciences, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Aug 3, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The bronchoscope and EndoSheath® Technology are designed to be used for upper airway management: including endoscopic treatment, diagnosis, and intubation of the airways. The digital video processor is intended for use with the VSI flexible video scope.
Device Story
The BRS-5000 is a flexible video bronchoscope used with a digital video processor and sterile, single-use EndoSheath® systems. The EndoSheath® covers the insertion tube to provide a protective barrier. The system captures visual data from the airway, which is transmitted to the digital video processor for display on a monitor. It is intended for use by clinicians in clinical settings for airway management, diagnosis, and treatment. The output allows the physician to visualize the airway anatomy in real-time to guide diagnostic or therapeutic procedures and facilitate intubation. The use of a disposable sheath aims to reduce the risk of cross-contamination between patients.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety, thermal, and EMC testing.
Technological Characteristics
Flexible video bronchoscope with digital video processor and disposable EndoSheath® protective barrier. Patient contact materials are identical to K072088. System includes image visualization and capture capabilities. Sterilization is not required for the single-use EndoSheath®; the scope is designed for compatibility with the sheath system.
Indications for Use
Indicated for patients requiring upper airway management, including endoscopic treatment, diagnosis, and intubation of the airways.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Flexible Trans-Nasal Esophagoscope with Digital Video Processor and Disposable EndoSheath® Systems (K072088)
Flexible Fiberoptic Bronchoscope with Disposable EndoSheath (K021344)
{0}------------------------------------------------
KO 91768
1100 0 700
510(k) Premarket Notification: Device Modification Video Bronchoscope with Digital Video Processor and BV Slide-On® EndoSheath® Systems
: . . . .
# 510(k) Summary
Date Prepared (21 CFR 807.92(a)(1): June 6, 2009
AUG 0 8 2009
Owner's Name (21 CFR 807.92(a)(1):
| | Vision-Sciences, Inc. |
|-------------------|-----------------------------------------------|
| Address: | 40 Ramland Road South<br>Orangeburg, NY 10962 |
| Telephone Number: | (845) 365-0600 |
| Fax Number: | (845) 365-0620 |
| Contact Person: | Lillian Quintero; Director QA/RA |
# Trade Name, Common Name, Classification (21 CFR 807.92(a)(2))
| Subject Device Name: | BRS-5000 Video Bronchoscope and BV Slide-On® EndoSheath®<br>Systems with Digital Video Processor |
|------------------------|--------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Flexible Video Bronchoscope with Sheath and Video Processor |
| Product Codes: | EOQ |
| FDA Regulations: | 21 CFR 874.4680 |
| Device Classification: | Class II |
## Predicate Device Names (21 CFR 807.92(a)(3))
K072088 Flexible Trans-Nasal Esophagoscope with Digital Video Processor and Disposable EndoSheath® Systems
K021344 Flexible Fiberoptic Bronchoscope with Disposable EndoSheath
Intubation of the airways for Bronchoscopes has received FDA marketing clearance in the following 510(k) Premarket Notifications:
K981543 Olympus Tracheal Intubation Fiberscope K082720 Olympus Tracheal Intubation Fibervideoscope
Common/Usual Name: Product Codes: FDA Regulations: Device Classification; Premarket Notification: Flexible fiberoptic endoscopes with sheaths and accessories EOB. EOO 21 CFR 874.4760 Class II K050972 / K040984 / K024095 / K942265
#### Device Description
The VSI flexible endoscopes are flexible endoscopes with connections to a video processor and display monitor. The EndoSheath® Systems are sterile, single-use protective sheath systems, with or without a working channel, that are intended to avver the entire insertion tube of the videoscope. The digital video processors are used with the flexible videoscopes for image visualization and capture.
#### Intended Use
The bronchoscope and EndoSheath® Technology are designed to be used for upper airway management: including endoscopic treatment, diagnosis, and intubation of the airways.
{1}------------------------------------------------
K091768
Vision-Sciences, Inc. June 9, 2009
#### 510(k) Premarket Notification: Device Modification Video Bronchoscope with Digital Video Processor and BV Slide-On® EndoSheath® Systems
#### Technological Characteristics
Vision Sciences believes that the subject device is substantially equivalent to the Vision Sciences' predicate devices. The subject device has very similar material composition and working dimensions; and identical viewing direction, image size, bending, re-processing/sterilization method and working dimensions as In addition, the subject device uses the same video processor as Vision Sciences other the predicate. videoscopes.
### Performance Testing
The subject device has been subjected to and passed electrical safety, thermal, and ECM testing requirements. The patient contact materials in the endoscope are identical to the materials used in predicate device (Vision Sciences' TNE-5000 (K072088)).
#### Conclusion
Based on the indications for use, technological characteristics, performance testing and comparison to predicate devices, the VSI flexible video endoscopes with digital video processors and disposable EndoSheath® Systems have been shown to be safe and effective for their intended use.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird figure.
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 0 3 2009
Vision Sciences, Inc c/o Stacie Geffner-Ativa 40 Ramland Road, South Orangeburg, NY 10962
Re: K091768
Trade/Device Name: Flexible Video Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope and accessories Regulatory Class: Class II Product Code: EOQ Dated: June 11, 2009 Received: June 16, 2009
Dear Ms. Geffner-Atiya:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Signature
Malvina B. Eydelman, M.D. · Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Vision-Sciences, Inc. June 9, 2009
510(k) Premarket Notification: Device Modification Video Bronchoscope with Digital Video Processor and BV Slide-On® EndoSheath® Systems
510(k) Number (if known):
K091718
Video Bronchoscope with Digital Video Processor and BV Slide-On® Device Name: EndoSheath® Systems
The bronchoscope and EndoSheath® Technology are designed to be Indications for Use: used for upper airway management: including endoscopic treatment, diagnosis, and intubation of the airways.
The digital video processor is intended for use with the VSI flexible video scope.
Prescription Use X OR Over-the -Counter Use (Per 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Karen H. Baker
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K091768
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.